Title: Chapter 19
1Chapter 19 20 Image Quality Techniques
- There are three geometric factors that affect
radiographic quality. - Magnification
- Distortion
- Focal Spot Blur
- We have explored these factors in the laboratory.
2Magnification
- All objects on the radiograph are larger that
their actual size. This is called magnification. - The magnification factor is the image size
divided by the object size. - At 40 (100 cm) factor is 1.1
- At 72 (180 cm) factor is 1.05
3Magnification
- Usually we do not know the size of the object so
we must determine the magnification factor
another way. - Image size S0D
- MF ----------------
------------ - Object Size SID
4Minimizing Magnification
- Large SID Use the less divergent beam.
- Chest X-rays are done at 72 SID to minimize
magnification. - Lateral C-spine done at 72
- Small OID Get patient as close to the film as
possible. - Basic principle for positioning.
5Distortion
- Distortion is the misrepresentation of the true
size and shape of the object being radiographed. - The amount of distortion depends upon the
thickness, position and shape.
6Thickness
- Thick objects are more distorted than thin
objects because of the greater change in Object
Image Distance.
7Thickness
- The position of the object relative to the
central axis will cause greater distortion with
thick and/or irregular shaped objects.
8Object Position
- If the object plane and image plane are parallel
the image will not be distorted. - If the object plane and image plane are not
parallel, distortion will occur.
9Spatial Distortion
- When multiple objects at different OIDs occur,
we get spatial distortion due to unequal
magnification. - Two arrows appear as one.
- When shifted laterally more distortion occurs
10Object Shape Distortion
- When the object plane is not parallel to the
image plane as when inclined, shape distortion
occurs. - This will result in foreshortening.
11Focal Spot Blur
- Focal spot blur is caused by the effective size
of the focal spot, which is larger at the cathode
side. - Focal spot blur is the most important factor in
determining spatial resolution.
12Focal Spot Blur
- Focal spot blur is impacted by the Object to
Image Distance.
13Focal Spot Blur Heel Effect
- There is more to the heel affect than just the
attenuation of the beam by the anode. - The focal spot blur is smaller at the anode side
and larger at the cathode side.
14Taking advantage of the Anode Heel Effect
Exam Toward Cathode Toward Anode
AP Full spine Abdomen Neck
Chest Abdomen Neck
AP T-spine Abdomen Neck
Lat T-spine Neck Abdomen
Humerus Shoulder Elbow
15Did you see a problem ?
- If the tube is mounted correctly for the AP Full
spine, Chest and A-P Thoracic Spine, the patient
must stand on his head for the lateral thoracic
spine!!!! - For erect radiography, the use of the anode heel
affect is limited.
16Object Factors that Affect Quality
- Subject Contrast
- Patient or part thickness
- Tissue mass density
- Affective atomic number
- Object shape
- kVp
17Radiographic Contrast
- Radiographic Contrast is how the film looks.
- It is the combination of receptor contrast and
subject contrast. - Contrast occurs between structures of different
densities.
18Thickness
- The thicker the body part or body section, the
greater attenuation of the beam. Contrast is
relative to the number of x-rays leaving the
body. - Remember that x-rays are merely shadows of the
anatomy based upon attentation of the beam.
19Thickness
- Radiographs of thin patients will have more
contrast than those of large patients. - Thicker object absorb more rays and will appear
lighter than thin objects.
20Tissue Mass Density
- Different sections of the body have may equal
thickness yet different mass density. - Two wrist may be the same thickness but the one
that is swollen will have greater mass density
due to water in the tissues.
21Effective Atomic Number
- While Compton interactions with tissues are not
impacted by the relative atomic number of
tissues, the photoelectric effect vary with the
cube of the atomic number. - When the effective atomic number of adjacent
tissues is very different, subject contrast is
very high.
22Object shape
- The object shape not only influences the geometry
but also through its contribution to subject
contrast. - A vessel on end has high contrast while other
have lower contrast.
23Varying tissue densities
- Bone absorbs most x-rays leaving a white shadow.
