Title: Topic 4' Problems in Radiation Detection and Measurement
1Topic 4. Problems in Radiation Detection and
Measurement
- Detection Efficiency
- Problems in the detection and measurement of beta
particles - Deadtime
2Detection Efficiency
- The Definition
- Geometric Efficiency
- Intrinsic Efficiency
- Energy Selective Counting
- Absorption and Scatter
3The Definition
- Emission Rate (assume ? rays per disintegration)
?(?rays/sec)3.7104(dis/µCi.sec) A(µCi) ?
(?rays/dis) - Detection Efficiency (with R the recorded
counting rate) DR/ ?
4Detection Efficiency
- Detection efficiency depends on a number of
factors absorption and scatter (F), geometric
efficiency (g), intrinsic efficiency (e) and
energy-selective counting (f). - It can be expressed as DFg e f
5Geometric Efficiency
- Inverse square law I?/4pr2 where I is the
intensity of radiation per unit area and ? is the
source emission rate. - For point source, geometric efficiency is
gpa/4pr2 where a is the detector surface area
and r is the distance from the point source.
6(No Transcript)
7Geometric Efficiency (small distance r)
- Assume the detector is close to the source (small
r), the area of the detector is SOr2 where
O2p(1-cos?). - The geometric efficiency is then
gpS/4pr2Or2/4pr2(1-cos?)/2 - When source in contact with the detector ?90o,
gp1/2 and when source is immersed in the
detector material, ?180o, gp1.
8(No Transcript)
9(No Transcript)
10Intrinsic Efficiency
- Intrinsic efficiency is defined as e(no. of
radiation interacting with detector)/(no. of
radiations striking the detector) - For ?-ray detector, eIo-Ioexp-µl(E)x/Io
1-exp-µl(E)x where Io is the incoming ?-ray
intensity and Ioexp-µl(E)x is the ?-ray
intensity that pass through the detector without
interaction.
11Effect of ? ray energy and detector thickness on
Intrinsic Efficiency
- For NaI(Tl) detector, the intrinsic efficiency
increases with the increase of thickness x and
decreases with the increase of the photon energy
(x5cm, for most nuclear medicine energy, e1) - For semiconductor detector, the intrinsic
efficiency is also energy dependent (comparison
with NaI(Tl) is complicated due to the coupled
atomic number and the crystal density)
12(No Transcript)
13Intrinsic Efficiency (gas filled detectors)
- For gas filled detectors, the intrinsic
efficiency is very small for photons (?, X, e
lt0.01) but good for particle radiations (a,ß,
e1). - Detection of ? rays by gas filled detectors is
mostly via electrons knocked out from the wall by
the ? rays.
14Energy Selective Counting
- Not all output signals are counted in energy
selective counting - Photo-fraction fp is defined as the ratio of the
counts in the photo-peak over the signal counts
produced by the detector. - Photo-fraction depends on ? energy, detector
material and detector size. - Full spectrum counting provides the maximum
possible counting rate (used for single
radionuclide with small scattering).
15(No Transcript)
16Non-uniform Detection Efficiency
17Detector Profiles
18Minimum Source-Detector Distances
19Coincident Detection(Multiple Emission)
- R1?1xD1xA
- R2 ?2xD2xA
- R12 ?1xD1x ?2xD2xA
- RR1R2-R12
20Absorption and Scatter (1)
- Absorption and scatter occur when radionuclide is
embedded within some media (tissue for instance). - Counting rate is decreased by absorption but may
be decreased or increased by scattering depending
on the numbers of the scattering from and into
the detector. - Absorption and scatter are dependent on the ?
energy, depth within the medium and if the energy
selective counting is used.
21(No Transcript)
22Absorption and Scatter (2)
- Absorption and scattering decreases with increase
of the ? energy - Absorption may predominate if the radionuclide is
embedded in greater depth. - Scattering is increased initially and reaching a
maximum and then decreased with the increase of
energy.
23Determination of Detection Efficiency
- Use known calibration source
- Calculation by definition equation DR/?
