Title: An introduction to Ultraviolet/Visible Absorption Spectroscopy
1An introduction to Ultraviolet/Visible Absorption
Spectroscopy
2- In this chapter, absorption by molecules, rather
than atoms, is considered. Absorption in the
ultraviolet and visible regions occurs due to
electronic transitions from the ground state to
excited state. Broad band spectra are obtained
since molecules have vibrational and rotational
energy levels associated with electronic energy
levels. The signal is either absorbance or
percent transmittance of the analyte solution
where
3An Introduction to Ultraviolet/VisibleMolecular
Absorption Spectrometry
- Absorption measurements based upon ultraviolet
and visible radiation find widespread application
for the quantitative determination of a large
variety species. - Beers Law
- A -logT logP0/P ?bc
- A absorbance
- ? molar absorptivity M-1 cm-1
- c concentration M
- P0 incident power
- P transmitted power (after passing through
sample)
4(No Transcript)
5(No Transcript)
6(No Transcript)
7- Measurement of Transmittance and Absorbance
- The power of the beam transmitted by the analyte
solution is usually compared with the power of
the beam transmitted by an identical cell
containing only solvent. An experimental
transmittance and absorbance are then obtained
with the equations. -
-
- P0 and P refers to the power of radiation after
it has passed through the solvent and the
analyte.
8Beers law and mixtures
- Each analyte present in the solution absorbs
light! - The magnitude of the absorption depends on its e
- A total A1A2An
- A total e1bc1e2bc2enbcn
- If e1 e2 en then simultaneous determination
is impossible - Need nls where es are different to solve the
mixture
9Limitations to Beers Law
- Real limitations
- Chemical deviations
- Instrumental deviations
10- 1.     Real Limitations
- Â
- a.      Beers law is good for dilute analyte
solutions only. High concentrations (gt0.01M) will
cause a negative error since as the distance
between molecules become smaller the charge
distribution will be affected which alter the
molecules ability to absorb a specific
wavelength. The same phenomenon is also observed
for solutions with high electrolyte
concentration, even at low analyte concentration.
The molar absorptivity is altered due to
electrostatic interactions.
11- b.     In the derivation of Beers law we have
introduced a constant (e). However, e is
dependent on the refractive index and the
refractive index is a function of concentration.
Therefore, e will be concentration dependent.
However, the refractive index changes very
slightly for dilute solutions and thus we can
practically assume that e is constant. - c.      In rare cases, the molar absorptivity
changes widely with concentration, even at dilute
solutions. Therefore, Beers law is never a
linear relation for such compounds, like
methylene blue.
12- 2.     Chemical Deviations
- Â
- This factor is an important one which largely
affects linearity in Beers law. It originates
when an analyte dissociates, associates, or
reacts in the solvent. For example, an acid base
indicator when dissolved in water will partially
dissociate according to its acid dissociation
constant
13(No Transcript)
14(No Transcript)
15(No Transcript)
16(No Transcript)
17- Chemical deviations from Beers law for
unbuffered solutions of the indicator Hln. Note
that there are positive deviations at 430 nm
and negative deviations at 570 nm. At 430 nm, the
absorbance is primarily due to the ionized In-
form of the indicator and is proportional to the
fraction ionized, which varies nonlinearly with
the total indicator concentration. At 570 nm, the
absorbance is due principally to the
undissociated acid Hln, which increases
nonlinearly with the total concentration.
18Calculated Absorbance Data for Various Indicator
Concentrations
19(No Transcript)
20- 3.     Instrumental Deviations
- Â
- a.      Beers law is good for monochromatic
light only since e is wavelength dependent. It is
enough to assume a dichromatic beam passing
through a sample to appreciate the need for a
monochromatic light. Assume that the radiant
power of incident radiation is Po and Po while
transmitted power is P and P. The absorbance of
solution can be written as
21(No Transcript)
22(No Transcript)
23- The effect of polychromatic radiation on Beers
law. In the spectrum at the top, the absorptivity
of the analyte is nearly constant over Band A
from the source. Note in the Beers law plot at
the bottom that using Band A gives a linear
relationship. In the spectrum, Band B corresponds
to a region where the absorptivity shows
substantial changes. In the lower plot, note the
dramatic deviation from Beers law that results.
