Title: Use of Ceramics in Total Hip Arthroplasty
1Use of Ceramics in Total Hip Arthroplasty
Begin
Arthur Mui, Cole Barthel, Kaizhen Chen
BME 215 Fall 2007
Duke University
2Navigational Guide
- This is an educational presentation about
ceramic hip implants. Please refer to the
following instructions for easy navigation of
this powerpoint. - Clicking on this button brings the
reader back to the main menu - Clicking on this button brings the
reader back to the sub menu - Clicking on this button brings the
reader to the previous slide - Clicking on this button brings the
reader to the next slide - Clicking on this button brings the
reader to a glossary of definitions of
medical terms
Learn More
Clicking on this button brings the reader to more
information on the topic
3Start by clicking on Why
ceramics? Then click on any of the four images
below to learn more
Click here to see references and surgeon and
vendor contacts
DISEASED STATE
DESIGN AND MATERIALS
SURGICAL IMPLEMENTATION
IMPLANT PERFORMANCE
4Why Ceramics?
Excellent Biocompatibility
Good Lubrication Properties
Abrasion Resistance
High Wear Resistance
Long Lifetime
Superior Corrosion Resistance
5Diseased State
Hip Anatomy
History of Hip Implants
Functions of the Hip
Forces on the Hip
Diseases of the Hip
Motivation for Replacement
Criteria for Hip Replacement
Chapter Summary
6History and Development of Ceramic Implants
Click on each button to learn more about a
particular time period
Second generation ceramics, zirconia introduced
Third generation ceramics
1970s
2003
1980s
Today
Use of alumina in total hip arthroplasty
FDA approved ceramic- on-ceramic hip joints
71970s
- The use of ceramic components in joint
replacement surgery was initiated in the 1970s
with the introduction of first generation alumina
products, when ceramics superior resistance to
wear in comparison to more traditional metal and
polyethylene materials became apparent. - The use of alumina ceramics in total hip
arthroplasty began in Europe. Professor Pierre
Boutin pioneered the use of ceramics in France in
1970, replacing traditional metal femoral heads
with alumina. However, early ceramic components
were prone to fracture as they had low densities
and a coarse microstructure.
http//www.ceramtec.com/index/divisions/medical_pr
oducts/ medical_professionals/history/04020,0125,0
341,4011.php
81980s
- Advances in material quality and processing
techniques and a better understanding of ceramic
design led to the introduction of second
generation alumina components in the 1980s that
offered even better performance than early
systems. The fracture rate of ceramic Biolox
femoral heads was 0.026 for first generation
alumina, 0.014 for second-generation alumina,
and 0.004 for femoral heads manufactured after
1994. -
- In 1985, the first zirconia femoral heads were
implanted.
92003
- The FDA approved the use of ceramic-on-ceramic
hip implants in the February of 2003. The first
ceramic-on-ceramic hip implant was implanted in
legendary golfer Jack Nicklaus in 1999 as part of
Strykers clinical study.
www.belfasttelegraph.co.uk/sport/article24642...
10Today
- Todays third generation alumina ceramic
material has a high density with a small grain
size, high chemical purity and a stable
crystalline structure. Additionally, the material
is hot isostatically pressed which decreases
grain size and in turn, the incidence of
fracture. Lubrication properties are also
improved with the decrease in grain size.
Individual ceramic components are also
extensively tested before they are shipped and
sold to the surgeons, which greatly decreases
risks of failure.
11Anatomy of a Normal Hip Joint
http//www.mylifeinaction.com/hip/arthritisandyour
hip/images/Normal_hip.gif
12Pelvis
- The bony pelvis is located at the base of the
spine and provides a strong and stable support
for the vertebral column and pelvic organs. It
consists of 2 large bones separated from each
other in the back by the sacrum, an extension of
the spinal column. The pelvis also accepts the
bones of the lower limb, connecting them to the
axial skeleton.
www.pdh-odp.co.uk/pelvis.htm
13Acetabulum
- The acetabulum is a deep fossa formed by the
ilium, ischium and pubis. It functions as the
socket that accepts the rounded head of the
femur. Together, the femoral head and the
acetabulum form the hip joint, which is
classified as a ball-and-socket joint. A
fibrocartilage lip called the labrum, which
extends around the rim of the cup-shaped cavity,
increases the depth of the acetabulum for
articulation with the femoral head and
contributes to hip joint stability. The
acetabular cavity is enclosed with a joint
capsule with a smooth synovial lining for
lubrication.
http//www.ori.org.au/bonejoint/hip/goodbad.htm
14Femur
- The femur is commonly called the thigh bone. The
upper side of the femur tapers into a neck that
is topped with a ball-shaped head capped with a
shiny, slippery tissue layer of articular
cartilage. The head of the femur fits snugly into
the cartilage-lined acetabulum, with just enough
space to rotate smoothly.
http//www.britannica.com/eb/art-101308/
15Function of the Hip
- Approximately 3 times the body weight is
distributed throughout the hip with routine
activities due to the muscle pull and joint
forces that occur. The range of movements of the
hip joint include
Flexion / Extension
Lateral / Medial Rotation
Abduction / Adduction
co-me.ch/projects/phase2/p07/p07_03.en.html
16Flexion / Extension
www.brianmac.co.uk/musrom.htm
17Lateral / Medial Rotation
www.brianmac.co.uk/musrom.htm
18Adduction / Abduction
www.brianmac.co.uk/musrom.htm
19Forces on the Hip
This movie shows the multi-directional forces
experienced at the hip joint during motions like
walking and sitting.
www.lifemodeler.com/LM_Manual/T_hip.htm
20Forces on the Hip (cont.)
