Chapter 31 Quality Control - PowerPoint PPT Presentation

1 / 70
About This Presentation
Title:

Chapter 31 Quality Control

Description:

Chapter 31 Quality Control Two areas of activity are designed to ensure the best possible image quality with the lowest possible exposure and minimum costs. – PowerPoint PPT presentation

Number of Views:246
Avg rating:3.0/5.0
Slides: 71
Provided by: w3PalmerE2
Category:

less

Transcript and Presenter's Notes

Title: Chapter 31 Quality Control


1
Chapter 31 Quality Control
  • Two areas of activity are designed to ensure the
    best possible image quality with the lowest
    possible exposure and minimum costs.
  • Quality Assurance deals with people
  • Quality Control deals with instrumentation and
    equipment.

2
Ten Step Approach to Quality Assurance
  • Assign responsibility
  • Delineate scope of care
  • Identify aspects of care
  • Identify outcomes that effect the aspects of
    care.
  • Establish limits of the scope of assessment.

3
Ten Step Approach to Quality Assurance
  • Collect and organize data.
  • Evaluate care when outcomes are reached.
  • Take action to improve care
  • Assess and document actions
  • Communicate information to organization-wide QA
    Program

4
QA Projects
  • Things that QA can evaluate includes
  • Scheduling of patients
  • Instructions given to patients
  • Wait times in the office
  • Interpretation of films
  • Retake analysis
  • Record accuracy

5
QA Program
  • Quality Assurance deals with people and processes
    used to complete tasks.
  • QA involves training and record keeping.
  • As the owner of the equipment, you will be
    responsible for your radiology services.
  • The State of California Department of Radiologic
    Health established the Standards of Good Practice
    that is the foundation of QA and QC in
    radiography.

6
QA and QC Requirements
  • Degree of requirements vary by state. California
    and New York have very tight standards for
    quality control of the radiographic and
    processing equipment.
  • We are required by statue to teach QA and QC in
    the radiology program. It is covered in detail in
    9th Quarter. My textboook covers QC in detail.

7
Quality Control
  • An acceptable QC program has three steps
  • Acceptance Testing
  • Routine performance monitoring
  • Maintenance

8
Acceptance Testing
  • The x-ray machine, cassettes and film processor
    or digital system are the largest capital expense
    you may experience.
  • It makes economic sense to make sure that the
    equipment meets the performance standards.
  • It is recommended that a third party such as a
    health physicist do the testing.

9
Acceptance Testing
  • Areas that should be tested include on the x-ray
    machine
  • Shielding of Room
  • Focal spot size
  • Calibration of mA, timer or mAs
  • Calibration of kVp
  • Linearity of exposure
  • Beam alignment
  • Grid centering
  • Collimation
  • Filtration (HVL)

10
Acceptance Testing
  • Areas that should be tested on the x-ray
    cassettes
  • Screen contact
  • Screen speed
  • Light leaks
  • Light spectrum matching

11
Acceptance Testing
  • Areas that should be tested on the x-ray film
    processor
  • Developer temperature
  • Replenishment rates
  • Travel time
  • Water flow
  • Hypo retention

12
Quality Control
  • The acceptance testing ensures that the machine
    was installed and calibrated properly.
  • The performance may drift or deteriorate over
    time. Consequently, periodic testing is required
    to monitor the performance.
  • With the exception of film processing most
    testing is annual or semiannual.

13
Quality Control
  • After a major repair, the machine should be
    retested to ensure that it was repaired properly.
  • When the testing shows that the machine is not
    performing properly, service or preventive
    maintenance is required.
  • Manufactures establish recommended preventive
    service schedules. When these are followed many
    repairs become unnecessary.

14
Radiographic Quality Control
  • Areas of concern in x-ray machine
  • Focal Spot Size will impact spatial resolution
  • Filtration will impact patient exposure
  • Collimation will impact patient exposure
  • kVp calibration will impact image quality and
    exposure.
  • Exposure timer accuracy will impact image quality
    and exposure

15
Radiographic Quality Control
  • Areas of concern in x-ray machine
  • Exposure linearity will impact exposure and image
    quality
  • Exposure reproducibility will impact exposure and
    image quality.
  • Alignment of tube and image receptor will impact
    exposure and image quality.

16
Focal Spot Testing
  • When the machine is installed or the tube is
    replaced, the focal spot size should be measured.
    Then annually thereafter.
  • A pin hole camera, star test pattern or line pair
    test tool.
  • As the tube ages, the focal spot tends to grow
    and spatial resolution is lost.

