Title: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
1RADIATION PROTECTION INDIAGNOSTIC
ANDINTERVENTIONAL RADIOLOGY
IAEA Training Material on Radiation Protection in
Diagnostic and Interventional Radiology
- L 22 Optimization of Protection in Dental
Radiology
2Introduction
- Dental radiology makes use of specific types of
equipment, needed for different purposes.
Frequent exposures though each with low dose
involve a risk for the practitioner and for the
patient - Background general principles of x-ray
diagnostic imaging
3Topics
- Dental X-ray equipment
- Radiation protection in dental radiology
- Quality control for dental equipment
4Overview
- To be able to apply the principle of radiation
protection to dental radiology system including
design and Quality Control.
5IAEA Training Material on Radiation Protection in
Diagnostic and Interventional Radiology
Part 22 Optimization of protection in dental
radiology
Topic 1 Dental x-ray equipment
6Types of units
- Intra-Oral units
- Standard dental tube
- uses an intra-oral image receptor
- has extra-oral x-ray tube
- Panoramic (OPG)
- Cephalometric ( Ceph)
7Intra-Oral Dental X-Ray Equipment
8Modern Dental X-Ray Unit
9Panoramic X-Ray Equipment
10Cephalometric X-Ray Equipment
11X-Ray Tube
- stationary Anode
- avoid overheating
- tube duty cycle
- typical 130 intaroral
- 110 OPG
- 420 mAs/hr intraoral
12Tube Head
13Generator Circuit
14Generators Pre-Heat
- Medium frequency - stable waveform
- Single phase (SP) - pulsed
- Pre-Heat separate circuit for heating filament
- Single Phase units without a pre-heat circuit
- initial pulses of variable kV
15Collimator
- 1. Lead Collimator
- with central hole
- 2. Spacer Tube
16Applicator Cones
Good Bad
Bad
17Cephalometric Holder
18Intra-Oral Dental X-Ray Equipment (technical data)
- Exposure time from 60 ms to 2.5 s
- Tube Min. 50 kV, 7mA
- Focal spot size ?1 mm
- Inherent filtration 2 mm Al equivalent
- Focus-skin distance 20 cm
- Irradiated field 28 cm2 with round section, 6
cm diameter collimator
19Panoramic X-Ray Equipment (technical data)
-
- Focal spot 0.5 mm
- kV 60 - 80 kV in 2 kV steps
- mA 4 - 10 mA steps 4, 5, 6, 8, 10
- Exposure time 12 s (standard projections) 0.16 -
3.2 s (cephalometric projections) - Flat panoramic cassette 15x30 cm (Lanex Regular
screens))
20Image Receptors in Dental Radiology
Intraoral Radiology
- Small films (2 x 3 or 3 x 4 cm) in light-tight
envelopes (no screen) - Digital intraoral sensors - compared with
category E film, the radiation dose is reduced by
60.
Panoramic Radiology and Cephalometry
- Film-screen combination
- Digital sensors - compared with film-screen
sensitivity class 200, the radiation dose is
reduced by 50-70.
21Dental Radiology Film Types
- Sensitivity class D
- Very good spatial resolution
- Typical delivered dose about 0.5 mGy
- Typical exposure times 0.3 - 0.7 s
- Sensitivity class E
- Good spatial resolution
- Typical delivered dose about 0.25 mGy
- Typical exposure times 0.1 - 0.3 s
22Part 22 Optimization of Protectionin Dental
Radiology
IAEA Training Material on Radiation Protection in
Diagnostic and Interventional Radiology
- Topic 2 Radiation Protection in Dental Radiology
23Radiation Protection in Dental Radiology
- Facts
- Very frequent examination (about 25 of all the
radiological examinations) - Delivered doses may differ of a factor 2 or 3.
(entrance doses between 0.5 and 150 mGy) - Image Quality often very low
- Organs at risk parathyroid, thyroid, larynx,
parotid glands
24Radiation Protection in Dental Radiology
Technical hints to reduce patient doses Quality
Control of Film Processing
- Keep under control time and temperature of the
developing process. - Do not use oxydized chemicals
- Do not adjust development time by viewing the film
25RP in Dental Radiology
Technical hints to reduce patient doses Privilege
the use of sensitive films
- Sensitivity class D
- Very good spatial resolution
- Typical delivered dose about 0.5 mGy
- Typical exposure times 0.3 - 0.7 s
- Sensitivity class E
- Good spatial resolution
- Typical delivered dose about 0.25 mGy
- Typical exposure times 0.1 - 0.3 s
26Radiation Protection in Dental Radiology
Technical hints to reduce patient doses Lead
apron and collar
Useful when the path of primary beam intercepts
the protected organs (downward bite-twin
projection).
27Radiation Protection in Dental Radiology
Panoramic examination
- Image quality not as good as in intra-oral films
- Important global information
- Relatively low dose(one panoramic examination ?
3?5 intra-oral films)
28 Part 22 Optimization of Protection in Dental
Radiology
IAEA Training Material on Radiation Protection in
Diagnostic and Interventional Radiology
- Topic 3 Quality Control for Dental Equipment
29Why Dental QC ?
- Widespread use of dental units
- Lack of QC history on most units
- Dental practitioners working in the primary
health care sector do not have the continuous
medical physics support available in a
hospital-based diagnostic imaging department
30What Tests ?
- Collimation
- Dose Evaluation
- Exposure Time
- Half Value Layer
- Kilovoltage (kVp)
- Leakage Radiation
31Quality Control for Dental Equipment
- The recommended tests are consequently divided
into - those simple tests which can be performed by
dental practice staff - those more complex tests which can be carried
out by medical physicists.
