Prsentation PowerPoint - PowerPoint PPT Presentation

1 / 92
About This Presentation
Title:

Prsentation PowerPoint

Description:

Prsentation PowerPoint – PowerPoint PPT presentation

Number of Views:69
Avg rating:3.0/5.0
Slides: 93
Provided by: biop4
Category:

less

Transcript and Presenter's Notes

Title: Prsentation PowerPoint


1
Light is a moving energy. Widefield, confocal,
spinning disc, advanced concepts Nathalie Garin,
Theo Lasser, EPFL
2
  • Magnification
  • Resolution
  • Working distance
  • Sampling

3
Working distance
Working distance in confocal, WD is an absolute
limit where the focal plan can be bellow the top
surface of the coverslip
Examples 63x/1.3 glycerol, wd0.28mm 63x/1.4
oil, wd0.1mm 63x/1.2 water, wd0.22mm
4
Comparison (Zeiss handbook)
5
(No Transcript)
6
Imaging one point
7
0.1um beads 63x/1.4 CY3 z step 280nm 235 images
Object 100nm bead
XZ image
8
E. Keller, handbook of biol. conf. mic.
9
E. Keller, handbook of biol. conf. mic.
10
Resolution Abbes law (or Rayleigh criterion)
Definition resolution of an optical system is
the smallest distance between 2 point sources
such that their diffraction patterns show a
detectable drop in intensity between them.
Lateral (XY) rairy 0.61
rairy the radius of the first dark around the
central disk of the Airy diffraction image
r
n refractive index of mounting medium
11
Rayleigh criteria
12
(No Transcript)
13
(No Transcript)
14
(No Transcript)
15
Illumination patterns and PSF
16
PSF loss of axial symetry
A. Egner and S. W. Hell
17
PSF loss of axial symetry
Beads in glycerol
Apochromat 63x/1.2 water imm
Apochromat 63x/1.3 glycerol imm
18
Refractive index mismatch
Objective 10x dry Sample matrigel in DABCO
19
Examples
Sample mammospheres in DABCO
Apochromat 20x glycerol
Apochromat 20x oil
20
Examples
Sample mouse embryo E11.5 in DABCO, draq5
Apochromat 63x/1.3 glycerol
Apochromat 63x/1.4 oil
21
Examples
Sample moss in water
xz
Apochromat 63x/1.2 water imm
Apochromat 63x/1.3 glycerol imm
22
Perfect Imaging
23
Confocal principle
L
a
s
e
r
P
i
n
h
o
l
e
-
P
i
n
h
o
l
e
o
b
j
e
c
t
i
v
e
b
e
a
m
d
e
t
e
c
t
o
r
s
p
l
i
t
t
e
r
S
c
a
n
m
i
r
r
o
r
T
u
b
e
-
l
e
n
s
l
i
g
h
t
d
e
f
o
c
u
s
e
d
p
l
a
n
e
l
i
g
h
t
c
o
n
j
u
g
a
t
e
d
O
b
j
e
c
t
i
v
e
p
l
a
n
e
O
b
j
e
c
t
-
Courtesy T. Lasser, EPFL
p
l
a
n
e
24
Point scanning (Leica)
Three independent galvos Low inertia ? high
speed Correct optics ? UV to IR No XY-Movement
in pupil
25
Leica TCS SP2 Innovation K-Scanner
  • Projection of single plane into objective pupil
    ensures perfect resolution
  • Low moving mass ensures high frame rates (which
    turned out to be wrong...)
  • Rotating complete assembly rotates scan w/o
    compromising speed
  • Large FOV (22mm)
  • Correct Illumination (low vignetation, Flat field)

26
correct illumination with beam scanning
galvo
optics
pupil
objective
focus
27
incorrect illumination with beam scanning
Pivot point
galvo
optics
pupil
objective
inhomogeneous field
focus
28
Pinhole
29
(No Transcript)
30
(No Transcript)
31
(No Transcript)
32
(No Transcript)
33
(No Transcript)
34
(No Transcript)
35
(No Transcript)
36
(No Transcript)
37
(No Transcript)
38
(No Transcript)
39
(No Transcript)
40
Spinning disc confocal
41
Spinning disc confocal
42
Structured illumination
Wide-field
Structured illumination
43
Structured illumination
In theory, structured illumination technic has
the same resolution as confocal microscope (Mitic
J. et al. Opt. Lett 2003) Limitation
inhomogenous illumination
44
Comparison
Wide field
ApoTome
Confocal
45
Deconvolution
  • Deconvolution is a mathematical operation used
    in image restoration to recover an object from an
    image that is degraded by blurring and noise.
  • In fluorescence microscopy the blurring is
    largely due to diffraction limited imaging by the
    instrument the noise is mainly photon noise.
  • Every element in the optical path affects the
    point spread function
  • Microscope type
  • Objective
  • Coverslip
  • Mounting medium
  • Sample (cell culture, slices)
  • External factors
  • (T, vibrations, etc)
  • The PSF is the geometrical signature of the
    microscope (and the sample)