17
Filtration
  • The filtration is measured but determining the
    half value layer of the beam at specific exposure
    levels. Minimum filtration is 2.5mm aluminum.
  • As a tube ages, tungsten will plate the x-ray
    port and increase filtration. This can cause
    technique problems. Inadequate filtration will
    significantly increase patient exposure.

18
Collimation
  • If the collimation is misaligned, intended
    anatomy can be missed.
  • It can be tested in many ways from using pennies
    to using test patterns.
  • Misalignment can not exceed 2 of the SID.
  • It is tested semiannually and after the
    replacement of the collimator lamp.

19
kVp Calibration
  • In diagnostic radiology, any change will impact
    patient exposure. A variation of about 3 will
    impact contrast and image density.
  • Can be tested with filtered ion chambers,
    filtered photodiodes or even a cassette with
    calibrated filters.
  • Tested annually.

20
Exposure Timer Accuracy
  • The exposure time is the responsibility of the
    operator. It will impact the density of the image
    and spatial resolution.
  • Tested with an ion chamber, multi-meter
    internally or even a spinning top.
  • Exposure time must be within 5 for exposure
    times greater than 10 ms and 20 less than 10 ms.

21
Exposure Linearity
  • Many combinations of mA and time will produce the
    same mAs value. The ability of the machine to
    produce a constant level of exposure with various
    combinations of mA and time is called exposure
    linearity.
  • Can be tested with a step wedge and densitometer
    or rate meter.
  • Should be within 10 for adjacent stations.

22
Exposure Reproducibility
  • Any exposure using the same factors should
    produce the same level of density and contrast on
    the image.
  • Sequential exposure should be reproducible to
    within 5
  • Can be tested with a rate meter or step wedge and
    densitometer.

23
Performance standards for x-ray equipment
24
Darkroom and Processing
  • The development of the image is dependent upon
    the temperature of the developer, its
    concentration and how long the film is in the
    developer.
  • The film is sensitive to variations in the
    environment and processing from the time it is
    manufactured until it is processed.
  • Darkroom and Processor QC is the key process of
    Quality Control.

25
Processing
  • Processor densitometry is performed daily before
    the first patient is exposed.
  • A sensitometer is used to produce a step wedge
    image on the film that is evaluated with a
    densitometer.
  • The densitometer reads the optical density of the
    processed image.
  • A digital thermometer is used to test the
    chemical temperatures in the processor.

26
Processing
  • Key densities on the processed film are measured
    and then graphed.
  • Base plus Fog is measured on an area of unexposed
    film to check the darkroom environment.
  • Speed is tested at the level of exposure that
    produces a density of 1.25OD
  • Contrast is tested at the level that produced a
    density of 0.40 OD and one that produced a
    density of 2.20.

27
Processing
  • By monitoring these densities, problems with film
    processing can be detected before image quality
    deteriorates.
  • In 9th Quarter we will cover how to perform
    processor QC and problem solve.

28
Waste Records
  • Since used fixer is classified as a hazardous
    waste material, it is important to maintain
    accurate records of usage and disposal.
  • The extent of records vary by city, county and
    state. You are responsible for the proper
    disposal of the waste. Some regions include
    developer as hazardous waste.

29
Silver Recovery
  • If the silver ions are removed from the fixer, it
    may be disposed of in the normal waste when
    diluted with water.
  • There are two primary types of silver recovery
    systems.
  • Metallic replacement uses steel wool and can
    recover 95 of the silver in the effluent
  • Electrolytic recovery passes direct current
    through the solution and nearly pure metallic
    silver is deposited on the cathode.

30
Silver Recovery
  • Old radiographic films and repeated films are
    retained for silver recovery. X-ray images can
    not be disposed of in normal trash.
  • They also can not be used to clean the processor
    rollers.
  • Waste recovery companies will either burn or
    chemically remove the silver from the film.

31
Performance standards for film processor and
darkroom equipment
32
Accessory QC
  • The cassettes and screens are the area of chief
    concern. Problems with either will result in
    artifacts on the images and increased retakes.
  • The screens need to be properly cleaned
    frequently. Mammography screens are cleaned
    daily. California recommends monthly cleaning.

33
Accessory QC
  • Dirty screens produce white artifacts on the
    image.
  • Multiple white artifacts indicate the need to
    replace the screens.
  • Screen contact is tested semiannually. A problem
    with screens contact will cause a loss of
    resolution.
  • As screens age, they loose speed so this is also
    tested.