32Quality Control for Dental Equipment
Tests which can be performed bydental practice
staff
33Quality Control for Dental Equipment
Tests performed by medical physicists
34Dental QC Methods
- Unit
- Intra-Oral Receptors
- (I/O)
- Cephalometric
- (Ceph)
- Panoramic
- (OPG )
- Test Method
- as for Radiology QC
-
- as for Radiology QC
- where possible
- immobilise unit
- remove slit collimator
35Test Equipment
- kVp meter
- measure kVp average
- Programmable delay 100 ms
- Range 50 to 120 kV
- Aluminum filters
- 4 x 1mm
- Grade 1100
- Dosimeter
- small large volume chambers
- 2mm wide detector for OPG
- Timer
- triggering at 75 peak kV
36Collimation
- Expose film or fluorescent screen
- Measure x-ray field image
37Collimator Light Intensity
- Ceph units
- Place external detector 1m from focus
- Measure illuminance in lux ( lumen/m2)
- Read each quadrant
- Limit gt100 lux at 1m
38Dead man Switch
- timer at 50 cm from focus
- set low kV, mA, long time
- start exposure
- release switch during exposure
- Require exposure cut-out when switch is
released. Check exposure time is less than set
time
39Kilovoltage
- Accuracy
- Set kVp meter to 100 msec delay
- Observe kVp waveform at 70 kV if poss.
- Limit measured kVp within 5 of set value
- Reproducibility
- Take 5 repeat exposures
- Limit coefficient of variation 2
40Dose Evaluation
- Skin dose from I/O units
- place cone 10 mm from dosimeter
- set maxillary molar/ bitewing setting
- Should be (65-70 kVp)
- 2-3 mGy for molar view
- lt 5 mGy for any view
41Output Reproducibility (1)
- Standard I/O units Ceph units
- Dosimeter position
- I/O units 10 mm from cone
- Ceph units 75 cm from focus
- or other recommended distance
- Five repeat exposures
- Limit coefficient of variation 5
42Output Reproducibility (2)
- Optional Method for OPG units
- Align detector on film cassette slit
- Measure dose rate
- Take 5 repeat exposures
- Limit coefficient of variation 5
43Exposure Time Accuracy
- Standard I/O units Ceph units
- Set timer to trigger at 75 peak kV
- Test times in the normal working range
-
- Limit 10 error for I/O units
- 5 error for all other units
44Timer Reproducibility
- Standard I/O units Ceph units
- Place timer in beam
- 5 repeat exposures
- Limit coefficient of variation 5
45Half Value Layer (HVL)
- Standard I/O units
- Position cone facing down
- Place dosimeter at 40 cm from focus
- Position Al filters near end of cone
- Measure dose
- measure with no added filters
- with 2,3,4 mm Al added, then again with no
filters - Plot on semi-log paper and find HVL
46HVL Set-up
47HVL Ceph OPG Units
- Position Al filters on collimator
- Cephalometric units
- Position dosimeter at 75 cm from focus
- OPG Units
- Position dosimeter on film cassette slit
- Measure dose rate, dose for fixed exposure time,
or dose for full scan - NB Test kVp accuracy before measuring HVL
48OPG Quality Control (kVp/HVL measurement)
49HVL- Minimum Values
- kVp HVL (mm Al)
- Intraoral Ceph/OPG
- 60 1.5 1.8
- 70 1.5 2.1
- 80 2.3 2.3
- 90 2.5 2.5
50Leakage Radiation
- Cover collimator with 2 mm lead
- Set large ionization chamber (180 cc) or solid
state detector at known distance d cm from focus - Expose on maximum kVp for 0.5 - 1 sec (but be
careful!) - Scale up the reading to 1 hour
51Leakage Radiation
- Choose the most likely points for leakage -
- where collimator joins to housing, lateral to the
tube, joints in housing etc. - Normalise measurement to 1 m from focus with
inverse square law - Dose1m dosed x (distanced/100)2
52Leakage Measurement
- In normal radiography, leakage measurements are
normalised to the maximum continuous rated tube
current R - Leakage Dose (for P mAs at 100 cm) x (R x
3600/P) - BUT...a dental unit has no continuous operation
rating
53Leakage Measurement
- Instead we normalize the dose reading to tube
duty cycle - Leakage Limit at 1 m from focus
- 0.25 mGy/hour - Standard I/O units
- 1.0 mGy/hour - Ceph OPG units
54Calculation of Leakage
- Measured leakage at 100 cm 2mGy/hr
- Duty cycle 1 20 (i.e. must wait 20 times
the exposure time between exposures to allow tube
cooling) - Thus actual leakage 2 / 20 mGy/hr
- 0.1 mGy/hr
- i.e. acceptable
55Radiation Protection in Dental Radiology
Quality Control of Film Processing
- Keep under control time and temperature of the
developing process. - Do not use oxidized chemicals
- Regularly check processing with phantom
56Dental Phantom
57Dental Phantom
58Where to Get More Information
- Code of practice for protection of persons
against ionising radiations arising from medical
and dental use, DHSS, HMSO, London, 1964 - Guidance notes for the protection of persons
against ionising radiations arising from medical
and dental use, DHSS, HMSO, London, 1988 - Radiation protection and quality assurance in
dental radiology. Radiation protection 81.
European Commission.(1995) CG-89-95-971-EN-C
59Summary
- General and specific types of x-ray equipment
(and image receptors) are reviewed (panoramic
and cephalometric), with technical data about
operating conditions - Although doses are generally low, the high
frequency of examinations requires radiation
protection (for the practitioner) in dental
radiology - Some tests are detailed for Quality control of
dental equipment.