Title: Advances in Bioscience Education Summer Workshop
1Advances in Bioscience Education Summer Workshop
- Immunolabeling for
- Fluorescence and Electron Microscopy
- June 27 - 29, 2006
- Biological Electron Microscope Facility
- Pacific Biosciences Research Center
- University of Hawaii at Manoa
2Biological Electron Microscope Facility
- Pacific Biosciences Research Center, University
of Hawaii at Manoa - Instrumentation, service and training
- State-of-the-art instruments for biological
microscopy - In operation since 1984
- Personnel
- Dr. Richard D. Allen, Director
- Dr. Marilyn F. Dunlap, Manager
- Tina M. (Weatherby) Carvalho, M.S., Supervisor
3Light and Electron Microscopy
- Light microscopy
- Glass lenses
- Source of illumination is usually light of
visible wavelengths - Tungsten bulb
- Mercury vapor or Xenon lamp
- Laser
- Electron microscopy
- Electromagnetic lenses
- Source of illumination is electrons
- Hairpin tungsten filament (thermionic emission)
- Pointed tungsten crystal (cold cathode field
emission) - Lanthanum hexaboride
4Epifluorescence Microscopy
- Olympus BX51 upright microscope
- Broad-band epifluorescence excitation and
detection - DIC optics
- Optronics scientific grade digital camera
5Epifluorescence
- Green photos courtesy Dr. Teena Michaels, KCC
- Red photo courtesy Dr. Claude Jourdan-LeSaux
6Common Fluorescence Applications
- Localize/identify specific organelles
- Detect live cells vs. dead cells, necrotic vs.
apoptotic cells - Determine cell membrane permeability
- Localize antigen-specific molecules
- Multiple labeling
7Laser Scanning Confocal Microscope
- Olympus Fluoview FV1000
- Three colors Trans-mitted simultaneously
- Excitation with 405, 458, 488, 515, 543, and 633
nm lasers - Various emission filters
- Optical sectioning
- 3-D reconstruction
- Stereo views
- Animations
8Laser Scanning Confocal Microscopy
- Adjustable pinhole aperture eliminates
out-of-focus glare - Better resolution
- Serial optical sections can be collected from
thick specimens - Live or fixed cell and tissue imaging
Drosophila eye
Photo courtesy of Gregg Meada Dr. Gert DeCouet,
UHM
9Epifluorescence vs. Confocal
Sample courtesy Gregg Meada Dr. Gert DeCouet,
UHM
10Field Emission Scanning Electron Microscopy
(FESEM)
- Hitachi S-800 FESEM
- High magnification (40x to 300,000x)
- High resolution (better than 2 nm)
- Easy to learn
- Hi-res digital images
- Prep equipment critical point dryer, sputter
coater
11SEM Images
12Transmission Electron Microscopy(TEM)
- Zeiss 10/A conventional TEM
- Excellent for training
- Film only
13LEO 912 Energy-Filtering TEM
- In-column energy filter (electromagnetic prism)
- Ultrathin to 0.5 µm sections
- Contrast tuning
- Elemental analysis with electron energy loss
spectroscopy (EELS) - Elemental mapping with electron spectrographic
imaging (ESI) - Eucentric goniometer stage
- Digital images
14Conventional TEM Micrographs
Bacteria in cell
Apoptosis
Skin
Chloroplast
Collagen
Virus in cell
15Negative Staining
- Viruses, small particles, proteins, molecules
- No sectioning
- Same day results
16EFTEM - Electron Spectrographic Imaging (ESI) -
elemental mapping
- Calcium in mitochrondria from ischemic brain
- Iron in liver
17EFTEM- Electron Energy Loss Spectroscopy (EELS)
18Ultramicrotomy
- Ultrathin (60-90 nm) sectioning of resin-embedded
specimens - Several brands/models available
- Cryoultramicrotomy
19Cryotechniques
- Ultrarapid cryofixation
- Metal mirror impact
- Liquid propane plunge
- Freeze fracture with Balzers 400T
- Cryosubstitution
- Cryoultramicrotomy Ultrathin frozen sections
(primarily for antibody labeling)
20Cryo Examples
- Freeze fracture, deep-etch, rotary shadow
- Cryosection/im-munogold label
- Cryosubstitution
21Image Manipulation and Analysis
- Soft Imaging System analySIS professional
software - EFTEM acquisition and analysis
- Light Microscopy
- Images from other sources
- Particle counting and analysis
- Feature extraction
- Image and results database
22Immunolocalization
- LM
- Fluor/confocal
- TEM
