ZOLL M Series/E Series Biphasic CO2 Self-Paced
Training Module 8
2
Objectives
  • At the end of this module the participant can
  • Apply an proper adapter for CO2 monitoring
  • Zero a sensor and/or adapter
  • Connect the CO2 cable to the back of the machine.

3
Selecting an Airway Adapter
  • Determine whether an adapter is needed with or
    without a mouthpiece

4
Sensor Re-Usable
Selecting an Airway Adapter
  • Based on an endotracheal tube diameter, use the
    following table.
  • Verify that the airway adaptor windows are clean
    and dry.
  • Mouthpiece Only Verify that adapter and
    mouthpiece are intact and securely fastened to
    each other.

5
Turn Selector Switch to Monitor (Gray) Defib
(Red) or Pacer (Green)
  • The EtCO2 display area
  • Symbols will represent the status of the EtCO2
    alarm
  • Numbers represent the EtCO2 and respiratory rate
    values

6
Zeroing the Equipment
  • To ensure the EtCO2 measurement is accurate, the
    sensor and the airway adapter must be zeroed
    before use.
  • The M Series/E Series automatically displays ZERO
    CO2 SENSOR or ZERO CO2 ADAPTER when zeroing is
    necessary.
  • The M Series/E Series retains the zero setting in
    memory, so zeroing is not necessary for later use
    of the same sensor, adapter and unit.

7
Zeroing the Sensor
  • Place the sensor in the zero cell located on the
    sensor cable.
  • ZEROING CO2 SENSOR will display and takes 15
    seconds to complete the process
  • Place the sensor in the REF cell located on the
    sensor cable
  • Confirm the ETC02 value displayed is 38mmhg
    (1/-2mmgh)

8
Zeroing the Airway Adapter
  • Press the Zero and the EtCO2 Zero menu displays
  • Press the Enter softkey
  • 15 seconds to complete the process

9
Connecting the Adapter
  • Insert the elbow tubing into the endotracheal
    tube
  • Hold airway adapter so side with two arrows is on
    the bottom with cables facing away from the
    patient

10
Connecting the Adapter
  • Insert end of elbow into the airway adapter
  • Insert end of ventilator wye into opposite end of
    airway adapter
  • Connect airway adapter to sensor cable

11
Attach Sensor Cable to C02 Cable
  • EtCO2 cable will connect into ETCO2 port at back
    of machine
  • Attach end of CO2 cable to the sensor cable and
    click into place

CO2 Sp02 Multifunction Cable ECG
12
Displaying the EtCO2 Menu
  • By accessing the Param softkey, the menu will
    display the CO2 option
  • Press the Select softkey until EtCO2 highlights
  • Press the Enter softkey

13
The EtCO2 Menu Displays at the Bottom of the
Screen
  • Scale softkey adjust the scale (size) of the
    waveforms
  • Average softkey adjust the length of breath
    measurement
  • Comp softkey adjust to compensate for excessive
    oxygen
  • Press the Enter key after any selection of
    softkey.

14
Factors That Can Affect EtCO2
  • Oxygen levels in excess of 60 (unless
    compensated for)
  • Nitrous oxide in the airway adapter (unless
    compensated for)
  • Desfluane in the exhaled breath beyond normal
    limits (5-6)
  • Halogenated agents

15
  • This concludes Module 8 EtCO2 Measurement.
  • Please continue on with other modules which may
    be required at your facility