1
RADIATION PROTECTION INDIAGNOSTIC
ANDINTERVENTIONAL RADIOLOGY
IAEA Training Material on Radiation Protection in
Diagnostic and Interventional Radiology
  • Part 16.5 Optimization of protection in
    fluoroscopy
  • Practical exercise

2
Overview / Contents
  • To become familiar with quality control test in
    fluoroscopy
  • To measure the imaging geometry and sizing
  • Interpretation of results

3
Part 16.5 Optimization of protection in
fluoroscopy
IAEA Training Material on Radiation Protection in
Diagnostic and Interventional Radiology
  • Image Distortion

4
Leeds Fluoroscopy Phantom Image distortion
  • Pincushion distortion results because the input
    phosphor is curved but the output phosphor is
    flat
  • This results in a warped image, which bends
    straight lines inwards
  • Pincushion distortion is reduced when
    magnification modes are used

5
Image distortion
Pin-cushion
S-distortion
6
Leeds Fluoroscopy Phantom Imaging geometry and
sizing
  • Incorporates
  • a wire matrix of 20 mm square spacing with 10 mm
    diametrical interpolations

7
Leeds Fluoroscopy Phantom Imaging geometry and
sizing
  • Interpretation of results
  • Measured filed size gt 0.85 nominal diameter of
    the I.I.
  • Integral distortion (mean diagonal of large
    square - 1)
    n x mean diagonal of central square
  • less than 10 .