34
Accessory QC
  • The other important accessory is the gonad
    protection devices and lead aprons.
  • Improper care of the apron can result in cracks
    and holes in the lead that reduces their
    effectiveness.
  • Aprons and shields are tested semiannually. The
    easiest was to test them is with
    video-fluoroscopy but film can be used.

35
Performance Standards for Accessories
36
Failed Lead Apron Test
37
Record keeping
  • All electromechanical devices need periodic
    service.
  • There are three types of maintenance.
  • Scheduled maintenance such as processor monthly
    or weekly service. It includes observing moving
    parts and lubrication.
  • Preventive maintenance is planned service and
    replacement of parts at regular intervals before
    they fail at inopportune times.

38
Record keeping
  • Non-scheduled maintenance is the worst type of
    service because it impacts patient service. It
    may also be very expensive.
  • With proper scheduled and preventive maintenance,
    non-scheduled service can be minimized.
  • All service schedules should meet manufacture
    recommendations.
  • All service should be documented as part of the
    quality assurance program.

39
Retake Analysis
  • Required part of a QA program in California.
  • Evaluation includes
  • View repeated
  • Cause of the repeat
  • Rate of retakes should be less than 5.
  • Information can be gathered from the log that the
    state mandates for patients being exposed to
    radiation.

40
Retake Analysis
  • Done every three months using a relatively large
    sample of data to see trends in
  • Type of examination being repeated.
  • Reasons for the repeated films.
  • Determine if additional training or review is
    needed.
  • Determine if equipment service might be required.

41
Repeat Analysis
  • Your patient log can be designed to capture both
    films usage and repeated films.
  • Data is gathered from log for analysis
  • Repeated films can be put into two main
    categories
  • X-ray Personnel Errors
  • Equipment malfunctions

42
X-ray Personnel Errors
  • Failure to measure patient.
  • Use of improper technical factors (mAs, kVp or
    distance)
  • Incorrect positioning
  • Improper Collimation
  • Improper use of accessories such as cassettes,
    grids or filters

43
X-ray Personnel Errors
  • Improper handling of exposed or unexposed films.
  • Failure to clearly communicate to the patient
    breathing instructions and to remain still.
  • Failure to observe patient during exposure.

44
X-ray Equipment and Accessory Failure or
Malfunctions
  • Inaccurate calibration of kVp and mA.
  • Inaccurate timer calibration
  • Dirty or damaged cassettes
  • Improperly labeled or damaged grids
  • Malfunctioning collimator
  • Improper film storage or processor function.

45
Reasons for Retake Films
  • Over or under exposure accounts for over 50 of
    retakes nationally.
  • Errors in positioning (25)
  • Patient motion 11
  • Processing errors 6
  • Wrong view, beam alignment,cassette screen or
    grid errors and artifacts.

46
Retake Films by Region
  • Cervical Spine Exams 7
  • Thoracic Spine Exams 17
  • Lumbar and Abdominal X-rays 40
  • Skull, Chest,Lower Extremities 15
  • Majority or retake results in unnecessary
    exposure to gonads or blood forming organs of the
    body.

47
Daily Log Design
  • Most regulatory agencies will require a log of
    patients having radiation exposure.
  • Columns can be added to capture film usage,
    repeats, views repeated and reason.
  • This data can be gathered and analyzed.

48
Retake Analysis
  • In this example, most of the retakes were of the
    T-spine. Potential reasons include
  • Improper use of filters
  • Incorrect measurements
  • Faulty technique chart

49
Retake Analysis
  • The reason of each retake is recorded and
    percentages are computed to determine the
    overall rate and rate by reason.
  • Less than 5 is ideal.

50
Quality Assurance and Digital Radiography
  • Information gathered for this lecture came from
    American College of Radiology Practice Guidelines
    for Digital Radiography 10/10/07
  • Quality Management in The Imaging Sciences
    Jeffrey Pap Mosby 1998

51
The Digital Radiography Advantage
  • Digital Radiography offers many advantages over
    film based radiography.
  • Film response to radiation follows a sigmoid
    response curve while digital response is linear.
    This results in a much wider dynamic response to
    exposure. The advantages include
  • Reduced retakes due to exposure factors.
  • Improved contrast resolution.
  • Improved latitude.

