Title: Atomic UV-Visible Spectroscopy
1Atomic UV-Visible Spectroscopy
Lecture Date January 28th, 2013
2Electronic Spectroscopy
- Spectroscopy involving energy level transitions
of the electrons surrounding an atom or a molecule
Atoms electrons are in hydrogen-like orbitals
(s, p, d, f)
Molecules electrons are in molecular orbitals
(HOMO, LUMO, )
From http//education.jlab.org
(The LUMO of benzene)
(The Bohr model for nitrogen)
3UV-Visible Spectroscopy
- Definition Spectroscopy in the optical
(UV-Visible) range involving electronic energy
levels excited by electromagnetic radiation
(often valence electrons). - Techniques discussed in this lecture are related
to the high-energy (non-optical) methods
covered in the X-ray spectroscopy lecture. - Methods discussed in this lecture
- Atomic absorption
- Atomic emission
- Laser induced breakdown spectroscopy
- Atomic fluorescence
4The Electromagnetic Spectrum
5Elemental Analysis
- Elemental analysis qualitative or quantitative
determination of the elemental composition of a
sample - Atomic UV-visible spectroscopic methods are
heavily used in elemental analysis - Other elemental analysis methods not discussed
here - Mass spectrometry (MS), primarily ICP-MS
- X-ray methods (XRF, SEM/EDXA, Auger spectroscopy,
XPS, etc) - Radiochemical or radioisotope methods
- Classical methods (e.g. color tests, titrations)
6Definitions of Electronic Processes
- Absorption radiation selectively absorbed by
molecules, ions, or atoms, accompanied by their
excitation (or promotion) to a more energetic
state. - Emission radiation produced by excited
molecules, ions, or atoms as they relax to lower
energy levels.
7The Absorption Process
- Electromagnetic radiation travels fastest in a
vacuum - When EM radiation travels through a substance, it
can be slowed by propagation interactions that
do not cause frequency (energy) changes - Absorption does involve frequency/energy changes,
since the energy of EM radiation is transferred
to a substance, usually at specific frequencies
corresponding to natural atomic or molecular
energies - Absorption occurring at optical frequencies
involves low to moderate energy electronic
transitions
c the speed of light (3.00 x 108 m/s) ?i
the velocity of the radiation in the medium in
m/s ni the refractive index at the frequency i
8Absorption and Transmission
- Absorbance
- A -log10 T log10 P0/P
b
A is linear vs. b! (A preferred over T)
Graphs from http//teaching.shu.ac.uk/hwb/chemistr
y/tutorials/molspec/beers1.htm
9The Beer-Lambert Law and Quantitative Analysis
- The Beer-Lambert Law (a.k.a. Beers Law)
- A ebc
- Where the absorbance A has no units, since A
log10 P0 / P - e is the molar absorbtivity with units of L mol-1
cm-1 - b is the path length of the sample in cm
- c is the concentration of the compound in
solution, expressed in mol L-1 (or M, molarity) - Beers law can be derived from a model that
considers infinitesimal portions of a block
absorbing photons in their cross-sections, and
integration over the entire block - Beers law is derived under the assumption that
the fraction of the light absorbed by each thin
cross-section of solution is the same - See pp. 302-303 of Skoog, et al. for details
-
10Deviations From the Beer-Lambert Law
- Deviations from Beers law (i.e. deviations from
the linearity of absorbance vs. concentration)
occur from - Intermolecular interactions at higher
concentrations - Chemical reactions (species having different
spectra) - Peak width/polychromatic radiation
- Beers law is only strictly valid with
single-frequency radiation - Not significant if the bandwidth of the
monochromator is less than 1/10 of the half-width
of the absorption peak at half-height.
For an alternative view, see Bare, William D. A
More Pedagogically Sound Treatment of Beer's
Law A Derivation Based on a Corpuscular-Probabil
ity Model, J. Chem. Educ. 2000, 77, 929.
11Deviations from the Beer-Lambert Law
- Intermolecular interactions at higher
concentrations cause deviations, because the
spectrum changes
Dimers, oligomers
Figure from Chapter 5 of Cazes, Analytical
Instrumentation Handbook 3rd Ed. Marcel-Dekker
2005.
