RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY - PowerPoint PPT Presentation

1 / 59
About This Presentation
Title:

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY

Description:

Background: general principles of x-ray diagnostic imaging. IAEA ... uses an intra-oral image receptor. has extra-oral x-ray tube. Panoramic (OPG) ... – PowerPoint PPT presentation

Number of Views:175
Avg rating:3.0/5.0
Slides: 60
Provided by: iaea4
Category:

less

Transcript and Presenter's Notes

Title: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY


1
RADIATION PROTECTION INDIAGNOSTIC
ANDINTERVENTIONAL RADIOLOGY
IAEA Training Material on Radiation Protection in
Diagnostic and Interventional Radiology
  • L 22 Optimization of Protection in Dental
    Radiology

2
Introduction
  • Dental radiology makes use of specific types of
    equipment, needed for different purposes.
    Frequent exposures though each with low dose
    involve a risk for the practitioner and for the
    patient
  • Background general principles of x-ray
    diagnostic imaging

3
Topics
  • Dental X-ray equipment
  • Radiation protection in dental radiology
  • Quality control for dental equipment

4
Overview
  • To be able to apply the principle of radiation
    protection to dental radiology system including
    design and Quality Control.

5
IAEA Training Material on Radiation Protection in
Diagnostic and Interventional Radiology
Part 22 Optimization of protection in dental
radiology
Topic 1 Dental x-ray equipment
6
Types of units
  • Intra-Oral units
  • Standard dental tube
  • uses an intra-oral image receptor
  • has extra-oral x-ray tube
  • Panoramic (OPG)
  • Cephalometric ( Ceph)

7
Intra-Oral Dental X-Ray Equipment
8
Modern Dental X-Ray Unit
9
Panoramic X-Ray Equipment
10
Cephalometric X-Ray Equipment
11
X-Ray Tube
  • stationary Anode
  • avoid overheating
  • tube duty cycle
  • typical 130 intaroral
  • 110 OPG
  • 420 mAs/hr intraoral

12
Tube Head
13
Generator Circuit
14
Generators Pre-Heat
  • Medium frequency - stable waveform
  • Single phase (SP) - pulsed
  • Pre-Heat separate circuit for heating filament
  • Single Phase units without a pre-heat circuit
  • initial pulses of variable kV

15
Collimator
  • 1. Lead Collimator
  • with central hole
  • 2. Spacer Tube

16
Applicator Cones
Good Bad
Bad
17
Cephalometric Holder
18
Intra-Oral Dental X-Ray Equipment (technical data)
  • Exposure time from 60 ms to 2.5 s
  • Tube Min. 50 kV, 7mA
  • Focal spot size ?1 mm
  • Inherent filtration 2 mm Al equivalent
  • Focus-skin distance 20 cm
  • Irradiated field 28 cm2 with round section, 6
    cm diameter collimator

19
Panoramic X-Ray Equipment (technical data)
  • Focal spot 0.5 mm
  • kV 60 - 80 kV in 2 kV steps
  • mA 4 - 10 mA steps 4, 5, 6, 8, 10
  • Exposure time 12 s (standard projections) 0.16 -
    3.2 s (cephalometric projections)
  • Flat panoramic cassette 15x30 cm (Lanex Regular
    screens))

20
Image Receptors in Dental Radiology
Intraoral Radiology
  • Small films (2 x 3 or 3 x 4 cm) in light-tight
    envelopes (no screen)
  • Digital intraoral sensors - compared with
    category E film, the radiation dose is reduced by
    60.

Panoramic Radiology and Cephalometry
  • Film-screen combination
  • Digital sensors - compared with film-screen
    sensitivity class 200, the radiation dose is
    reduced by 50-70.

