Title: Design Case Studies
1Design Case Studies
2Multidetector Brain Scanning System Development
- Problem given point source response, how can one
improve net spatial sensitivity of a device? - Pre-CAT scanning and other current technologies
3Detector Point source response
Lead collimator cross section detector are on
the left on this diagram, iso-response profiles
are to the right
4Problem Statement
- "Given one or more focusing collimators (and
related electronics), develop a brain scanning
system that shows increased sensitivity compared
to a rectilinear scan using a single detector,
and which gives a cross sectional image of the
brain more suited for surgical planning than
rectilinear scans."
5Solution 1 Multicrystal Tomographic Scanner
Schematic of early tomographic scanner, U.S.
Patent Number 3,591,806, eight detectors arranged
in a circle with a common intercept point.
6Solution 2 Mutually Orthogonal Multicrystal
Tomographic Scanner
Mutually orthogonal schematic arrangement one
quadrant of a 12-detector system with 4 sets of
three mutually orthogonal detectors
7TRIZ principle 5
- "combining in space homogenous objects destined
for contiguous operations " - Both systems worked, one was patented, both
formed the basis for other technologies (see
reference patents.)
8Testing of Anesthetists Hypotheses
- A challenging protocol may be developed using a
simulator that tests anesthetists' skills at
maintaining patient homeostasis within limits,
and - An analytical technique may be demonstrated that
will suggest that "skill level" may be inferred
from the data collected from the simulator.
9(No Transcript)
10Simulation Method
- Inform examinee who the patient is (Stan, normal
young male) - Operation type low anterior bowel resection
- SOP please
- Inform re stage of surgery
- Start!
11And we are off
12The protocol (Stable Anesthesia)
- Induction ? Intubation (epi) ? Maintenance
- ? Incision (epi) ? Fluid loss ( 3L) ?
- ? Maintenance ? Ischemia Desaturation (? lung
changes) - ? Maintenance ? Emergence
- ? Extubation (? adequacy)
13Trained Physician Results
14Novice Results
15Conclusion
- The human patient simulator may be used as a
testing device to do inter-individual comparison
of anesthetist response to simulated stresses
during anesthetic procedures. - A simple measure of competency of intervention
may be derived by a time out of range measure
-
- King, P.H., Pierce, D., Higgins, M., Beattie, C.,
Waitman, L. C. A Proposed Method for the
Measurement of Anesthetist Care Variability,
Journal of Clinical Monitoring and Computing, 16
(2) 121-125, 2000.
16Apnea Detection System
- A means of monitoring patients for potential
respiratory depression needs to be found, such
that the patient in a step-down unit may be
better monitored than just with simple periodic
visits by hospital personnel.
17A partial solution included the following items
- CO2 monitoring -too low a CO2 level, or no
waveform (apnea) as detected by a commercial unit
Sampling was achieved with a single capillary
tube placed near the patient's nostrils. The
units used had an alarm level that could be set
manually the alarm signal could be accessed for
use in other devices. - An autodialing motion detection alarm system was
modified and connected to the capnometer. The
system would then dial the charge a nurses'
beeper when the system alarmed. The beeper was
unique to the patient being monitored. - The basic system had to be modified with an
on/off switch that could be activated when the
patient was talking (seen as a high respiration
rate) or eating. - A special cannula had to be obtained to sample
air from both nostrils and from near the mouth.
This was necessary for patients who mouth
breathed due to snoring and/or stuffy noses.
18Cancer Clinic Charting
- While charting the pathway of patients through a
screening clinic (breast cancer screening) a
means had to be found to display not only the
process but also the patient perception of the
process.
19Cancer clinic diagnosis/discharge/follow-up
section
20EKG Analysis Techniques
- 2 -3 hours of ekg sampled at 1000 samples/sec
analyze report on 10-20 averaged ekgs at
multiple points in time (mouse, implanted
transmitter) - Desire r-r interval analysis also
- Desire arrhythmia analysis
21Alternatives considered
- Access Excel Visual Basic
- IDL and Visual C
- PV-Wave C
- MATLAB Visual C Excel Visual Basic
- Others approaches, ease of use?
22Current solution
- Initial developmental work done in Excel with
visual basic (debug test) - Migration to MATLAB as soon as feasible
- Currently in use
23EKG Analysis Module
- Given the mandate to do so, design an
introductory freshman module (1 credit hour) to
allow students potentially interested in your
department to sample what they might experience
in their remaining three years should they elect
your department's major.
24Electrocardiogram Capture and Analysis
- data analysis techniques in electrocardiography,
- medical and engineering nomenclature,
- engineering and engineering applied to medicine,
- technologies involved in cardiology and
electrocardiography, - and societal ramifications of heart-related
research
25Basics
- EKG lectures, basic ekg recognition
- Student ekg sampled, analyzed with Excel (for R-
wave analysis HR determination) - Same ekg analyzed with MATLAB
- http//vubme.vuse.vanderbilt.edu/King/ekg2000.htm
for layout, etc.
26Choosing the Correct Plastic Material
- The purpose of the test is to provide rough
insight into whether plastic materials subjected
to stress are compatible with common cleaning
substances. The materials being considered are
Valox, Zytel, Cycoloy, and Thermocomp.
27Test dogbones of material
- Kleen-Aseptic
- Sporicidin
- Cidex
- Aldiced
- Virex 256
- Wescodyn
- Acetone
- Bleach
- Isopropyl alcohol.
28Testing -
- Tensile tests
- Flex tests
- Time/exposure tests
- Inspect visually
- Test to failure
29Similar tests procedures
- Materials tests for autoclaving
- Choosing the right cleaning material