12Deviations from the Beer-Lambert Law
- Deviations caused by use of polychromatic light
on a spectrum in which e changes a lot over the
bandwidth of the light. - Consider two wavelengths a and b with ?a and ?b
? 1000, 1000
? 1500, 500
? 1750, 250
Absorbance (A)
Concentration (M)
13Atomic Emission
- Two types of emission spectra
- Continuum
- Line spectra
- Examples
- ICP-OES (inductively-coupled plasma optical
emission spectroscopy), also known as ICP-AES
(atomic emission spectroscopy) - LIBS (laser-induced breakdown spectroscopy)
14The Emission Process
- Atoms/molecules are driven to excited states (in
this case electronic states), which can relax by
emission of radiation.
M heat ? M
Higher energy
?E hn
Lower energy
- Other process can happen instead of emission,
such as non-radiative relaxation (e.g. transfer
of energy by random collisions).
M ? M heat
15Atomization The Dividing Line for Atomic and
Molecular Optical Electronic Spectroscopy
- Samples used in optical atomic (elemental)
spectroscopy are usually atomized - This destroys molecules (if present) and leaves
just atoms and atomic ions - The UV-visible spectrum of the atoms is of
interest, not the molecular spectrum.
16Atomic Electronic Energy Levels
- Electronic energy level transitions in hydrogen
the simplest of all! - Balmer series (visible)
- Transitions start (absorption) or end (emission)
with the first excited state of hydrogen - Lyman series (UV)
- Transitions start (absorption) or end (emission)
with the ground state of hydrogen
Diagrams from http//csep10.phys.utk.edu/astr162/l
ect/light/absorption.html
17Atomic Electronic Energy Levels
- Term symbols and electronic states used to
precisely define the state of electrons
spin multiplicity
s total spin quantum number j total
angular momentum quantum number l orbital
quantum number (s,p,d,f) mj state
s,p,d,f,g (l value)
2P
Term
2P3/2
Level
2j1
2P3/2-1/2
State
- Used to denote energy levels, and label Grotrian
(or term) diagrams for the hydrogen atom
Figure from the Sapphire Electronic Spectroscopy
Software Package, Cavendish Instruments Limited.
18Energy Levels for Different Atoms
- Atomic absorption and emission are generally
selective and specific for different elements on
the periodic table, allowing for qualitative
identification of elements
Diagrams from http//csep10.phys.utk.edu/astr162/l
ect/light/absorption.html
19Atomic Electronic Energy Levels
- Term (Grotrian) diagram for the sodium atom
each transition on the diagram can be linked to a
peak in the UV-visible spectrum - The number of lines can approach 5000 for
transition-metal elements. - Line broadening can be caused by
- Doppler effects
- pressure broadening (collisions)
- Lifetime of state (uncertainty)
Figure from H. A. Strobel and W. R Heineman,
Chemical Instrumentation A Systematic Approach,
Wiley, 1989.
20The Simulated UV-Visible Spectrum of Na0
From http//www.nist.gov/pml/data/asd.cfm
21Intensity of Atomic Electronic Energy Levels
- The population of energy levels partly determines
the intensity of an emission peak - The Boltzmann distribution relates the energy
difference between the levels, temperature, and
population
E energy of state P number of states
having equal energy at each level N number of
atoms in state
- Key point to get more atoms into excited states,
you need higher temperatures.