21
Dental Radiology Film Types
  • Sensitivity class D
  • Very good spatial resolution
  • Typical delivered dose about 0.5 mGy
  • Typical exposure times 0.3 - 0.7 s
  • Sensitivity class E
  • Good spatial resolution
  • Typical delivered dose about 0.25 mGy
  • Typical exposure times 0.1 - 0.3 s

22
Part 22 Optimization of Protectionin Dental
Radiology
IAEA Training Material on Radiation Protection in
Diagnostic and Interventional Radiology
  • Topic 2 Radiation Protection in Dental Radiology

23
Radiation Protection in Dental Radiology
  • Facts
  • Very frequent examination (about 25 of all the
    radiological examinations)
  • Delivered doses may differ of a factor 2 or 3.
    (entrance doses between 0.5 and 150 mGy)
  • Image Quality often very low
  • Organs at risk parathyroid, thyroid, larynx,
    parotid glands

24
Radiation Protection in Dental Radiology
Technical hints to reduce patient doses Quality
Control of Film Processing
  • Keep under control time and temperature of the
    developing process.
  • Do not use oxydized chemicals
  • Do not adjust development time by viewing the film

25
RP in Dental Radiology
Technical hints to reduce patient doses Privilege
the use of sensitive films
  • Sensitivity class D
  • Very good spatial resolution
  • Typical delivered dose about 0.5 mGy
  • Typical exposure times 0.3 - 0.7 s
  • Sensitivity class E
  • Good spatial resolution
  • Typical delivered dose about 0.25 mGy
  • Typical exposure times 0.1 - 0.3 s

26
Radiation Protection in Dental Radiology
Technical hints to reduce patient doses Lead
apron and collar
Useful when the path of primary beam intercepts
the protected organs (downward bite-twin
projection).
27
Radiation Protection in Dental Radiology
Panoramic examination
  • Image quality not as good as in intra-oral films
  • Important global information
  • Relatively low dose(one panoramic examination ?
    3?5 intra-oral films)

28
Part 22 Optimization of Protection in Dental
Radiology
IAEA Training Material on Radiation Protection in
Diagnostic and Interventional Radiology
  • Topic 3 Quality Control for Dental Equipment

29
Why Dental QC ?
  • Widespread use of dental units
  • Lack of QC history on most units
  • Dental practitioners working in the primary
    health care sector do not have the continuous
    medical physics support available in a
    hospital-based diagnostic imaging department

30
What Tests ?
  • Collimation
  • Dose Evaluation
  • Exposure Time
  • Half Value Layer
  • Kilovoltage (kVp)
  • Leakage Radiation

31
Quality Control for Dental Equipment
  • The recommended tests are consequently divided
    into
  • those simple tests which can be performed by
    dental practice staff
  • those more complex tests which can be carried
    out by medical physicists.

32
Quality Control for Dental Equipment
Tests which can be performed bydental practice
staff
33
Quality Control for Dental Equipment
Tests performed by medical physicists
34
Dental QC Methods
  • Unit
  • Intra-Oral Receptors
  • (I/O)
  • Cephalometric
  • (Ceph)
  • Panoramic
  • (OPG )
  • Test Method
  • as for Radiology QC
  • as for Radiology QC
  • where possible
  • immobilise unit
  • remove slit collimator

35
Test Equipment
  • kVp meter
  • measure kVp average
  • Programmable delay 100 ms
  • Range 50 to 120 kV
  • Aluminum filters
  • 4 x 1mm
  • Grade 1100
  • Dosimeter
  • small large volume chambers
  • 2mm wide detector for OPG
  • Timer
  • triggering at 75 peak kV

36
Collimation
  • Expose film or fluorescent screen
  • Measure x-ray field image

37
Collimator Light Intensity
  • Ceph units
  • Place external detector 1m from focus
  • Measure illuminance in lux ( lumen/m2)
  • Read each quadrant
  • Limit gt100 lux at 1m

38
Dead man Switch
  • timer at 50 cm from focus
  • set low kV, mA, long time
  • start exposure
  • release switch during exposure
  • Require exposure cut-out when switch is
    released. Check exposure time is less than set
    time

39
Kilovoltage
  • Accuracy
  • Set kVp meter to 100 msec delay
  • Observe kVp waveform at 70 kV if poss.
  • Limit measured kVp within 5 of set value
  • Reproducibility
  • Take 5 repeat exposures
  • Limit coefficient of variation 2

40
Dose Evaluation
  • Skin dose from I/O units
  • place cone 10 mm from dosimeter
  • set maxillary molar/ bitewing setting
  • Should be (65-70 kVp)
  • 2-3 mGy for molar view
  • lt 5 mGy for any view

