Title: Quality Control in Diagnostic Radiology
1- Quality Control in Diagnostic Radiology
2Factors driving Q.C.Why do we do it?
- Legal Requirements
- Accreditation
- JCAHO
- ACR
- Clinical improvement
- equipment performance
- image quality
Medical Physicists at Work
3Q.C. Goals
- Minimize dose to
- patients
- staff
- Optimize image quality
- Establish baselines
- More on this in a moment
4Why is Q.C. Important?
Without a QC program the only way to identify
problems is on patient images. And some
problems dont show up on images.
Yeah, thats what I always say.
5QC can detect
- Malfunctions
- Unpredictability
- may be hard to isolate clinically
- Inefficient use of Radiation
- high fluoroscopic outputs
- Radiation not reaching receptor
- inadequate filtration
- oversized collimation
6Goals of a Q.C. Program
- Obtain acceptable image with least possible
radiation exposure to - patients
- staff
- Attempt to identify problems before they appear
on patient films - without QC problems only detected on patient films
7Acceptable Image
- Image containing information required by
radiologist for correct interpretation - goal minimize exposure while maintaining
acceptability - high exposure films often have excellent
appearance - cardboard cassettes
8Q.C. Baselines
- Baselines
- quantitative data on equipment obtained during
normal operations - Baselines useful for troubleshooting
- isolating problem component, for example
- generator
- processor
- Allows efficient use of engineering / repair
personnel
9X-Ray Quality Control
- Processor Sensitometry
- Filtration
- Focal Spot Size
- Collimation
- Maximum Fluoroscopic Output
- Calibration Verification
- Phototimer Performance
10Photographic Density
- Optical density
- measure of film blackness or opacity
IoD log --- ItwhereD
densityIo light incident on filmIt light
transmitted by film - D1 gt 1/10 of light transmitted
- D2 gt 1/100 of light transmitted
- D3 gt 1/1000 of light transmitted
Io
Film
It
11H D Curve
- Shows relationship between radiation striking
film/screen and optical density - Function of
- film
- screen
- kVp
- developer temp
- chemistry condition
12Sensitometry Parameters
- Base Fog
- O.D. of unexposed portion of film
Base Fog
13Sensitometry Parameters
- Speed
- O.D. of selected step about 1.0 above base fog
O.D.
H D Curve
1.0
Speed
Log Relative Exposure(step)
14Sensitometry Parameters
- Contrast
- slope of H D curve
- Difference between optical densities of two
selected steps - Higher step O.D. - Lower step O.D.
15Temperature Problems
- developer (critical)
- affects
- contrast
- speed
- base fog
- must be controlled within about /- .5o
- wash water
- on older model processors, wash water was
pre-heat for developer - dryer
16Film Processor Sensitometry
- Sensitometer
- flashes calibrated step wedge on film
- Densitometer
- reads optical density (O.D.) of selected steps
- Plot results
- Notes
- Use Control Film
- Perform before first clinical useof processor
17(No Transcript)
18(No Transcript)
19Why is Filtration Important?
- Tube emits spectrum of x-ray energies
- Filtration preferentially attenuates low energy
photons - low energy photons expose patients
- do not contribute to image
- low penetration
20Half Value Layer (HVL)
- We dont measure filtration
- We measure HVL
- HVL amount of absorber that reduces beam
intensity by exactly 50
21Half Value Layer
- Depends upon
- kVp
- waveform (single/three phase)
- inherent filtration
- Minimum HVL regulated by law
- Maximum HVL regulated only in mammography
Georgia State Rules Regulations for X-Ray
22Radiographic HVL Setup
23Checking HVL Compliance(Radiographic)
- How much aluminum must be placed in beam to
reduce intensity by exactly 50?
90 kVp Measurements 2.5 mm Al minimum HVL
filter mR (mm Al) -------------------
0 250 2.5 133
filter mR (mm Al) -------------------
0 250 2.5 125
filter mR (mm Al) -------------------
0 250 2.5 111
Not OK! Must remove Al to reduce beam to exactly
50
OK! Must add Al to reduce beam to exactly 50
Acceptable HVL gt 2.5 mm
Marginal HVL 2.5 mm
Unacceptable HVL lt 2.5 mm
24Checking HVL Compliance(Radiographic)
- Is this machine legal?
- 2.5 mm Al minimum filtration at 90 kVp
90 kVp Measurements
filter mR (mm Al) -------------------
0 450 2.5 205
25Fluoroscopic HVL Setup
26Fluoroscopic HVL
- Set desired kilovoltage manually
- measure exposure rates instead of exposure
- Move absorbers into beam as needed
27Focal Spot Size
- We measure apparent focal spot
- Trade-off
- smaller spot reduces geometric unsharpness
- larger spot improves heat ratings
28Focal Spot Size (cont.)
