Diapositiva 1 - PowerPoint PPT Presentation

1 / 27
About This Presentation
Title:

Diapositiva 1

Description:

Mass Spectrometry Mass spectrometry (MS) differs from other common forms of organic spectral analysis in that the sample does not absorb radiation such as infrared ... – PowerPoint PPT presentation

Number of Views:34
Avg rating:3.0/5.0
Slides: 28
Provided by: sil47
Category:

less

Transcript and Presenter's Notes

Title: Diapositiva 1


1
Mass Spectrometry Mass spectrometry (MS) differs
from other common forms of organic spectral
analysis in that the sample does not absorb
radiation such as infrared, ultraviolet, or radio
waves from the electromagnetic spectrum. In
contrast to infrared (IR) or nuclear magnetic
resonance (NMR) spectrometry, both of which
identify compounds with specificity comparable to
that of mass spectrometry, MS is a destructive
method of analysisthat is, the sample cannot be
recovered after mass spectral analysis. Mass
spectrometers are typically not standalone
instruments. Most often they are connected
physically and electronically to a
chromatograph the chromatograph separates
mixtures and introduces the sample into the mass
spectrometer the mass spectrometer ionizes
analyte molecules, then separates and detects the
resulting ions. In order to be analyzed by mass
spectrometry, sample molecules must be ionized.
2
(No Transcript)
3
(No Transcript)
4
(No Transcript)
5
(No Transcript)
6
(No Transcript)
7
(No Transcript)
8
The ion that is formed first results directly
from ejection of a single electron from the
neutral molecule (product c). This molecular ion
(M) is very important because it has virtually
the same mass as that of the analyte molecule
(the small mass of the lost electron can be
ignored). Since mass spectrometry actually
measures the mass-tocharge ratio (m/z) of an ion,
not its mass, an ion having a charge greater than
1 is found not at the m/z value corresponding to
its mass (m), but rather at m/2, m/3, or m/4,
depending on the number of charge states.
9
Electron Ionization Source The EI source is
most commonly a small chamber about 1 cc in
volume, in which analyte molecules interact with
a beam of highly energetic electrons that have
typically been accelerated through a potential
difference of 5070 volts (V) across the volume
of the ion source.
10
(No Transcript)
11
Chemical Ionization Unlike EIMS, in which
molecules are ionized through interaction with
high-energy electrons, ionization in chemical
ionization mass spectrometry (CIMS) depends on
collisions of ions and molecules. In positive ion
CIMS the sample is ionized by reaction with ions
generated within a large excess of a relatively
low molecular mass reagent gas such as methane
(as CH5), isobutane as (CH3)3C, or ammonia.
12
In CIMS the concentration of analyte molecules is
small compared to that of reagent gas molecules.
Thus, the electron beam, which is more energetic
than that used in EIMS (200 eV), preferentially
ionizes the reagent gas. Analyte molecules are
ionized through reaction with reagent gas ions,
rather than by the electron beam. Most reagent
gas ions are strong proton donors and form
protonated molecules (sometimes incorrectly
called pseudomolecular ions) that have a mass 1 u
greater than that of the molecular mass of the
original compound
M CH5 MH CH4
13
(No Transcript)
14
Electrospray Ionization Basically ESI works by
converting the HPLC effluent, already containing
the sample in ionic form, into an aerosol and
subjecting the resulting spray to high voltage in
a chamber held near atmospheric pressure. This
process creates a mist of charged droplets that
flow toward the orifice of the capillary. In the
configuration shown, the nebulizing needle, which
creates the aerosol, is orthogonal
(perpendicular) to the eventual direction of ion
flow toward the m/z analyzer. Other geometric
configurations are possible and have been
used. As the charged droplets travel toward the
capillary opening, they are subjected to the
counterflow of a drying gas, such as nitrogen
(N2), which causes evaporation of solvent
molecules from the droplets. Evaporation, charge
concentration, and droplet disintegration
continue until the analyte ions are finally
desorbed into the vapor phase, passed into the
sampling capillary, then on into the high vacuum
of the m/z analyzer.
15
(No Transcript)
16
(No Transcript)
17
(No Transcript)
18
(No Transcript)
19
(No Transcript)
20
(No Transcript)
21
Trappola Ionica Può essere considerato una
variante dell'analizzatore a quadrupolo. Anzichè
permettere agli ioni di attraversare il campo
quadrupolare, la trappola ionica trattiene tutti
gli ioni al suo interno. Lo spettro di massa è
generato variando il potenziale elettrico in modo
da espellere in sequenza dalla trappola verso il
rivelatore gli ioni secondo un valore m/z
crescente.
22
(No Transcript)
23
(No Transcript)
24
(No Transcript)
25
(No Transcript)
26
Analizzatore a quadrupolo E costituito da
quattro barre cilindriche metalliche, lunghe ca.
20 cm., che delimitano il "cammino" percorso
dagli ioni provenienti dalla camera di
ionizzazione e diretti al detector. Le barre sono
mantenute ad un potenziale elettromegnetico
oscillante, in modo che quando le due sbarre
verticali hanno potenziale positivo quelle
orrizzontali lhanno negativo, e viceversa. Gli
elettroni, accelerati dalle piastre
acceleratrici, entrano nel tunnel delimitato
dalle barre e vengono respinti dai poli positivi
ed attratti dai negativi. Tuttavia, a causa
delloscillazione del quadrupolo gli ioni
assumono una traiettoria a zig zag e finiscono
con lo scaricarsi su una delle barre, tranne
quelli che, per una certo valore di frequenza di
oscillazione, hanno unenergia cinetica tale per
cui il moto diventa sinusoidale e riecono ad
uscire dal tunnel ed entrare nel sistema di
rivelazione (fotomoltiplicatore). Operando quindi
una scansione di frequenza di oscillazione del
campo è possibile far uscire ioni a massa
molecolare crescente.
27
(No Transcript)
Write a Comment
User Comments (0)
About PowerShow.com