Quality Control in Diagnostic Radiology - PowerPoint PPT Presentation

1 / 60
About This Presentation
Title:

Quality Control in Diagnostic Radiology

Description:

Factors driving Q.C. Why do we do it? Legal Requirements Accreditation JCAHO ACR Clinical improvement equipment performance image quality Q.C. Goals Minimize dose to ... – PowerPoint PPT presentation

Number of Views:921
Avg rating:3.0/5.0
Slides: 61
Provided by: George57
Category:

less

Transcript and Presenter's Notes

Title: Quality Control in Diagnostic Radiology


1
  • Quality Control in Diagnostic Radiology

2
Factors driving Q.C.Why do we do it?
  • Legal Requirements
  • Accreditation
  • JCAHO
  • ACR
  • Clinical improvement
  • equipment performance
  • image quality

Medical Physicists at Work
3
Q.C. Goals
  • Minimize dose to
  • patients
  • staff
  • Optimize image quality
  • Establish baselines
  • More on this in a moment

4
Why is Q.C. Important?
Without a QC program the only way to identify
problems is on patient images. And some
problems dont show up on images.
Yeah, thats what I always say.
5
QC can detect
  • Malfunctions
  • Unpredictability
  • may be hard to isolate clinically
  • Inefficient use of Radiation
  • high fluoroscopic outputs
  • Radiation not reaching receptor
  • inadequate filtration
  • oversized collimation

6
Goals of a Q.C. Program
  • Obtain acceptable image with least possible
    radiation exposure to
  • patients
  • staff
  • Attempt to identify problems before they appear
    on patient films
  • without QC problems only detected on patient films

7
Acceptable Image
  • Image containing information required by
    radiologist for correct interpretation
  • goal minimize exposure while maintaining
    acceptability
  • high exposure films often have excellent
    appearance
  • cardboard cassettes

8
Q.C. Baselines
  • Baselines
  • quantitative data on equipment obtained during
    normal operations
  • Baselines useful for troubleshooting
  • isolating problem component, for example
  • generator
  • processor
  • Allows efficient use of engineering / repair
    personnel

9
X-Ray Quality Control
  • Processor Sensitometry
  • Filtration
  • Focal Spot Size
  • Collimation
  • Maximum Fluoroscopic Output
  • Calibration Verification
  • Phototimer Performance

10
Photographic Density
  • Optical density
  • measure of film blackness or opacity
    IoD log --- ItwhereD
    densityIo light incident on filmIt light
    transmitted by film
  • D1 gt 1/10 of light transmitted
  • D2 gt 1/100 of light transmitted
  • D3 gt 1/1000 of light transmitted

Io
Film
It
11
H D Curve
  • Shows relationship between radiation striking
    film/screen and optical density
  • Function of
  • film
  • screen
  • kVp
  • developer temp
  • chemistry condition

12
Sensitometry Parameters
  • Base Fog
  • O.D. of unexposed portion of film

Base Fog
13
Sensitometry Parameters
  • Speed
  • O.D. of selected step about 1.0 above base fog

O.D.
H D Curve
1.0
Speed
Log Relative Exposure(step)
14
Sensitometry Parameters
  • Contrast
  • slope of H D curve
  • Difference between optical densities of two
    selected steps
  • Higher step O.D. - Lower step O.D.

15
Temperature Problems
  • developer (critical)
  • affects
  • contrast
  • speed
  • base fog
  • must be controlled within about /- .5o
  • wash water
  • on older model processors, wash water was
    pre-heat for developer
  • dryer

16
Film Processor Sensitometry
  • Sensitometer
  • flashes calibrated step wedge on film
  • Densitometer
  • reads optical density (O.D.) of selected steps
  • Plot results
  • Notes
  • Use Control Film
  • Perform before first clinical useof processor

17
(No Transcript)
18
(No Transcript)
19
Why is Filtration Important?
  • Tube emits spectrum of x-ray energies
  • Filtration preferentially attenuates low energy
    photons
  • low energy photons expose patients
  • do not contribute to image
  • low penetration

20
Half Value Layer (HVL)
  • We dont measure filtration
  • We measure HVL
  • HVL amount of absorber that reduces beam
    intensity by exactly 50

21
Half Value Layer
  • Depends upon
  • kVp
  • waveform (single/three phase)
  • inherent filtration
  • Minimum HVL regulated by law
  • Maximum HVL regulated only in mammography

Georgia State Rules Regulations for X-Ray
22
Radiographic HVL Setup
23
Checking HVL Compliance(Radiographic)
  • How much aluminum must be placed in beam to
    reduce intensity by exactly 50?

90 kVp Measurements 2.5 mm Al minimum HVL
filter mR (mm Al) -------------------
0 250 2.5 133
filter mR (mm Al) -------------------
0 250 2.5 125
filter mR (mm Al) -------------------
0 250 2.5 111
Not OK! Must remove Al to reduce beam to exactly
50
OK! Must add Al to reduce beam to exactly 50
Acceptable HVL gt 2.5 mm
Marginal HVL 2.5 mm
Unacceptable HVL lt 2.5 mm
24
Checking HVL Compliance(Radiographic)
  • Is this machine legal?
  • 2.5 mm Al minimum filtration at 90 kVp

90 kVp Measurements
filter mR (mm Al) -------------------
0 450 2.5 205
25
Fluoroscopic HVL Setup
26
Fluoroscopic HVL
  • Set desired kilovoltage manually
  • measure exposure rates instead of exposure
  • Move absorbers into beam as needed

27
Focal Spot Size
  • We measure apparent focal spot
  • Trade-off
  • smaller spot reduces geometric unsharpness
  • larger spot improves heat ratings