- SEM with backscatter detector
23Approaches to Immunolabeling
- Direct Method Primary antibody contains label
- Indirect Method Primary antibody followed by
labeled secondary antibody - Amplified Method Methods to add more reporter
to labeled site - Protein A Method May be used as secondary
reagent instead of antibody
24Direct Labeling Method
- Labeled primary antibody reacts directly with the
antigen in the histological or cytological
preparation
25Two-step Indirect Method
- Fluorescent-conjugated secondary antibody
attaches to primary antibody that is bound to
antigen
26Amplified Method
- If the antibody reporter signal is weak, the
signal can be amplified by several methods, e.g.,
streptavidin-biotin complex
27Double-labeling Method
- Use primary antibodies derived from different
animals (e.g., one mouse antibody and one rabbit
antibody) - Then use two secondary antibodies conjugated with
reporters that can be distinguished from one
another
28Immunolabeling for Transmission Electron
Microscopy
- Normally do Two-Step Method
- Primary antibody applied followed by colloidal
gold-labeled secondary antibody - May also be enhanced with silver
- Can also do for LM
29Preparation of Biological Specimens for
Immunolabeling
- The goal is to preserve tissue as closely as
possible to its natural state while at the same
time maintaining the ability of the antigen to
react with the antibody - Chemical fixation of whole mounts prior to
labeling for LM - Chemical fixation, dehydration, and embedment in
paraffin or resin for sectioning for LM or TEM - Chemical fixation for cryosections for LM
- Cryofixation for LM or TEM
30Chemical Fixation
- Antigenic sites are easily denatured or masked
during chemical fixation - Glutaraldehyde gives good fixation but may mask
antigens, plus it is fluorescent - Paraformaldehyde often better choice, but results
in poor morphology , especially for electron
microscopy - May use e.g., 4 paraformaldehyde with 0.5
glutaraldehyde as a good compromise
31Preembedding or Postembedding Labeling
- May use preembedding labeling for surface
antigens or for permeabilized cells - The advantage is that antigenicity is more likely
preserved - Postembedding labeling is performed on sectioned
tissue, on grids, allowing access to internal
antigens - Antigenicity probably partially compromised by
embedding
32Steps in Labeling of Sections
- Chemical fixation
- Dehydration, infiltration, embedding and
sectioning - Optional etching of embedment, permeabilization
- Blocking
- Incubation with primary antibody
- Washing
- Incubation with secondary antibody congugated
with reporter (fluorescent probe, colloidal gold) - Washing, optional counterstaining
- Mount and view
33Controls! Controls! Controls!
- Omit primary antibody
- Irrelevant primary antibody
- Pre-immune serum
- Perform positive control
- Check for autofluorescence
- Check for non-specific labeling
- Dilution series
34Dilutions are Important
- Typically should do an extensive dilution series
to determine best concentration of both primary
and secondary antibodies - This shows an antibody at concentrations of 1100
and 12000
35Know Your Artifacts
- And use them to your advantage!
- Green is label orange-red is autofluorescence
- Acts as counterstain
36Autofluorescence
- Need to select label that will be readily
distinguished from autofluorescence - Several techniques to quench autofluorescence
37What is a Microscope?
- A tool that magnifies and improves resolution of
the components of a structure - Has three components one or more sources of
illumination, a magnifying system, and one or
more detectors - Light microscopes use a beam of light for
illumination and include fluorescence and
confocal microscopes - Electron microscopes use electrons as a source of
illumination and include transmission and
scanning electron microscopes
38Light and Electron Microscopes
- Lenses are used to control a beam of
illumination, magnify, and direct an image to a
detector
39Light Microscopes
40Objective Lenses
- Objective lens choice is important!
