Nuclear Magnetic Resonance II - PowerPoint PPT Presentation

1 / 41
About This Presentation
Title:

Nuclear Magnetic Resonance II

Description:

Emphasizing the differences among spin density, T1, and T2 relaxation time ... Steady-state precession with short TR ( 50 msec) ... – PowerPoint PPT presentation

Number of Views:65
Avg rating:3.0/5.0
Slides: 42
Provided by: michae354
Category:

less

Transcript and Presenter's Notes

Title: Nuclear Magnetic Resonance II


1
Nuclear Magnetic Resonance II
  • Pulse sequences

2
Pulse sequences
  • Emphasizing the differences among spin density,
    T1, and T2 relaxation time constants of the
    tissues is the key to contrast sensitivity in MR
    images
  • Tailoring the pulse sequences timing, order,
    polarity, and repetition frequency of the RF
    pulses and applied magnetic field gradients
    makes the emitted signals dependent on T1, T2 or
    spin density relaxation characteristics

3
Pulse sequences (cont.)
  • MR relies on three major pulse sequences
  • Spin echo
  • Inversion recovery
  • Gradient recalled echo
  • Used in conjunction with localization methods
    (i.e., the ability to spatially encode the signal
    to produce an image), contrast-weighted images
    are obtained

4
Spin echo
  • Spin echo describes the excitation of the
    magnetized protons in a sample with an RF pulse
    and production of the FID, followed by a second
    RF pulse to produce an echo
  • Timing between the RF pulses allows separation of
    the initial FID and the echo and the ability to
    adjust tissue contrast

5
(No Transcript)
6
(No Transcript)
7
Time of repetition
  • The standard spin echo pulse sequence uses a
    series of 90-degree pulses separation by a period
    known as the time of repetition (TR), which
    typically ranges from about 300 to 3,000 msec
  • A time delay between excitation pulses allows
    recovery of the longitudinal magnetization

8
Partial saturation
  • After the TR interval, the next 90-degree pulse
    is applied, but usually before the complete
    longitudinal magnetization recovery of the
    tissues
  • The FID generated is less than the first FID
  • After the second 90-degree pulse, a steady-state
    longitudinal magnetization produces the same FID
    amplitude from each subsequent 90-degree pulse
  • Tissues become partially saturated, with the
    amount of saturation depending on the T1
    relaxation time

9
(No Transcript)
10
Spin echo contrast weighting
  • Contrast in an image is proportional to the
    difference in signal intensity between adjacent
    pixels in the image
  • The signal, S, produced by an NMR system is
    proportional to other factors as follows
  • ?H is the spin (proton) density
  • f(v) is the signal arising from fluid flow

11
(No Transcript)
12
Spin echo contrast weighting
  • The equation shows that for the same values of TR
    and TE (i.e., for the same pulse sequence),
    different values of T1 or T2 (or of spin density
    or flow) will change the signal S
  • The signal in adjacent voxels will be different
    when T1 or T2 changes between these voxels
  • By changing the pulse sequence parameters TR and
    TE, the contrast dependence in the image can be
    weighted toward T1 or toward T2

13
T1 weighting
  • A T1-weighted spin echo sequence is designed to
    produce contrast chiefly based on the T1
    characteristics of the tissues by de-emphasizing
    T2 contributions
  • This is done using relatively short TR to
    maximize the differences in longitudinal
    magnetization during the return to equilibrium,
    and a short TE to minimize T2 dependency during
    signal acquisition
  • Tissues with shorter T1 have higher image
    intensity

14
(No Transcript)
15
(No Transcript)
16
Spin (proton) density weighting
  • Image contrast with spin density weighting relies
    mainly on differences in the number of
    magnetizable protons per volume of tissue
  • To minimize T1 differences of the tissues, a
    relatively long TR is used
  • Signal amplitude differences in the FID are
    preserved with a short TE, so the influences of
    T2 are minimized
  • Tissues with higher spin density (e.g., fat, CSF)
    have higher image intensity

17
(No Transcript)
18
(No Transcript)
19
T2 weighting
  • T2 weighting follows directly from the spin
    density weighting sequence reduce T1 effects
    with a long TR, and accentuate T2 differences
    with a longer TE
  • Compared with a T1-weighted image, inversion of
    tissue contrast occurs, because short-T1 tissues
    usually have a short T2, and long-T1 tissues
    usually have a long T2