This finite element analysis (FEM) model shows
the stress distribution in a ceramic ball of the
hip implant.
www.endolab.de/computer/computersimulation_e.htm
21Diseases Affecting the Hip
Osteoarthritis
Rheumatoid Arthritis
Avascular Necrosis
Hip Dysplasia
Post-traumatic Arthritis
- Intense chronic pain is often experienced, with
impairment of day-to-day activities like running
or walking. Total hip replacement is usually
considered only if normal functions are still
impaired even with the help of anti-inflammatory
drugs or pain medication.
22Osteoarthritis
- Progressive degenerative joint disease
- Most common cause for hip replacement surgery
- Often a result of aging, a congenital abnormality
of the hip joint or trauma - Cartilage in the hip joint becomes cracked and
pitted - Movement of the joint becomes difficult and
painful
http//www.highlands-ortho.com/degenhip.gif
23Osteoarthritis
- Cartilage between the femoral head and the
acetabulum progressively wears away - Bones rub against each other, resulting in severe
pain, deformity and loss of mobility - Bone spurs or osteophytes may develop,
resulting in severe pain and limitation in
mobility
http//www.oagb.net/education/hipresurface/images/
hip_compare.jpg
24Rheumatoid Arthritis
- Chronic, inflammatory autoimmune disorder
- Inflammation and soft tissue swelling
- Chemicals produced in the joint space cause it to
become thickened and inflamed - Synovial fluid destroys the cartilage, leading to
cartilage loss, pain and stiffness
http//www.gaortho.com/Portals/0/RheumatoidArthrit
isHip.jpg
25Post-traumatic Arthritis
- Fractures to the hip bones can result from trauma
such as a fall or blow to the hip - Post-traumatic arthritis may develop after joint
injury if the bone and cartilage do not heal
properly - The joint is no longer smooth, which leads to
increased wear on the joint surfaces, resulting
in severe pain
http//medicalimages.allrefer.com/large/hip-fractu
re.jpg
26Avascular Necrosis
- Fractures or dislocations of the hip can result
in avascular necrosis if the arteries supplying
this area are damaged - It can also arise due to blockage of the blood
vessels caused by agents such as sickle cell
anemia, abnormal red blood cells and fat
particles - Alcoholism and steroids also increase risks of
avascular necrosis - Without blood, bone tissue dies and collapses,
leading to collapse of the joint surface
http//www.mylifeinaction.com/hip/arthritisandyour
hip/
27Hip Dysplasia
- Abnormal formation of the hip joint in which the
femoral head is not stable in the acetabulum - Refers to a hip that is subluxatable
- or dislocatable
- Acetabular dysplasia - acetabulum is shallow,
rendering the hip joint unstable - Developmental dysplasia - child has apparently
normal hips at birth, but develops problems in
his/her first year of life
www.msnyuhealth.org/.../body_ganz_osteotomy.html
28Motivation for Total Hip Replacement
One of the most reliable operations in
orthopedic surgery
Allows patient participation in gentle leisure
activities
Improves quality of life, mobility and
independence
Vastly reduces or eliminates pain of disease
www.walgreens.com/library/contents.html?docty...
29Criteria for Hip Replacement
- Pain experienced is enough to restrict work,
recreation and ordinary daily activities - Pain is not relieved with the use of
anti-inflammatory medicine - Patients X-ray shows advanced arthritis, or
other problems like avascular necrosis - Age, overall health and bone density are also
important factors to be considered before the
operation
orthopedics.about.com/.../a/arthritis_2.htm
30Summary Diseased State
- The hip joint is the joint between the femur and
the acetabulum of the pelvis - Its range of motion include abduction/adduction,
lateral and medial rotation and flexion/extension - Some diseases affecting the hip include
osteoarthritis, rheumatoid arthritis, avascular
necrosis, hip dysplasia and post-traumatic
arthritis - Total hip replacement is only considered if the
pain is debilitating enough to interfere with
daily activities, and if pain persists despite
the use of medication - Total hip replacement vastly improves the quality
of life and mobility of patients
www.dreamscape.com/cnytc/ELA.html
31Design and Materials
Overall Schematic
Commonly Used Materials
Ceramic-Ceramic Joint
Materials Comparison
Chapter Summary
32Design Overall Schematic
- There are 4 components to a hip replacement
prosthesis - 1) The Acetabular Shell2) The Cup3) The
Ball4) The Stem - While the stem is always made out of metal, there
are 4 commonly used combinations for the cup and
the ball - 1) Metal Ball on Polyethylene Cup
- 2) Ceramic Ball on Polyethylene Cup
- 3) Metal Ball on Metal Cup
- 4) Ceramic Ball on Ceramic Cup
- This section will focus on the design of the
ceramic on ceramic joint
Illustration of total hip prosthesis (click for
details)
Novartis Ceramic Hip Brochure
33Illustration of a Total Hip Prosthesis
- The following illustration shows all the parts in
a total hip prosthesis
Stryker Ceramic Hip Brochure
34Design Ceramic-Ceramic Joint
- There are a few distinctive advantages of a
ceramic-ceramic joint - Low Wear
- Low Friction
- High Hardness / Low Grain Size
- High Lubrication
- High Biocompatibility
- There are a few concerns to address when using a
ceramic-ceramic joint - Brittleness / Fracture
- Cost / Patient Factors
- Dislocation is a major problem that occurs in all
hip joints. The chance of dislocation is related
to - Jump Distance
- Range of Motion
- Diameter of Femoral Head
35Wear
- Wear of prosthetic joints is a significant
problem, because the debris produced by wear can
cause adverse tissue reactions that may lead to
inflammation and massive bone loss around the
implant and consequently loosen the implant - The following are wear rates of joints made of
different materials. As shown, the
ceramic-ceramic joint has the lowest wear rate
ceramic-ceramic1 microns / year
metal-metal4.2 microns / year
ceramic-polyethylene20 microns / year
metal-polyethylene220 microns / year
36Wear (cont.)
