FRCR: Physics Lectures Diagnostic Radiology - PowerPoint PPT Presentation

1 / 58
About This Presentation
Title:

FRCR: Physics Lectures Diagnostic Radiology

Description:

Diagnostic Radiology Lecture 4 Film-screen radiography Dr Tim Wood Clinical Scientist Overview Film-screen radiography Processing Intensifying screens and the film ... – PowerPoint PPT presentation

Number of Views:214
Avg rating:3.0/5.0
Slides: 59
Provided by: orgu1249
Category:

less

Transcript and Presenter's Notes

Title: FRCR: Physics Lectures Diagnostic Radiology


1
FRCR Physics LecturesDiagnostic Radiology
  • Lecture 4
  • Film-screen radiography
  • Dr Tim Wood
  • Clinical Scientist

2
Overview
  • Film-screen radiography
  • Processing
  • Intensifying screens and the film cassette
  • The characteristic curve and sensitivity
  • Image quality

3
The story so far
4
What is image quality
  • Image quality describes the overall appearance of
    the image and its fitness for purpose
  • Remember There is always a play-off between
    image quality and patient dose
  • We only need images that are of diagnostic
    quality (fit for purpose) not pretty pieces of
    art!
  • The main factors to consider are
  • Contrast
  • Spatial resolution
  • Noise

5
Image contrast
  • The final contrast in the image will depend on a
    number of factors, such as
  • Subject contrast an inherent property of the
    patient being imaged that will depend on the
    attenuation coefficients of the tissues (or
    contrast media), the thickness of structures, the
    nature of any overlapping tissues and the
    incident X-ray spectrum (kVp, filtration, etc
    discussed previously)
  • Detector properties film and digital detectors
    each have different implications for the contrast
    in the final image (these will be discussed in
    the next lectures)
  • Scattered radiation scatter can degrade image
    contrast if it reaches the detector as it conveys
    no information about where it came from. Scatter
    rejection techniques may be used to remove this.

6
Contrast and scatter
  • Primary radiation carries the information to be
    imaged
  • Scatter obscures this as it carries no
    information about where it came from
  • The amount of scatter (S) may be several times
    the amount of primary (P) in the same position
  • The ratio S/P depends on thickness of patient
  • For typical PA chest 41 (20 of photons carry
    useful information)
  • For typical lateral pelvis 91 (10 of photons)

7
Contrast and scatter
  • Scatter is generally quite uniform across the
    image
  • Reduces contrast that would otherwise be produced
    by the primary by a factor of (1S/P) up to 10
    times!

Intensity
SP
10
contrast reduced
S
8
Scatter 4 x Primary (PA Chest)
SP contrast (10-9)/10 x 100 10
Primary contrast (2-1)/2 x 100 50
P
2
Position
8
Scatter reduction
  • Scatter and patient dose may be reduced by
  • Reducing the field area with collimation (reduce
    volume of tissue generating scatter)
  • Compress the tissue to minimise overlying
    structures (reduce volume of tissue generating
    scatter)
  • Scatter may be reduced at the expense of
    increased patient dose by
  • Reducing kVp less forward scatter is produced
    and is much less penetrating (less scatter
    produced)
  • Use an anti-scatter grid to remove scatter from
    the X-ray beam
  • Use an air-gap to reduce the intensity of scatter
    reaching the detector

9
Spatial Resolution
  • Spatial resolution describes the ability to see
    fine detail within an image
  • Fine detail is clearer when the contrast is high
  • e.g. microcalcifications
  • May be expressed as the smallest visible detail,
    but most common descriptor is the highest
    frequency of lines that can be resolved in a
    high-contrast bar pattern

10
Limiting factors
  • Geometric Unsharpness due to finite focal spot
    size (discussed previously)
  • Movement Unsharpness imaging moving structures
    may result in blur e.g. heart, lungs, etc
  • Minimise by immobilisation e.g. compression in
    mammography
  • Breath hold techniques
  • Use shortest possible exposure times (may be at
    the expense of high mA)
  • Absorption Unsharpness gradual changes in
    absorption near tapered/rounded structures
  • Minimise by careful patient positioning
  • Detector resolution will be discussed in
    following lectures

11
Noise
  • Noise is a random, usually unwanted, variation in
    brightness or colour information in a visible
    image
  • Noise is one of the most important limiting
    factors to contrast and spatial resolution
  • The most significant source is quantum noise (or
    mottle) due to the low levels of radiation used
    to form an image
  • Other sources include film grain and electronic
    noise in the image receptor

12
Noise
  • Image noise is most apparent in image regions
    with low signal level, such as shadow regions or
    underexposed images
  • Noise gives a grainy, mottled, textured or snowy
    appearance to an image
  • Noise can mask fine detail in a radiograph
  • Noise reduces visibility of parts of a
    radiographic image
  • Noise is particularly an issue with image details
    that are already of low contrast