28
Focal Spot Size (cont.)
  • Focal spot size changes with technique
  • Standard technique required
  • 75 kV (typical)
  • 50 maximum mA for focal spot at kV used
  • direct exposure (no screen)
  • NEMA Standardsdefines tolerances

Nominal Size Tolerance ------------------------
------------- gt1.5 mm
30 gt0.8 and lt1.5 mm 40 lt0.8 mm
50
29
Focal Spot Measuring Tools
  • Direct MeasurementPin Hole Camera
  • Slit Camera
  • Indirect Measurement of Resolving Power
  • Star Test Pattern
  • Bar Phantom

30
Direct Focal Spot Measurement
  • Measure focal spot directly in each direction
  • Use triangulation to correct for distances
  • formula corrects for finite tool size
  • two exposures required for slit

Slit Camera
Pinhole Camera
31
Star Test Pattern
  • Measures resolving power
  • infers focal spot size
  • Dependent on focal spot energy distribution
  • measure
  • largest blur diameter (in each direction)
  • magnification
  • use equation to calculate focal spot size

32
Bar Phantom
  • Measures resolving power
  • Find smallest group where you can count three
    bars in each direction

33
Bar Phantom Setup
34
Radiographic Collimation
  • X-Ray / Light Field Alignment
  • Beam Central Axis
  • should be in center of x-ray beam
  • Collimator field size indicators
  • PBL (automatic collimation)
  • field automatically limited to size of receptor
  • Bucky Alignment
  • Using longitudinal bucky light transverse
    detent, x-ray field should be centered on bucky
    film

35
X-Ray / Light Field Alignment
  • Mark light field on table top with pennies

36
Radiographic X-Ray / Light Field Alignment
37
Fluoroscopic Collimation
  • image field is scale seen on monitor
  • expose film on table above scale
  • compare visual field (monitor) with x-ray field
    on film
  • must check all magnification modes

38
Fluoroscopic Collimation
39
Fluoroscopic Collimation
40
Maximum Fluoro Output
  • put chamber in beam on tabletop
  • block beam with lead above chamber
  • fools generator into providing maximum output
  • 10 R/min. limit for ABS fluoro

41
Maximum Fluoro Output
Lead
42
Calibration Performance Parameters
  • Timer Accuracy
  • Repeatability
  • Linearity/Reciprocity
  • Kilovoltage accuracy
  • mA
  • must be measured invasively

43
Non-invasive Calibration Tools
  • Fancy
  • Ion chambers
  • Electronic Black Boxes
  • Not as fancy
  • Wisconsin Test Cassette
  • Pocket Dosimeter
  • Spin Top

44
Timing Review
  • Single Phase
  • Full-wave rectified
  • 120 pulses / second
  • Half-wave rectified
  • 60 pulses / second
  • Self rectified
  • 60 pulses second
  • Three Phase / Constant Potential / Medium or High
    Frequency
  • continuous output (not pulsed)

45
Time Measurement
  • Digital Black Box Meter
  • Spin Top
  • spins manually
  • one dot per pulse (single phase)
  • can not use with 3 phase
  • Synchronous Spin Top
  • spins at a set rate
  • can use for single or 3 phase
  • measure angle for 3 phase

One hole in solid disk
46
Spin Top
  • If this is a half-wave rectifier, what is
    exposure time?

47
Spin Top
  • If this is a full-wave rectifier, what is
    exposure time?

48
Synchronous Spin Top
  • What is this waveform?
  • What is the exposure time?

90o
1 revolution / sec.
49
kVp Measurement
  • Waveform represented by a single number
  • Wisconsin Test Cassette
  • Digital beam analyzers
  • use differential filtration
  • Invasive measurements
  • dynalyzer

50
Calibration
120 kVp
mA time mAs mR mR / mAs
(msec) ------------------------------------------
------------ 100 .1 10 240
24 200 .05 10 ?
? 50 .2 10
? ?
Constant mAs
  • mR/mAs should stay constant for all combinations
    of mA kVp at any particular kVp

51
Calibration
120 kVp
mA time mAs mR mR / mAs
(msec) ------------------------------------------
----------- 100 .1 10 240
24 200 .1 20 ?
? 100 .4 40 ?
?
Double mAs
Double mAs again
  • mR/mAs should stay constant for all combinations
    of mA time at any particular kVp

52
Phototiming(check with output or film)
  • Reproducibility
  • Density Controls
  • Field Placement
  • Field Balance

Phototiming Operation should be Predictable
53
Phototimer Density Control Settings
R
R
T
a
b
l
e
t
o
p
54
Phototiming Density Steps should be predictable
approximately even
55
Phototimer Field Placement / Balance
  • Placement
  • cover desired field with lead
  • select field as indicated
  • Balance
  • no fields covered
  • select field as indicated

56
Phototimer Field Placement / Balance
57
Phototimingchecked with film density
  • kV Response
  • phototimer pick-up attenuation may vary with kV
  • phototimer must track kV response of rare-earth
    film
  • Rate Response
  • Check with varying
  • phantom (lucite) thickness
  • mA

58
kV/Rate Response
kV
70
81
90
Lucite
17.5
4.5
4.9
5.2
Depth
12.5
4.7
(cm)
7.5
4.7
Thickness Tracking
4
Optical
Density
2
0
17.5
12.5
7.5
Lucite Thickness
59
Phototiming(etc.)
  • Minimum response time
  • Limitations of phototimer with rare earth systems
  • Back-up time function
  • Maximum mAs
  • 600 mAs (above 50 kVp)
  • 2,000 mAs (below 50 kVp)

60
The End
Any questions, you varmints?
Write a Comment
User Comments (0)
About PowerShow.com