- A ceramic-ceramic joint has a wear rate of
approximately 1 micron / year, which is up to 200
times less than the rate of wear of a
metal-polyethylene joint - Although a metal-metal joint has similar rates of
wear, there are concerns that this coupling can
result in potentially harmful concentrations of
metal ions being released into the bloodstream - High concentrations of metal ions have been
suspected to cause kidney and liver diseases
37Friction
- The friction coefficient of the ceramic-ceramic
combination is 0.09 compared to 0.21 with
metal-polyethylene coupling - Using 1000 N alternate loading with water
lubrication, an initial running-in-wear is
observed due to a self-polishing effect of the 32
mm diameter ceramic head against the ceramic cup
followed by a decease of the friction torque to a
steady state of 0.6 Nm
Frictional torque versus time for three materials
combinations
38Hardness / Grain Size
- The hardness of alumina is second only to a
diamond (Alumina has a microhardness of 23,000
Vickers) - Ceramics small grain size and grain distribution
also increases its scratch resistance. Ceramic
grains lie between 1.5 5 microns
Illustration of Microstructure of Alumina (x1000
magnification)
Illustration of Grain Size Distribution
39Hardness / Grain Size (cont.)
- Comparatively, metal-metal joints can be
scratched causing an abrasive surface. Foreign
debris in the joint may also accelerate implant
wear - A minor drawback of ceramic-ceramic joints is
that they are reported to make a squeaking noise
occasionally. Nonetheless, the squeaking noise
does not affect the implant function
40Lubrication
- The high density of ceramics allow a very smooth
surface finish - The surface roughness of alumina (Ra) is 0.02 µm,
which is superior to any metallic finish available
Trident Ceramic Acetabular System Brochure
41Lubrication (cont.)
- The hydrophilic nature of alumina also affords
better lubrication in an aqueous environment. The
contact angle of water is 44 degrees for Alumina,
72 degrees for 316 L Stainless Steel, 80 degrees
for UHMWPE, and 87 degrees for Cr-Co alloys - A lower contact angle promotes better
lubrication, which in turn produces less wear
particles
42Biocompatibility
- Alumina, the most commonly used ceramic material,
is known to be very biocompatible. Laboratory
analysis has shown that ceramic debris may be
better tolerated by the body - Clinical studies have also shown that the alumina
surface is completely covered with protein
molecules immediately after implantation, and as
a result, the body does not recognize it as
foreign
43Biocompatibility (cont.)
- Tissue biopsies show that in well fixed alumina
(ceramic-ceramic) prostheses, alumina wear
particles increased with time however, only a
low macrophagic reaction has occurred, and no
necrosis (death of tissue) was observable - Tissue biopsies from metal-polyethylene total hip
prostheses showed that polyethylene wear
particles elicited a stronger body reaction when
compared to alumina particles. Polyethylene
particles were bigger, with numerous giant
macrophages directly in contact
44Brittleness and Fracture
- Back in 1974, as many as 1 in 300 ceramic
components fractured however, increased ceramic
quality and advanced cup design have reduced that
to as low as 1 in 20,000 today - The first generation ceramic joints sometimes
failed as a result of manufacturing defects,
causing the implants to crack and shatter.
Furthermore, the grain size was 40x larger than
that of modern ceramics, which have grain sizes
on the order of 1um
Fracture of Ceramic Heads in Total Hip
Replacement by B. Habermann
45Brittleness and Fracture (cont.)
- However, recent technological advancements have
led to the manufacture of alumina ceramic
components with reduced grain size, fewer
inclusions and limited grain boundaries, which in
turn leading to a much tougher and stronger
material
46Brittleness and Fracture (cont.)
- One of the most common causes of fracture results
from impingement of the femoral stem neck and the
cup with the hip in hyperflexion and external
rotation. Consequently, the cup becomes loose and
dislocates, resulting in a large force exerted on
the head and taper, bending the taper and
fracturing the head - Strykers developed a solution to this problem by
adding an elevating sleeve rim between the cup
and the shell. Such a design shifts the
impingement point from the stem neck and the cup
to the stem neck and shell, which has a lower
chance of damaging the ceramic insert
Illustration of the old design
Strykers New Design
Trident Ceramic Acetabular System Brochure
47Brittleness and Fracture (cont.)
- When the cup is vertically tilted or becomes
vertically tilted after loosening, the contact
stresses become high enough to generate
subcritical cracks in the ceramic. Therefore the
material will have an increased chance of
fracturing
Illustration of contact stress distribution based
on cup orientation The left diagram illustrates
a normal cup and the right diagram illustrates a
dislocated cup
48Cost / Patient Factors
- Ceramic on ceramic implant configurations command
premium prices, 7500 for implant components as
compared to 5500 for metal on metal components
- Patient Factors When choosing a material for a
hip joint, sex, age, and diagnosis should be
taken into consideration. Males and younger
patients demonstrate higher average activity, and
thus a higher risk for problems related to wear,
such as osteolysis. Categorically, patients with
medical comorbidities that limit activity
demonstrate lower average wear rates and have a
reduced risk for failure due to wear and
osteolysis
49Jump Distance
- Jump distance is the distance a femoral head must
travel to dislocate. The greater the jump
distance, the less risk of dislocation - Typically, jump distance is defined as the
vertical distance the femoral head must travel to
dislocate after impingement. For the Stryker
designed head, it incorporates an additional
dislocation safety factor of 2.7 mm, which
decreases the risk of dislocation - The following are the jump distances
corresponding to different head diameters
Trident Ceramic Acetabular System Brochure
50Jump Distance (cont.)