24(No Transcript)
25- Therefore, the linearity between absorbance and
concentration breaks down if incident radiation
was polychromatic. In most cases with UV-Vis
spectroscopy, the effect small changes in
wavelengths is insignificant since e differs only
slightly especially at the wavelength maximum.
26- b.     Stray Radiation
- Â
- Stray radiation resulting from scattering or
various reflections in the instrument will reach
the detector without passing through the sample.
The problem can be severe in cases of high
absorbance or when the wavelengths of stray
radiation is in such a range where the detector
is highly sensitive as well as at wavelengths
extremes of an instrument. The absorbance
recorded can be represented by the relation - A log (Po Ps)/(P Ps)
- Where Ps is the radiant power of stray radiation.
27(No Transcript)
28Instrumental Noise as a Function in Transmittance
29(No Transcript)
30(No Transcript)
31(No Transcript)
32- Therefore, an absorbance between 0.2-0.7 may be
advantageous in terms of a lower uncertainty in
concentration measurements. At higher or lower
absorbances, an increase in uncertainty is
encountered. It is therefore advised that the
test solution be in the concentration range which
gives an absorbance value in the range from
0.2-0.7 for best precision. - However, it should also be remembered that we
ended up with this conclusion provided that sT is
constant. Unfortunately, sT is not always
constant which complicates the conclusions above.
33EFFECT OF bandwidth WIDTH
- Effect of bandwidth on spectral detail for a
sample of benzene vapor. Note that as the
spectral bandwidth increases, the fine structure
in the spectrum is lost. At a bandwidth of 10 nm,
only a broad absorption band is observed.
34- Effect of slit width (spectral bandwidth) on peak
heights. Here, the sample was s solution of
praseodymium chloride. Note that as the spectral
bandwidth decreases by decreasing the slit width
from 1.0 mm to 0.1 mm, the peak heights increase.
35Effect of Scattered Radiation at Wavelength
Extremes of an Instrument
- Wavelength extremes of an instrument are
dependent on type of source, detector and optical
components used in the manufacture of the
instrument. Outside the working range of the
instrument, it is not possible to use it for
accurate determinations. However, the extremes of
the instrument are very close to the region of
invalid instrumental performance and would thus
be not very accurate. An example may be a visible
photometer which, in principle, can be used in
the range from 340-780 nm. It may be obvious that
glass windows, cells and prism will start to
absorb significantly below 380 nm and thus a
decrease in the incident radiant power is
significant.
36B UV-VIS spectrophotometer A VIS
spectrophotometer
EFFECT OF SCATTERED RADIATION
- Spectrum of cerium (IV) obtained with a
spectrophotometer having glass optics (A) and
quartz optics (B). The false peak in A arises
from transmission of stray radiation of longer
wavelengths.
37- The output from the source at the low wavelength
range is minimal. Also, the detector has best
sensitivities around 550 nm which means that away
up and down this value, the sensitivity
significantly decrease. However, scattered
radiation, and stray radiation in general, will
reach the detector without passing through these
surfaces as well as these radiation are
constituted from wavelengths for which the
detector is highly sensitive. In some cases,
stray and scattered radiation reaching the
detector can be far more intense than the
monochromatic beam from the source. False peaks
may appear in such cases and one should be aware
of this cause of such peaks.
38- Instrumentation
- Light source
- ? - selection
- Sample container
- Detector
- Signal processing
- Light Sources (commercial instruments)
- D2 lamp (UV 160 375 nm)
- W lamp (vis 350 2500 nm)
39SourcesDeuterium and hydrogen lamps (160 375
nm)
Excited deuterium molecule with fixed quantized
energy
Dissociated into two deuterium atoms with
different kinetic energies
Ee ED2 ED ED hv
Ee is the electrical energy absorbed by the
molecule. ED2 is the fixed quantized energy of
D2, ED and ED are kinetic energy of the two
deuterium atoms.