- Not all objective lenses are created equal
- The more correction a lens has, the less
transmission - Resolution is dictated by Numerical Aperture (NA)
- Talk to your microscope company representative
41Light Microscopes - Resolution
- Resolution depends on the light gathering of the
objective, which depends on the NA, and on the
light path, which includes the slide, sample,
mounting medium, coverslip, and air or immersion
oil
42Light Path in Fluorescence
- Light delivered through excitation filter and
then objective lens to specimen where it is
absorbed emitted light goes back through
objective lens through barrier filter and
emission filter and then to detector.
43Fluorescence Microscopes
- Illumination light path is the same as the
sampling light path - Need to maximize the light throughput in both
directions no more than 22 of light will be
detected on a good day - Need to match refractive indices (RI)
- Use the best optics with the fewest elements
44Optical Choices for Fluorescence
- Minimize the number of lens elements to increase
light throughput, but correct for spherical
aberration - Optimize magnification and NA best choice often
a 60X 1.4NA plan objective - Only use magnification required to collect the
information needed - Use a mercury lamp for normal work and a xenon
lamp for quantitative studies
45Kohler Illumination
- Kohler illumination is essential for good
transmitted light contrast - Focus slide
- Close field diaphragm
- Focus diaphragm in field by adjusting condenser
height - Center diaphragm in field
- Open diaphragm to fill field and recheck
centration - Adjust iris diaphragm (on condenser) to taste
(affects contrast and depth of focus)
46Elements of Fluorescence Microscope
- Light source
- Mercury vapor
- Xenon
- Laser
- Optical lenses
- Optical filters
- Detection system
- Eye
- Film camera
- Digital camera
- Photomultiplier tube (PMT)
47Fluorescence
- Photons of a certain energy excite the
fluorochrome, raising it to a higher energy
state, and as it falls back to its original
state it releases energy in the form of a photon
of lower energy than the excitation energy.
48Fluorescence
- Fluorochromes are excited by specific wavelengths
of light and emit specific wavelengths of a lower
energy (longer wavelength)
49Filter Cubes for Fluorescence
- Filter cubes generally have an excitation filter,
a dichroic element, and an emission filter - The elements of a cube are selected for the
excitation and fluorescence detection desired
50Classification of Filters
- Long pass passes longer wavelengths
- Short pass passes shorter wavelengths
- Band pass passes defined wavelengths
- Dichromatic mirror transmits long wavelengths,
reflects shorter wavelengths
51Choose Fluorochrome/Filter Combos
52Spectral Characteristics of Probes
- Omega Filters Curv-o-Matic
- http//www.omegafilters.com/front/curvomatic/spect
ra.php - Other filter and microscope companies
53Ideal Fluorochrome
- Small size must get into cell
- High absorption maximum sensitive to excitation
- Narrow absorption spectrum excited by a narrow
wavelength - High quantum efficiency likely to fluoresce
- Narrow emission spectrum so you can find it
specifically - Large Stokes shift emission curve far enough
away from excitation curve to minimize
bleedthrough
54Types of Fluorochromes
- Simple dyes
- Acridine orange, DAPI, Propridium iodide, Lucifer
yellow - Physiological probes
- Calcium green, Rhodamine 123, Fluorescein
diacetate - Specific probes
- Phalloidin, Lectins, GFP, Primary and secondary
antibodies
55Laser Scanning Confocal Microscopy
- Fluorescence technique
- Uses laser light for excitation
- Improves image resolution over conventional
fluorescence techniques - Optically removes out-of-focus light and detects
only signal from focal plane - Can construct an in-focus image of considerable
depth from a stack of images taken from different
focal planes of a thick specimen - Can then make a 3-D image that can be tilted,
rotated, and sliced
56Principal Light Pathway in Confocal Microscopy
- Laser light is scanned pixel by pixel across the
sample through the objective lens - Fluorescent light is reflected back through the
