IEC SC62A MT 23Liaison Report to C63 SC84 May
2011
  • Jeffrey L. Silberberg
  • FDA/CDRH

2
IEC SC62A MT23 Liaison Report
  • MT23 met February 2011 in Carlsbad, CA
  • Prepared second CD of IEC 60601-1-2 Ed. 4 for
    circulation
  • IEC 60601-1-2 Ed. 4 circulated 2011-03-04 as
    62A/746/CD
  • Comment period closes 2011-06-10
  • US SC62A TAG meeting 2011-05-13 to prepare US
    comments

3
Changes made from previous draft
  • Performance (EMC) requirements were moved to an
    informative annex
  • Several of the proposed new test methods were
    deleted
  • Low-frequency magnetic field emissions
  • Low-frequency magnetic field immunity
  • Close-distance RF immunity
  • Under consideration

4
Changes made from previous draft (contd)
  • Several requirements were deleted
  • Harmonics and flicker emissions for equipment
    rated at gt 16 A
  • Harmonics and interharmonics immunity
  • Patient cable conducted emissions test is now
    informative
  • Deleted requirements and published standards
    listed as things that should be considered in the
    Risk Management process

5
Changes made from previous draft (contd)
  • Some general immunity test levels were lowered
  • Radiated RF
  • Home and Transport ESD
  • Higher immunity test levels in frequency bands
    where sources are
  • ISM
  • Amateur radio
  • Portable communications equipment

6
Changes made from previous draft (contd)
  • The close distance RF test was replaced with an
    interim test using IEC 61000-4-3 at 30 V/m to
    approximate portable transmitters
  • Recommended separation distances, but smaller
    distances
  • Warning symbol required if you dont do the test

7
Changes made from previous draft (contd)
  • An MT23 member is writing proposals for SC77A for
    circulation as an NWIP
  • Proximity RF immunity
  • Proximity magnetic field immunity
  • Normative references are now dated unless there
    is a reason otherwise
  • PORT is now a defined term, figure
  • Clinic environment renamed to Small clinic
  • Categories Hospital, Small clinic, Home,
    Transport, Special

8
Example test levels for Enclosure port
Phenomenon Immunity test levels Immunity test levels Immunity test levels Immunity test levels
Phenomenon Hospital Small clinic Home Transport
Electrostatic discharge  8 kV contact 2 kV, 4 kV,  8 kV,  15 kV air  8 kV contact 2 kV, 4 kV,  8 kV,  15 kV air  8 kV contact 2 kV, 4 kV,  8 kV,  15 kV air  8 kV contact 2 kV, 4 kV,  8 kV,  15 kV air
EM Field a) 3 V/m f) g) 80 MHz 1000 MHz, 1,4 GHz - 2,7 GHz and 5 GHz 6 GHz b) 80 AM _at_ 1 kHz c) 3 V/m f) g) 80 MHz 1000 MHz, 1,4 GHz - 2,7 GHz and 5 GHz 6 GHz b) 80 AM _at_ 1 kHz c) 10 V/m f) 80 MHz 1000 MHz b) 3 V/m 1,4 GHz - 2,7 GHz and 5 GHz 6 GHz 80 AM _at_ 1 kHz c) 20 V/m f) 80 MHz 1000 MHz b) 10 V/m f) 1,4 GHz - 2,7 GHz 3 V/m f) 5 GHz 6 GHz) 80 AM _at_ 1 kHz c)
9
MT23 Next steps
  • Finish (revising) responses to 460 comments from
    first CD
  • Respond to 217 comments on questionnaire
    (62A/700/Q)
  • MT23 meeting scheduled for August 2011
  • To respond to comments on 62A/746/CD and prepare
    CDV
  • Probably in White Oak / Silver Spring, MD
  • Goal for CDV November 2011