46
PSF
Confocal PSF
Widefield PSF
47
What restore means?
  • Reattribute the out of focus light to the point
    source
  • Raise resolution of small objects
  • Increase definition of structure in multiple
    dimensions
  • Improve signal to noise (SNR)
  • Simplest processing for segmentation
  • More accurate localization of intensity for
    quantification, ratiometry and colocalization
  • Only images acquired with the correct sampling
    parameters
  • (Nyquist, cf. calculator) can be deconvolved
  • Bad acquisitions are not improved

48
(No Transcript)
49
(No Transcript)
50
(No Transcript)
51
2-photon confocal microscopy
PSF
Brad Amos
52
(No Transcript)
53
STED STimulated Emission Depletion
Stefan Hell
54
The target of STED microscopy
  • Increase xy resolution in fluorescence
    microscopy over classical Abbé Limits
  • Inventor Stefan Hell, Max Planck Institute for
    Nanobiophotonics, Goettingen, Germany
  • xy resolution 200nm
  • z resolution (confocal) 500nm

55
STED Principle

Courtesy Stefan Hell, MPI for Nanobiophotonics,
Goettingen, Germany
  • Two superimposed picosecond pulsed beams
    excitation beam and ring- shaped red-shifted
    depletion beam, tightly synchronized
  • Ring-shaped beam depletes excited molecules in
    outer focal area before fluorescence is
    emitted ? sharpens up focus
  • Resolution is not limited by wavelength any
    more

56
STED Resolution
Point Spread Function Confocal vs
STED measured with fluorescent nanoparticles
under the same conditions
  • Typical lateral (X-Y) resolution in a
    Confocal 200x200 nm
  • Typical lateral (X-Y) FWHM in STED is 90x90
    nm
  • STED z resolution is confocal (500 nm)
  • STED enables separation of structures even
    smaller than its FWHM due to the sharp
    peak Actual resolution is lt 72 nm

57
65 nm fluorescent beads are not resolved in a
Confocal


Confocal
58
65 nm fluorescent beads are resolved with STED


STED
59
STED Application
Enhanced detail on PTK2 cells, microtubules


Confocal
STED
60
Confocal Application Neuromuscular Synapses
Substructuresare not resolved
1 micrometer
61
STED Application Neuromuscular Synapses
STED resolves substructures of presynaptic active
zones. Images are taken from Drosophila
neuromuscular synapses. Liprin protein, stained
with ATTO 647N 1024x1024 pixels Courtesy
Stephan SigristWuerzburg
1 micrometer
62
STED Application Neuromuscular Synapses after
linear deconvolution
STED resolves substructures of presynaptic active
zones. Images are taken from Drosophila
neuromuscular synapses. Liprin protein, stained
with ATTO 647N 1024x1024 pixels, Courtesy
Stephan SigristWuerzburg
1 micrometer
63
Confocal Application Neuromuscular Synapses
Substructures are not resolved
1 micrometer
64
STED Application Neuromuscular Synapses
STED resolves substructures of presynaptic active
zones (Ca channels) Images are taken from
Drosophila neuromuscular synapses. Bruchpilot
protein stained withATTO 647N. 2048x2048
pixels Courtesy Stephan SigristWuerzburg
1 micrometer
65
Functional imaging (FRAP, FRET, FCS, )
66
FRAP Fluorescence Recovery After Photobleaching
Axelrod  2D model of diffusion of membrane-bound
molecules the diffusion coefficient D for a
particle in a free volume depends on the
Boltzmann constant (k), the absolute temperature
(T), the viscosity of the solution (h), and the
hydrodynamic radius (R) of the particle. It is
affected by the following parameters The
size of the molecule an eightfold increase of
the size of a soluble sperical protein decreases
D by factor 2. the viscosity of the
cellular environment e.g. membranes have a much
higher viscosity than cytoplasm
protein-protein-interactions and binding to
macromolecules can also slow down the diffusion
if flow or active transport is involved in
the movement of the probed molecule, the measured
movement rate can become significantly higher
than the theoretical diffusion rate- Assuming a
gaussian profile for the bleaching beam, the
diffusion constant D can be simply calculated
from D w2 / 4t1/2 where w is the width of
the beam and t1/2 is the time required for the
bleach spot to recover half of its initial
intensity.
Tutorial http//www.embl.de/eamnet/frap/index.htm
l
67
FRAP
68
FRAP
Ref B. L. Sprague and J. G. McNally, TRENDS in
Cell Biology, 2005.
69
FRET Fluorescence Resonance Energy Transfert
70
Fluorescence Resonance Energy Transfer
  • Laurent Gelman
  • Facility for Advanced Imaging and Microscopy