Element/Line (nm) Ne/Ng at 2000 K Ne/Ng at 3000 K Ne/Ng at 10000 K
Na 589.0 9.9 x 10-6 5.9 x 10-4 2.6 x 10-1
Ca 422.7 1.2 x 10-7 3.7 x 10-5 1.0 x 10-2
Zn 213.8 7.3 x 10-15 5.4 x 10-10 3.6 x 10-3
(Values from Cazes pg 79, Table 1)
22Basic Instrument Design for Atomic UV-Visible
Spectrometers
Radiation Source (Selective spectral lines)
Sample (in torch)
Wavelength Selector (can be before sample)
Detector (photoelectric transducer)
Source (sample in torch)
Wavelength Selector
Detector (photoelectric transducer)
- Wavelength selector is a mono- or polychromator
23Sources for Atomic Emission
- History Emission came first (study of sunlight
by Fraunhofer in 1817, identification of spectral
lines), studied throughout the 1800s and early
1900s
- Before the use of the plasma for OES in 1964, the
flame/gas torch (or arc/spark, etc) had the
following problems - Temperature instability
- Not hot enough to excite/decompose all materials
Atomizer/ Emission Source Temperature (C)
Flame 1700-3150
Plasma (e.g. ICP) 4000-8000
Electric arc 4000-5000
Electric spark gt10000
- Today The plasma has become the almost
universally-preferred method - History atomic emission placed demands on
monochromators - Today Technology has led to polychromators/detect
ors with sufficient resolution
24Plasma Torch Sources
- Plasma a low-density gas containing ions and
electrons, controlled by EM forces
25Plasma Torch Sources
- In the inductively-coupled plasma (ICP) torch,
the sample will reside for several milliseconds
at 4000-8000K. - Other designs direct current plasma, microwave
induced plasma
- An argon ICP torch in action
Photo by Steve Kvech, http//www.cee.vt.edu/progr
am_areas/environmental/teach/smprimer/icpms/icpms.
htmArgon20Plasma/Sample20Ionization
26More on Plasma Torches
- Another view of an argon ICP torch
Diagram from Lagalante, Appl. Spect. Reviews.
34, 191 (1999)
27Arc and Spark Sources for Atomic Emission
- Arc and spark sources used for qualitative
analysis of organic and geological samples - Only semi-quantitative because of source
instability - Spark sources achieve higher energies
- Several mg of solid sample is packed between
electrodes, 1-30 A of current is passed achieving
several hundred volts potential. - Applications include metals analysis or cases
where solids must be analyzed.
28Designs for Monochromators and Polychromators
Paschen-Runge design, shown as a polychromator
Czerny-Turner design, shown as a monochromator
- Polychromators
- High sample throughput rate
- Spectral interference can be an issue if the
interfering spectral line is not included on the
detector array - Monochromators
- Flexibility to access any wavelength within the
dimensions of the monochromator - Good for applications requiring complex
background corrections - Less sensitive lower radiation throughput
(because light blocked by slits)
29Atomic Emission Diffraction Gratings
- Diffraction gratings are used to select
wavelengths (in combination with collimating
lens, and slits) - Echelle (ladder) gratings high dispersion and
high resolution (a two-step system with a
cross-disperser standard grating or prism) - 1000-1500 grooves/mm typical for UV-Vis work
- Require filters to isolate orders (i.e. n1)
Figure from T. Wang, in J. Cazes, ed, Ewings
Analytical Instrumentation Handbook
30Atomic Emission Detectors
- At the end of the spectrometer, photons are
detected. - Commonly used detectors
- Photomultiplier tubes (PMT) dynamic range 109
- Solid-state detectors
- Charge-coupled devices (CCD) 1D or 2D arrays
(charge readout or transfer devices) - Silicon photodiodes with thousands of individual
addressable elements - These detectors are very sensitive, very
well-suited to 2D echelle grating polychromators,
very fast
31Example Detector Photomultiplier Tubes
- A PMT is a vacuum tube that contains a
photosensitive material, called the photocathode - The photocathode ejects electrons when it is
struck by light. These ejected electrons are
accelerated towards a dynode which ejects two to
five secondary electrons for every electron that
strikes its surface. - The secondary electrons strike another dynode,
ejecting more electrons which strike yet another
dynode, and so on (electron multiplication). - The electrical current measured at the anode is
then used as a relative measure of the intensity
of the radiation reaching the PMT.
32Modern ICP-OES Spectrometers
- Example system
- Varian Vista PRO
- Features
- 1. Axial flame view
- 2. Echelle grating polychromator (note the design
is like a Czerny-Turner monochromator) - 3. CCD detector
- CCD chips are made of sub-arrays matched to
emission lines.
Figure from Varian Vista PRO sales literature.
33Detection Limits of ICP-OES
- Typical detection limits (for a Varian Vista MPX)
- Considerations include the number of emission
lines, spectral overlap - Linearity can span several orders of magnitude.