13
So what does this all really mean?
Something with lots of contrast is easier to see
than
Signal on detector
something with little contrast
Position on detector
14
So what does this all really mean?
But there is some loss of sharpness due to
various process, so the sharp edge
Signal on detector
actually becomes blurred, which may not matter
for large structures
Position on detector
15
So what does this all really mean?
Signal on detector
but if the structure is small
Position on detector
Signal on detector
it may start to disappear into the background
Position on detector
16
So what does this all really mean?
Up to now we havent considered the third
component, noise
which may not be a problem if there isnt too
much
17
So what does this all really mean?
but if there is lots of noise in the system,
the detail may be obscured
Signal on detector
whilst the low level of noise may be
problematic if contrast is low to start with
Position on detector
18
So what does this all really mean?
or resultion is poor
Signal on detector
Position on detector
19
(No Transcript)
20
(No Transcript)
21
(No Transcript)
22
Pause
23
Film-screen radiography
24
Film-Screen Imaging
  • Traditionally, all X-ray image capture has been
    through X-ray film

Emulsion
Protective layer
Adhesive layer
Film base
Emulsion
25
Film
  • Polyester film base gives mechanical strength to
    the film does not react to X rays
  • Emulsion consists of silver halide grains (AgBr)
  • The image is formed by the reaction of AgBr
    grains to X-ray photons
  • The sensitivity of the film depends on number of
    grains
  • Must be evenly distribution
  • Typically each crystal is about 1 µm in size
  • larger grains more sensitive (contrast),
  • smaller grain better resolution
  • Adhesive layer ensures emulsion stays firmly
    attached to base
  • Protective layer prevents mechanical damage

26
Film
  • Film is actually much more sensitive to visible
    light and UV than it is to X-rays
  • Hence, use a fluorescent screen to convert X-ray
    photons to light photons
  • Enables lower patient dose!
  • A latent image is formed upon exposure, which
    cannot be seen unless the film undergoes chemical
    processing
  • Mobile silver ions are attracted to electrons
    liberated by light photons, forming a speck of
    silver metal on the surface

27
Processing
  • The invisible latent image is made visible by
    processing
  • There are three stages to this process
  • Development
  • Fixing
  • Washing

28
Processing
  • First stage is development
  • Film is immersed in an alkaline solution of a
    reducing agent (electron donor)
  • Reduces positive silver ions to metallic grain of
    silver (black specks)
  • Unexposed crystals are unaffected by the
    developer bromide ions repel the electron donor
    molecules
  • However, given sufficient time, the developer
    will penetrate the unexposed crystals
  • The amount of background fog is dependent upon
    the time, strength and temperature of the
    developer

29
Processing
  • Second stage is fixing
  • If the film is exposed to light after the first
    stage, the whole film becomes black
  • To fix the film, unaffected grains are
    dissolved by an acid solution, leaving the X-ray
    image in the form of black silver specks
  • Final stage is washing
  • The film is washed in water and dried with hot
    air
  • Inadequate washing would result in a brown/yellow
    film over time (from excess acid) and smell

30
Processing
  • Automatic processors use a roller system to
    transfer the film through the different solutions
  • Regular Quality Assurance of the processor is
    vital for producing good quality radiographs
  • Image is then viewed by transmission of light
    from a light box with uniform brightness
  • Dark lots of X-rays
  • Light relatively few X-rays e.g. through bone

31
Production of a Radiograph
Process Time What Happens
1. Manufacture Crystals of a suitable size are made and suspended in gelatine
2. Exposure 0.01 10 sec Latent image created
3. Wetting 10 sec Wet film so that subsequent development is uniform
4. Development 3 10 min Convert latent image to silver
5. (Acid) wash 1 min Stop development and remove excess developer
6. Fixing and hardening 10 30 sec Dissolve out remaining AgBr and harden gelatine
7. Washing 30 sec Remove products of developer and fixer
8. Dry 30 sec Remove water
32
Logarithms
  • A logarithm is an exponent the exponent to
    which the base must be raised to produce a given
    number
  • 104 10x10x10x10 10,000
  • log1010000 4
  • i.e., 4 is the logarithm of 10000 with base 10
  • Seen in many applications
  • Richter earthquake scale
  • Sound level measurements (decibels dB)
  • Optical Densities blackness on film (OD)
  • Written as log10x or if no base specified in
    physics texts as log x it is interpreted as the
    same

33
Properties of logs
  • log101 0
  • log1010 1
  • log10xy log10x log10y
  • log10x/y log10x - log10y

34
Optical Density
  • Optical Density the amount of blackening in the
    film
  • Defined as the log of the ratio of the
    intensities of the incident and transmitted light
  • log is used as the eyes response is logarithmic

35
Optical Density
  • Optical density can be measured with a
    densitometer
  • From the definition, if 1 of light is
    transmitted, D 2.0
  • If 10 is transmitted, D 1.0
  • The density of an area of interest on a properly
    exposure film should be about 1.0
  • Lung field may be 2.0
  • Areas with Dgt3.0 too dark to see any detail on a
    standard light box

36
Contrast
  • Contrast is the difference in optical densities
  • Contrast OD1 OD2
  • High contrast - e.g. black and white
  • Low contrast e.g. grey and grey!

