Title: Chem. 231
1Chem. 231 3/4 Lecture
2Announcements I
- Set 1 Lab Reports
- Both Lab Reports due 3/13
- Only one of the lab reports needs to be detailed,
the other can be a summary - Homework Set 3 due Wednesday
- Quiz 3 (last quiz) next Monday
- Set 2 Labs (two need to do one of them)
- SPE/HPLC lab
- SPME/GC lab
- More instructions coming
3Announcements II
- Set 3 Labs
- Hope to have instructions completed by Monday
- Todays Lecture
- SPE Lab Information
- SPME Lab Information
- Practical Applications of HPLC
4SPE/HPLC LabInformation
- Goal Determine concentration of unknown phenol
in aqueous solution - Background
- Phenols are possible soil contaminants (e.g.
pentachlorophenol and nitrophenols) and can
contaminate water supplies - Phenols are also in some beverages (e.g.
vanillin, other guaiacols in mostly aqueous
environment) - Solid Phase allows trapping of moderately
non-polar compounds needed for sensitive analysis
OH
guaiacols
R
OCH3
5SPE/HPLC LabInformation Methodology
- Overview
- Unknowns will need to be trapped on the SPE
cartridge, then eluted (with smaller volume) to
concentrate samples - Analysis by HPLC (will need to develop method and
determine elution order as in set 1 lab) - SPE Procedure
- Clean SPE cartridges (first with removal eluent,
then with sample solvent) - Flow rate must be slow enough for transfer
- Apply sample to SPE cartridges
6SPE/HPLC LabInformation Methodology
- SPE Procedure cont.
- Wash sample with sample solvent (removes polar
contaminants from pores) - Remove analytes with stronger solvent
- If removing into volumetric flask, need to fill
to line (more quantitative method) - Unknown Analysis
- Will want to use SPE method on unknown to
determine unknown compounds
7SPE/HPLC LabInformation Methodology
- SPE Efficiency Testing
- To determine if unknown compound losses (in
passing through SPE cartridge or in removal from
SPE cartridge) are significant - Done by keeping and analyzing multiple fractions
8SPME/GC LabInformation
- Goal
- Determine the concentration of two flower scent
monoterpene unknowns in a sample - Background
- Many of the scent compounds in flowers are
monoterpenes (consist of two isoprene units) - Traditional analysis would involve extraction and
liquid injection - Alternative of SPME is solvent free
myrcene
9SPME/GC LabInformation - Methodology
- GC Methodology
- Requires splitless injection (so cant use Buck
GCs) - We will be using an HP6890 and an Agilent 7890
GC-MS (this will also give qualitative analysis) - These have software control closer to the Agilent
HPLCs - See 2/4 lecture, slide 16 for procedure using SPME
10SPME/GC LabInformation - Methodology
- GC Methodology cont.
- SPME equipment is fragile
- SHEATH ALWAYS ON WHEN GOING IN OUR OUT OF SEPTA
OR DAMAGE OCCURS - NO CONTACT WITH DICHLOROMETHANE
- SPME fibers have limited lifetimes under normal
use and may need replacing - Besides what is shown in past slide, SPME must be
pre- and/or post- heated (can use injector for
this if conditions are o.k.)
11SPME/GC LabInformation - Methodology
- GC Methodology cont.
- Quantitative Analysis by SPME/GC works best when
- SPME conditions are kept as consistent as
possible (same exposure conditions in vial and
same desorption conditions in GC inlet) - An internal standard can help control some
variables so that if the transfer of two
compounds is consistent, good results will occur
12Practical Applications of HPLC
- Review of Instrument Components
Chromatograph Components
Chromatographic Column
Detector
Flow Control
Mobile Phase Reservoir
Injector
Waste or fraction collection
Chromatogram
Signal to data recorder
13Practical Applications of HPLC
- Mobile Phase Selection
- HPLC class will determine main mobile phase
requirements - mostly non-polar solvent for normal phase
- polar solvents for reversed phase
- water plus ion exchanger for ion chromatography
- Other Major Factors
- desired retention (adjusting polarity of solvent
in HPLC to get desired retention or eluent ionic
strength in IC) - need for pH adjustment (to reduce or promote ion
formation)
14Practical Applications of HPLC
- Mobile Phase Selection Cont.
- Additional Considerations
- Selectivity (different solvents will have
different solvent analyte interactions best to
choose solvent that emphasizes analyte
differences) - Solvent viscosity (low viscosity means smaller
back pressure for given flow rate) - Solvent miscibility
- Sample solubility
- Detector limitations (e.g. wavelength cut-offs
for UV detection) - Compatibility with column packing and tubing
15Practical Applications of HPLC
To column
- Solvent Selection
- Single Pump
- Manual Solvent Selection
- manual valve selection or replacement of mobile
phase - not capable of gradient runs
- Low Pressure Mixing
- controlled valve for selection of eluents (4
possible with our Agilent system) - fast proportional opening of valves allows good
control for gradients - Dual Pump Systems
- each pump runs independently
- better gradient control and retention time
stability vs. low pressure mixing - also smaller dwell time (difference in time
between switching valves and effects at columns)
Mixing chamber
pump
To column
pumps
gradient starts
Pressure
dwell time
Effect seen
time
16Practical Applications of HPLC
- Pump Requirements
- High pressures
- Small packing material diameters require high
pressures (up to 300-400 bar) so that one can
obtain reasonable flow rates - Newest packing material diameters (sub 5 mm)
require even higher pressures or shorter column
lengths) in UPLC which can go to 1000 bar - Pumps require specific components capable of
withstanding such high pressures (e.g. use of
sapphire piston rods) - Stable Flow rates
- Usually involve feedback control
17Practical Applications of HPLC
- Pump how they work
- Most pumps use two piston heads 180º out of phase
to reduce pressure fluctuations - Solvents go into and out of piston heads through
one-way check valves - Exit check valve closes on in stroke and
entrance check valve closes on out stroke
In Stroke
Check valves
Out Stroke
open
closed
closed
open
pistons
18Practical Applications of HPLC
- Pump Failures
- Common for check valves to fail
- This results in a drop in pressure during one of
the cycles resulting in an alternating pressure
Bad check valve leaking
19Practical Applications of HPLC
6 port valve
- Injection Valves
- Fixed loop injector is by far the most common
type - small loop volume with manual injection
- with autosamplers, it is common to use partial
filling (e.g. fill 10 mL of a 250 mL loop) - For large sample injections, the loop can be
replaced with a trap - Specialized injectors
- used for injecting very small samples in
microbore columns - SPME injectors
From pump
To column
Waste
sample in
LOAD POSITION
INJECT POSITION
20Practical Applications of HPLC
- Sample Injection
- Injection Volumes
- With sample solvent eluent, volume injected
shouldnt contribute significantly to extra
column broadening (rule of thumb injection volume
lt 0.25(early eluting peak volume)) - Sample Solvent
- Ideal solvent is solvent weaker than eluent (or
initial eluent in a gradient) - Then can use larger injection volume since
analytes will trap on column - Stronger solvents can result in distorted peak
shapes or require even smaller injection volumes
width 0.084 min. (or 5.1 s) _at_ flow 1.2
mL/min or peak volume 100 mL, so use vol lt 25 mL