- Focal spot size changes with technique
- Standard technique required
- 75 kV (typical)
- 50 maximum mA for focal spot at kV used
- direct exposure (no screen)
- NEMA Standardsdefines tolerances
Nominal Size Tolerance ------------------------
------------- gt1.5 mm
30 gt0.8 and lt1.5 mm 40 lt0.8 mm
50
29Focal Spot Measuring Tools
- Direct MeasurementPin Hole Camera
- Slit Camera
- Indirect Measurement of Resolving Power
- Star Test Pattern
- Bar Phantom
30Direct Focal Spot Measurement
- Measure focal spot directly in each direction
- Use triangulation to correct for distances
- formula corrects for finite tool size
- two exposures required for slit
Slit Camera
Pinhole Camera
31Star Test Pattern
- Measures resolving power
- infers focal spot size
- Dependent on focal spot energy distribution
- measure
- largest blur diameter (in each direction)
- magnification
- use equation to calculate focal spot size
32Bar Phantom
- Measures resolving power
- Find smallest group where you can count three
bars in each direction
33Bar Phantom Setup
34Radiographic Collimation
- X-Ray / Light Field Alignment
- Beam Central Axis
- should be in center of x-ray beam
- Collimator field size indicators
- PBL (automatic collimation)
- field automatically limited to size of receptor
- Bucky Alignment
- Using longitudinal bucky light transverse
detent, x-ray field should be centered on bucky
film
35X-Ray / Light Field Alignment
- Mark light field on table top with pennies
36Radiographic X-Ray / Light Field Alignment
37Fluoroscopic Collimation
- image field is scale seen on monitor
- expose film on table above scale
- compare visual field (monitor) with x-ray field
on film - must check all magnification modes
38Fluoroscopic Collimation
39Fluoroscopic Collimation
40Maximum Fluoro Output
- put chamber in beam on tabletop
- block beam with lead above chamber
- fools generator into providing maximum output
- 10 R/min. limit for ABS fluoro
41Maximum Fluoro Output
Lead
42Calibration Performance Parameters
- Timer Accuracy
- Repeatability
- Linearity/Reciprocity
- Kilovoltage accuracy
- mA
- must be measured invasively
43Non-invasive Calibration Tools
- Fancy
- Ion chambers
- Electronic Black Boxes
- Not as fancy
- Wisconsin Test Cassette
- Pocket Dosimeter
- Spin Top
44Timing Review
- Single Phase
- Full-wave rectified
- 120 pulses / second
- Half-wave rectified
- 60 pulses / second
- Self rectified
- 60 pulses second
- Three Phase / Constant Potential / Medium or High
Frequency - continuous output (not pulsed)
45Time Measurement
- Digital Black Box Meter
- Spin Top
- spins manually
- one dot per pulse (single phase)
- can not use with 3 phase
- Synchronous Spin Top
- spins at a set rate
- can use for single or 3 phase
- measure angle for 3 phase
One hole in solid disk
46Spin Top
- If this is a half-wave rectifier, what is
exposure time?
47Spin Top
- If this is a full-wave rectifier, what is
exposure time?
48Synchronous Spin Top
- What is this waveform?
- What is the exposure time?
90o
1 revolution / sec.
49kVp Measurement
- Waveform represented by a single number
- Wisconsin Test Cassette
- Digital beam analyzers
- use differential filtration
- Invasive measurements
- dynalyzer
50Calibration
120 kVp
mA time mAs mR mR / mAs
(msec) ------------------------------------------
------------ 100 .1 10 240
24 200 .05 10 ?
? 50 .2 10
? ?
Constant mAs
- mR/mAs should stay constant for all combinations
of mA kVp at any particular kVp
51Calibration
120 kVp
mA time mAs mR mR / mAs
(msec) ------------------------------------------
----------- 100 .1 10 240
24 200 .1 20 ?
? 100 .4 40 ?
?
Double mAs
Double mAs again
- mR/mAs should stay constant for all combinations
of mA time at any particular kVp
52Phototiming(check with output or film)
- Reproducibility
- Density Controls
- Field Placement
- Field Balance
Phototiming Operation should be Predictable
53Phototimer Density Control Settings
R
R
T
a
b
l
e
t
o
p
54Phototiming Density Steps should be predictable
approximately even
55Phototimer Field Placement / Balance
- Placement
- cover desired field with lead
- select field as indicated
- Balance
- no fields covered
- select field as indicated
56Phototimer Field Placement / Balance
57Phototimingchecked with film density
- kV Response
- phototimer pick-up attenuation may vary with kV
- phototimer must track kV response of rare-earth
film - Rate Response
- Check with varying
- phantom (lucite) thickness
- mA
58kV/Rate Response
kV
70
81
90
Lucite
17.5
4.5
4.9
5.2
Depth
12.5
4.7
(cm)
7.5
4.7
Thickness Tracking
4
Optical
Density
2
0
17.5
12.5
7.5
Lucite Thickness
59Phototiming(etc.)
- Minimum response time
- Limitations of phototimer with rare earth systems
- Back-up time function
- Maximum mAs
- 600 mAs (above 50 kVp)
- 2,000 mAs (below 50 kVp)
60The End
Any questions, you varmints?