THERAC 25 - PowerPoint PPT Presentation

1 / 9
About This Presentation
Title:

THERAC 25

Description:

December 1985 patient in Yakima Wa receives overdose. 21 March 1986- Tyler TX accident ... 17 January 1987- Second Yakima WA Accident. Why? The turntable was ... – PowerPoint PPT presentation

Number of Views:184
Avg rating:3.0/5.0
Slides: 10
Provided by: ideaEn
Category:
Tags: therac | yakima

less

Transcript and Presenter's Notes

Title: THERAC 25


1
THERAC 25
  • RAM MOHAN BOPPUDI
  • EE 585 FAULT TOLERANT
  • COMPUTING

2
BACKGROUND
  • Therac 25 was a medical linear accelerator , a
    linac developed by Atomic Energy Of Canada Ltd.
  • Therac 25 is a radio therapy machine used to
    destroy tumours using high energy beams.
  • 11 Therac 25s were installed 5 in US , 6 in
    Canada.

3
Background ( contd)
  • Therac 25 was derived from its previous version
    Therac 6 and Therac 20.
  • Differences from Therac 20
  • - Uses double pass technique which is absent
    in previous versions
  • - Software is responsible for safety
  • - Hardware safety interlocks removed
  • - Less space and economic

4
Therac-25 Turntable
Field Light Mirror
Counterweight
Beam Flattener (X-ray Mode)
Turntable
Scan Magnet (Electron Mode)
5
Accidents
  • 3 June 1985 patient at Marietta GA received
    overdose
  • 26 July 1985 Hamilton ONT patient severely
    burned , died November 1985
  • December 1985 patient in Yakima Wa receives
    overdose
  • 21 March 1986- Tyler TX accident
  • 11 March 1986 2nd Tyler TX accident
  • 17 January 1987- Second Yakima WA Accident

6
Why?
  • The turntable was in the wrong position.
  • Patients were receiving x-rays without
    beam-scattering.
  • No hardware safety interlocks
  • Non descriptive error messages
  • User override able error modes
  • Software designed by only one person

7
Cost of the Bug
  • To users (patients)
  • Four deaths, two other serious injuries.
  • To developers (AECL)
  • One lawsuit
  • Settled out of court
  • Time/money to investigate and fix the bugs
  • To product owners (11 hospitals)
  • System downtime

8
Corrective Action Plan
  • Numerous hardware and software changes
  • All interruptions related to dosimetry not
  • continuable
  • independent hardware software shutdowns
  • potentiometer on turntable
  • hardware interlocks
  • dead man switch motion enable
  • Fix documentation, messages, user manuals

9
REFERENCES
  • An investigation of the Therac-25 Accidents
  • Nancy Leveson
  • Clark S.Turner
  • Wikipedia on Therac 25
Write a Comment
User Comments (0)
About PowerShow.com