37
Intensifying screens
  • Film is relatively insensitive to X-rays directly
  • Only about 2 of the X-rays would interact with
    the emulsion
  • Requires unacceptably high doses to give a
    diagnostic image
  • An intensifying screen is a phosphor sheet the
    same size as the film, which converts the X-rays
    to a pattern of light photons
  • The intensity of the light is proportional to the
    intensity of X-rays
  • The pattern of light is then captured by the film
  • One exception is intraoral dental radiography,
    where screens are not practical

38
Intensifying screens
  • Modern intensifying screens use rare earth
    materials, which emit light that is matched to
    the sensitivity of the film being used
  • Spectral match between the emission of the screen
    and the absorption in the film e.g. blue or green
  • K-edges clinically relevant (39-61 keV)
  • Rare earth screens used as they very efficient at
    converted absorbed X-ray energy into light
  • Results in a faster (more sensitive) system
  • The sensitive emulsion of the film must be in
    close contact with the screen

39
Intensifying screens
  • General radiography film usually double coated
    with emulsion on each side of the base
  • The front screen absorbs 1/3 of X-rays and 1/2
    light travels forward and is absorbed by front
    layer of emulsion
  • Rear screen absorbs 1/2 of X-rays transmitted
    through the front and exposes the rear emulsion
  • 2/3 of total X-ray fluence absorbed in screens
  • Mammography only uses a single screen to maximise
    spatial resolution (more on this later)
  • Screen materials chosen to have no
    phosphorescence (delayed fluorescence) to avoid
    ghost images

40
The film-cassette
  • Flat, light tight box with pressure pads to
    ensure film in good contact with the screen(s)
    mounted on the front (and back)
  • The tube side of the cassette is low atomic
    number material (Z6) to minimise attenuation
  • Rear of cassette often lead backed to minimise
    back scatter (not in mammo)

41
The characteristic curve
Optical density
  • Plotting OD against log exposure gives the
    Characteristic Curve of the X-ray film
  • Different types of film subtle differences but
    all basically the same

Saturation
Linear region, gradient gamma
Solarisation
Fog
Log exposure
42
The characteristic curve
  • Depends on type of film, processing and storage
  • Fog Background blackening due to manufacture and
    storage (undesirable)
  • Generally in the range 0.15-0.2
  • Linear portion useful part of the curve in which
    optical density (blackening) is proportional to
    the log of X-ray exposure
  • The gradient of the linear portion determines
    contrast in an image and patient exposures must
    lie within this region
  • Need to match this to the clinical task!
  • Hence, film suffers from a limited and fixed
    dynamic range

43
The characteristic curve
  • Gradient of linear region
  • Gamma, ? OD2 OD1 log E2-log E1
  • Gamma depends on
  • Emulsion
  • Size and distribution of grains
  • Film developing
  • Gamma Contrast
  • Latitude useful range of exposures

Optical density
Linear region
Latitude
Log exposure
44
The characteristic curve
  • Gamma and latitude are inversely related
  • High gamma low latitude
  • Wide latitude (low gamma) for chests
  • High gamma (low latitude) for mammography
  • At doses above the shoulder region, the curve
    flattens off at D3.5
  • Saturation, whereby all silver bromide crystals
    have been converted to silver
  • At extremely high exposures density will begin to
    fall again due to solarization
  • Not relevant to radiography

45
Film Speed
  • Definition 1 / ExposureBF1
  • Reciprocal of Exposure to cause an OD of 1 above
    base plus fog
  • Speed of film sensitivity amount of radiation
    required to produce a radiograph of standard
    density
  • Speed shifts H-D curve left and right
  • Fast film requires less radiation (lower patient
    dose)
  • Speed is generally used as a relative term
    defined at a certain OD one film may be faster
    than another at a certain point on the curve

46
Factors affecting speed
  • Size of grains larger means faster
  • This is the main factor and conflicts with the
    need for small crystals to give good image
    sharpness.
  • Fast films are grainier but reduce patient dose
  • Thickness of emulsion
  • Double layers of emulsion give faster films
  • Radiosensitisers added
  • (X-ray energy)

