Title: Bioavailability
1(No Transcript)
2Bioavailability
- Bioavailability means the rate and extent to
which the active substance is adsorbed from a
pharmaceutical product and become available at
the site of action
3Bioequivalence
- Two medical products are bioequivalent if they
are pharmaceutical equivalent or pharmaceutical
alternatives and if their bioavailabilities after
administration in the same molar dose are similar
to such degree that their effects, with respect
to both efficicy and safety, will be essentially
the same
4Design and conduct of studies
- The study should be designed in such a way that
the formulation effect can be distinguished from
other effects. - Most common is a two-period, two-sequence
crossover design, if the formulations to be
compared is two - Single dose studies
- Steady-state studies
5Design and conduct of studies
- Adequate wash out periods between treatments
- Sampling schedule
- to provide an adequate estimation of Cmax
- to cover the plasma concentration time curve long
enough, 80 of AUC - 24 hours cycle at steady state?
- drugs with long half-life?
6Subjects
- Healthy volunteers
- The inclusion/exclusion criteria should be
clearly stated in the protocol - Both sex
- 18-55 years old
- Normal weight
- Screened for
- laboratory test
- medical history
- medical examination
- preferable non-smokers and without a history of
alcohol or drug abuse
7Chemical analysis
- The bioanalytical part of bioequivalence trials
should be conducted according to the applicable
principle of Good Laboratory Practice (GLP)
8Good Laboratory Practice (GLP)
- Test plan (Analytical protocol)
- Sample traceability
- Documentation, possible to reconstruct the study
- Analytical method validation report
- Analytical report signed by responsible
investigator
9Pre-study phase
- The method used must be well characterised
- Stability
- Specificity
- Accuracy
- Precision
- Limit of quantitation
- Response function
10Validation objective
- To demonstrate that the analytical procedure is
suitable for its intended purpose
11Analytical method validation
Analytical Procedure
Stability
Selectivity
Robustness
Validation
Accuracy
Calibration curve
Precision - within-run- between-run
Limit of Quantitation LOQ
Recovery
12Analytical procedure
Separation
Detection
Sample preparation
13Specificity (selectivity)
- Ability of an analytical method to measure only
what it is intended to measure - Blank samples from six different subjects
- Will other drugs, metabolites or endogenous
components interfere in the measurements?
14Accuracy
- The closeness of mean test results obtained by
the analytical method to the true value
(concentration) of the analyte.
15Accuracy
- Accuracy should be measured at minimum 3 levels
- At least 5 determinations per concentration
- The mean value should be within 15 of the actual
value - At the lower limit of quantitation level within
20 is accepted
x
x
x
x
x
x
16Precision
- The closeness of individual measurements of an
analyte when the procedure is applied repeatedly
to multiple aliquotes of a single homogenous
volume of biological matrix
17Precision
- Precision should be measured at minimum 3 levels
- At least 5 determinations per concentration
- The calculated CV should not exceed 15
- At the lower limit of quantitation level, CV
should not exceed 20 - Subdivided into within-run and between-run
18Precision and Accuracy
Conc.
Accuracy
Precision
n
nmol/l
()
Within-run
Between-run
0.76
0.6
5.1
6.1
18
23
3.6
1.7
1.8
18
122
3.9
1.3
1.3
18
19Recovery
- The extraction efficiency of an analytical method
- Recovery of an analyte need not be 100
20Lower limit of quantitation
- The lowest standard on the calibration curve
should be accepted as the lower limit of
quantitation (LLOQ) - if
21Lower limit of quantification
- The analyte responce at LLOQ is at least 5 times
the blank response - The peak should be identifiable and discrete
- Precision within 20 CV
- Accuracy of 80-120
22LLOQ (1.50 nmol/l) for morphine
23Calibration/Standard curve
- A calibration curve is the relation between
instrument response and known concentrations of
the analyte - Should be prepared in the same biological matrix
as the samples - Should consist of 6-8 samples covering the
expected range - Should include LLOQ and a blank sample
- Should include a zero sample (with internal
standard) - Same curve fitting, weighting in prestudy and
study - Any changes should be documented
24Calibration curve
25Sample dilution
- Any required sample dilutions should use like
matrix - Dilution QC sample should be used
26Robustness
- How many samples can be analysed in one run?
27Robustness
115
110
105
100
Found concentration
0
10
20
30
40
50
60
70
80
90
100
110
95
90
85
Sample No.
28Stability of your substance
In Freeze/Thaw tests
In room temperature (4 h)
In the automatic injector
In stock solutions
In plasma during storage
29Analytical method validation
Analytical Procedure
Stability
Selectivity
Robustness
Validation
Accuracy
Calibration curve
Precision - within-run- between-run
Limit of Quantitation LOQ
Recovery
30References
- Guidance for IndustryBioanalytical Method
Validation FDA, May 2001 - Workshop Report Shah, V.P. Et al.,
Pharmaceutical Research 1992 9588-592. - Workshop Report Shah, V.P. et al.,
Pharmaceutical Research 2000 171551-1557
31Costs
- Validation 130-180.000 SEK
- Stability 15-20.000 SEK for each time point
- QA 11.000 SEK/study
32The study phase (1)
- ...in which the validated bioanalytical method is
applied to the actual analysis of samples from
the biostudy mainly in order to confirm the
stability, accuracy and precision.
33The study phase (2)
- Calibration curve in each run
- Six Quality Control samples in each run
- Pre-stablished SOPs for procedures (method)
- Acceptance criteria for a run- accuracy and
precision of the calibration curve- accuracy and
precision of the QC samples- repeat analysis - It is preferable to analyse all study samples
from a subject in a single run
34The study phase (3)
- The QC samples should be used to accept or reject
the run (2 samples at 3 levels) - Four QC samples out of six should be within 15
of their nominal value - Two QC samples can be outside 15 but not both
at the same concentration
35System suitability test
The lowest calibration sample is injected before
each run. The system is accepted if
- Signal to noise ratio is above 5 for the
substance. - The peak shape is acceptable after visual
inspection of the chromatogram - The retention times are within 10 of the
previous run.
36The analytical report should include
- Results for all calibration curves
- Results for all quality control samples
- Representative number of chromatograms
- Should include data from subjects who eventually
dropped-out - Reanalysed samples and the reason for reanalyses
- The analytical validation report
- The responsible investigator(s) should sign for
their respective section of the report
37Chiral active substances
- The bio-analytical method should be enantiomeric
- Unless
- Both products contain the same stable singel
enantiomer - Both products contain the racemate and both
enantiomers show linear pharmacokinetics
38Also guidance for
- Reference and test product
- Data analysis
- In vitro dissolution comlementary to a
bioequivalence study - Reporting of results
- Application for products containing new active
substances - Application for products containing approved
active substances -
- is given