- The jump distance measurement is more complicated
than it seems since the implant cup is usually
embedded 45 degrees into the pelvis - The formula used for the vertical jump distance
is r - rcos(45) 2sin(45) - The following illustrations show the jump
distance calculation in greater detail
Trident Ceramic Acetabular System Brochure
51Range of Motion
- Range of motion is critical for the patient
because it enhances optimal movement and activity
post-operatively - Clinical studies have shown that a greater range
of motion was observed for larger heads compared
with smaller ones
Trident Ceramic Acetabular System Brochure
52Diameter of Femoral Head
- Typical ceramic ball head sizes range from 28 mm
to 36 mm in diameter - While a larger femoral head diameter is known to
provide a more favorable jump distance and range
of motion, it also reduces the chance of fracture
due to better distribution of stress - An area of concern is that in some laboratory
studies, the range of motion prior to neck-socket
impingement leading to dislocation was increased
as ball diameter increased. Moreover, very large
diameter bearings, 40 mm or greater, have been
used to stabilize hips with a history of
recurrent instability
www.devicelink.com/mtm/archive/07/05/005.html
53Materials Commonly Used for Balls and Cups
Click on a material to learn more
Titanium
UHMWPE
Alumina
Alumina/Zirconia
54Titanium
- Pros
- High biocompatibility
- High strength
- Ductile
- Cons
- Undesirable wear rate
Learn More
Learn More
Learn More
Learn More
Uses Popular orthopedic implant material (hips,
knees, shoulders, etc.) First material
used for hip balls and still a very popular
option Ball/Cup Pairs Titanium ball with UHMWPE
cup
Titanium Ball
www.onlinetmd.com
55High Biocompatibility of Titanium
- Titanium has extremely low corrosion rates making
it ideal for implantation - Titanium is also bio-inert which means that it
will not react with the surrounding environment
www.timet.com/specialized.html
56High Strength of Titanium
- A hip joint can be loaded with multiple times
body weight depending on the activity. As such,
the materials used must be of high strength. - Titanium is a very strong metal (also used in
aircraft parts) and can support much higher loads
than what is seen in a hip joint
www.tirings.com/titanium_facts.php
57Ductility of Titanium
- As with all metals titaniums crystal structure
is held together with metallic bonds - These bonds enable the atoms in the crystal to
slide past each other when subjected to stress.
This creates a ductile material that will deform
before it fractures - Ductility is important because it greatly reduces
the risk of fracture
Titanium Unit Crystal each atom is held to the
other through sharing of electrons (metallic
bonds)
www.msm.cam.ac.uk
58Titanium Wear
- The production of wear particles is a major issue
because they can cause osteolysis (resorption of
bone around the implant) - Wear occurs at the interface between the hip ball
and acetabular cup due to the rubbing motion
created during normal movement - Wear particles are responsible for most titanium
artificial hip failures
Stryker Orthopedics Brochure
59UHMWPE (Ultra High Molecular Weight Polyethylene)
- Pros
- Low friction/hard
- Biocompatible
- Cons
- Undesirable wear rate
Learn More
Learn More
Learn More
Uses Low friction bearing surface in
manufacturing and biomedical applications
Used exclusively as an acetabular cup in
artificial hip joints Ball/Cup Pairs Titanium or
Alumina/Zirconia ball with UHMWPE cup
Hip and Knee bearing surfaces
www.disanto.com, www.onlinetmd.com
60Hardness and Low Friction of UHMWPE
- UHMWPE has a low friction coefficient and is very
hard making it an ideal bearing surface for the
articulation between the ball and the cup.
www.ticona.com
61Biocompatibility of UHMWPE
- UHMWPE consists of very large molecules and there
are many cross-links in its chemical structure.
Therefore, the material does not react with the
surrounding biological environment.
www.msm.cam.ac.uk, http//www.zimmer.co.za/z/ctl/o
p/global/action/1/id/7867/template/MP/prcat/M2/pro
d/y
62Wear of UHMWPE
- The wear of a joint is dependent on the
combination of the ball and acetabular
components. In Titanium-UHMWPE combinations the
wear rate (220 microns/year) is a major issue.
However, in Alumina/Zirconia-UHMWPE joints the
wear rate is greatly reduced.
Stryker Orthopedics Brochure
63Alumina
- Cons
- Susceptible to cracking / slow crack propagation
- Pros
- Less wear
- High strength
- Excellent corrosion resistance
- Good biocompatibility
- Low friction coefficient
Learn More
Learn More
Learn More
Learn More
Learn More
Learn More
Uses One common use of alumina is in billiard
balls Introduced as an artificial hip
component in 1972 in an effort to reduce the wear
produced from titanium/UHMWPE
pairs Ball / Cup Pairs Alumina balls are almost
always paired with alumina cups Learn More
Manufacturing alumina Crystal Structure
Learn More
Learn More
64Wear of Alumina
- Alumina on alumina joints (1 micron/year) have a
much lower wear rate than that of titanium on
UHMWPE joints (220 microns/year)
Stryker Orthopedics Brochure
65High Strength of Alumina
- Alumina is an extremely strong material. It has
a tensile strength and tensile modulus greater
than even titanium.
66Corrosion Resistance of Alumina
- As with most ceramics alumina practically does
not corrode at all. This increases its
biocompatibility and means that it will not
interact with the biological fluids surrounding
it.
