ENTC 4350 - PowerPoint PPT Presentation

About This Presentation
Title:

ENTC 4350

Description:

The more common breathing-assist units use a mask that fits over the patient's nose and mouth. ... dosing a patient with a gas like halothane via a nose cone. ... – PowerPoint PPT presentation

Number of Views:112
Avg rating:3.0/5.0
Slides: 103
Provided by: HughBl
Learn more at: http://faculty.etsu.edu
Category:
Tags: entc | nose

less

Transcript and Presenter's Notes

Title: ENTC 4350


1
ENTC 4350
  • Ventilators

2
  • Sometimes the outcome of the pulmonary function
    tests may dictate a need for assisted breathing,
    which is provided by the use of ventilatory
    assist devices or respirators.
  • The most critical test in determining whether a
    patient requires respirator therapy is the blood
    gas situation, particularly regarding oxygen.

3
  • Excessive or inadequate CO2 can be compensated by
    drugs but there is no substitute or compensation
    for low oxygen levels.

4
  • In extreme cases, when the usual methods of
    oxygen enrichment and ventilatory assistance are
    inadequate or cannot be used, artificial blood
    oxygenation may be required.

5
  • The techniques of anesthesiology are related to
    those of ventilatory assistance
  • In fact, ventilatory assistance may be required
    as the anesthesiologist maintains the surgical
    patient on that razors edge between life and
    death.

6
ASSISTED BREATHING
  • Under normal conditions, breathing is pretty much
    of an automatic thing.
  • The rib cage moves outward, the lungs expand, and
    air enters the newly opened lung spaces.
  • Oxygen diffuses through the alveoli into the
    blood, and carbon dioxide diffuses out.

7
  • As an electrical analog to the pulmonary
    mechanism, we can think of the lungs as a
    capacitor that is filled with electricity passing
    through a resistance network from a voltage
    source

8
  • The pressure of the atmosphere is analogous to
    the voltage source or battery, and the nose.
    mouth, trachea, bronchi, and so on provide the
    resistances.
  • If every thing is normal, the atmosphere provides
    plenty of pressure, the tracheobronchial tree
    does not induce significant resistance, and the
    resuIting system is a low-impedance source that
    delivers adequate air to the lungs.

9
  • If we climb a mountain, the atmospheric pressure
    drops.
  • This, in effect, is like weakening the battery or
    raising its output impedance. and we feel
    altitude sickness.

10
  • The same thing can occur if the air passages are
    blocked the body yells about a shortage of
    oxygen.
  • Once again, a system which normally has low
    output impedance has been converted to a
    high-impedance system.

11
SOURCES OF RESPIRATORY DIFFICULTY
  • Some of the more common problems include the
    following
  • Deterioration of lung tissue may interfere with
    the gas-exchange processes in the lung.
  • Infection or mucus can obstruct the trachea or
    bronchial tubes and reduce the air flow through
    these passages.
  • Neural or muscular control of the nh cage may be
    impaired.
  • Lung diseasese.g.. pulmonary fibrosis or
    pneumoniamay reduce the lung elasticity and
    produce stiff lung.

12
  • The term compliance is often used in cases of
    stiff lung, and the thing to remember is that
    compliance is the opposite of stiffness.
  • A normal lung has good elasticity, it is
    compliant.

13
  • A stiff lung is just the opposite.
  • That is why compliance measurements are an
    important guide to patient health.

14
  • All these problems are to some extent related to
    aging, and, as our population grows older, we can
    expect to see more widespread use of
    breathing-assist devices.
  • Problems 1 and 2 are examples of what happens
    when the resistance in the circuit goes up less
    current flows.

15
  • Problems 3and 4 are of a different nature.
  • In these, the input impedance of the system,
    i.e., the lung, increases because the lung cannot
    expand to the proper volume.

16
  • With a transducer or a recorder, high input
    impedance was what we wanted, but in this case,
    we must get a maximum quantity of air into the
    lung to provide for O2 diffusion and CO2 removal.
  • The lung as a capacitor has become smaller, and
    that is bad news.
  • In terms of our example, pulmonary compliance is
    analogous to increased electrical capacitance.

17
MECHANICAL VENTILATORY ASSIST DEVICES
  • The objective here is to force air, oxygen or a
    controlled mixture of the two into the patients
    lungs.
  • Too little air, or oxygen. will cause lung
    deterioration too much will be just as bad.

18
  • The earliest mechanical breathing system was the
    famous (or infamous) iron lung.
  • This boiler-shaped device applied a negative
    pressure to the outside of the chest to induce
    rib cage expansion it then reversed and applied
    a positive pressure to compress the lungs and
    force air out.