40(No Transcript)
41Sources
Deuterium lamp UV region
- (a) A deuterium lamp of the type used in
spectrophotometers and (b) - its spectrum. The plot is of irradiance E?
(proportional to radiant power) versus - wavelength. Note that the maximum intensity
occurs at 225 m.Typically, - instruments switch from deuterium to tungsten at
350 nm.
42Visible and near-IR region
- (a) A tungsten lamp of the
- type used in spectroscopy
- and its spectrum (b).
- Intensity of the tungsten
- source is usually quite low
- at wavelengths shorter than about 350 nm. Note
that the intensity reaches a maximum in the
near-IR - region of the spectrum
- (1200 nm in this case).
43- The tungsten lamp is by far the most common
source in the visible and near IR region with a
continuum output wavelength in the range from
350-2500 nm. The lamp is formed from a tungsten
filament heated to about 3000 oC housed in a
glass envelope. The output of the lamp approaches
a black body radiation where it is observed that
the energy of a tungsten lamp varies as the
fourth power of the operating voltage.
44- Tungsten halogen lamps are currently more popular
than just tungsten lamps since they have longer
lifetime. Tungsten halogen lamps contain small
quantities of iodine in a quartz envelope. The
quartz envelope is necessary due to the higher
temperature of the tungsten halogen lamps (3500
oC). The longer lifetime of tungsten halogen
lamps stems from the fact that sublimed tungsten
forms volatile WI2 which redeposits on the
filament thus increasing its lifetime. The output
of tungsten halogen lamps are more efficient and
extend well into the UV.
45SourcesTungsten lamps (350-2500 nm)
- Why add I2 in the lamps?
- W I2 ? WI2
- Low limit 350 nm
- Low intensity
- Glass envelope
46- 3. Xenon Arc Lamps
- Â
- Passage of current through an atmosphere of high
pressured xenon excites xenon and produces a
continuum in the range from 200-1000 nm with
maximum output at about 500 nm. Although the
output of the xenon arc lamp covers the whole UV
and visible regions, it is seldom used as a
conventional source in the UV-Vis. The radiant
power of the lamp is very high as to preclude the
use of the lamp in UV-Vis instruments. However,
an important application of this source will be
discussed in luminescence spectroscopy which will
be discussed later
47(No Transcript)
48Sample Containers
- Sample containers are called cells or cuvettes
and are made of either glass or quartz depending
on the region of the electromagnetic spectrum.
The path length of the cell varies between 0.1
and 10 cm but the most common path length is 1.0
cm. Rectangular cells or cylindrical cells are
routinely used. In addition, disposable
polypropylene cells are used in the visible
region. The quality of the absorbance signal is
dependent on the quality of the cells used in
terms of matching, cleaning as well as freedom
from scratches.
49- Instrumental Components
- Source
- ? - selection (monochromators)
- Sample holders
- Cuvettes (b 1 cm typically)
- Glass (Vis)
- Fused silica (UVVis)
- Detectors
- Photodiodes
- PMTs
50- 1. Single beam
- Place cuvette with blank (i.e., solvent) in
instrument and take a reading ? 100 T - Replace cuvette with sample and take reading ?
T for analyte (from which absorbance is calcd)
51Instrumentation
- Most common spectrophotometer Spectronic 20.
- On/Off switch and zero transmission adjustment
knob - Wavelength selector/Readout
- Sample chamber
- Blank adjustment knob
- Absorbance/Transmittance scale
52(No Transcript)
53- End view of the exit slit of the Spectronic 20
- spectrophotometer pictured earlier
54- Single-Beam Instruments for the
Ultraviolet/Visible Region
55- Single-Beam Computerized Spectrophotometers
Inside of a single-beam spectrophotometer
connected to a computer.
56Types of Instruments
- Instrumental designs for UV-visible photometers
- or spectrophotometers. In (a), a single-beam
instrument is shown. Radiation from the filter
or monochromator passes through either the
reference cell or the sample cell before
striking the photodetector.