- Take into account the emission frequency if the
calibration source is differing from the
radionuclide concerned (see next table).
24(No Transcript)
25(No Transcript)
26Problems in the detection and measurement of beta
particles
- Liquid Scintillation Detectors are generally used
for short ranged ß particle detection. - A survey meter may be used to detect surface
contamination by ß particles provided it has an
entrance window sufficiently thin to permit the ß
particles to enter the sensitive volume of the
detector
27(No Transcript)
28Gas Filled Detectors in ß Assay
- Sample self absorption (before reaching the
detector) and backscattering (from sample and
sample holder) are the problems. - Self absorption depends on sample thickness and ß
energy (strong absorption for low energy and
thicker samples). - Backscattering increases the sample counting rate
(20-30).
29(No Transcript)
30(No Transcript)
31NaI(Tl) For High Energy ß
- NaI(Tl) detectors can be used to detect higher
energy ß emitting radionuclides (32P, Emax1.7
Mev, not 14C, Emax0.156MeV) by counting the
Bremsstrahlung radiation. - Greater activities are required (1000 times more
than ? emitters) because of the low production
efficiency of Bremsstrahlung radiation.
32Deadtime
- Cause of Deadtime
- Mathematical Models
- Window Fraction Effects
- Deadtime Correction Methods
33Causes of Deadtime (1)
- Deadtime is the time required for the counting
system to process an individual event (it is also
called pulse resolving time t). - Several components (detector, pulse amplifiers,
pulse high analyzer, scaler and computer
interface) could contribute to the deadtime and
the longest component determines the system
deadtime.
34Causes of Deadtime (2)
- The deadtime of NaI(Tl) or semiconductor
detectors are mainly caused by pulse amplifiers
(pulse pileup and baseline shift, typical 0.5-5
µsec) - GM tubes are mainly caused by detectors pulse
overlap (typical 50-200 µsec). - Deadtime results in counting losses (overlap loss
or window shift loss).
35Mathematical Models
- Counting systems can be classified as paralyzable
and non-paralyzable type and most of nuclear
medicine systems are paralyzable type. - A non-paralyzable system is one for which, if an
event occurs during the deadtime t of a preceding
event, then the second event is simply ignored. - A paralyzable system is one for which each event
introduces a deadtime t whether or not that event
actually was counted.
36(No Transcript)
37Nonparalyzable Systems(1)
- The relationship between observed count rate Ro
and true count rate Rt for the nonparalyzable
system is given by
38Nonparalyzable Systems(2)
- The observed counting rate Ro has a maximum
value. For nonparalyzable system, it is given by
39(No Transcript)
40Paralyzable Systems(1)
- The relationship between observed count rate and
true count rate for the paralyzable system is
given by
41Paralyzable Systems(2)
- The observed counting rate Ro also has a maximum
value. For paralyzable system, it is given by
42Deadtime Losses
- Deadtime losses is defined as (Rt-Ro) and the
percentage losses (PL) is given by
43Window Fraction Effects
- Deadtime losses depend on total counting
spectrum. - Deadtime per event in the selected window depends
on the fraction of the window ta t/wf - Window fraction effect must be considered in
comparing deadtime of different systems using
energy selective counting
44Deadtime Correction Methods
- True counting rate can be calculated if we know
the deadtime and the observed counting rate
(either use the equation for nonparalyzed system
directly or use graphic or approximation for
paralyzed system). - Deadtime can be calculated by using the maximum
observed counting rates - Deadtime can also be measured by using two source
method
45The Two Source Method
- Two equal sources and three measurements, R1,
R12, R2. Equal time should be used in short lived
radionuclide to avoid decay correction. - The dead times are calculated by
46(No Transcript)
47Fixed Rate Pulser Method
- A fixed rate pulser is connected to the
preamplifier of the radiation detector. - Different input pulse rates (Pt) are generated
(pulse hight larger than the photo-peak) and the
observed counting rates (Po) are recorded. - Real detector system is then connected to the
preamplifier and the observed counting rates (Ro)
are measured. The true counting rate (Rt) is then
calculated by
48(No Transcript)