Title: EUROMET Project 692
1EUROMET Project 692
Volume calibration of a 100 ml Gay-Lussac
Pycnometer
Portuguese Institute for uality
2 Run in parallel and in close co-operation
EUROMET Project 691
Calibration Inter-comparison of a 5-litre volume
glass standard
SP - Swedish Testing and Research Institute
3EUROMET Project 691 and 692
Projects purpose
- Compare experimental methods
- Compare uncertainties calculation in the
calibration of the pycnometer and flask
- Will be representative for all type of laboratory
glassware
4EUROMET Project 691 and 692
Projects Details
The instruments
100 ml Pycnometer
5 litre Flask
5EUROMET Project 691 and 692
Projects Details
The instruments
- The volume can not be changed without breaking it
6EUROMET Project 691 and 692
The suggest method
- Volume calculation formula for a temperature of
20ºC
7EUROMET Project 691 and 692
Results and uncertainty
- A spreadsheet will be supplied
- The uncertainty budget according to GUM
- Discussion of results and uncertainty components
8EUROMET Project 691 and 692
Preliminary results for the Pycnometer
/-0,01
Mean
9EUROMET Project 691 and 692
Time Table
10EUROMET Project 691 and 692
Transportation
- The transportation of the instruments are done
simultaneously
- The carnet is paid by the pilot lab
- Each laboratory are requested to take care of the
transportation to the next one
11EUROMET Project 691 and 692
Problems
12EUROMET Project 691 and 692
Resolution
- A new pycnometer was sent as soon as the damage
occurred
- Find and Correct the difference found due to the
damage, if needed, at the end of the measurements
13EUROMET Project 691 and 692
14EUROMET Project 691 and 692
15The coming key comparison
- At least 2 volume standards.
- In discussion a 100 ml pycnometer.
- Tested a 20-l vessel for delivered volume
- in the state of cloning organised by CENAM in
Mexico.
16Key Comparison Project
17Key Comparison Project
18Key Comparison Project
20-l volume standard
? 0,0057
19Key Comparison Project