Stryker Orthopedics Brochure
67Biocompatibility of Alumina
- Because alumina has a lower wear rate it is
considered to be more biocompatible than titanium
on UHMWPE joints - It is also believed that the particles produced
from alumina have less of a biological reaction
than those from titanium
68Low Friction Coefficient of Alumina
- Alumina has a very low friction coefficient which
makes it an excellent choice as a bearing
material - This also helps to reduce the amount of wear
69Fracture Toughness of Alumina
- One problem with ceramics in general is that they
are brittle materials and therefore susceptible
to fracture if stresses are high enough. Alumina
is also susceptible to slow crack propagation in
which a small defect can grow through continued
loading until it is large enough that the
material will fracture. -
www.cases.bham.ac.uk/bio/femoral.htm
70Manufacture of Alumina
- Alumina is most commonly manufactured using
powder processing techniques. The alumina powder
is isostatically compressed and then shaped by
grinding while being sintered at 1600 to 1800oC.
The process produces a nearly pore-free
polycrystalline solid with grain sizes of 3 to
5µm. - The techniques are used partly because the
surface of alumina must be highly controlled down
to the sub-micron level. Because alumina is
susceptible to cracking any microcracks in the
surface must be either eliminated or greatly
reduced. The grain size must also be controlled
in order to give the best fracture toughness.
71Manufacture of Alumina
Click on any process to learn more
72Alumina Powder
- The ore bauxite, which consists of Al2O3, Fe2O3
and SiO2 is purified using the Bayer process to
achieve aluminum powder. - The powder is used to make aluminum metal and can
be used as an abrasive
www.ltdceramics.com
73Cold Isostatic Pressing
- The alumina powder is poured into a mold and
compressed from all sides using static pressure
from a fluid - This compacts the powder and allows manufacture
of highly controlled shapes
205.209.102.103
74Sintering
- Sintering is the process of heating a material to
below its melting temperature until the particles
begin to adhere to one another - Alumina is pre-sintered before machining and then
sintered again before use to further set the
material
www.esrf.eu
75CNC Tuning and Taper Cutting
- After pre-sintering, the component is shaped
using CNC (computer numerical control) and other
traditional machining techniques.
http//www.computernumericalcontrol.net/CNC-lathe.
jpg, www.penntoolco.com/catalog/products/products.
...
76Grinding and Polishing
- After sintering at 1600oC, the part is ground and
polished to achieve a smooth surface - This is very important since the part will be
used as a bearing surface.
www.ukam.com/products.htm
77Crystal Structure of Alumina
- Alumina is also known as aluminum oxide and has a
corundum crystal structure (rhombohedral).
Alumina has the chemical formula Al2O3 and in
commercial use it has less than 0.4 other
alcolide metals added such as titanium dioxide
and magnesium. - Aluminas highly organized crystal structure
gives it its strength and toughness but the ionic
bonds that connect the molecules together make it
susceptible to fracture. The molecules cannot
slide past one another as in metallic bonds so
when stresses are high enough the molecules break
apart from one another.
Aluminum Oxide Unit CellGrey Aluminum, Red -
Oxygen
videoinside.org/show/Aluminium_oxide
78Alumina / Zirconia
- Pros
- Increased hardness / fracture strength
- Cons
- More difficult to manufacture
Learn More
Learn More
Uses Zirconia was introduced as a hip joint
material because of its higher strength
and toughness compared to alumina
It was discovered later that the zirconia
components had a high failure rates
Alumina/zirconia parts combine the properties of
both alumina and zirconia and the
failure rate is much reduced. Ball/Cup Pairs
Alumina/zirconia balls are always paired with
UHMWPE cups Learn More Alumina/zirconia material
properties
79Fracture Strength of Alumina / Zirconia
- By adding relatively small amounts of zirconia
(ZrO2) to alumina (Al2O3) the hardness and
fracture strength of the material can be
increased. - Zirconia derives its fracture strength from its
crystal structure. If zirconia is manufactured
correctly it is in a stable tetragonal state. In
order for the material to fracture it must
transition into a monoclinic state. The energy it
takes for this process to happen increases the
resistance to cracking.
www.chemistry.pomona.edu/.../Zirconium.htm
80Manufacture of Alumina / Zirconia
- Alumina / Zirconia composites are manufactured
using the same powder processing techniques as
pure alumina, but the manufacturing of these
composites must be controlled even more. If the
grain size of the zirconia grains are too large
or too small and if the percentage of zirconia is
too large the zirconia can transition to
monoclinic when the material is cooled, which can
cause cracking. Commercial alumina/zirconia
composites have 25 zirconia.
81Comparison of Material Properties
- Some notable comparisons
- The hardness increases from titanium to the
ceramic materials (zirconia hardness is slightly
higher than alumina) - The ceramic materials have very high compressive
yield strengths and flexural strengths - The fracture toughness of zirconia is slightly
higher than that of alumina
82Summary Design and Materials
- The hip implant consists of 4 parts the
acetabular shell, the ball, the cup and the stem - Commonly used materials for the ball and cup
include titanium, UHMWPE, alumina and
alumina/zirconia - Ceramic-on-ceramic joints have the lowest
recorded wear rate among all the possible
couplings between ball and cup - Alumina has excellent biocompatibility, high
corrosion resistance, a low friction coefficient
and low wear rate but is susceptible to cracking
and slow crack propagation
www.dreamscape.com/cnytc/ELA.html
83Surgical Implementation
Hip Replacement Surgery
Pre-Operative Procedures
Surgical Protocol
Hip Implant Sterilization
Rehabilitation
Chapter Summary
84Hip Replacement Surgery
- Removes diseased femoral head and damaged
cartilage from the hip socket - Femoral head is replaced by a ball fixed to a
stem - Stem is inserted into the hollow part of the
femur - Socket is replaced with an acetabular shell with
a lining cup
www.memagazine.org/.../hipnew/hipnew.html
85Hip Replacement Surgery (cont.)