52
The Digital Radiography Advantages
  • Much of the quality assurance for film based
    radiography is to assure proper photographic
    processing of the image. This is eliminated
    unless a film printer or multi-format camera is
    used.
  • Less space and perhaps less staff is required
    because of reduced storage space requirements.
    Images can be stored on a hard disk or CD.

53
The Digital Radiography Advantages
  • If the images are left in digital format, no
    darkroom or darkroom personnel is needed.
  • No film, processing chemistry and waste recovery
    costs.
  • Computed Radiography can be added to existing
    equipment. Direct Digital requires a total
    replacement of the image receptor.

54
Computed Radiography and Quality Control
  • Computed radiography still uses conventional
    radiographic equipment so basic radiography
    quality control testing is still needed. This
    would include
  • Calibration of the generator
  • Collimation
  • Beam Alignment
  • Collimation
  • Linearity of exposure
  • Spatial resolution (focal spot)

55
Computed Radiography and Quality Control
  • Calibration of the generator.
  • The linear dose response can be a double edged
    sword. It can mask calibration problem that can
    result in higher patient exposure.
  • Over exposed images really look good so this has
    resulted in dose creep.
  • Collimation Beam Alignment
  • Collimation is critical for the computer to
    determine the proper contrast and density scale.
  • Spatial resolution (focal spot)
  • Film can record greater spatial resolution than
    digital radiography so resolution should be
    monitored more often than film based systems.

56
Computed Radiography and Quality Control
  • We have discussed much of the radiographic
    quality control earlier.
  • The end product in radiography is the image and
    accurate report of findings.
  • Therefore for film, the processor function,
    screen contract and artifact avoidance was
    critical.
  • With digital radiography the computer monitor,
    image transfer and image receptor plate become
    the critical areas for quality assurance.

57
Monitor Testing
  • The acquisition workstation and secondary
    displays should be tested frequently using a
    standard image test pattern.
  • The SMPTE RP-133 (Society of Motion Picture and
    Television Engineers) or AAPM TG-18 QC test
    patterns are suitable.

58
(No Transcript)
59
Using the SMPTE Test Pattern
  • As a image fidelity test, the test pattern data
    files should be sized to occupy the full area
    used to display images.
  • The overall SMPTE image appearance should be
    inspected to assure that absence of gross
    artifacts.

60
Using the SMPTE Test Pattern
  • There should be no blurring or bleeding of bright
    display areas into dark areas or aliasing of the
    spatial imaging patterns.
  • All monitors used for interpretation should be
    tested at least monthly.

61
Using the SMPTE Test Pattern
  • For the dynamic range test, both the 5 and 95
    contrast areas should be visible and distinctly
    different than the adjacent 0 and 100 areas.
  • Therefore

62
Using the SMPTE Test Pattern
  • The brightness and contrast of the monitor is
    adequate if the 5 squares at both ends of the
    gray scale are visible.
  • The gray scale is shown as a series of squares in
    the center of the image that range from black
    (0) to white (100).

63
Using the SMPTE Test Pattern
  • It may be difficult to get the monitor to show
    both of these inner squares perfectly.
  • Most monitors do better showing the 95 square
    than the 5 square.
  • However, you might try adjusting the ambient
    light to improve the visibility of the 5 square.

64
Spatial Resolution Aliasing
  • The spatial resolution (linearity) and aliasing
    (distortion) of the monitor are within acceptable
    limits if the high contrast bar patterns in the
    test image are distinct as simple black and white
    pairs.

65
Spatial Resolution Aliasing
  • In each corner and the very center of the image,
    inspect the six squares filled with varying
    widths of alternating black/white horizontal and
    vertical lines.

66
Spatial Resolution Aliasing
  • You should be able to differentiate all the lines
    from fat to narrow (6 pixels, 4 pixels and 2
    pixels) and both horizontally and vertically.

67
BWH Pattern
  • The BWH pattern is a test of the continuous gray
    scale from the center to outside of the pattern.
  • It was developed at the Brigham Women's Hospital
    Radiology Department.

68
BWH Pattern
  • No concentric ring like features should be
    present in the image.
  • If they are present, the images displayed on your
    system will be of less than optimum quality.

69
New TG-18 Monitor QC
  • The TG-18 Monitor QC test pattern combines
    features of the SMPTE pattern with resolution
    tests.
  • New patterns with chest, breast and knee images
    are coming available soon.

70
End of Lecture
Write a Comment
User Comments (0)
About PowerShow.com