57(No Transcript)
58- 2. Double beam (most commercial instruments)
- Light is split and directed towards both
reference cell (blank) and sample cell - Two detectors electronics measure ratio (i.e.,
measure/calculate absorbance) - Advantages
- Compensates for fluctuations in source intensity
and drift in detector - Better design for continuous recording of spectra
59General Instrument Designs Double Beam In - Space
Needs two detectors
60General Instrument Designs Double Beam In - Time
61(No Transcript)
62- Merits of Double Beam Instruments
- Compensate for all but the most short term
fluctuation in radiant output of the source - Compensate drift in transducer and amplifier
- Compensate for wide variations in source
intensity with wavelength
63Location of Sample cell
- In all photometers and scanning
spectrophotpmeters described above, the cell has
been positioned after the monochromators. This is
important to decrease the possibility of sample
photodecomposition due to prolonged exposure to
all frequencies coming from the source. However,
the sample is positioned before the monochromator
in multichannel instruments like a photodiode
array spectrophotometer. This can be done without
fear of photodecomposition since the sample
exposure time is usually less than 1 s.
Therefore, it is now clear that in UV-Vis where
photodecomposition of samples can take place, the
sample is placed after the monochromators in
scanning instruments while positioning of the
sample before the monochromators is advised in
multichannel instruments.
64- 3. Multichannel Instruments
- Photodiode array detectors used (multichannel
detector, can measure all wavelengths dispersed
by grating simultaneously). - Advantage scan spectrum very quickly snapshot
lt 1 sec. - Powerful tool for studies of transient
intermediates in moderately fast reactions. - Useful for kinetic studies.
- Useful for qualitative and quantitative
determination of the components exiting from a
liquid chromatographic column.
65Multi-channel Design
66(No Transcript)
67- A multichannel diode-array spectrophotometer, the
Agilent Technologies 8453.
684. Probe Type Instruments
- These are the same as conventional single beam
instruments but the beam from the monochromators
is guided through a bifurcated optical fiber to
the sample container where absorption takes
place. The attenuation in reflected beam at the
specified wavelength is thus measured and related
to concentration of analyte in the sample. - A fiber optic cable can be referred to as a light
pipe where light can be transmitted by the fiber
without loss in intensity (when light hits the
internal surface of the fiber at an angle larger
than a critical angle). Therefore, fiber optics
can be used to transmit light for very long
distances without losses. A group of fibers can
be combined together to form a fiber optic cable
or bundle. A bifurcated fiber optic cable has
three terminals where fibers from two separate
cables are combined at one end to form the new
configuration.
69(No Transcript)
70Fiber optic probe
71(No Transcript)
72Double Dispersing Instruments
- The instrument in this case has two gratings
where the light beam leaving the first
monochromators at a specified wavelength is
directed to the second grating. This procedure
results in better spectral resolution as well as
decreased scattered radiation. However, double
dispersing instruments are expensive and seem to
offer limited advantages as compared to cost
especially in the UV-Vis region where exact
wavelength may not be crucial.
73- Optical diagram of the Varian Cary 300
double-dispersing spectrophotometer. A second
monochromator is added immediately after the
source.
74(No Transcript)
75Molar absorptivities
- e 8.7 x 10 19 P A
- A cross section of molecule in cm2 (10-15)
- P Probability of the electronic transition (0-1)
- Pgt0.1-1 ? allowable transitions
- Plt0.01 ? forbidden transitions
76Molecular Absorption
- M hn ? M (absorption 10-8 sec)
- M ? M heat (relaxation process)
- M ? ABC (photochemical decomposition)
- M ? M hn (emission)
77Visible Absorption Spectra
78- The absorption of UV-visible radiation generally
results from excitation of bonding electrons. - can be used for quantitative and qualitative
analysis
79- Molecular orbital is the nonlocalized fields
between atoms that are occupied by bonding
electrons. (when two atom orbitals combine,
either a low-energy bonding molecular orbital or
a high energy antibonding molecular orbital
results.) - Sigma (?) orbital
- The molecular orbital associated with single
bonds in organic compounds - Pi (?) orbital
- The molecular orbital associated with parallel
overlap of atomic P orbital. - n electrons
- No bonding electrons
80Molecular Transitions for UV-Visible Absorptions
- What electrons can we use for these transitions?
