Title: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
1 RADIATION PROTECTION INDIAGNOSTIC
ANDINTERVENTIONAL RADIOLOGY
IAEA Training Material on Radiation Protection in
Diagnostic and Interventional Radiology
- L10 Patient dose assessment
2Introduction
- A review is made of
- The different parameters influencing the patient
exposure - The problems related to instrument calibration
- The existing dosimetric methods applicable to
diagnostic radiology
3Topics
- Parameters influencing patient exposure
- Dosimetry methods
- Instrument calibration
- Dose measurements
4Overview
- To become familiar with the patient dose
assessment and dosimetry instrument
characteristics.
5Part 10 Patient dose assessment
IAEA Training Material on Radiation Protection in
Diagnostic and Interventional Radiology
- Topic 1 Parameters influencing the patient
exposure
6Essential parameters influencing patient exposure
Tube voltage Tube current Effective filtration
Kerma rate mGy/min
Kerma ?Gy
Exposure time
min
Area exposure product ?Gy m2
Field size
m2
7Factors in conventional radiography beam,
collimation
- Beam energy
- Depending on peak kV and filtration
- Regulations require minimum total filtration to
absorb lower energy photons - Added filtration reduces dose
- Goal should be use of highest kV resulting in
acceptable image contrast - Collimation
- Area exposed should be limited to area of
CLINICAL interest to lower dose - Additional benefit is less scatter, netter
contrast
8Factors in conventional radiography grid,patient
size
- Grids
- Reduce the amount of scatter reaching image
receptor - But at the cost of increased patient dose
- Typically 2-5 times Bucky factor or grid ratio
- Patient size
- Thickness, volume irradiatedand dose increases
with patient size - Except for breast (compression) no control
- Technique charts with suggested exposure factor
for various examinations and patient thickness
helpful to avoid retakes
9Factors affecting dose in fluoroscopy
- Beam energy and filtration
- Collimation
- Source-to-skin distance
- Inverse square law maintain max distance from
patient - Patient-to-image intensifier
- Minimizing patient-to- I I will lower dose
- But slightly decrease image quality by increased
scatter - Image magnification
- Geometric and electronic magnification increase
dose - Grid
- If small sized patient (les scatter) perhaps
without grid - Beam-on time!
10Factors affecting dose in CT
- Beam energy and filtration
- 120-140 kV shaped filters
- Collimation or section thickness
- Post-patient collimator will reduce slice
thickness imaged but not the irradiated thickness - Number and spacing of adjacent sections
- Image quality and noise
- Like all modalities dose increasegtnoise
decreases
11Factors affecting dose in spiral CT
- Factors for conventional CT also valid
- Scan pitch
- Ratio of couch travel in 1 rotation dived by
slice thickness - If pitch 1, doses are comparable to
conventional CT - Dose proportional to 1/pitch
12Part 10 Patient dose assessment
IAEA Training Material on Radiation Protection in
Diagnostic and Interventional Radiology
- Topic 2 Patient dosimetry methods
13How to measure doses
Calorimetry
Chemical (Fricke dosimeter)
They need to know a characteristic parameter
Absolute methods
Ionometry (ionization chamber)
Photography
Scintillation
Relative methods
TL
Ionometry
14Patient dosimetry
- Radiography entrance surface dose ESD
- By TLD
- Output factor
- Fluoroscopy Dose Area Product (DAP)
- CT
- Computed Tomography Dose Index (CTDI)
15From ESD to organ and effective dose
- Except for invasive methods, no organ doses can
be measured - The only way in radiography measure the Entrance
Surface Dose (ESD) - Use mathematical models to estimate internal
dose. - The physical methods similar to those used in
radiotherapy can be used but not accurate - Mathematical models based on Monte Carlo
simulations the history of thousands of photons
is calculated - Dose to the organ tabulated as a fraction of the
entrance dose for different projections - Since filtration, field size and orientation play
a role long lists of tables (See NRPB R262 and
NRPB SR262)
16From DAP to organ and effective dose
- In fluoroscopy moving field, measurement of
Dose-Area Product (DAP) - In similar way organ doses calculated by Monte
Carlo modelling - Based on mathematical model
- Conversion coefficients were estimated as organ
doses per unit dose-area product - Again numerous factors are to be taken into
account as projection, filtration, - Once organ doses are obtained, effective dose is
calculated following ICRP60
17Part 10 Patient dose assessment
IAEA Training Material on Radiation Protection in
Diagnostic and Interventional Radiology
- Topic 3 Instrument calibration
18Calibration of an instrument
- Establish Calibration Reference Conditions (CRC)
type and energy of radiation, SDD, rate, ... - Compare response of your instrument with that of
another instrument (absolute or calibrated) - Get the calibration factor
f the reference instrument
Response o
appropriate unit
F
Response of the instrument to be calibrated
19Range of use
Hypothesis the instrument reading is a known
monotonic function of the measured quantity
(usually linear within a specified range)
Instrument Reading
1/F tg ?