47
Effect of developing conditions
  • Increasing developer temperature, concentration
    or time increases speed at the expense of fog
  • Developer conditions should be optimised for
    maximum gamma, and minimum fog
  • Automatic processor has temperature controls and
    time maintained by roller speed
  • Concentration is controlled by automatic
    replenishment of the chemicals

48
Film-screen sensitivity
  • Intensification factor
  • Each X-ray photon generates 1000 light photons
  • Just under half of these will reach the film
  • 100 light photons to create a latent image
  • Hence, more efficient process
  • Intensification factor is the ratio of air KERMA
    to produce D 1 for film alone, to that with a
    screen
  • Intensification factor typically 30-100
  • Speed class
  • Most common descriptor of sensitivity
  • Speed 1000/K, where K is air KERMA (in µGy) to
    achieve D 1
  • Typically 400 speed (K 2.5 µGy)

49
Image quality
  • Contrast
  • Contrast in film-screen radiography is due to
    both subject contrast, scatter and gamma
  • Remember, high gamma high contrast low
    latitude (and vice-versa)
  • Contrast is fixed for any given film and
    processing conditions
  • Image detail may be lost if contrast is too high
    as it may be lost in the saturated or fog regions
  • Hence, vital to match gamma to the clinical task
  • Ambient light conditions and viewing box
    uniformity may also impact on the subjective
    contrast presented to the Radiologist
  • Use a darkened room, mask off unused areas of
    lightbox, etc

50
Image quality
  • Screen-unsharpness
  • The film-screen system has inherent unsharpness
    additional to geometric, motion and absorption
  • Only partly due to finite size of the emulsion
    crystals
  • Most significant effect is due to spread of light
    from the point of X-ray absorption in the
    phosphor, to detection by the film
  • Depends on the point in the phosphor where the
    interaction occurs
  • Thicker phosphor layers more sensitive (absorb
    more X-rays), but result in more blurring allow
    lower patient doses

51
Screen-unsharpness
Object
Phosphor
Film
52
Screen unsharpness
  • Speed class should be chosen carefully to match
    the application
  • e.g. 400-speed (thick phosphor) for thick
    sections of the body (abdo/pelvis),
  • e.g. 100-150-speed (thin phosphor) for
    extremeties (require detail)
  • Also may have reflective layer on top of phosphor
    to increase sensitivity (reflect light photons
    back to the film) at the expense of resolution
  • Colour dyes to absorb light photons at wider
    angles (longer path lengths) at the expense of
    sensitivity

53
Screen unsharpness
  • Crossover light photons from the front screen
    may be absorbed by the rear emulsion (and
    vice-versa)
  • Crossover is a significant contributor to overall
    unsharpness
  • Reason for only using one screen in mammography
    where resolution is critical
  • Minimise screen-unsharpness by ensuring good
    contact between the screen and film
  • Poor contact may result from damage to the film
    cassette

54
Film-screen in clinical practice
  • Kilovoltage Increased kV gives
  • Increased penetration lower patient dose
  • Increased exposure latitude larger range of
    tissues displayed, BUT lower radiographic
    contrast
  • Reduction in mAs shorter exposures less
    motion blur
  • mAs
  • Correct mAs must be chosen to ensure the correct
    level of blackening on the film avoid under or
    overexposing the film
  • Too much saturation, too little thin image
  • Produce standard protocols that can be adapted
    for patient size

55
Exposure Control
  • For an acceptable image, require a dose at the
    image receptor of about 3 µGy for film-screen
    radiography
  • This is the exit dose from the patient after
    attenuation
  • Entrance surface dose (ESD) is much higher than
    this
  • 10 times greater than exit dose for PA chest
  • 100 times greater for skull
  • 1000 times greater for AP pelvis
  • 5000 times greater for lateral lumbar spine

56
Automatic Exposure Control (AEC)
  • Limited latitude of film makes it difficult to
    choose correct mAs skill and experience of
    radiographer
  • Alternative is to use an AEC to terminate the
    exposure when enough dose has been delivered to
    the film
  • AEC is a thin radiation detector (ionisation
    chamber) behind the grid, but in front of the
    film (though in mammo it is behind to avoid
    imaging the chamber on the film)
  • Usually three chambers that can be operated
    together or individually

57
Automatic Exposure Control (AEC)
  • When a predetermined level of radiation is
    detected, the exposure terminates
  • Choice of chambers determined by clinical task
  • e.g. left and right for lungs in PA chest, but
    central if looking at spine
  • Also has a density control that can increase or
    decrease exposure where necessary
  • AEC limited to exposures in the Bucky system

58
Modern Day
  • Film is dying out
  • Across most (but not all) of the country film is
    no longer used for General X-ray imaging
  • Only mammography (breast imaging), where very
    high resolution specialist film is used
  • This Trust no longer uses film for mammography,
    and is on the verge of being fully digital
Write a Comment
User Comments (0)
About PowerShow.com