- Water absorbs less x-rays leaving a light gray
shadow - Fat absorbs fewer x-rays leaving a dark gray
shadow. - Air absorbs little x-rays and is black on the
film.
24Varying densities in the Body
- Air, oil, water and metal (natural) absorb
different degrees of the x-rays and produce
contrast. - Heavy metals are used as contrast media to
enhance contrast in the body in medical
radiology. Principle ones are Barium and Iodine.
25kVp
- We have little control over the previous factors
controlling subject contrast. - BUT!!!!!
- The absolute magnitude of subject contrast is
greatly influenced by the kVp of operation.
26kVp
- kVp also influences film contrast but not to the
extend it controls subject contrast. - Low kVp high contrast short scale
- High kVp low contrast broad scale
27Motion Blur
- If any element of the x-ray moves during
exposure, we get motion blur. - Patient motion is the most common cause of motion
blur. - Motion blur is more common in erect radiography.
28Ways to Control Motion Blur
- Use the shortest possible exposure time
- Restrict patient motion by instructions or
restraining device. - Use a large SID
- Use a small OID
29Tools to Improve Quality
- Patient Positioning
- Get the patient close to the film reduce OID.
- Center the beam to the area of interest.
- Get the area of interest parallel to the beam or
film. - Restrain motion and communicate with the patient.
- Use short exposure times.
30Tools to Improve Quality
- Image receptors
- Use the correct film screen combination for the
examination. Extremity for wrist. Regular for
spines. - Intensifying screens reduce patient exposure at
least 20 times. - As the speed of the image receptor increases,
radiographic noise and contrast resolution
decreases.
31Tools to Improve Quality
- Image receptors
- Low contrast imaging procedures have wider
latitude and a larger margin of error in
producing acceptable radiographs. - Use the highest speed system that will provide
adequate contrast and density over the entire
spectrum of examinations.
32Tools to Improve Quality
- Technique selection
- We must select the optimum technical factors.
- We must get the exposure time as low as possible
so high frequency machine are important. - Contrast controlled by the kVp used
- Density controlled by the mAs used.
33Tools to Improve Quality
- kVp has a greater influence than mAs.
- Technique selection is a balancing act. Balancing
contrast, density and exposure.
34Chapter 20 Radiographic Technique
- Several factors influence the selection of
technical factors. The primary factors that
impact exposure and image quality are - SID
- mAs
- kVp
35Patient Factors
- The anatomic thickness and body composition
greatly impact the technical factors. - The technique chart is based upon the Sthenic
Body Type.
36Patient Factors
- Sthenic is strong active
- Hyposthenic is thin but healthy
- Hyperstenic is obese
- Astenic is small, frail sometime emaciated, and
often elderly
37Patient Thickness
- The thickness of the patient should not be
guessed. - It should be measured with calipers.
- Patient thickness is measured in cm.
38Body composition
- The type of tissue in the area of exposure will
impact the technical factors. - The tissue types in the chest are different from
the abdomen. - Disease processes will also impact the exposure
factors. Obtaining a good clinical history is
important. History must be communicated to the
radiographer.
39Classifying Pathology
- Radiolucent (Destructive)
- Active TB
- Atrophy
- Bowel obstruction
- Cancer
- Degenerative arthritis
- Emphysema
- Osteoporosis
- Pneumothorax
- Radiopaque Constructive)
- Aortic aneurysm
- Ascites
- Atelectasis
- Cirrosis
- Hypertrophy
- Metastasis
- Pleural Effusion
- Pneumonia
- Sclerosis
40Image Quality Factors
- Image quality factors include
- OD
- Contrast
- Image Detail
- Image Distortion
- OD is the optical density or radiographic
density. OD is controlled by the mAs and SID.
41Optical Density
- Numerically low OD is a low number like 0.25.
- Dark is a high number like 2.20 to 4.0
- Light is underexposed
- Dark is over exposed
- If density is the only factor that needs to be
changed, change the mAs.
42Optical Density
- A 30 change in mAs is needed to make a
perceptible change in optical density. - Usually when a change in optical density is
needed, the mAs is either doubled or halved. - kVp must be changed by 4 to produce the same
change in optical density. - Changing kVp will also impact penetration and
contrast.