Response at calibration
?
MeasuredQuantity
Calibration Value
20Use of a calibrated instrument
- Under the same conditions as the CRC
- Within the range of use
Q (dosimetric quantity) F x R (reading of the
instrument)
21Correction factors for use other than under the
CRC
A. Energy correction factor
Correction
Factor
1.06
1.04
1.02
1
0.98
0.96
0.94
0.92
1
2
3
4
HVL(mm Al)
22Correction factors for use other than under the
CRC
B. Directional correction factor
23Correction factors for use other than under the
CRC
C. Air density correction factor (for ionization
chambers)
t
p
)
273
(
0
K
D
t
p
)
273
(
0
p
t
,
calibration values
0
0
24Accuracy and precision of a calibrated instrument
(1)
A
C
Readings
B
True value
Curve A Instrument both accurate and
precise Curve B Instrument accurate but not
precise Curve C Instrument precise but not
accurate
25 Accuracy and precision of a calibrated instrument
(2)
26Requirements on Diagnostic dosimeters
Traceability
Well defined reference X Ray spectra not
available
Accuracy
At least 10 - 30
27Limits of error in the response of diagnostic
dosimeters
Parameter
Range of values
Reference condition
Deviation ()
Radiation quality
According to manufacturer
70 kV
5-8
Dose rate
According to manufacturer
--
4
Direction of radiation incidence
5
Preference direction
3
Atmospheric pressure
80-106 hPa
101.3 hPa
3
Ambient temperature
15-30
20 C
3
28Part 10 Patient dose assessment
IAEA Training Material on Radiation Protection in
Diagnostic and Interventional Radiology
- Topic 4 Dose measurements how to measure dose
indicators ESD, DAP,CTDI
29What we want to measure
- The radiation output of X Ray tubes
- The dose-area product
- The computed-tomography dose index (CTDI)
- Entrance surface dose
30Measurements of Radiation Output
X Ray tube
Filter
SDD
Ion. chamber
Lead slab
Table top
Phantom (PEP)
31Measurements of Radiation Output
- Operating conditions
- Consistency check
- The output as a function of kVp
- The output as a function of mA
- The output as a function of exposure time
32Dose Area Product (DAP)
Transmission ionization chamber
33Dose Area Product (DAP)
0.5 m
1 m
2 m
Air Kerma Area Areaexposure product
2.5103 ?Gy 4010-3 m2 100 ?Gy m2
40103 ?Gy 2.510-3 m2 100 ?Gy m2
10103 ?Gy 1010-3 m2 100 ?Gy m2
34Calibration of a Dose Area Product (DAP)
35Computed Tomography Dose Index (CTDI)
TLD dose (mGy)
50
Nominal slice width
3 mm
40
30
CTDI 41.4
20
10
0
1
2
3
4
5
6
7
8
9
10
11
12
36Computed Tomography Dose Index (CTDI)
CTDI
Dose
Dose profile
Nominal slice width
37TLD arrangement for CTDI measurements
X Ray beam
Gantry
Support jig
X Ray beam
Capsule
Axis of rotation
axis of rotation
Capsule
Couch
Gantry
LiF -TLD
38Measurement of entrance surface dose
TLD
39Summary
- In this lesson we learned the factors influencing
patient dose, and how to have access to an
estimation of the detriment through measurement
of entrance dose, dose area product or specific
CT dosimetry methods.
40Where to Get More Information
- Equipment for diagnostic radiology, E. Forster,
MTP Press, 1993 - The Essential Physics of Medical Imaging,
Williams and Wilkins. Baltimore1994 - Leitz, W., Axiesson, B., Szendro, G. Computed
tomography dose assessment - a practical
approach. Nuclear Technology 37 1-4 (1993) 377-80