Title: ZOLL M SeriesE Series Biphasic: SpO2 SelfPaced Training Module


1
ZOLL M Series/E Series Biphasic SpO2 Self-Paced
Training Module 7
2
Objectives
  • At the end of this module the participant can
  • Place an SpO2 device.
  • Configure alarms for the SpO2 option.
  • Connect the SpO2 cable to the back of the
    machine.

3
Select a Sensor Site
  • Choose a well-profused site
  • Ring or middle finger of non-dominant hand
  • Long or great toe
  • Avoid extremity with a blood pressure cuff,
    arterial catheter or intravenous line

4
Sensor Re-Usable
  • Place the patients digit inside the sensor
  • The top and bottom half of sensor should be
    parallel for an accurate measurement

5
Sensor Re-Usable
The emitter and detector of the infrared light
should be vertically aligned Remove wrapping and
folk over finger, securing the wings in place
6
Sp02 Cable
  • Verify that the SpO2 cable is connected to the
    SpO2 port on the back of the
  • M Series/E Series

Co2 SpO2 Multifunction Cable ECG
7
Connecting the Sp02 Cable
  • Whether using the disposable sensor or reusable
    sensor
  • Align the Masimo logo from the sensor cable to
    the Masimo logo on the SpO2 cable and click
    together

8
Turn Selector Switch to Monitor (Gray) Defib
(Red) or Pacer (Green)
  • Once the M Series/E Series is turned on, the SpO2
    measurement will appear in the left corner of the
    screen.
  • To adjust the SpO2 alarm, access the Alarms
    softkey and use the Sp02 softkey to scroll to the
    desired setting

9
Reading the Display
  • Pulse Oxymetry information is displayed in the
    SpO2 area as a number
  • Also shown are the state of the alarm and the
    signal strength indicator

10
Displaying the SpO2 Menu
  • By accessing the Param softkey, the menu will
    display the SpO2 option
  • Press Select softkey until SpO2 is highlighted.
    Press the Enter softkey.

11
Changing the Sensitivity
  • When the patient is exhibiting low blood flow,
    the sensitivity can be increased.
  • Press the Sens. softkey and then use the Sens.
    softkey to scroll through the options. Pressing
    enter will select the option.

12
Printing the SpO2 Waveform
  • Using the Wave2 softkey, press until the SpO2
    waveform displays under the ECG display
  • Press the Recorder button to print a strip chart
  • Press the Recorder button again to halt the strip
    chart printing

13
Factors That Can Affect SpO2
  • Incorrect sensor application or use
  • Dysfunctional hemoglobin (meth-hemoglobin)
  • Excessive patient movement
  • Hypotension, vasoconstriction, hypothermia
  • Nail polish, fungus, acrylic nails

14
  • This concludes Module 7 SpO2 Measurement. Please
    continue on with other modules which may be
    required at your facility