41
Output Reproducibility (1)
  • Standard I/O units Ceph units
  • Dosimeter position
  • I/O units 10 mm from cone
  • Ceph units 75 cm from focus
  • or other recommended distance
  • Five repeat exposures
  • Limit coefficient of variation 5

42
Output Reproducibility (2)
  • Optional Method for OPG units
  • Align detector on film cassette slit
  • Measure dose rate
  • Take 5 repeat exposures
  • Limit coefficient of variation 5

43
Exposure Time Accuracy
  • Standard I/O units Ceph units
  • Set timer to trigger at 75 peak kV
  • Test times in the normal working range
  • Limit 10 error for I/O units
  • 5 error for all other units

44
Timer Reproducibility
  • Standard I/O units Ceph units
  • Place timer in beam
  • 5 repeat exposures
  • Limit coefficient of variation 5

45
Half Value Layer (HVL)
  • Standard I/O units
  • Position cone facing down
  • Place dosimeter at 40 cm from focus
  • Position Al filters near end of cone
  • Measure dose
  • measure with no added filters
  • with 2,3,4 mm Al added, then again with no
    filters
  • Plot on semi-log paper and find HVL

46
HVL Set-up
47
HVL Ceph OPG Units
  • Position Al filters on collimator
  • Cephalometric units
  • Position dosimeter at 75 cm from focus
  • OPG Units
  • Position dosimeter on film cassette slit
  • Measure dose rate, dose for fixed exposure time,
    or dose for full scan
  • NB Test kVp accuracy before measuring HVL

48
OPG Quality Control (kVp/HVL measurement)
49
HVL- Minimum Values
  • kVp HVL (mm Al)
  • Intraoral Ceph/OPG
  • 60 1.5 1.8
  • 70 1.5 2.1
  • 80 2.3 2.3
  • 90 2.5 2.5

50
Leakage Radiation
  • Cover collimator with 2 mm lead
  • Set large ionization chamber (180 cc) or solid
    state detector at known distance d cm from focus
  • Expose on maximum kVp for 0.5 - 1 sec (but be
    careful!)
  • Scale up the reading to 1 hour

51
Leakage Radiation
  • Choose the most likely points for leakage -
  • where collimator joins to housing, lateral to the
    tube, joints in housing etc.
  • Normalise measurement to 1 m from focus with
    inverse square law
  • Dose1m dosed x (distanced/100)2

52
Leakage Measurement
  • In normal radiography, leakage measurements are
    normalised to the maximum continuous rated tube
    current R
  • Leakage Dose (for P mAs at 100 cm) x (R x
    3600/P)
  • BUT...a dental unit has no continuous operation
    rating

53
Leakage Measurement
  • Instead we normalize the dose reading to tube
    duty cycle
  • Leakage Limit at 1 m from focus
  • 0.25 mGy/hour - Standard I/O units
  • 1.0 mGy/hour - Ceph OPG units

54
Calculation of Leakage
  • Measured leakage at 100 cm 2mGy/hr
  • Duty cycle 1 20 (i.e. must wait 20 times
    the exposure time between exposures to allow tube
    cooling)
  • Thus actual leakage 2 / 20 mGy/hr
  • 0.1 mGy/hr
  • i.e. acceptable

55
Radiation Protection in Dental Radiology
Quality Control of Film Processing
  • Keep under control time and temperature of the
    developing process.
  • Do not use oxidized chemicals
  • Regularly check processing with phantom

56
Dental Phantom
57
Dental Phantom
58
Where to Get More Information
  • Code of practice for protection of persons
    against ionising radiations arising from medical
    and dental use, DHSS, HMSO, London, 1964
  • Guidance notes for the protection of persons
    against ionising radiations arising from medical
    and dental use, DHSS, HMSO, London, 1988
  • Radiation protection and quality assurance in
    dental radiology. Radiation protection 81.
    European Commission.(1995) CG-89-95-971-EN-C

59
Summary
  • General and specific types of x-ray equipment
    (and image receptors) are reviewed (panoramic
    and cephalometric), with technical data about
    operating conditions
  • Although doses are generally low, the high
    frequency of examinations requires radiation
    protection (for the practitioner) in dental
    radiology
  • Some tests are detailed for Quality control of
    dental equipment.
Write a Comment
User Comments (0)
About PowerShow.com