MEEPS - PowerPoint PPT Presentation

About This Presentation
Title:

MEEPS

Description:

4 Channel EEG Recorder. Low Noise Amplifier. Digitized output (NRZ ... Future Directions. Complete this and send the chip!!! Integration with Telemetry System ... – PowerPoint PPT presentation

Number of Views:12
Avg rating:3.0/5.0
Slides: 16
Provided by: biolog
Category:

less

Transcript and Presenter's Notes

Title: MEEPS


1
MEEPS
  • Mohsen Mollazadeh
  • Mixed Signal VLSI Systems
  • Prof. Cauwenberghs

2
EEG Recording
  • Why?
  • Slow waves (application BCI)
  • Brocas Area(audiotory Cortex)
  • Eplilepsy

3
System Design
  • 4 Channel EEG Recorder
  • Low Noise Amplifier
  • Digitized output (NRZ and Miller)
  • Fully Differential Pathway

4
OpAmp
5
OpAmp
6
High Resistance Generation
7
OP Amp Results
8
OpAmp Results
Corner Frequency 100 Hz Thermal Noise level
25 nV/sqrt(Hz)
9
OpAmop Results
  • Gain 40 db
  • Power Consumption 60 uW
  • Input Referred Noise 4.5 uVrms
  • Bandwidth .8 10 KHz
  • Resistance Achieved gt5Tohms

10
OpAmp Layout
11
Delta Sigma AtoD
12
Delta Sigma results
13
EEG Channel
14
Future Directions
  • Complete this and send the chip!!!
  • Integration with Telemetry System
  • FIEEP (A Fully Implantable elctro-Electrochemical
    Probe)

15
Acknowledgements
  • Thanks to
  • Prof. Cauwenberghs
  • Kartik, Chris, Yunbin, Abdullah and Milutin
Write a Comment
User Comments (0)