20
(No Transcript)
21
(No Transcript)
22
Spin echo parameters
  • T1 contrast TR 400-600 msec, TE 5-30 msec
  • Spin density contrast TR 1,500-3,000 msec, TE
    5-30 msec
  • T2 contrast TR 1,500-3,000 msec, TE 60-150
    msec
  • For conventional spin echo sequences, both a spin
    density and a T2-weighted contrast signal are
    acquired during each TR by acquiring two echoes
    with a short TE and a long TE

23
(No Transcript)
24
Inversion recovery
  • Inversion recovery, also known as inversion
    recovery spin echo, emphasizes T1 relaxation
    times of the tissues by extending the amplitude
    of the longitudinal recovery by a factor of two
  • An initial 180-degree RT pulse inverts the Mz
    longitudinal magnetization to Mz
  • After a delay time known as the time of inversion
    (TI), a 90-degree RF pulse rotates the recovered
    fraction of Mz spins into the transverse plane to
    generate the FID

25
Inversion recovery (cont.)
  • A second 180-degree pulse at time TE/2 produces
    an echo signal at TE
  • The echo amplitude of a given tissue depends on
    TI, TE, TR, and the magnitude of Mz
  • The signal intensity is approximated as follows

26
(No Transcript)
27
(No Transcript)
28
Inversion recovery (cont.)
  • Inversion recovery sequence produces negative
    longitudinal magnetization that results in
    negative (in phase) or positive (out of phase)
    transverse magnetization when a short TI is used
  • Actual signals encoded in one of two ways
  • A magnitude (absolute value) signal flips
    negative Mz values to positive values
  • A phase signal preserves the full range of the
    longitudinal magnetization from Mz to Mz

29
Short tau inversion recovery
  • Short tau inversion recovery, or STIR, is a pulse
    sequence that uses a very short TI and magnitude
    signal processing
  • Materials with short T1 have a lower signal
    intensity (the reverse of a standard T1-weighted
    image)
  • All tissues pass through zero amplitude (Mz 0),
    known as the bounce point or tissue null
  • Judicious TI selection can suppress a given
    tissue signal (e.g., fat)

30
Fluid attenuated inversion recovery
  • The use of longer TI reduces the signal levels of
    CSF and other tissues with long T1 relaxation
    constants
  • Fluid attenuated inversion recovery, or FLAIR,
    reduces CSF signal and other water-bound anatomy
    in the MR image by using a TI selected at or near
    the bounce point of CSF

31
(No Transcript)
32
(No Transcript)
33
(No Transcript)
34
(No Transcript)
35
Gradient recalled echo
  • The gradient recalled echo (GRE) technique uses a
    magnetic field gradient to induce the formation
    of an echo, instead of the 180-degree pulses
    discussed earlier
  • The GRE is not a true spin echo technique but a
    purposeful dephasing and rephasing of the FID
  • Magnetic field inhomogeneities and tissue
    susceptibilities are emphasized in GRE

36
(No Transcript)
37
GRE and flip angles
  • Depending on the desired contrast, flip angles of
    a few degrees to more than 90 degrees are applied
  • With short TR, smaller flip angles are used
    because there is less time to excite the spin
    system, and more transverse magnetization is
    actually generated with smaller flip angles
  • This occurs because magnetization quickly builds
    up in the tissue sample
  • Tissue contrast in GRE pulse sequences depends on
    TR, TE, and the flip angle

38
(No Transcript)
39
GRE with long TR (gt200 msec)
  • For the GRE sequence with long TR and flip
    angles greater than 45 degrees, contrast
    behaviour is similar to that of spin echo
    sequences
  • Major difference is the generation of signals
    based on T2 contrast rather than T2 contrast
  • In terms of clinical interpretation, the
    mechanisms of T2 contrast are different from
    those of T2, particularly for MRI contrast agents
  • A long TE tends to show the differences between
    T2 and T2

40
Steady-state precession with short TR (lt50 msec)
  • Flip angle has the major impact on the contrast
    weighting of the resultant images
  • Small flip angles spin density weighting is the
    single most important factor
  • Moderate flip angles contrast depends on the
    difference in T2/T1 ratio between the tissues
  • Large flip angles contrast depends on T2 and
    T1 weighting an increase in TE contributes to
    the T2 contrast in the image

41
(No Transcript)
Write a Comment
User Comments (0)
About PowerShow.com