Title: Magnetic Resonance Imaging Part 1 The Science Bit
1Magnetic Resonance Imaging Part 1The Science
Bit
- Lynn Graham DCR Msc
- Clinical Specialist in MRI
2Particle Physics!
3OUTLINE ( part 1)
- History Local origins of MRI
- Fundamental Physics of MRI
- Tissue contrast Versatility
4History Lesson
- Carl Fredrich Gauss. (1777-1855)
- German Physicist
- Findings led to a knowlegdge of magnetism and its
quantifiation - Gauss- unit of measurement of magnetism
- Nikola Tesla (1856 1943)
- Serbian Electrical Engineer
- Work in electromagnetic induction
- Tesla unit of measurement for Magnetic Field
strength
5Sir Joseph Larmor FRS.MA.DSCMathematician
Physicist 1857-1942
- Born 11th July 1857 at Magheragall, Co Antrim
- Educated at RBAI, Belfast
- Graduated from Queens 1877
- Appointed Professor _at_ St Johns College Cambridge
1903 - Knighted 1909
6The Larmor Equation
Key to Nuclear Magnetic Resonance
Clinical Magnetic Resonance Imaging
7Nuclear Magnetic Resonance
- NUCLEAR ATOMIC SPIN
- ve electric charge
- Intrinsic spin/ Precession
- nuclear magnetic moment
8PRECESSION
9Nuclear Magnetic Resonance
- MAGNETIC MOMENTS ALIGN WITH B0
No B0 random motion
B0 alignment
10Nuclear Magnetic Resonance
Out of phase
In phase
11OUT OF PHASE
12IN PHASE
13The Larmor Equation
- ?L Larmor frequency (MHz)
- B0 magnetic field (Tesla)
- gyromagnetic ratio
Key to Nuclear Magnetic Resonance
14NMR CLINICAL MRI
Fat Water 99 body tissue
H ALIGNMENT PRESCESSION
15NMR CLINICAL MRI
- Apply the Larmor equation
H1 _at_ 1 T ? 42.58 MHz T-1
_at_ 1.5 T Larmor frequency 63.87 MHz
16Electromagnetic Spectrum
Precessional Frequency of H
17Resonance Excitation
- Energy in the form of an RF pulse
- Leads to misalignment with B0 antiparallel
- Also leads to phase coherence.
This is Excitation
18Resonance / Excitation
B0
RF pulse
19Relaxation
- Remove the RF and the spins will loose their
energy. - Realign with B0 relax
- Loose phase coherance decay
Energy loss is variable
20Relaxation / Decay
B0
21Excitation Relaxation MR Signal
Bo
NMV
90 RF pulse
Current induced in RF coil due to alternating B
field MRI signal
22Particle Physics almost done!
23Image Formation
X
Z
Y
24Spatial Localisation
B0
B0
Gradient
B0 -
25Image Formation
XX
Y
Z
26Pixel Mapping
Each line of data is stored as the Image is
built up gradually Fourier transform decodes
data forms the image
Phase
Frequency
27X
Z
Y
28Resolution
Few pixels Short scan time
Many pixels Long scan time
29The MR effect!
30Differing MR Images
31Relaxation Free Induction Decay (FID)
- The spins will loose their energy in two ways
Energy decays slowly Relaxing back to B0
T1 Recovery
T2 decay
Loose phase coherance
32T1 Recovery
B0
Realign with
33T2 Decay
In phase
Out of phase
Loose phase
34Brownian Motion
- FAT
- Large , slow molecules
- Lots of bumps
- Fast energy loss
- Short T1 Short T2
- WATER
- Small , fast molecules
- Fewer bumps
- Slow energy loss
- Long T1 long T2
Mr Blobby Vs Speedy Gonzalez!
35Typical T1 T2 values for tissues (_at_1.5T)
Tissue T1 value T2 value
Distilled water Cerebro Spinal Fluid Gray Matter White Matter Fat Muscle Liver Kidney 3000 2400 900 780 260 750 500 760 3000 160 100 90 80 50 40 30
36Pulse Sequences
- Pre-set sequences of excitation, relaxation and
signal organization that vary tissue contrast and
image quality.
37T1 and T2 weightings
- Sag spine T2W
- Fluid bright
38Tissue differentiation
- gt 99 body tissues produce MR signal
- Each tissue has unique properties
- - molecular
structure - - number of H
ions - -
moving/stationary - Each tissue behaves differently in the MR
environment - ?
- Unique MR signals from normal abnormal
tissues - ?
- Excellent disease
diagnosis.
39Tissue contrast Versatility of MRI
T1 SE
T2 SE
T1 SE gad
GE brain
40Tissue contrast Versatility of MRI
Fat sat orbits
FLAIR
Black blood
Angio
41Coming up Next !!!
- Clinical Applications of MRI
- MRI Equipment
- Safety issues of MRI
- Advantages Disadvantages of MR
-
- MRI vs Other imaging modalities ( CT/ USS)
- Clinical Images
-