4330 - 50 Rule
- If the film is under exposed, double the mAs.
- If the film is over exposed, cut the mAs in half.
- If the film is slightly underexposed, increase
the mAs 30. - If the film is slightly overexposed, reduce the
mAs 30.
4430 Density Change
- The lower image was the first image taken. It was
dark but normally would be acceptable. - The top image was the mAs reduced 30. The air
fluid levels in the sinus is easier to see.
4515 Rule
- The OD can be changed with kVp but it will also
impact exposure and contrast also. - Increase of 15 in kVp is equal to cutting the
mAs in half. - Decrease of 15 in kVp is equal to doubling the
mAs. - If the film is underexposed, increase kVp 15.
- If the film is overexposed, decrease kVp 15.
46Contrast
- The function of contrast is to make the anatomy
more visible. - Contrast is the difference in density of adjacent
structures. - The relative penetrability of the x-ray through
different tissues determines the image contrast.
47Contrast
- Contrast can be measured as the Gray Scale of
Contrast. It is the range of optical density from
white to black on the image. - Contrast is controlled by kVp.
48Adjusting Contrast with 15 Rule
- An 15 increase in kVp and a reduction of mAs by
50 will produce the same OD but lower contrast. - Used to reduce exposure or reduce exposure time/
- An 15 decrease in kVp and doubling the mAs will
produce the same OD but higher contrast.
49Image Detail
- The sharpness of image detail refers to the
ability to see structural lines or borders of
tissue in the image. - The visibility of image detail is best measured
by the contrast resolution. - The geometric factors of focal spot selection,
SID and OID will impact sharpness.
50Image Detail
- Visibility of image detail is impacted by
factors such as image fog. - Scatter radiation reduces the ability to
visualize lines of detail. - Light fog or processing can impact the visibility
of structures. - Collimation, screen combination and the use of a
grid are other factors that impact image detail.
51Distortion
- The position of the x-ray tube greatly impacts
distortion of the image. The image may be
elongated or foreshortened. - The proper Positioning of the tube, anatomic part
and image receptor greatly impacts distortion.
52Types of Technique Charts
- There are four primary means to establish
techniques. - Variable kVp Fixed mAs
- Fixed kVp with varying mAs.
- High kVp with varying mAs
- Automatic Exposure Charts when AEC is used.
53Variable kVp Charts
- The mAs is fixed and the kVp is varied based upon
patient thickness. - Usually by a formula such as 2 x thickness 30
kVp for single phase - 24 cm patient 2423078kVp
- For high frequency use 23 and for three phase
use 25. - Small patient used low kVp high contrast
- Large patient used high kVp low contrast
54Variable kVp Charts
- Contrast was very inconsistent.
- Very little latitude on smaller patients.
- Higher radiation exposure
- This type of chart should be avoided.
55Fixed kVp Technique
- kVp is fixed and mAs varies by patient thickness.
Usually 30 per two cm. - Uses Optimum kVp for the body part
- Contrast is constant.
- Wider latitude
- Lower exposure
56Fixed kVp Technique Variations
- High kVp technique uses over 100 kVp
- No longer used for bone.
- Long ago used for spine but images are too gray.
Low contrast - Mostly used for chest and barium contrast studies.
57Fixed kVp Technique Variations
- Automatic Exposure Technique Charts
- Uses optimum kVp and high backup mAs.
- Ion chamber or photo cell determines when correct
density is achieved on film and terminates
exposure. - Proper positioning is critical to get the area of
interest over the ion chamber.
58Using the Technique Chart
- The chart is not the Bible but is a guide.
- Works about 85 of the time so it is a great
starting point. - Lists factors used for each view based upon
measurement of the patient. - Can include as much as you want to include.