71
Overview of the presentation
Basic principles of FRET Techniques to measure
FRET
Courtesy of Laurent Gelman, FMI
72
Fluorescence Resonance Energy Transfer
No interaction
Dimerization
emission
excitation
emission
excitation
D
A
D
A
FRET is the non-radiative transfer of excitation
energy from a donor fluorophore to a nearby
acceptor. (Förster, Ann. Phys. 1948 55-75)
Donor Fluorophore
Acceptor Fluorophore
S1
S1
absorption
emission
emission
S0
S0
73
Energy Transfer Efficiency
- The energy transfer efficiency E is related to
the distance r separating a given donor and
acceptor pair.
1 1 (r/R0)6
D
A
E
r
  • R0 is the distance at which 50 of FRET occurs.
    The higher, the better.
  • Distance between fluorophores must be between 1
    and 10 nm.
  • - For a given donor and acceptor pair, E is a
    function of
  • the overlap between the donor emission and
    acceptor excitation spectra
  • the absorption coefficient for the acceptor
  • the quantum yield of the donor
  • the distance and the relative orientation of the
    donor and acceptor.

74
FRET and molecular interactions
  • FRET occurs over a maximal distance of 10nm
  • ? FRET increases the resolution of light
    microscopy (250nm).
  • FRET occurs over distances similar to the size
    of proteins.
  • ? Detection of a FRET signal is almost equivalent
    to interaction.
  • ? No FRET signal does not mean no interaction.

75
Donor and Acceptor Em/Ex spectra I
- The overlap between the donor emission and
acceptor excitation spectra is crucial. - The
possibility to resolve the emission spectra from
both fluorophores may be an issue too (sensitized
emission)
Calculation websites http//www.mcb.arizona.edu/i
pc/fret/index.html http//home.earthlink.net/pubs
pectra/
EYFP
ECFP
New pairs CyPet-Ypet, Miyawaki's CY11.5,
Cerulean-mCitrine Cy3-Cy5, Alexa488-Alexa555,
Alexa488-Cy3 GFP-Cy3
76
How to measure FRET?
3 Techniques
? Acceptor photobleaching quenching of donor
fluorescence (fixed cells)
? Sensitized emission of acceptor fluorescence
(living cells)
? Fluorescence lifetime microscopy (FLIM-FRET,
requires special equipment)
A plethora of methods
etc See Berney C, Danuser G, Biophysical J 2003
3992-4010
77
Acceptor photobleaching
No FRET
FRET

A
A
D
D
IDA
detection
detection
IDA
excitation
excitation
78
Acceptor photobleaching
FRET
No FRET
ECFP
A
D
A
D
EYFP
79
Acceptor photobleaching
IDA - IDA IDA
E
  • - Requires photodestruction of the acceptor.
  • - Best applied to fixed samples or single
    time-point measurements. Not appropriate for
    continuous monitoring of FRET over time or for
    fast diffusing molecules.
  • - Application of a high laser power to bleach
    acceptor may be deleterious for the sample.
  • There is a risk to bleach the donor.
  • Excess of acceptor needed.
  • Very sensitive to incomplete bleaching of the
    acceptor.
  • See also Kenworthy et al., Methods (2001), p.289
    (review including a troubleshooting section)

