Title: Design and simulation of micro-SPECT:
1- Design and simulation of micro-SPECT
- A small animal imaging system
-
Freek Beekman and Brendan Vastenhouw Section
tomographic reconstruction and instrumentation Im
age Sciences Institute University Medical Center
Utrecht
2- PRESENTATION OUTLINE
- Introduction in tomography
- Tomography with labeled molecules (tracers).
- Principles of SPECT
- Image reconstruction
- Ultra-high resolution SPECT for imaging small
laboratory animals gt Need for high resolution
gamma detectors
3 Computed Tomography
- Cross-sectional images of the local X-ray
attenuation in an object are reconstructed from
line integrals of attenuation (projection data)
using a computer
1979 Hounsfield and Cormack share Nobel Prize..
4Why Computed Tomography ?
5We are curious how we, other people, animals,
etc, look inside...
but we dont like to (be) hurt !
6- Examples of Tomography
- Anatomy
- X-ray Computed Tomography
- Magnetic Resonance Imaging (MRI)
- Molecule distributions
- Positron Emission Tomography (PET)
- Single Photon Emission Computed Tomography
(SPECT)
7- X-ray CT Cross-sectional images of X-ray
attenuation provide knowledge about anatomy
8We are also curious how organs...
..are functioning in vivo
9- Molecular imaging
- Emission tomographs (PET and SPECT) are suitable
in vivo imaging of functions (blood perfusion,
use of oxygen and sugar, protein concentrations)
- Uses low amounts of injected radiolabeled
molecules
10What area in the brain is responsible for a task?
PET and SPECT imaging enables mapping of of
radiolabeled molecule distributions
11SPECT Single Photon Emission Computed
Tomography
- Patient is injected with a molecule labeled with
a gamma emitter. - For determination of travel direction detectors
are equipped with a lead collimator.
12Collimated gamma-camera
IIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIII
lt Lead collimator
Detector gt
- To form an image, the travel direction of
detected photons must be known. - The collimator selects ?-quanta which move
approximately perpendicular to the detector
surface.
13lt Slice of Tc-99m distribution
Slice of SPECT image gt
- Slices are reconstructed (Filtered Back
Projection (FBP) or Iterative Reconstruction). - Resolution in humans 6-20 mm
- Resolution can be much better in small animals (lt
1 mm)
14SPECT Technetium-99m Cardiac Perfusion Image
15IMAGE RECONSTRUCTION FROM PROJECTIONS Analytical
(Radon Inversion) Discrete (Statistical) Methods
16SPECT reconstruction problem
p M a n b ltgt p j Mjiai nj
bj ai activity in voxel i pj projection data
in pixel j bj back-ground in pixel j (e.g.
scatter) nj noise in pixel j Mji probability
that photon is emitted in voxel I is detected in
pixel j. Attenuation, detector blur and
scatter can be included. Estimate a from above
equation
17 - SPECT reconstruction matrix
- is complicated by
- Detector blurring
- Attenuation
- Scatter
- 3D reconstruction
18 Iterative Reconstruction illustrated
Object space
Projection space
Estimated projection
Current estimate
- Simulation (or
- re-projection)
Compare e.g. - or /
Measured projection
Update
Error projection
Object error map
Back- projection
19Example iteration process ML-EM reconstruction
brain SPECT
0 iterations
10 iterations
30 iterations
60 iterations
20 line integral model
accurate PSF-model
21Small animal molecular imaging using single
photon emitters (micro-SPECT)
22Expected contribution of micro-SPECT to science
- Partly replacement of sectioning, counting and
autoradiography. - Reduction of number of animals required
- Dynamic and longitudinal imaging in intact
animals - Contribution to understanding of gene functions
- Acceleration of pharmaceutical development
- Breakthroughs in areas like cardiology,
neurosciences, and oncology - Extension of micro-SPECT technology to clinical
imaging (2006)
23In Vivo Nuclear Microscopy
- (Eur J. Nucl. Med and Mol. Im., in press)
- Golden micro-pinholes
- gt Super High Resolution
SEM image of gold alloy pinhole
24Mouse thyroid
Microscopic slide
I-125 pinhole image
1 mm
20 min. acquisition
arrows indicate locations parathyroid glands
25Pinhole imaging geometries for small animal
imagingSPECT(micro-SPECT)
26- Spatial resolution clinical SPECT 15 mm
- Spatial resolution current small animal SPECT
and PET 1.0-2.5 mm - Micro-SPECT dedicated small animal SPECT.
- with resolution 0.2-0.4 mm
-
- Effect of Resolution on Rat Brain phantom
2 mm 1 mm 0.5 mm 0.25 mm
0 mm
27State-of-the-art pinhole SPECT
A-SPECT two pinholes. Mouse rotates in tube
Thyroid of mouse (I-125)
Mouse bone scan (Tc-99m)
28Micro-SPECT
29Simulations A-SPECT vs. Micro-SPECT
Â
Truth
A-SPECT
Micro-SPECT
30Finally We need a ready set of detectors plus
associated electronicsSolid state?
30 mm
- SPECIFICATIONS
- Energies of 30-140keV
- Counting mode
- Capture efficiency gt80 _at_140keV
- Spatial resolution 200 microns
- Energy resolution (10-20)
10 mm
- Contact Freek beekman f.beekman_at_azu.nl
- 31 30 250 7779
- We need approx. 40 detector elements.