Title: Scanning Electron Microscope
1Scanning Electron Microscope
2(No Transcript)
3The basic premise of an SEM is that signal
produced from a scanned area of the specimen is
displayed as an image with the exact same scan
pattern on a CRT
4The scan pattern on the specimen is created by a
set of deflection coils in the column that move
the beam in a coordinated X/Y pattern. This is
referred to as a scan or raster pattern
5Cathode Ray Tube accelerates electrons towards
the phosphor coated screen where they produce
flashes of light upon hitting the phosphor.
Deflection coils create a scan pattern forming an
image in a point by point manner
6Color CRTs usually have three separate e-guns,
one each for red, green, and blue (RGB)
7(No Transcript)
8The scan generator coordinates the movement of
the primary beam with the movement of the e-gun
in the back of the CRT
9Magnification is accomplished by scanning a
progressively smaller portion of the specimen and
displaying the image on the CRT. Thus total
magnification is square area of CRT divided by
area scanned.
10In contrast focus is accomplished by bringing the
beam to its crossover point on the surface of the
specimen. In this way focus and magnification
are completely separate from one another in the
SEM.
11In the TEM the specimen lies very close to the
objective lens resulting in a relatively large
half angle of illumination. In SEM since the
image is not formed by an objective lens the half
angle can be very small resulting in a large
depth of field.
1210X
An SEM focused at high magnification will still
be in focus at low magnification
13110X
14200X
15400X
164K
1716K
1845K
19Weak Lens Larger probe size, low resolution,
long working distance, and larger depth of field
Strong Lens Small probe size, high resolution,
short working distance and shallow depth of field
20A smaller final lens aperture can reduce the half
angle and therefore increase the depth of field.
This is true on a relatively strong lens which
has a fairly short working distance and therefore
high resolution.
21The SEM forms an image by generating a number of
signals as a result of the beam interacting with
the specimen.
22The SEM is a probe forming (e- beam) and signal
detecting device. By developing an image created
in a point by point fashion an important factor
is the signal to noise (S/N) ratio. The signal
being the result of the beam interacting with the
specimen and the noise being the result of
imperfections in the electronics of the detector
and display systems as well as spurious signal.
23Signal can be increased by Creating more beam
specimen interactions Noise can be reduced
by Cooling electronics Keeping detectors
settings to a minimum Signal/noise ratio can be
increased by Placing detector closer to source
of signal Slowing down the scan (collect more
signal per unit time)
24Although the same amount of signal is produced
throughout the specimen the topography of the
surface will allow differing amounts of signal to
reach a detector placed off to the side.
25A number of different detectors can be
incorporated into the chamber surrounding the
specimen.
26The shadow produced in an SEM is determined by
the position of the detector but the view is a
beams eye view as if column
one were looking down the
27LEO Gemini Column
A detector placed within the column is known as
an in-lens detector and produces a very
different image compared to a conventionally
located detector
28Secondary Electron Detector
29Secondary Electron Detector
Side Mounted In-Lens
30MIRA SEM
31MIRA SEM