34Atomic Absorption Spectroscopy (AAS)
- In the beginning atomic emission was the only
way to do elemental analysis via optical
spectroscopy - Bunsen and Kirchhoff (1861) invented a
non-luminous flame to study emission. Showed
that alkali elements in the flame removed lines
from a continuous source. - Walsh (1955) notices that molecular spectra are
often obtained in absorption (e.g. UV-Vis and
IR), but atomic spectra are always obtained in
emission. Proposes to use atomic absorption (AA
or AAS) for elemental analysis - Advantages over emission far less interference,
avoids problems with flame temperature
35Atomic Absorption Spectroscopy Instruments
- Atomic absorption spectrometry is one of the most
widely used methods for elemental analysis. - Basic principles of AA
- The sample is atomized via
- A flame (methane/H2/acetylene and air/oxygen)
- An electrothermal atomizer (an electrically-heated
graphite tube or cup) - UV-Visible light is projected through the flame
- The atoms absorb light (electronic excitation),
reducing the beam - The difference in intensity is measured by the
spectrometer
Source
P0
Sample/Flame
P
Monochromator
Detector
Images are of Aurora AI1200, http//www.spectronic
.co.uk
36Atomic Absorption Sources
- Hollow cathode lamps sputtering of an element
of interest, generating a line emission spectrum
- Typical linewidths of 0.002 nm (0.02Ã…)
- Single and multi-element lamps are available
- Other AA Sources electrode-less discharge lamp
(EDL) see Skoog Ch 9B-1
37Atomic Absorption Monochromators
- The monochromator filters out undesired light in
AA (typical bandwidths are 1 angstrom/0.1 nm) - This differs from ICP-OES, where the
monochromator actually analyzes the frequency. - In other words there is no need to scan the
grating, just set (aimed through a slit) and run - Echelle (ladder) gratings (combined with a
cross-disperser) are popular
Figure from T. Wang, in J. Cazes, ed, Ewings
Analytical Instrumentation Handbook
38Other Features of Atomic Absorption Systems
- Sample nebulizers Produces aerosols of samples
to introduce into the flame (oxyacetylene is the
hottest) - Detectors Common examples are photomultiplier
tubes, CCD (charge-coupled devices), and many
more. - Monochromator removes emissions from the flame
(flame is often kept cool just to avoid emission) - Modulated source (chopper) also removes the
remaining emissions from the flame. The signal
of interest is given an AC modulation and passed
through a high-pass filter. - Spectral interferences
- Absorption from other things (besides the element
of interest) other flame components,
particulates, etc Scattering can cause similar
problems - Background correction can help
39Graphite Furnace and Hydride AAS
- Graphite furnace and electrothermal AAS
- Analyze solutions, solids, slurries, by placing a
small amount (uL) of sample on a support for
evaporation and them atomization - More efficient atomization (entire sample
atomized at once) leads to smaller sample
quantity requirements or better sensitivity, but
reproducibility can be an issue
- Hydride generation AAS
- Efficiently volatilizes hydride forming elements
(As, Se, Tl, Pb, Bi, Sb, Te) by making their
hydrides via pre-reaction with sodium borohydride
and HCl - Inexpensive method of increasing sensitivity of
an AAS to ppt levels for these elements - Mercury cold-vapor AAS (Hg only)
40Detection Limits of Atomic Absorption Systems
- Detection limits in ppb (µg/L) for a selection of
elements - Individual results can vary depending on system,
matrix, etc
AAS (Flame) AAS (Electrothermal) ICP-OES
Al 30 0.005 2
As 100 0.02 40
Cd 1 0.0001 2
Hg 500 0.1 1
Mg 0.1 0.00002 0.05
Pb 10 0.002 2
Sn 20 0.1 30
Zn 2 0.00005 2
Values from D. A. Skoog, et al., Principles of
Instrumental Analysis, 5th Ed., Orlando,
Harcourt Brace and Co. 1998, pg. 225.
41How Are Elements Actually Analyzed?
