Title: Slayt 1
1Lecture 2 Light scattering and determination
of the size of macromolecules
2- Theory
- Static Light scattering(SLS) (static" or
"Rayleigh" scattering or MALLS) - (molecular weight, hydrodynamic size)
-
- Dynamic Light scattering(DLS) (photon
correlation spectroscopy (PCS) - or quasi-elastic light scattering (QELS))
(polydispersity) - Electrophoretic Light scattering(ELS)
Zeta potential - Application examples
-
- Molecular weight
- Sizing
- Polydispersity
-
3(No Transcript)
4(No Transcript)
5(No Transcript)
6--gt Fraunhofer Theory (diffraction)
--gt Mie Theory (diffraction - diffusion)
The Fraunhofer theory is applicable for large
particles compared to the wavelength l
(diffusion and absorption are not considered).
For smaller particles, it is appropriate to use
Mie Theory.
7(No Transcript)
8(Rayleigh orani)
990o (1 Cos2 ?) 1
10(No Transcript)
11(No Transcript)
12(No Transcript)
13In polymer physics, the radius of gyration is
used to describe the dimensions of a polymer
chain. The radius of gyration of a particular
molecule at a given time is defined as
where
is the mean position of the
monomers. As detailed below, the radius of
gyration is also proportional to the root mean
square distance between the monomers
The theoretical hydrodynamic radius Rhyd arises
in the study of the dynamic properties of
polymers moving in a solvent. It is often similar
in magnitude to the radius of gyration.
14The radius of gyration for this case is given by
aN represents the contour length of the polymer
contour length (in polymers) The maximum
end-to-end distance of a linear polymer chain.For
a single-strand polymer molecule, this usually
means the end-to-end distance of the chain
extended to the all-trans conformation. For
chains with complex structure, only an
approximate value of the contour lengthmay be
accessible. IUPAC Compendium of Chemical
Terminology, 2nd Edition, 1997
15(No Transcript)
16(No Transcript)
17(No Transcript)
18(No Transcript)
19Debye plots are most accurate when applied to any
macromolecule with Rg lt 12 nm, including globular
proteins and dendrimers. In addition, such plots
are generally accurate for random coil polymers
with Mw lt 100 kDa.
20(No Transcript)
21(No Transcript)
22(No Transcript)
23(No Transcript)
24(No Transcript)
25(No Transcript)
26(No Transcript)
27(No Transcript)
28(No Transcript)
29(No Transcript)
30(No Transcript)
31(No Transcript)
32(No Transcript)
33(No Transcript)
34(No Transcript)
35(No Transcript)
36Static LS
Dynamic LS
Particle Sizing in Concentrates by Dynamic Light
Scattering
Static LS Dynamic LS
37(No Transcript)
38(No Transcript)
39(No Transcript)
40(No Transcript)
41NORMALIZATION N(q)-1 ( I ray.scatt.(q) I
solvent(q)) ( Iray.scatt.(90o)
Isolvent(90o)) Rq I x cal.cte. x N(q)
Rq soln Isoln x cal.cte. x N(q)
Rq solvent Isolvent x cal.cte. x N(q)
?Rq Rq, soln Rq, solvent (
Isoln Isolvent) x cal.cte. x N(q)
scattering intensity of toluene at 90o
............................I toluene 0.964
q scattering vector(angle) rayleigh ratio of
toluene _at_ 660nm 1.183E-5 cm-1 cal. cte.
rayleigh ratio of toluene at 660nm I
toluene (90o) web adress to find out (dn/dc)
values for the polymers www.ampolymer.com/FRD/dn
dc.htm
42to find out the scattering intensities of the
samples at each angle, we have to divide the
intensity that is read by the instrument for that
angle by the referance intensity again read by
the instrument.
For PS ntoluen 1.4903 (dn/dc)PS 0.1050
ml/g K (4p2 n02 (dn/dc)2) / (NA ?4)
( ? 660nm) Cal Cte 1.2271 x 10 -5
4 x (3.14)2 (1.4903)2(0.1050)2 ml2/g2 K
----------------------------------------------
6.02 x10 23mol-1 x (660)4 nm4
0.965706797 K -----------------------------
8.454 x 10 -36 ml2 mol /g2 nm4 1.1422791 x
10 35 K 8.4 x 10 -36 cm6 mol /g2 10-28cm4 K
8.4 x 10 -8 cm2 mol /g2 R? 1.183E-5 cm-1 Kc
(cm2 mol /g2) g/cm3 mol ___
-------------------- ------- R?
cm-1 g
43- Experimental procedure
- Preparation of Rayleigh scatter
- Preparation of polymer/protein solutions
3) Determination of calibration constant of
Instrument (cal. Cte) 4) Measuring of scattering
intensity of normalization solution. 5)
Measuring of scattering intensity of solvent and
solutions
44Exercise 1
- The Rayleigh ratio for a series of dilute
solutions of polymethyl methacrylate(PMMA) in - ethylene dichloride at 25 oC was determined in a
light scattering photometer at various - angles ?. The table shows values of C/?R? for
the various concentrations (c) and - scattering angles (?).
- __________________________________________________
____________________ - c
- __________________________________________________
____________________ - ? 0.0096 0.0048 0.0024 0.0012
- 30 56.3 35.9 26.4 21.4
- 45 57.1 36.4 26.7 21.5
- 60 57.5 36.8 26.8 21.8
- 75 58.3 37.5 27.6 22.6
- 90 59.1 38.4 28.3
23.6 - __________________________________________________
___________________ - Given n 1.5 , dn/dc 0.11 cm3 g-1 , ? 436
nm and Avagadros Number 6.03 x 1023, - calculate Mw and Rg of PMMA.
45Exercise 2
46(No Transcript)
47Small-angle scattering
- Small-angle scattering (SAS) is a scattering
technique based on the deflection of a beam of
particles, or an electromagnetic or acoustic
wave, away from the straight trajectory after it
interacts with structures that are much larger
than the wavelength of the radiation. The
deflection is small (0.1-10) hence the name
small-angle. SAS techniques can give information
about the size, shape and orientation of
structures in a sample. - SAS can refer to
- Small angle neutron scattering (SANS)
- Small-angle X-ray scattering (SAXS)
- Biological small-angle scattering, SAXS or SANS
applied to biological systems
48Small angle neutron scattering (SANS)
- Small angle neutron scattering (SANS) is a
laboratory technique, similar to the often
complementary techniques of small angle X-ray
scattering (SAXS) and light scattering. - While analysis of the data can give information
on size, shape, etc., without making any model
assumptions a preliminary analysis of the data
can only give information on the radius of
gyration for a particle using Guinier's
equation.1
49Technique
- During a SANS experiment a beam of neutrons is
directed at a sample, which can be an aqueous
solution, a solid, a powder, or a crystal. The
neutrons are elastically scattered by changes of
refractive index on a nanometer scale inside the
sample which is the interaction with the nuclei
of the atoms present in the sample. Because the
nuclei of all atoms are compact and of comparable
size neutrons are capable of interacting strongly
with all atoms. This is in contrast to X-ray
techniques where the X-rays interact weakly with
hydrogen, the most abundant element. - In zero order dynamical theory of diffraction the
refractive index is directly related to the
scattering length density and is a measure of the
strength of the interaction of a neutron wave
with a given nucleus.
50Guinier law
51Guinier law
52Small Angle X-ray Scattering (SAXS)