Title: Chapter 31 Quality Control
1Chapter 31 Quality Control
- Two areas of activity are designed to ensure the
best possible image quality with the lowest
possible exposure and minimum costs. - Quality Assurance deals with people
- Quality Control deals with instrumentation and
equipment.
2Ten Step Approach to Quality Assurance
- Assign responsibility
- Delineate scope of care
- Identify aspects of care
- Identify outcomes that effect the aspects of
care. - Establish limits of the scope of assessment.
3Ten Step Approach to Quality Assurance
- Collect and organize data.
- Evaluate care when outcomes are reached.
- Take action to improve care
- Assess and document actions
- Communicate information to organization-wide QA
Program
4QA Projects
- Things that QA can evaluate includes
- Scheduling of patients
- Instructions given to patients
- Wait times in the office
- Interpretation of films
- Retake analysis
- Record accuracy
5QA Program
- Quality Assurance deals with people and processes
used to complete tasks. - QA involves training and record keeping.
- As the owner of the equipment, you will be
responsible for your radiology services. - The State of California Department of Radiologic
Health established the Standards of Good Practice
that is the foundation of QA and QC in
radiography.
6QA and QC Requirements
- Degree of requirements vary by state. California
and New York have very tight standards for
quality control of the radiographic and
processing equipment. - We are required by statue to teach QA and QC in
the radiology program. It is covered in detail in
9th Quarter. My textboook covers QC in detail.
7Quality Control
- An acceptable QC program has three steps
- Acceptance Testing
- Routine performance monitoring
- Maintenance
8Acceptance Testing
- The x-ray machine, cassettes and film processor
or digital system are the largest capital expense
you may experience. - It makes economic sense to make sure that the
equipment meets the performance standards. - It is recommended that a third party such as a
health physicist do the testing.
9Acceptance Testing
- Areas that should be tested include on the x-ray
machine - Shielding of Room
- Focal spot size
- Calibration of mA, timer or mAs
- Calibration of kVp
- Linearity of exposure
- Beam alignment
- Grid centering
- Collimation
- Filtration (HVL)
10Acceptance Testing
- Areas that should be tested on the x-ray
cassettes - Screen contact
- Screen speed
- Light leaks
- Light spectrum matching
11Acceptance Testing
- Areas that should be tested on the x-ray film
processor - Developer temperature
- Replenishment rates
- Travel time
- Water flow
- Hypo retention
12Quality Control
- The acceptance testing ensures that the machine
was installed and calibrated properly. - The performance may drift or deteriorate over
time. Consequently, periodic testing is required
to monitor the performance. - With the exception of film processing most
testing is annual or semiannual.
13Quality Control
- After a major repair, the machine should be
retested to ensure that it was repaired properly. - When the testing shows that the machine is not
performing properly, service or preventive
maintenance is required. - Manufactures establish recommended preventive
service schedules. When these are followed many
repairs become unnecessary.
14Radiographic Quality Control
- Areas of concern in x-ray machine
- Focal Spot Size will impact spatial resolution
- Filtration will impact patient exposure
- Collimation will impact patient exposure
- kVp calibration will impact image quality and
exposure. - Exposure timer accuracy will impact image quality
and exposure
15Radiographic Quality Control
- Areas of concern in x-ray machine
- Exposure linearity will impact exposure and image
quality - Exposure reproducibility will impact exposure and
image quality. - Alignment of tube and image receptor will impact
exposure and image quality.
16Focal Spot Testing
- When the machine is installed or the tube is
replaced, the focal spot size should be measured.
Then annually thereafter. - A pin hole camera, star test pattern or line pair
test tool. - As the tube ages, the focal spot tends to grow
and spatial resolution is lost.
17Filtration
- The filtration is measured but determining the
half value layer of the beam at specific exposure
levels. Minimum filtration is 2.5mm aluminum. - As a tube ages, tungsten will plate the x-ray
port and increase filtration. This can cause
technique problems. Inadequate filtration will
significantly increase patient exposure.