19
  • A simple (simple to discuss, that is) system of
    this type can simulate normal breathing for an
    indefinite period of time.

20
  • The older, boiler-type systems had to be provided
    with glove ports so that the nurses could perform
    necessary patient services.
  • They were clumsy, and bedsores were a continual
    problem.

21
  • The more modern systems are mounted on the chest
    wall, and in some cases, the patients can sit up
    or even walk for limited periods of time.
  • One major problem with many breathing-assist
    systems is their need for electrical power.
  • Any loss of power or other breakdown is
    immediately life-threatening.

22
  • Some systems have provisions for emergency
    operation, and it is well to find and practice
    using them before they are needed.
  • Frayed and damaged electrical wires or a smell of
    hot insulation is a sign that something is
    definitely wrong.

23
  • If you can reach the drive motor on the machine,
    make it a practice to touch it now and then to
    get a feeling for its normal temperature.
  • If the motor seems warmer than usual, it is time
    to call the maintenance department.

24
  • If your unit has no emergency power supply, you
    will have to be prepared to ventilate patients by
    hand (or mouth) in case of power outage.
  • It is also vital that you know just what alarms
    the machine provides for various problems.

25
  • The more common breathing-assist units use a mask
    that fits over the patients nose and mouth.
  • There are two types of machines the
    constant-pressure and the constant-volume units.

26
  • The constant-volume or volume-regulated
    respirator delivers a given volume of air
  • The constant-pressure or pressure-regulated
    respirator delivers air until a predetermined
    pressure is reached.
  • Both units should allow a measurement of the
    actual volume delivered and the proper gauges to
    measure the pressure applied to the lungs.

27
  • Both of these systems are triggered by the
    patient upon inhalation.
  • Patient inspiration pulls a small vacuum on one
    side of a diaphragm
  • The diaphragm then moves and closes a switch to
    the machine.
  • On some units, the switch can be set to trigger
    on every breath, on every third or fourth breath,
    or on any breath in which the patient does not
    inspire a sufficient quantity of air.

28
  • If natural breathing stops, the machine will
    operate continuously at some fixed rate.
  • For patients who have no normal breathing reflex,
    the machine is set to cycle at some 12 to 15
    breaths per minute.

29
Constant-Pressure Machines
  • The constant-pressure respirator delivers air (or
    air plus oxygen) to the patient until the
    pressure reaches some preset level, at which time
    the pressure is released.
  • The volume delivered by a constant-pressure
    machine will vary with the patients position.

30
  • Any sputum or other plug in the airway system can
    stop the machine before it delivers a significant
    volume of air.
  • This problem is particularly serious when the
    patient resents the machine the patient can
    effectively prevent the delivery of air by
    blocking the flow with his tongue.

31
  • In fact, some patients will actually eject the
    mouthpiece.
  • An air flow sensor to detect this loss of contact
    should be available, and if you have one, test
    it.

32
  • Most sensors are of the heated thermistor type
  • If air flow stops they overheat and sound an
    alarm.

33
  • Another problem with the constant-pressure units
    is poor control of the oxygen-to-air ratio.
  • If the O2 level goes too high, it will result in
    patient distress.

34
  • In most cases, the constant-pressure machines are
    used for intermittent assistance only.
  • Regular checking of the oxygen-to-air ratio is
    important.

35
  • The fact that many constant-pressure machines
    operate on a purely pneumatic basis from gas
    bottles and do not need electrical power makes
    them suitable for use during patient transfer or
    in situations when the electrical power is
    uncertain.

36
Constant-Volume Machines
  • The constant-volume machine is designed to
    deliver a preset quantity of air by means of a
    piston and cylinder.
  • The machine will, within limits, raise the
    pressure as high as is necessary to deliver the
    proper quantity of air.

37
  • Constant-volume systems have the advantage of
    delivering a known amount of gas regardless of
    the patients position.
  • Airway obstructions and patient resistance can be
    overridden by the constant-volume system, but
    good drainage must be provided to insure that
    sputum is not forced down into the lung passages.

38
  • Constant-volume systems are more complex than
    constant-pressure systems, but they allow better
    control over gas delivery and they are, in
    general, more suitable for long-term
    applications.

39
  • Most constant-volume units provide a spirometer
    to allow the nurse to see how much air the
    patient is taking in on an unassisted breath.
  • The volume delivered by the machine can then be
    adjusted to provide more ventilation if
    necessary.

40
  • Many constant-volume machines have alarms for a
    variety of emergencies, such as loss of power,
    mucus blockage, breakage or kinks in the tubing,
    and so on.
  • If you turn off the alarms while aspirating the
    patient, be sure to turn them on again when the
    machine is reactivated.
  • Failure to do this has resulted in some nasty
    accidents.