- Femoral and acetabular components of the
prosthesis are attached to the bone by creating a
space slightly smaller than the prosthesis and
then pushing it into this tight space - The prosthesis may be attached to the bone by
bone cement - Implant components are pre-sterilized prior to
usage, usually by gamma radiation in an air
environment, with doses ranging from 2.5 - 4 MRad
www.dartmouth.edu/.../corr5.html
86Hip Replacement Surgery (cont.)
All pictures from Strykers Surgical Protocol
Brochure
87Pre-Operative Procedures
- Pre-operative planning and X-ray evaluation aids
in selection of the most favorable implant style
and optimal size for the patients hip pathology.
- Selecting potential implant styles and sizes
ahead of time facilitates operating room
preparation and assures availability of an
appropriate selection. - X-ray detection helps detect anatomic
abnormalities that could prevent successful
implantation.
88Sterilization of Implant
Hip replacement components come to the surgeon
packaged and pre-sterilized Sterilization
techniques
UHMWPE gas plasma
Ceramics Metals gamma radiation
Other popular sterilization techniques Ethylene
Oxide, Heat
89Gamma Radiation
- Cons
- More expensive than ethylene oxide
- Not an in-house operation
- Pros
- Rapid
- No residues
- Good for convoluted shapes
- Penetrates the material well
- Easy to control the dosage
- Can be sterilized in its packaging
How it works The bacterias DNA is degraded
through ionization
How its done The part is passed by radiating
rods on a conveyer belt
www.medicaldesign.com, www.medicaldesign.com
90Ethylene Oxide
- Pros
- Compatibility with many materials
- Good for heat sensitive products
- Low cost
- Does not compromise the mechanical properties of
UHMWPE the way radiation does
- Cons
- Penetration can be difficult
- Residues are present
- Long process
How it works Alkylates proteins and DNA
How its done EO gas is introduced to a chamber
around the part for a specific amount of time and
then flushed away.
img.trade.tootoo.com
91Heat (Autoclaving)
- Pros
- Fast
- No residues
- Very low cost
- Can be used in-house
- Cons
- Heat sensitive parts like many polymers can not
be sterilized this way
How it works Oxidizes and denatures enzymes
How its done The part is placed in a very hot
wet environment for a specific amount of time
www.fragoimpex.com
92Gas Plasma
- Cons
- A new process that does not have in-vivo data to
back it up
- Pros
- Does not compromise the mechanical properties of
UHMWPE the way radiation does - Can be used for materials sensitive to
temperature, radiation, and chemicals
How it works Uses ionized gas to deactivate
bacteria on the surface
How its done The gas passes over the part and
two electrodes create the plasma that kills the
bacteria
www.torontosurplus.com
93Surgical Protocol
- The hip is dislocated and the femoral head is
removed - Soft tissue from the acetabulum is removed to
gain adequate exposure for reaming - Excision of labrum and osteophytes
www.elib.gov.ph/edatabase/elibgetdb.php/http/...,
www.rcsed.ac.uk/journal/svol1_6/10600004.html
94Surgical Protocol (cont.)
- Reamer handle is oriented at 45o of abduction
- Reaming progresses in 1mm increments until final
sizing is achieved - Care is taken not to enlarge or distort the
acetabulum by eccentric reaming - Final state ideally shows the hemispherical
acetabulum denuded of cartilage with the
subchondral plate intact and anterior acetabular
wall preserved
www.elib.gov.ph/edatabase/elibgetdb.php/http/...,
www.rcsed.ac.uk/journal/svol1_6/10600004.html
95Surgical Protocol (cont.)
- Appropriately sized acetabular component is
selected - The metal shell is impacted into the acetabulum
using a mallet until a tight fit is achieved - Screws may be used to secure the shell
- Ceramic/polyethylene insert is carefully
introduced - Insert is turned into final pre-locking position
- Insert is seated firmly by firm mallet blows
www.elib.gov.ph/edatabase/elibgetdb.php/http/...,
www.rcsed.ac.uk/journal/svol1_6/10600004.html
96Surgical Protocol (cont.)
- The hollow center portion of the femur is cleaned
and enlarged - A cavity matching the shape of the implant stem
is created - Top end of femur is planed and smoothed so that
stem can be inserted flush with the bone surface - Stem is impacted into place using the mallet
www.elib.gov.ph/edatabase/elibgetdb.php/http/...,
www.rcsed.ac.uk/journal/svol1_6/10600004.html
97Rehabilitation Process
- Physical therapy and exercises to strengthen the
artificial hip - Aided walking with crutches for up to 4-6 weeks
- Use of cane for another 4-6 weeks before being
able to walk unaided - High impact sports like running and aerobics are
strongly not recommended - Full recovery varies from 3-6 months
www.hipsandknees.com/hip/bhr/rehab.htm
98Summary Surgical Implementation
- Total hip replacement involves replacing the
diseased femoral head with a ball fixed to a
stem, and the socket with an acetabular shell
with a lining cup - During surgery, soft tissue from the acetabulum
is removed in preparation for reaming, and the
acetabular shell is then impacted into the cavity
using a mallet - The center portion of the femur is removed for
insertion of the femoral stem - Sterilization of ceramic hip implants is
commonly done using gamma radiation and ethylene
oxide - The rehabilitation process involves physical
therapy and use of a cane or stick until full
recovery at 3-6 months
www.dreamscape.com/cnytc/ELA.html
99Implant Performance
Clinical Trial 1
Implant Performance
Clinical Trials Summary
Clinical Trial 2
Failure Modes
Chapter Summary
100Implant Performance
- The best way to evaluate the performance of an
implant is to collect data through clinical
trials - We will look into the results of 2 FDA approved
clinical trials on a cementless
ceramic-on-ceramic total hip arthroplasty design
101Methods of Clinical Trial 1
- Objective
- Analyze the failure of a ceramic-on-ceramic hip
implant - Time frame
- Oct 1999 (rolling participation) to Apr 2005
- Samples
- 282 patients (315 hips) are treated with a
ceramic-on-ceramic implant - Median weight of patients was 83.4 kg
- Clinical data
- Harris hip score (indicates range of motion)
- Radiographs
- Retrieval analysis (on failed explanted implants
only) - Definition of failure
- Failure was defined as fracture or displacement
of the ceramic liner
Trial Results
- Robert A. Poggie, Thomas R. Turgeon and Richard
D. Coutts, Failure Analysis of a Ceramic Bearing
Acetabular Component. The Journal of Bone
Joint Surgery. 2007.