18Collimation
- If the collimation is misaligned, intended
anatomy can be missed. - It can be tested in many ways from using pennies
to using test patterns. - Misalignment can not exceed 2 of the SID.
- It is tested semiannually and after the
replacement of the collimator lamp.
19kVp Calibration
- In diagnostic radiology, any change will impact
patient exposure. A variation of about 3 will
impact contrast and image density. - Can be tested with filtered ion chambers,
filtered photodiodes or even a cassette with
calibrated filters. - Tested annually.
20Exposure Timer Accuracy
- The exposure time is the responsibility of the
operator. It will impact the density of the image
and spatial resolution. - Tested with an ion chamber, multi-meter
internally or even a spinning top. - Exposure time must be within 5 for exposure
times greater than 10 ms and 20 less than 10 ms.
21Exposure Linearity
- Many combinations of mA and time will produce the
same mAs value. The ability of the machine to
produce a constant level of exposure with various
combinations of mA and time is called exposure
linearity. - Can be tested with a step wedge and densitometer
or rate meter. - Should be within 10 for adjacent stations.
22Exposure Reproducibility
- Any exposure using the same factors should
produce the same level of density and contrast on
the image. - Sequential exposure should be reproducible to
within 5 - Can be tested with a rate meter or step wedge and
densitometer.
23Performance standards for x-ray equipment
24Darkroom and Processing
- The development of the image is dependent upon
the temperature of the developer, its
concentration and how long the film is in the
developer. - The film is sensitive to variations in the
environment and processing from the time it is
manufactured until it is processed. - Darkroom and Processor QC is the key process of
Quality Control.
25Processing
- Processor densitometry is performed daily before
the first patient is exposed. - A sensitometer is used to produce a step wedge
image on the film that is evaluated with a
densitometer. - The densitometer reads the optical density of the
processed image. - A digital thermometer is used to test the
chemical temperatures in the processor.
26Processing
- Key densities on the processed film are measured
and then graphed. - Base plus Fog is measured on an area of unexposed
film to check the darkroom environment. - Speed is tested at the level of exposure that
produces a density of 1.25OD - Contrast is tested at the level that produced a
density of 0.40 OD and one that produced a
density of 2.20.
27Processing
- By monitoring these densities, problems with film
processing can be detected before image quality
deteriorates. - In 9th Quarter we will cover how to perform
processor QC and problem solve.
28Waste Records
- Since used fixer is classified as a hazardous
waste material, it is important to maintain
accurate records of usage and disposal. - The extent of records vary by city, county and
state. You are responsible for the proper
disposal of the waste. Some regions include
developer as hazardous waste.
29Silver Recovery
- If the silver ions are removed from the fixer, it
may be disposed of in the normal waste when
diluted with water. - There are two primary types of silver recovery
systems. - Metallic replacement uses steel wool and can
recover 95 of the silver in the effluent - Electrolytic recovery passes direct current
through the solution and nearly pure metallic
silver is deposited on the cathode.
30Silver Recovery
- Old radiographic films and repeated films are
retained for silver recovery. X-ray images can
not be disposed of in normal trash. - They also can not be used to clean the processor
rollers. - Waste recovery companies will either burn or
chemically remove the silver from the film.
31Performance standards for film processor and
darkroom equipment
32Accessory QC
- The cassettes and screens are the area of chief
concern. Problems with either will result in
artifacts on the images and increased retakes. - The screens need to be properly cleaned
frequently. Mammography screens are cleaned
daily. California recommends monthly cleaning.
33Accessory QC
- Dirty screens produce white artifacts on the
image. - Multiple white artifacts indicate the need to
replace the screens. - Screen contact is tested semiannually. A problem
with screens contact will cause a loss of
resolution. - As screens age, they loose speed so this is also
tested.