41
0ther Features of Breathing-Assistance Systems
  • There are other breathing-assistance systems that
    provide positive pressure during the patients
    normal inspiration.
  • In many cases, the added air or oxygen can serve
    as a carrier for water vapor or for drugs
    dispersed by means of a nebulizer.

42
  • Some dispute exists about the value of
    intermittent positive-pressure breathing (IPPB)
    systems and their use should be controlled by the
    policy of the hospital.
  • Some patients use positive-pressure units for
    home therapy, and again, the rationale for this
    is the physicians decision.

43
  • Another factor in the employment of
    breathing-assistance systems is the use of
    differential inspiration-expiration periods.
  • Usually, the time allowed for expiration is twice
    that allowed for inspiration.

44
  • This additional time is needed for the patients
    muscular system to exhale the air that the
    constant-volume or constant-pressure machine has
    forced into the lungs.

45
  • A feature of these systems is the application of
    periodic, extra large breaths.
  • This procedure is called sighing the patient.

46
  • Sighing is thought to be necessary to prevent
    deterioration of the lung tissue, which may lead
    to atelectasis.
  • The exact mechanism of atelectasis is not yet
    under- stood, and there are various opinions
    about the need for sighing.

47
  • We suggest you consult a local pulmonary
    specialist if the question arises.

48
Power source Electric blower Inspiratory flow rate control Delivery tube from blower to plastic cylinder Inlet valve to plastic cylinder Outlet valve from plastic cylinder Sensitivity control Inspiratory signal lamp J. Inlet valve to bellows K. Outlet valve from bellows M. Rubber bellows N. Plastic cylinder P. Humidifier chamber R. Foam of bubbles S. Heater with thermoswltch T. Non-rebreathing valve W. Patient connection
49
  • Some breathing-assistance machines maintain a
    small positive pressure on the patients
    respiratory system at all times.
  • This is called positive-expiratory pressure
    (PEP), and it seems to be an important factor in
    weaning the patient off the respiratory-assist
    system.

50
  • The proper pressure level (about 25 mm Hg) should
    not be exceeded
  • Too great a pressure would excessively increase
    the work of exhalation.

51
Oxygen Enrichment Methods and Problems
  • Many machines provide facilities for oxygen
    enrichment and humidification as necessary.
  • If significant quantities of O2 are used,
    humidification is always necessary.

52
  • Humidity plays an important role in loosening
    sputum.

53
  • When O2 is used with assisted breathing, the
    apparatus for controlling the mixture of O2 and
    air should be carefully checked.
  • Just because the mixer worked last year does not
    mean it will work tomorrow.
  • Some systems use an O2 sensor, whereas others mix
    predetermined volumes of O2 and air.

54
  • Excessive O2 can be dangerous, especially to the
    newborn.
  • It is worth emphasizing the hazard of excessive
    O2
  • Experiments with normal individuals indicate that
    an O2 level of 35 to 55 can be tolerated
    indefinitely, but with 100 O2 severe pain and
    respiratory disturbances occur after 6 to 30
    hours of exposure.

55
  • If your breathing-assistance machine is driven by
    compressed O2 with a venturi for air mixture,
    there is no way of knowing exactly what the
    O2-to-air ratio is.
  • In many cases where compressed O2 was used to
    drive the machine, the gas inhaled by the patient
    was found to be over 40 O2 .

56
  • This suggests that compressed air should be used
    to drive the machine and the O2 be added as
    necessary.
  • Periodic checks with an O2 analyzer might well be
    in order.

57
Weaning the Patient from the Machine
  • This becomes a problem whenever the patient gets
    lazy and does not put out enough effort to
    breathe without the machine.
  • In this case, you have to be kind but firm and
    adjust the machine so that it only triggers on
    every third or fourth breath.

58
  • Some machines can be set to trigger only if the
    patient fails to inspire enough air.
  • In many cases, the machine is simply disconnected
    and the patient is encouraged, or forced, to walk
    about.

59
Humidity Control
  • As ambient air enters the nasal passages, it is
    humidified to avoid drying the trachea and lung
    tissue.
  • Under normal circumstances, the nasal passages
    produce an adequate amount of moisture to handle
    the driest of air.

60
  • This may not be the case when assisted breathing
    is employed, because commercial bottled air or O2
    is sold in bone dry condition to avoid
    condensation and rusting of the tanks.
  • If a patient is to breathe bottled gas for any
    period of time, it should be humidified with a
    bubbler to avoid excessive drying of nasal tissue.