102Results of Clinical Trial 1
- 14 out of 315 (4.4) ceramic-on-ceramic hips
failed - Time to failure ranged from 8 to 42 months (avg
25 months) - Harris hip scores indicated that the patients
with and without failure were quite active with
an essentially unrestricted range of hip motion - Retrieval analysis demonstrated stripe and rim
wear with evidence of adhesive wear, indicating a
potentially high-friction interaction at the
articulation
Results Analysis
Comparison Table
Retrieval Analysis
103Comparison of Failed vs. Non-Failed Implant
104Analysis of Clinical Trial 1
- Based on the results, the biggest differential
factor between failed and non-failed implants is
the weight of the patient - Median weight of the patients with a failed
implant was 102.5 kg, while median weight of the
patients with no failure was 83.4 kg - Based on the results, patients with a body weight
of gt91 kg had a 4.76 times greater odds of
failure - No significant association was found between
failure and age, range of motion, acetabular cup
size, stem size, stem type, or cup abduction
angle
105Retrieval Analysis
- Among the 14 failed implants, 12 of the bearing
surfaces were found to have fractured, the
remaining 2 implants were dislocated - Scanning electron microscopy identified abrasion
of the ceramic surface with grain pull-out within
the striped wear area of the ceramic heads
106Methods of the Clinical Trial 2
- Objective
- Observe the performance of a ceramic-on-ceramic
hip implant - Time frame
- Nov 1997 (rolling participation) to 2005
- Samples
- 79 patients (93 hips) are treated with a
ceramic-on-ceramic implant - Mean weight of patients was 63.7 kg
- Clinical data
- Harris hip score (indicates range of motion)
- Radiographs
- Minimum duration of follow-up of five years
- Definition of failure
- Failure was defined as fracture or displacement
of the ceramic liner
Trial Results
- Jeong Joon Yoon, young-Min Kim, et al.
Alumina-on-Alumina Total Hip Artroplasty. A
Five-Year Minimum Follow-up Study. The Journal
of Bone Joint Surgery. 2005.
107Results of Clinical Trial 2
- 1 out of 93 (1.0) ceramic-on-ceramic hips failed
- The failed hip in the patient was observed after
a motor cycle accident - The mean Harris hip score was 97 points at the
time of the latest follow-up evaluation - Ceramic wear was not detectable in the
thirty-seven hips in which the femoral head could
be differentiated from the cup on radiographs
108Summary of Clinical Trials
- The findings in both studies indicated that
contemporary ceramic-on-ceramic hip
arthroplasties performed with use of a cementless
stem are associated with excellent clinical
results and implant stability at five years - It was discovered that higher body weight of
patient leads to a higher chance of implant
failure - The relatively higher failure rate in Clinical
Trial 1 could be attributed to its higher
patient weight - See Journal References for more detail regarding
discussed clinical trials
109Failure Modes of Ceramic Hips
- Here are some failure modes specific to ceramic
total hips - Impingement
- Fracture
- Loosening, post-operative infection and
dislocation - Stripe Wear
- A possible concern not leading to failure is
- Noises from ceramic components
110Impingement
- In extreme hip joint positions like too much
bending, the neck of the femoral component may
impinge against the rim of the cup component.
www.totaljoints.info/ceramic_total_hips.htm
111Impingement (cont.)
- The edge of the ceramic cup receives a blow with
every such extreme movement. - Eventually, continuing blows fracture the cup
which splinters into many fragments. - The use of computer navigated insertion of
components reduces the risk of faulty positioning
of the cup and diminishes the risk of
impingement. - Patient awareness of risks of extreme hip
movements and adequate surgeon advice also help
to reduce such possibilities of failure.
www.totaljoints.info/ceramic_total_hips.htm
112Fracture of Ceramic Balls
- Fractures of modern ceramic balls are extremely
rare because the modern medical grade ceramic has
very fine structure, produced by HIP (hot
isostatic pressing). - Ceramic components are also individually tested
before use with weights 60x patient body
weights. - The reported fracture rate of modern ceramic
balls is exceedingly small 0.004 or 4 in 100
000.
www.totaljoints.info/ceramic_total_hips.htm
113Fracture of Ceramic Liners
- The fracture often starts off as a failure of the
binding between the polyethylene sleeve and
ceramic liner which is caused by impingement of
the neck against the rim of the liner. - They appear as hair-fine fracture lines at the
beginning and repeated small traumas will
exacerbate the fractures till the liner
eventually splinters and results in patient pain
and discomfort.
www.totaljoints.info/ceramic_total_hips.htm
114Loosening, post-operative infection and
dislocation
- Loosening of ceramic components has been reported
for 0.5 of components in the Lineage System. - Post-operative infection rate has been reported
to be 0.7 with ceramic total hips in the USA
study. This is similar to the rates of
postoperative infection observed in operations
with other total hip systems. - Dislocation rate has been reported for 2.4 of
ABC systems and for 1 of the Transcend and
Lineage systems. The rates of dislocations for
other total hip systems varies from 0 up to 7,
so that the dislocation rates of ceramic hips are
not especially low, but not exceptionally high
either.