34Accessory QC
- The other important accessory is the gonad
protection devices and lead aprons. - Improper care of the apron can result in cracks
and holes in the lead that reduces their
effectiveness. - Aprons and shields are tested semiannually. The
easiest was to test them is with
video-fluoroscopy but film can be used.
35Performance Standards for Accessories
36Failed Lead Apron Test
37Record keeping
- All electromechanical devices need periodic
service. - There are three types of maintenance.
- Scheduled maintenance such as processor monthly
or weekly service. It includes observing moving
parts and lubrication. - Preventive maintenance is planned service and
replacement of parts at regular intervals before
they fail at inopportune times.
38Record keeping
- Non-scheduled maintenance is the worst type of
service because it impacts patient service. It
may also be very expensive. - With proper scheduled and preventive maintenance,
non-scheduled service can be minimized. - All service schedules should meet manufacture
recommendations. - All service should be documented as part of the
quality assurance program.
39Retake Analysis
- Required part of a QA program in California.
- Evaluation includes
- View repeated
- Cause of the repeat
- Rate of retakes should be less than 5.
- Information can be gathered from the log that the
state mandates for patients being exposed to
radiation.
40Retake Analysis
- Done every three months using a relatively large
sample of data to see trends in - Type of examination being repeated.
- Reasons for the repeated films.
- Determine if additional training or review is
needed. - Determine if equipment service might be required.
41Repeat Analysis
- Your patient log can be designed to capture both
films usage and repeated films. - Data is gathered from log for analysis
- Repeated films can be put into two main
categories - X-ray Personnel Errors
- Equipment malfunctions
42X-ray Personnel Errors
- Failure to measure patient.
- Use of improper technical factors (mAs, kVp or
distance) - Incorrect positioning
- Improper Collimation
- Improper use of accessories such as cassettes,
grids or filters
43X-ray Personnel Errors
- Improper handling of exposed or unexposed films.
- Failure to clearly communicate to the patient
breathing instructions and to remain still. - Failure to observe patient during exposure.
44X-ray Equipment and Accessory Failure or
Malfunctions
- Inaccurate calibration of kVp and mA.
- Inaccurate timer calibration
- Dirty or damaged cassettes
- Improperly labeled or damaged grids
- Malfunctioning collimator
- Improper film storage or processor function.
45Reasons for Retake Films
- Over or under exposure accounts for over 50 of
retakes nationally. - Errors in positioning (25)
- Patient motion 11
- Processing errors 6
- Wrong view, beam alignment,cassette screen or
grid errors and artifacts.
46Retake Films by Region
- Cervical Spine Exams 7
- Thoracic Spine Exams 17
- Lumbar and Abdominal X-rays 40
- Skull, Chest,Lower Extremities 15
- Majority or retake results in unnecessary
exposure to gonads or blood forming organs of the
body.
47Daily Log Design
- Most regulatory agencies will require a log of
patients having radiation exposure. - Columns can be added to capture film usage,
repeats, views repeated and reason. - This data can be gathered and analyzed.
48Retake Analysis
- In this example, most of the retakes were of the
T-spine. Potential reasons include - Improper use of filters
- Incorrect measurements
- Faulty technique chart
49Retake Analysis
- The reason of each retake is recorded and
percentages are computed to determine the
overall rate and rate by reason. - Less than 5 is ideal.
50Quality Assurance and Digital Radiography
- Information gathered for this lecture came from
American College of Radiology Practice Guidelines
for Digital Radiography 10/10/07 - Quality Management in The Imaging Sciences
Jeffrey Pap Mosby 1998
51The Digital Radiography Advantage
- Digital Radiography offers many advantages over
film based radiography. - Film response to radiation follows a sigmoid
response curve while digital response is linear.
This results in a much wider dynamic response to
exposure. The advantages include - Reduced retakes due to exposure factors.
- Improved contrast resolution.
- Improved latitude.
52The Digital Radiography Advantages
- Much of the quality assurance for film based
radiography is to assure proper photographic
processing of the image. This is eliminated
unless a film printer or multi-format camera is
used. - Less space and perhaps less staff is required
because of reduced storage space requirements.