61
Role of Carbon Dioxide
  • Carbon dioxide displays an interesting aspect of
    lung physiology in that the air entering the
    lungs must have a low level of CO2 (below 1) in
    order for the CO2 to diffuse out of the blood.
  • If the CO2 level rises above 2, or about 10 mm
    Hg, there will be changes in the patients
    behavior.

62
  • At the 3 level, the disturbance is severe, and
    stupor and death ensue at the 6 CO2 level.
  • This dramatic effect of CO2 is due to the direct
    influence of this gas on respiration and heart
    rate.

63
  • An increase in the CO2 level in the lungs is a
    signal to the body that a more rapid metabolic
    rate is required.
  • This allows CO2 to be used as a respiratory
    stimulant, but only up to a point.

64
  • If the CO2 level goes above about 4, there will
    be patient distress, regardless of how much O2 is
    available.

65
  • The CO2 level is one reason why medical gases are
    handled differently than the gases used for
    welding.
  • If a tank of welding oxygen is 3 CO2, the buyer
    might get a lower price, and the gas would be
    quite suitable for welding.

66
  • In hospital use, however, this could be a
    tragedy.
  • Other impurities that are kept out of
    breathing-grade gases are oil vapor (because it
    causes lung problems) and carbon monoxide.
  • The effects of carbon-monoxide poisoning are so
    serious and so well known that no discussion is
    necessary.

67
  • The fact that CO2 normally stimulates the
    respiratory process is the basis for the tendency
    to use CO2 whenever there is depression of the
    respiratory process.
  • The danger here is that such patients are already
    suffering from excess CO2 or hypercapnia, and the
    respiratory control center has become immune to
    this stimulus.

68
  • At this point, the lack of O2, or hypoxia, is the
    only thing that maintains the breathing process.
  • Administration of CO2 will not be of any help,
    because the response center is already saturated,
    and the sudden admission of high concentrations
    of O2 is equally dangerous, because it will
    remove the only stimulus for breathing (hypoxia)
    and respiration will cease.

69
  • Patients with hypercapnia, or, as it is sometimes
    called, carbon-dioxide narcosis, must be brought
    out very carefully.
  • Excessive loss of CO2, or hypocapnia, may be due
    to a variety of factors associated with
    hyperventilation.

70
  • In this case, CO2 is administered in spite of the
    fact that it is usually considered a respiratory
    stimulus.
  • The patients blood level of CO2 must be raised
    to the proper level, and this is usually best
    done by CO2 administration.

71
  • The point is that the effects of CO2 are not
    necessarily simple, and when dealing with
    problems of ventilation, both thought and
    knowledge are needed.

72
BLOOD OXYGENATION SYSTEMS
  • We know that oxygen is carried by the blood to
    various parts of the body.
  • The carrier system for O2 involves the formation
    of a chemical compound, oxyhemoglobin, which
    serves as the actual transport medium in the
    blood.

73
  • When there is severe impairment or failure of the
    blood-gas exchange that normally occurs in the
    lungs, a blood oxygenation machine can be used.

74
  • The blood is pumped from the body through a
    gas-exchange system. then to a heating or cooling
    system and, finally, through a bubble separator
    and filter and back to the body.

75
  • The gas-exchange system provides for diffusion of
    O2 into the blood to produce oxyhemoglobin.
  • It must also encourage CO2 (carbon dioxide) and
    CO to diffuse out of the blood.

76
  • These processes can be accomplished by bubbling
    airor more commonly, oxygenthrough the blood
    while at the same time rotating a series of discs
    in the blood to increase the contact area between
    the gas and the blood.
  • In some systems, a complex of narrow passages is
    produced by a membrane system.

77
  • The membrane (a special plastic) permits O2 to
    diffuse into the blood while allowing CO and CO2
    to diffuse out.

78
The Bubble Problem
  • The main problem with blood oxygenation systems
    is that O2 can enter the blood as a solution of a
    gas in a liquid.
  • The danger of this gas-in-solution business is
    that the gas can come back out of solution and
    appear as bubbles.

79
  • If the gas bubbles at the wrong time (e.g., after
    the blood has returned to the patient), there
    will be dire consequences.

80
  • The physics of the bubble problem depends upon
    two effects
  • Frothing of the blood in the gas-exchange system.
  • This produces large bubbles that are easily
    removed by the usual bubble separation system.
  • True solution problems based on the effects of
    Henrys law.
  • This is by far the most serious problem.

81
  • Henrys law states that the quantity of gas that
    dissolves in a fluid goes up as a function of the
    gas pressure in the system.
  • In the older blood oxygenation systems, the O2
    pressure was held at a value that was high enough
    to insure the rapid diffusion of O2, but this
    meant that there was some dissolution of O2 in
    the blood.