80
Sensitized emission FRET
A
D
Detection
Excitation
Main problem results from channel cross-talk ?
Select filters, microscope settings and
fluorophores that minimize these channel
cross-talks or "bleed-throughs".
FRET IFRET - BTdonor emission - BTacceptor
excitation
81
Bleed-through (BT) calculation
Requires to analyze 2 sets of cells expressing
the donor or the acceptor alone
Cells expressing ECFP alone
FRET setting
ECFP setting
525-545
465-485
IFRET
BTECFP
IECFP
405
405
Cells expressing EYFP alone
FRET setting
EYFP setting
525-545
525-545
IFRET
BTEYFP
IEYFP
405
514
82
FRET measurement and calculation
Cells expressing A-ECFP and B-EYFP alone
FRET channel
EYFP channel
ECFP channel
525-545
525-545
465-485
405
514
405
FRET IFRET - (IECFP x BTECFP) - (IEYFP x
BTEYFP)
83
FRET normalization
FRET channel
ECFP channel
EYFP channel
FRET
ECFP EYFP
ECFP fused to EYFP
FRET IECFP
FRET (IECFP x IEYFP)
FRET (IECFP x IEYFP)1/2
Xia Z, Liu Y. Biophys J. 2001 2395-402.
84
More corrections
IFRET - (IECFP x BTECFP) - (IEYFP x
BTEYFP) (IECFP x IEYFP)1/2
NFRET
  • Not necessary for the ECFP/EYFP pair if
    detection windows are small enough.
  • Not enough to be quantitative
  • See also
  • Chen et al, Biophys J 2006 L39-41
  • Van Rheenen et al, Biophys J 2004 2517-29
  • Gordon GW et al, Biophys J 1998 2702-13

85
Other controls
- The ratio between the fluorescently labeled
proteins is crucial and should never exceed 5
to10 (sensitized emission FRET). - Before the
FRET experiment, the amount of vectors giving
similar expression levels should be determined by
western blot.
A-ECFP
B-EYFP
EYFP-B
ng of vector transfected
10 20 30 40
10 20 10 20
- Expression vectors for ECFP fused to EYFP,
giving FRET and no FRET are useful controls
86
Image processing
87
FRET biosensors
? Calcium signaling (Chameleon and related
Calcium-sensors)
? Spatiotemporal monitoring of RhoA activation at
the sub-cellular level in living cells (Olivier
Pertz et al. Nature 2006)
Model and system Mouse Embryonic Fibroblasts
expressing a RhoA biosensor (tet-inducible) RhoA
(full length) - YFP - Linker - CFP - Rhotekin
(Rho-binding domain)
Method Sensitized emission FRET
NFRET IFRET / ICFP
Olivier Pertz et al. Nature 2006
88
Acknowledgements
Jens Rietdorf Patrick Schwarb
Center for Integrative Genomics
Jérôme Feige Béatrice Desvergne Walter Wahli
89
FCS Fluorescence Correlation Spectroscopy
  • Functional imaging
  • single protein detection

FCS (fluorescent correlation spectroscopy) is
based on the measurement of fluorescence
fluctuations. This fluctuations can be due to the
diffusion of the fluorophore in the excitation
volume or a change of fluorescence quantum yield
due to chemical reactions.
90
FCS
91
Haustein, E. Schwille, P. Curr Opin Struct Biol,
2004
92
References
  • D. Axelrod, D.E. Koppel, J. Schlessinger, E.
    Elson, W.W. Webb (1976), Biophysical Journal 16,
    1055-1069.
  • M. J. Booth and T. Wilson Journal of Biomedical
    Optics - July 2001 - Volume 6, Issue 3, pp.
    266-272
  • Kevin Braeckmans, Liesbeth Peeters, Niek N.
    Sanders, Stefaan C. DeSmedt, and Joseph Demeester
    Biophys J. 2003 October 85(4) 2240?2252.
    Three-Dimensional Fluorescence Recovery after
    Photobleaching with theConfocal Scanning Laser
    Microscope
  • A. Egner and S. W. Hell, Aberrations in confocal
    and multi-photon fluorescence microscopy induced
    vy refractive index mismatch (Handbook of Biol.
    Conf. Microscopy)
  • Haustein, E. Schwille, P. Curr Opin Struct Biol,
    2004
  • F. Helmchen and W. Denk, 2005, Nature methods,
    Deep tissue two-photon microscopy
  • L. Landmann Journal of Microscopy, Vol. 208, Pt.
    2 November 2002, pp. 134-147 - Deconvolution
    improves colocalization more than filtering
    techniques
  • J. Pawley, Handbook of Biological Confocal
    Microscopy
  • T. Staudt et al., 2007, Microscopy Research and
    Technique, 2,2-Thiodiethanol a new water
    soluble mounting medium for high resolution
    optical microscopy
  • H. Zhao et al. 2005, Journal ord Biomedical
    Optics, Emission spectra of bioluminescent
    reporters and interaction with mammalian tissue
    determine the sensitivity of detection in vivo
  • R. M. Zucker et al. - 1998- Cytometry - confocal
    laser scanning microscopy of apoptosis in
    organogenesis-stage mouse embryos
Write a Comment
User Comments (0)
About PowerShow.com