81MO Diagram for Formaldehyde (CH2O)
H
C
O
H
s
p
n
82Singlet vs. triplet
- In these diagrams, one electron has been excited
(promoted) from the n to ? energy levels
(non-bonding to anti-bonding). - One is a Singlet excited state, the other is a
Triplet.
83Type of Transitions
- s ? s
- High energy required, vacuum UV range
- CH4 ? 125 nm
- n ? s
- Saturated compounds, CH3OH etc (? 150 - 250 nm)
- n ? ? and ? ? ?
- Mostly used! ? 200 - 700 nm
84Examples of UV-Visible Absorptions
LOW!
85UV-Visible Absorption Chromophores
86Effects of solvents
- Blue shift (n- p) (Hypsocromic shift)
- Increasing polarity of solvent ? better solvation
of electron pairs (n level has lower E) - ? peak shifts to the blue (more energetic)
- 30 nm (hydrogen bond energy)
- Red shift (n- p and p p) (Bathochromic shift)
- Increasing polarity of solvent, then increase the
attractive polarization forces between solvent
and absorber, thus decreases the energy of the
unexcited and excited states with the later
greater - ? peaks shift to the red
- 5 nm
87UV-Visible Absorption Chromophores
88Typical UV Absorption Spectra
Chromophores?
89The effects of substitution
Auxochrome function group
Auxochrome is a functional group that does not
absorb in UV region but has the effect of
shifting chromophore peaks to longer wavelength
as well As increasing their intensity.
90Now solvents are your container
- They need to be transparent and do not erase the
fine structure arising from the vibrational
effects
Polar solvents generally tend to cause this
problem
Same solvent must be Used when comparing absorptio
n spectra for identification purpose.
91(No Transcript)
92Summary of transitions for organic molecules
- s ? s transition in vacuum UV (single bonds)
- n ? s saturated compounds with non-bonding
electrons - l 150-250 nm
- e 100-3000 ( not strong)
- n ? p, p ? p requires unsaturated functional
groups (eq. double bonds) most commonly used,
energy good range for UV/Vis - l 200 - 700 nm
- n ? p e 10-100
- p ? p e 1000 10,000
93List of common chromophores and their transitions
94Organic Compounds
- Most organic spectra are complex
- Electronic and vibration transitions superimposed
- Absorption bands usually broad
- Detailed theoretical analysis not possible, but
semi-quantitative or qualitative analysis of
types of bonds is possible. - Effects of solvent molecular details complicate
comparison
95Rule of thumb for conjugation
If greater then one single bond apart - e are
relatively additive (hyperchromic shift) - l
constant CH3CH2CH2CHCH2 lmax 184 emax
10,000 CH2CHCH2CH2CHCH2 lmax185 emax
20,000 If conjugated - shifts to higher ls
(red shift) H2CCHCHCH2 lmax217 emax
21,000
96Spectral nomenclature of shifts
97What about inorganics?
- Common anions n?p nitrate (313 nm), carbonate
(217 nm) - Most transition-metal ions absorb in the UV/Vis
region. - In the lanthanide and actinide series the
absorption process results from electronic
transitions of 4f and 5f electrons. - For the first and second transition metal series
the absorption process results from transitions
of 3d and 4d electrons. - The bands are often broad.
- The position of the maxima are strongly
influenced by the chemical environment. - The metal forms a complex with other stuff,
called ligands. The presence of the ligands
splits the d-orbital energies.
98Transition metal ions
99Charge-Transfer-Absorption
- A charge-transfer complex consists of an
electron-donor group bonded to an electron
acceptor. When this product absorbs radiation, an
electron from the donor is transferred to an
orbital that is largely associated with the
acceptor. - Large molar absorptivity (emax gt10,000)
- Many organic and inorganic complexes
100(No Transcript)