59Using the Technique Chart
- Recommended factors for chart
- optimum kVp for view
- mAs based upon cm measurement
- filters used
- SID tube angle used
- Bucky or non-Bucky
- Cassette film type
60Using the Technique Chart
- Charts should be
- accessible
- easy to read
- not hand written
- based upon the type of machine and machine
controls. - mAs or mA and time
61Technique Variables
- Variable machine electrical output
- Incoming power and ability of machine to
compensate for variations in incoming power. - Type of High-voltage Power
- Single phase to High Frequency reduce mAs 50
- High Frequency to Single phase double mAs
- Grid ratio
- Non-Bucky Holder
62Technique Variables
- Variable machine electrical output
- Relative Speed Value of cassettes film
combination. - 400 speed to 200 speed double mAs
- 200 speed to 400 speed reduce mAs 50
63mAs kVp Relationship
- There are some basic rules for mAs and kVp that
are used to adjust the technical factors. - Remember x-rays are like toast.
- Dark is too dense
- Light has inadequate density
- This tells you which was to go.
64mAs Rules
- Since mAs controls density, it is usually used to
adjust density. - 30 increase needed to make a noticeable change
in density. - 50 mAs reduction will reduce density 50
- Doubling mAs will doubles density.
65mAs Rules
- If image is too dark reduce mAs 50.
- If image is too light double mAs.
- Doubling mAs can be done by doubling mA or time.
- Doubling time increases chance for motion blur.
66kVp Rule
- kVp will also change density.
- A light film from low kVp is called under exposed
or under penetrated. - Very white image because no x-rays reached the
film. - Too dark is over exposed, some say over
penetrated. They are different.
67kVp Rule
- Over penetrated will result only if the kVp used
is too high for the view. It will be dark and
very flat (lacking contrast) - Density is very sensitive to changes in kVp.
- A 2 kVp (HF) to 4 kVp change is noticeable. About
4. - The 15 rule works with density adjustment.
68kVp Rule
- 15 increase in kVp will double density.
- 15 decrease in kVp will reduce density 50.
- 15 increase in kVp doubling mAs
- 15 decrease in kVp half the mAs
- 10 kVp 15 change in the 60 to 90 kVp range.
69Optimum kVp
- Optimum kVp will provide the best contrast with
the least amount of radiation. - If using the optimum kVp you should not need to
adjust kVp. - kVp can be changed based upon body habitus and
disease.
70Optimum kVp
- Small Extremity
- Large Extremity
- Cervical Spine AP or Lat
- APOM
- Thoracic AP
- Thoracic Lat
- Lumbar AP
- Lumbar Oblique
- Lumbar Lateral
- Pelvis
- Abdomen
- Ribs
- Chest
- 55-65
- 65-70
- 70-74
- 75-78
- 75
- 80
- 74
- 80
- 90
- 80
- 70
- 70
- 110
71Dark film
- Dark No contrast reduce kVp no change to mAs
- Black no structures seen reduce both.
- Dark look at the optimum kVp range.
- If reducing kVp goes beyond optimum kVp
- reduce mAs 50
72Light Film
- If film is so light that no structures are seen
then it is under penetrated so increase kVp. - kVp controls penetration.
- If structures seen but lacks density increase
double mAs.
73Patient Factors
- Very muscular or large boned increase mAs by 50
- Very muscular and large boned increase mAs 50
and kVp 4 to 6 kVp. - Obese increase mAs 50
- Edema increase mAs 30
- Frail decrease kVp 5 to 15
74Patient Factors
- Osteoporotic patient over 60 years old decrease
mAs 30 to 50 - 6 to 12 years old reduce mAs 30 to 50
- Infants to 6 years old decrease mAs 75
75Item that Affect Detail
- Spatial Resolution controlled by focal spot size
and image receptor. - Detail influenced by
- SID
- OID
- Motion Blur
- Density Contrast of Image
76Items that Affect OD
- Optical density is controlled by mAs
- OD influenced by
- kVp
- SID
- Thickness
- Density
- Collimation
77Items that Affect OD
- OD influenced by
- Grid Ratio
- Development time and temperature
- Image receptor speed
78Items that Affect Contrast
- Contrast controlled by kVp
- Contrast influenced by
- mAs
- Development Time Temperature
- Collimation
- Grid ratio
- Image receptor
79End of Lecture