objective and filters (dichroic mirrors) - Adjustable pinhole apertures for PMTs eliminate
out-of-focus flare - Image is detected by photomultiplier(s) and
digitized on computer
57Compressed Z-stack Image
- 3-D reconstruction
- Tilt and rotate
- Stereo projection
- Animation
- Montage
- Image enhancement
Photo courtesy Dr. Alex Stokes, Queens Medical
Center
58Confocal Movies
Photo courtesy Dr. Alex Stokes, Queens Medical
Center
59Confocal Projects
- Investigation of Wnt pathways in sea urchin
gastrulation (Dr. Christine Byrum/Dr. Athula
Wikramanayake) - Localization of transmembrane proteins in airway
smooth muscle cells (Dr. Lynn Iwamoto, Kapiolani) - GFP in drosophila (Gregg Meada/Dr. Gert deCouet)
- Neurohormones (Dr. Ian Cooke/Toni Hsu)
- IL-10 receptors of lung fibroblasts (Dr. Claude
Jourdan-LeSaux) - Aggregation of acetylcholine receptors in muscle
cells (Drs. Jes Stollberg, UHM, and Michael
Canute, HPU)
60Differential Interference Contrast (Nomarski)
61Digital Imaging
- Digital advantages include sensitivity, speed,
quantitation, feature extraction and image
analysis - CCD cameras - High resolution, slow
- Video cameras Low resolution, fast
- Photomultiplier tubes (PMTs) point recorders,
used for confocal
62Digital Cameras
- Need enough sensitivity for signal you want to
detect - Need enough speed for event you want to detect
- Need enough grayscales 8 bits for
documentation, 12 bits for quantitation - Need enough resolution - the number of of pixels
must be sufficient to distinguish features of
interest, but too many pixels is a waste of data
space - Color is simply three black and white images
combined and useful primarily for image
processing
63Optronics MacroFire Digital Camera
- Extremely sensitive
- 2048 x 2048 pixels
- Millisecond exposures
- Firewire
- Fits on both Olympus compound and stereo zoom
microscopes - Suitable for BF, DF, and Fluorescence
- Also Optronics MagnaFire SP 1280 x 1024 pixels
and Nikon Coolpix cameras
64TEM
- Transmission Electron Microscope
- Illumination source is beam of electrons from
tungsten wire - Electromagnetic lenses perform same function as
glass lenses in LM - Higher resolution and higher magnification of
thin specimens
65Specimen Preparation for TEM
- Chemical fixation with buffered glutaraldehyde
- Or 4 paraformaldehyde with gt1 glutaraldehyde
- Postfixation with osmium tetroxide
- Or not, or with subsequent removal from sections
- Dehydration and infiltration with liquid epoxy or
acrylic resin - Polymerization of hard blocks by heat or UV
- Ultramicrotomy 60-80nm sections
- Labeling and/or staining
- View with TEM
66Colloidal Gold Immunolabeling for TEM
- Colloidal gold of defined sizes, e.g., 5 nm, 10
nm, 20 nm, easily conjugated to antibodies - Results in small, round, electron-dense label
easily detected with EM - Can be enhanced after labeling to enlarge size
for LM or EM
67Colloidal Gold in TEM
68Colloidal Gold in TEM
69Double Immunogold Labeling of Negatively Stained
Specimens
- Bacterial pili serotypes dried onto grid and
sequentially labeled with primary antibody, then
Protein-A-5nm-gold and Protein-A-15-nm-gold
before negative staining
70TEM Grids
- TEM grids are 3 mm supports of various meshes
- You will handle them by the edges with fine
forceps
71Colloidal Gold in SEM
- Gold particles are often difficult to see against
the membrane with secondary electron detection - Gold particles show up brighter with
backscattered electron detection
72Preparation of Images for Publication
- Microscopy Images are your data!
- Adjustment and labeling of images for figure
plates with Adobe Photoshop
73How to Contact the BEMF
- Location Snyder Hall 118 University of Hawaii
at Manoa - Phone 808 956-6251
- FAX 808 956-5043
- URL http//www.pbrc.hawaii.edu/bemf
- E-mail dunlap_at_pbrc.hawaii.edu
- tina_at_pbrc.hawaii.edu
74Acknowledgments
- We thank all of the researchers who agreed to let
us use their images for this presentation
75Microscopy Microanalysis 2005
- July 31 - August 4, 2005
- Hawaii Convention Center
- Over 1100 talks and posters
- Huge trade show featuring the latest in
microscopes and related instrumentation,
software, and support - Pre-meeting workshops
- http//mm2005.microscopy.org