Images can be stored on a hard disk or CD.
53The Digital Radiography Advantages
- If the images are left in digital format, no
darkroom or darkroom personnel is needed. - No film, processing chemistry and waste recovery
costs. - Computed Radiography can be added to existing
equipment. Direct Digital requires a total
replacement of the image receptor.
54Computed Radiography and Quality Control
- Computed radiography still uses conventional
radiographic equipment so basic radiography
quality control testing is still needed. This
would include - Calibration of the generator
- Collimation
- Beam Alignment
- Collimation
- Linearity of exposure
- Spatial resolution (focal spot)
55Computed Radiography and Quality Control
- Calibration of the generator.
- The linear dose response can be a double edged
sword. It can mask calibration problem that can
result in higher patient exposure. - Over exposed images really look good so this has
resulted in dose creep. - Collimation Beam Alignment
- Collimation is critical for the computer to
determine the proper contrast and density scale. - Spatial resolution (focal spot)
- Film can record greater spatial resolution than
digital radiography so resolution should be
monitored more often than film based systems.
56Computed Radiography and Quality Control
- We have discussed much of the radiographic
quality control earlier. - The end product in radiography is the image and
accurate report of findings. - Therefore for film, the processor function,
screen contract and artifact avoidance was
critical. - With digital radiography the computer monitor,
image transfer and image receptor plate become
the critical areas for quality assurance.
57Monitor Testing
- The acquisition workstation and secondary
displays should be tested frequently using a
standard image test pattern. - The SMPTE RP-133 (Society of Motion Picture and
Television Engineers) or AAPM TG-18 QC test
patterns are suitable.
58(No Transcript)
59Using the SMPTE Test Pattern
- As a image fidelity test, the test pattern data
files should be sized to occupy the full area
used to display images. - The overall SMPTE image appearance should be
inspected to assure that absence of gross
artifacts.
60Using the SMPTE Test Pattern
- There should be no blurring or bleeding of bright
display areas into dark areas or aliasing of the
spatial imaging patterns. - All monitors used for interpretation should be
tested at least monthly.
61Using the SMPTE Test Pattern
- For the dynamic range test, both the 5 and 95
contrast areas should be visible and distinctly
different than the adjacent 0 and 100 areas. - Therefore
62Using the SMPTE Test Pattern
- The brightness and contrast of the monitor is
adequate if the 5 squares at both ends of the
gray scale are visible. - The gray scale is shown as a series of squares in
the center of the image that range from black
(0) to white (100).
63Using the SMPTE Test Pattern
- It may be difficult to get the monitor to show
both of these inner squares perfectly. - Most monitors do better showing the 95 square
than the 5 square. - However, you might try adjusting the ambient
light to improve the visibility of the 5 square.
64Spatial Resolution Aliasing
- The spatial resolution (linearity) and aliasing
(distortion) of the monitor are within acceptable
limits if the high contrast bar patterns in the
test image are distinct as simple black and white
pairs.
65Spatial Resolution Aliasing
- In each corner and the very center of the image,
inspect the six squares filled with varying
widths of alternating black/white horizontal and
vertical lines.
66Spatial Resolution Aliasing
- You should be able to differentiate all the lines
from fat to narrow (6 pixels, 4 pixels and 2
pixels) and both horizontally and vertically.
67BWH Pattern
- The BWH pattern is a test of the continuous gray
scale from the center to outside of the pattern. - It was developed at the Brigham Women's Hospital
Radiology Department.
68BWH Pattern
- No concentric ring like features should be
present in the image. - If they are present, the images displayed on your
system will be of less than optimum quality.
69New TG-18 Monitor QC
- The TG-18 Monitor QC test pattern combines
features of the SMPTE pattern with resolution
tests. - New patterns with chest, breast and knee images
are coming available soon.
70End of Lecture