- For AA and ICP-OES, samples are dissolved or
digested into solution, flowed into the
flame/plasma and analyzed. - Two methods for quantitative analysis
calibration - Standard calibration the unknown samples
absorbance/emission is compared with several
references which bracket the expected
concentration assuming a linear relationship. - Standard addition the unknown sample is divided
into several portions. One portion is directly
analyzed, the others have the reference material
added in varying amounts. The linear
relationship is determined, and the intercept is
used to calculate the real concentration of the
unknown. - Speciated analysis may be needed. The analysis of
atomic species, elements in chemically
distinguishable environments, usually by
hyphenation (e.g. ICP-OES coupled to a HPLC, AA
coupled to a GC) or offline extraction. - At the end the results yield elements in ppm,
ppb, mg/mL, or below LOQ or LOD
42Laser-Induced Breakdown Spectroscopy (LIBS)
- A focused laser can be used to create a plasma
(usually a pulsed Q-switched NdYAG laser)
- Portable systems capable of standoff analysis are
now available applications in the detection of
explosives, chemical warfare agents,
environmental analysis, etc
Figure from D. A. Cremers, R. C. Chinni,
Laser-induced breakdown spectroscopy -
Capabilities and limitations, Appl. Spectrosc.
Rev., 2009, 44, 457-506, http//dx.doi.org/10.1080
/05704920903058755.
43A Typical LIBS Spectrum
- The LIBS spectrum of ibuprofen drug substance
- Emission lines used for C, H, O, and N analysis
were 247.9, 656.3, 777.2 (triplet), and 746.8 nm,
along with the molecular band of C2 at 516 nm.
Figure from J. Anzano et al., Rapid
characterization of analgesic pills by
laser-induced breakdown spectroscopy (LIBS),
Med. Chem. Res. 2009, 18, 656664.
44Atomic Fluorescence
- Developed as an alternative to AA and ICP-OES,
with potentially greater sensitivity. - Has not yet achieved widespread use but cheaper
tunable lasers may change this. - Laser stimulated emission (coherent emission
from an excited state induced by a second photon) - Processes that emit a fluorescent photon
Non-radiative
hv
Thermal
hv
Non-radiative
hv
hv
Resonance
Direct Line
Stepwise
Thermally-assisted
45Atomic Fluorescence
- Basic AF instrument design
Sample
Wavelength Selector
Detector (photoelectric transducer)
Radiation source
(90 angle)
- AF sources include hollow-cathode lamps,
electrodeless discharge tubes (brighter), and
lasers (brightest)
Picture of HCT lamps from Perkin-Elmer
46A Comparison of Atomic Fluorescence with Other
Techniques
Plasma Emission (ICP-OES) AA (Flame) Atomic Fluorescence
Dynamic Range Wide Limited Wide
Qualitative Analysis Good Poor Poor
Multielement Scan? Good Poor Poor
Trace Analysis Good Good Good
Small samples Good Good Good
Matrix interferences Low High Low
Spectral interferences High Low Low
Cost Moderate (100K USD) Low (50K USD) Moderate
47Further Reading
- Required
- A. F. Lagalante, Atomic absorption spectroscopy
A tutorial review. Appl. Spectrosc. Rev. 1999,
34, 173-189. - A. F. Lagalante, Atomic emission spectroscopy
A tutorial review. Appl. Spectrosc. Rev. 1999,
34, 191-207. - Optional
- J. Cazes, Ed. Ewings Analytical Instrumentation
Handbook, 3rd Edition, 2005, Marcel Dekker,
Chapters 3 and 4. - D. A. Skoog, F. J. Holler and S. R. Crouch,
Principles of Instrumental Analysis, 6th Edition,
2006, Brooks-Cole, Chapters 8, 9, and 10. - N. Lewen, The use of atomic spectroscopy in the
pharmaceutical industry for the determination of
trace elements in pharmaceuticals, J. Pharm.
Biomed. Anal. 2011, 55, 653-661,
http//dx.doi.org/10.1016/j.jpba.2010.11.030. - H. A. Strobel and W. R. Heineman, Chemical
Instrumentation A Systematic Approach, 3rd
Ed., Wiley (1989). - D. A. Cremers, R. C. Chinni, Laser-induced
breakdown spectroscopy - Capabilities and
limitations, Appl. Spectrosc. Rev., 2009, 44,
457-506, http//dx.doi.org/10.1080/057049209030587
55.