82
  • If the pressure was reduced even slightly, O2
    could come back out as bubbles, like removing the
    cap from a bottle of soda.
  • This problem could be compounded if the blood is
    cooled before it enters the oxygenation system,
    as, say, in hypothermic therapy.

83
  • Gases are more soluble in cooled blood, but they
    tend to come out of solution when the blood warms
    up.
  • Boiled water tastes flat because it has no air
    dissolved in it.

84
  • The blood oxygenation system always includes a
    bubble and froth remover, but it might not always
    get all the dissolved gas out.
  • If this unit is inadequate,
  • If the blood is too cold,
  • If the residence time (i.e., the time the blood
    spends in the oxygenator) is too short, or
  • If the O2 pressure in the oxygenator is too high,
  • The bubbles will then reappear after the blood
    has entered the connecting tubes or, worse yet,
    the patients body.

85
  • This has led to be a severe problem with many
    blood oxygenation systems.
  • A useful technique for detecting bubbles in
    oxygenated blood is ultrasonic monitoring.

86
ANESTHESIOLOGY AND ASSOCIATED PROBLEMS
  • In most hospitals, the giving of anesthetics is a
    medical specialty.
  • The anesthesiologist is responsible for watching
    the patient, providing the proper dosage, and
    advising the surgeon of any changes in patient
    condition.

87
  • There are three basic types of anesthesia
    systems
  • open,
  • partially open, and
  • closed.

88
  • The open system is the simplest one.
  • The liquid anesthetic is dripped onto a gauze
    cone that covers that patients nose and mouth.
  • The anesthetic vaporizes and is mixed with the
    air that enters the patients lungs.

89
  • One advantage of this system is its insensitivity
    to excessive use of anesthetic
  • It is almost impossible to give the patient too
    much.

90
  • The disadvantages are
  • The need for excessive quantities of anesthetic,
  • Contamination of the operating room, and
  • Possible fire or explosion hazards if ether is
    used.

91
  • The partially open system is really a generic
    term for several kinds of apparatus.
  • In most units, the anesthetic gases are provided
    in a bag reservoir so that they do not mix with
    air.

92
  • As the patient breathes, he or she receives a
    controlled mixture of air and anesthetic.
  • In some cases, the air will be enriched with O2
    or c to assist the patients respiration
    processes.
  • If bottled gases are used, appropriate
    humidification is necessary to avoid patient
    injury.

93
  • This system makes it convenient to use gases that
    are not liquids at room temperature.
  • You would have great difficulty in dosing a
    patient with a gas like halothane via a nose
    cone.

94
  • The partially open systems allow the gas mixture
    that leaves the patients lung to exhaust into
    the operating room.
  • This provides for good control of the patients
    CO2 level, but there is some contamination of the
    OR.

95
  • Some machines provide for partial rebreathing of
    the expired air, thereby saving anesthetic and
    reducing OR contamination.
  • With this system, the patient must be watched
    carefully, because of the possibility that excess
    anesthetic will build up in the system.

96
  • This is one place where an ear Oximeter or
    indwelling O2 sensor can be helpful
  • Any change in the patients condition would be
    detected almost immediately.

97
  • The last and most common systemthe closed
    systemprovides for 100 rebreathing of the
    air-anesthetic mixture.
  • As the patient exhales, the excess CO2 and water
    vapor are absorbed in appropriate canisters.

98
  • Any necessary O2 or additional anesthetic is
    added to the mixture before the gases are cycled
    back to the patient.
  • This system prevents any contamination of the OR,
    but the canisters for water vapor and CO2 must be
    watched and changed regularly.

99
  • The canisters usually contain a chemical
    indicator that tells the operator when a change
    is needed by turning color.
  • The canisters also pick up CO2 and water from the
    ambient air whenever they have been opened.
  • Most hospitals put on new canisters for every
    operation.

100
  • If ventilatory assistance is needed during
    anesthesia, the constant-volume or
    constant-pressure machines discussed previously
    can be used.
  • In emergencies, the rebreathing bag, which is
    used to catch the expired gas, can be used to
    ventilate the patient
  • The operator periodically squeezes and then lets
    go of the bag.

101
  • This activates the absorption and supply system,
    and it allows patient maintenance in case of
    machine failure.

102
  • Improper gas mixtures must be watched for,
    because many anesthesia systems do not have
    integral gas analyzers.
  • You just have to adjust the appropriate
    flowmeters and observe the patients condition.
Write a Comment
User Comments (0)
About PowerShow.com