www.totaljoints.info/ceramic_total_hips.htm,
www.hrorthopaedics.co.uk/Hip20Replacement.htm
115Stripe Wear
- Stripe wear refers to the long, narrow area of
wear damage observed on some alumina balls in
ceramic total hips. - This results from the line contact between the
head and the edge of the liner. - Another cause of stripe wear is the pistoning of
hip bearings during walking. - Micro-separation of the bearing centers occurs
during the swing phase of normal walking and
subsequent edge loading with the heel strike
causes the stripe.
www.totaljoints.info/ceramic_total_hips.htm,
www.hrorthopaedics.co.uk/Hip20Replacement.htm
116Noises from Ceramic Components
- Many patients have reported clicking or squeaking
noises from their new hips when they change hip
positions. - Noises may be caused by a tendon or scar tissue
streak which glides over the protruding portion
of the new total hip joint. - They may also be caused by small pistoning
movements during walking. - These noises most likely result from faulty
positioning of the ceramic cup.
www.totaljoints.info/ceramic_total_hips.htm,
www.hrorthopaedics.co.uk/Hip20Replacement.htm
117Summary Implant Performance
- Contemporary ceramic-on-ceramic total hip
arthroplasties have produced excellent results
and implant stability at 5 years - Higher body weights in patients may lead to
higher chances of implant fracture - Some of the failure modes of ceramic hips include
fracture of balls and liners, stripe wear,
impingement, implant loosening and post-operative
infection and dislocation
www.dreamscape.com/cnytc/ELA.html
118Glossary of Terms
- Dislocatable Femoral head can come out of the
socket when stressed - Labrum A ring of fibrocartilage around the edge
of an articular surface of a bone - Osteophyte Also known as bone spurs. These are
bony projections that form all joints and are
often seen in conditions like arthritis - Reaming To form, shape, taper, or enlarge (a
hole or bore, for example) with or as if with a
reamer - Subluxatable Unstable when stressed
- Click here to go back to where you were
119Web References
- http//www.spinecarehelp.com/hip-replacement-surge
ry.htm - http//www.medicinenet.com/total_hip_replacement/p
age2.htm - http//www.pdrhealth.com/patient_education/BHG01RH
08.shtml - http//www.oagb.net/education/hipresurface/images/
hip_compare.jpg - http//en.wikipedia.org/wiki/Hip
- http//www.hipsandknees.com/hip/hipanatomy.htm
- www.lasvegaspaininstitute.com/pain.htm
- www.4newjoints.com/Total_Hip.htm
- http//hipreplacement.co.uk/Primary/Advantag.html
- http//www.williambkurtz.com/Hip_Folder/Total20Hi
p20Replacement.htm - http//www.devicelink.com/emdm/archive/06/11/019.h
tml - http//www.coa-aco.org/library/clinical_topics/cer
amic_bearing_surfaces.html - http//www.touchbriefings.com/pdf/1857/hawes.pdf
- http//www.morganadcanvedceramics.com/articles/med
ical_apps.htm - http//www.patentstorm.us/patents/5370694-descript
ion.htm - www.totaljoints.info/ceramic_total_hips.htm
- http//www.mghp.com/services/procedure/hipreplacem
ent.shtml
120Journal References
- R. Nizard, L. Sedel, D. Hannouche, M. Hamadouche,
P. Bizot, Alumina Pairing in Total Hip
Replacement, The Journal of Bone and Joint
Surgery - Trident Ceramic Acetabular System The Path to
Approval Brochure - Matthew Sloanaker, T. Goswani, Review of Wear
Mechanisms in Hip Implants, Science Direct - Thomas P. Schmalzried, M.D.. How I choose a
bearing surface for my patients. The Journal of
Arthroplasty Volume 19, Issue 8, Supplement 1,
December 2004, Pages 50-53 - Samir Sodha, M.D., Johnation P. Garinao, M.D, et
al. Concepts of the Modern Ceramic on Ceramic
Total Hip Arthroplasty and Early Results. UPOJ
Volume 14, Spring 2001, Pages 1-4 - P. Boutin, P. christel, et al. The use of dense
alumina-alumina ceramic combination in total hip
replacement. France, 1988 - Kevin B. Fricka, M.D., Amanda Marshall, M.D., et
al. Constrained Liners in revision Total Hip
Arthroplasty An Overuse Syndrome. The Journal
of Arthroplasty Volume 21, Issue 4, Supplement
1, June 2006, Pages 121-125 - A. Prof Sunil Kumar, Dr Andrew Lewies, The
influence of the surface chemistry of metallic
and ceramic implants wear debris particles on the
cellular response - Robert A. Poggie, Thomas R. Turgeon and Richard
D. Coutts, Failure Analysis of a Ceramic Bearing
Acetabular Component. The Journal of Bone
Joint Surgery. 2007. - Jeong Joon Yoon, young-Min Kim, et al.
Alumina-on-Alumina Total Hip Artroplasty. A
Five-Year Minimum Follow-up Study. The Journal
of Bone Joint Surgery. 2005. - Click here to go back to summary of clinical
trials
121Surgeon and Vendor Contacts
Surgeon Contact Michael Bolognesi, MD Clinical
Faculty, Division of Orthopaedic Surgery Duke
University Email contact michael.bolognesi_at_duke.
edu Vendor Contact Brian Daigle Stryker
Orthopaedics Email contact Brian.Daigle_at_stryker.
com