Medical Physicist Requirements - PowerPoint PPT Presentation

1 / 5
About This Presentation
Title:

Medical Physicist Requirements

Description:

Table 3- Medical Physicist s Radiographic QC Survey Requirements Standard Test Item As required at N.J.A.C. 7:28-15.3 Radiographic Unit Assembly Evaluation – PowerPoint PPT presentation

Number of Views:93
Avg rating:3.0/5.0
Slides: 6
Provided by: porl7
Category:

less

Transcript and Presenter's Notes

Title: Medical Physicist Requirements


1
Table 3- Medical Physicists Radiographic QC Survey Requirements Table 3- Medical Physicists Radiographic QC Survey Requirements
Item Test Standard
1 Radiographic Unit Assembly Evaluation As required at N.J.A.C. 728-15.3
2 Collimation Assessment As required at N.J.A.C. 728-15.3
3 Collimator Illumination As required at N.J.A.C. 728-15.3
4 Half Value Layer As required at N.J.A.C. 728-15.3
5 MA Exposure Linearity As required at N.J.A.C. 728-15.3
6 KVp Accuracy/Reproducibility As required at N.J.A.C. 728-15.3
7 Timer Accuracy/Reproducibility As required at N.J.A.C. 728-15.3
8 AEC, Reproducibility, Tracking, Density Control As required at N.J.A.C. 728-15.3
9 Entrance Skin Exposure (ESE) Measurement Determine ESE for common exam and
10 Image Quality Evaluation (Recommendation) Established standard for phantom test tool used
11 Review Facility/Technologist QC Test Records Review QC tests for proper procedure and corrective action
12 Physicist Report and Recommendations Communicate results and recommendations to registrant
2
Medical Physicist Requirements
  • Pursuant to N.J.A.C. 728-22.8(f)5i, the
    medical physicist shall prepare a report that
    reviews the overall quality assurance program
    being carried out by the registrant and contains
    raw data, results and recommendations of the
    medical physicist's equipment tests performed in
    accordance with items 1 through 10 in Table 4,
    Medical Physicist's Radiographic QC Survey.

3
Translation
  • The medical physicist is in violation of N.J.A.C.
    N.J.A.C. 728-22.8(f)5i if any MPQC tests are not
    performed or are performed improperly
  • 1st offense Administrative Order
  • 2nd offense AO and penalty assessment of 300.00

4
Medical Physicist Requirements
  • Pursuant to N.J.A.C. 728-22.8(f)5ii, the
    medical physicist shall prepare a report that
    reviews the overall quality assurance program
    being carried out by the registrant and contains
    results and recommendations of the medical
    physicist's review performed in accordance with
    item 11 in Table 4, Medical Physicist's
    Radiographic QC Survey.

5
Translation
  • Medical physicist is in violation of N.J.A.C.
    728-22.8(f)5ii. if they fail to included in
    their summary and deficiencies in facility's QA
    program and recommended corrective actions
  • 1st offense Administrative Order
  • 2nd offense AO and penalty assessment of 500.00
Write a Comment
User Comments (0)
About PowerShow.com