Title: IEC SC62A MT 23 Liaison Report to C63
1IEC SC62A MT 23Liaison Report to C63 SC84 May
2011
- Jeffrey L. Silberberg
- FDA/CDRH
2IEC SC62A MT23 Liaison Report
- MT23 met February 2011 in Carlsbad, CA
- Prepared second CD of IEC 60601-1-2 Ed. 4 for
circulation - IEC 60601-1-2 Ed. 4 circulated 2011-03-04 as
62A/746/CD - Comment period closes 2011-06-10
- US SC62A TAG meeting 2011-05-13 to prepare US
comments
3Changes made from previous draft
- Performance (EMC) requirements were moved to an
informative annex - Several of the proposed new test methods were
deleted - Low-frequency magnetic field emissions
- Low-frequency magnetic field immunity
- Close-distance RF immunity
- Under consideration
4Changes made from previous draft (contd)
- Several requirements were deleted
- Harmonics and flicker emissions for equipment
rated at gt 16 A - Harmonics and interharmonics immunity
- Patient cable conducted emissions test is now
informative - Deleted requirements and published standards
listed as things that should be considered in the
Risk Management process
5Changes made from previous draft (contd)
- Some general immunity test levels were lowered
- Radiated RF
- Home and Transport ESD
- Higher immunity test levels in frequency bands
where sources are - ISM
- Amateur radio
- Portable communications equipment
6Changes made from previous draft (contd)
- The close distance RF test was replaced with an
interim test using IEC 61000-4-3 at 30 V/m to
approximate portable transmitters - Recommended separation distances, but smaller
distances - Warning symbol required if you dont do the test
7Changes made from previous draft (contd)
- An MT23 member is writing proposals for SC77A for
circulation as an NWIP - Proximity RF immunity
- Proximity magnetic field immunity
- Normative references are now dated unless there
is a reason otherwise - PORT is now a defined term, figure
- Clinic environment renamed to Small clinic
- Categories Hospital, Small clinic, Home,
Transport, Special
8Example test levels for Enclosure port
Phenomenon Immunity test levels Immunity test levels Immunity test levels Immunity test levels
Phenomenon Hospital Small clinic Home Transport
Electrostatic discharge 8 kV contact 2 kV, 4 kV, 8 kV, 15 kV air 8 kV contact 2 kV, 4 kV, 8 kV, 15 kV air 8 kV contact 2 kV, 4 kV, 8 kV, 15 kV air 8 kV contact 2 kV, 4 kV, 8 kV, 15 kV air
EM Field a) 3 V/m f) g) 80 MHz 1000 MHz, 1,4 GHz - 2,7 GHz and 5 GHz 6 GHz b) 80 AM _at_ 1 kHz c) 3 V/m f) g) 80 MHz 1000 MHz, 1,4 GHz - 2,7 GHz and 5 GHz 6 GHz b) 80 AM _at_ 1 kHz c) 10 V/m f) 80 MHz 1000 MHz b) 3 V/m 1,4 GHz - 2,7 GHz and 5 GHz 6 GHz 80 AM _at_ 1 kHz c) 20 V/m f) 80 MHz 1000 MHz b) 10 V/m f) 1,4 GHz - 2,7 GHz 3 V/m f) 5 GHz 6 GHz) 80 AM _at_ 1 kHz c)
9MT23 Next steps
- Finish (revising) responses to 460 comments from
first CD - Respond to 217 comments on questionnaire
(62A/700/Q) - MT23 meeting scheduled for August 2011
- To respond to comments on 62A/746/CD and prepare
CDV - Probably in White Oak / Silver Spring, MD
- Goal for CDV November 2011