Title: Therac 25 Incident


1
Therac 25 Incident
  • By

2
Introduction
  • Background
  • The Therac 25 machine was a linear accelerator or
    lineac. This is a medical machine that fires high
    energy radiation to destroy tumors.
  • Therac 25 was the successor to the Therac 6 and
    20. It began development in 1976.
  • Therac 25 was created by the Canadian firm Atomic
    Energy Comission Limited (AECL)

3
Background to accidents
  • Eleven Therac 25 devices were sold, 6 to Canada
    and 5 to USA
  • Between 1985 and 1987 6 recorded deaths were
    attributed to the machine
  • Deaths caused by massive radiation overdoses
  • In 1987 the Therac 25 was recalled by AECL after
    several US and Canadian federal investigations.

4
How Therac 25 Worked
  • Therac 25 could operate in two modes
  • Electron
  • X-Ray

Electron Mode
X-Ray Mode
5
How Therac 25 Worked
  • In Electron mode the gun did not need shielding.
  • In X-Ray mode a shield was needed due to the
    higher energy.
  • In each mode bending magnets had to be set before
    it fired

6
Problems with Therac 25
  • Problems in the design process
  • Software reuse management
  • Failure of testing
  • Failings of the risk assessments
  • Irresponsible usage

7
Therac 25 design problems
  • Attempted to combine functionality from previous
    versions for lower cost
  • Apparently no consultation with stakeholders -
    such as medical technicians
  • Over confidence in the previous designs
    resulting in complacency

8
Software Reuse
  • Therac 25 borrowed heavily from previous Therac
    designs.
  • Therac 20 and Therac 6 relied on hardware safety
    systems
  • Therac 25 removed the hardware safety
  • Therac 25 software had the same bugs as found in
    Therac 20

9
Testing of Therac 25
  • As software came from Therac 20 it was assumed to
    be bug free
  • AECL assumed only hardware failures were possible
    as software doesnt degrade over time
  • AECL performed a safety analysis on the
    Therac-25 and apparently excluded the
    software.(Leveson et al, 1993)

10
Failure of risk assessment
  • AECL did not perform a proper risk assessment of
    Therac 25
  • Hardware would be the primary cause of failure
  • Believed overdosing of patients was impossible

11
Software issues
  • Used flags to synchronise processes
  • Used 8-bit flags which were prone to overflowing
  • Created race condition between magnet control and
    keyboard control
  • The error messages were indecipherable to users
    technicians

12
Implications
  • If the user entered any data while the gun was
    being set it did not react
  • However the rest of the system did react,
    allowing unshielded X-Rays to be fired
  • A race condition in a real-time system!

13
Implications
  • The use of flags that were too small allowed
    overflow
  • This overflow allowed unshielded X-Rays to be
    fired
  • This was and is fatal!

14
Error Messages
  • Error messages gave an error number with no
    explanation
  • Error message documentation was incomplete
  • Only AECL technicians understood the documented
    errors
  • Frequently appeared so lost their value as a
    warning

15
Consequences
  • Users frequently continued passed errors without
    understanding them
  • Errors did not alarm users and so were not
    reported
  • This happened in the majority of accidents

16
AECLs Failure of Care
  • AECL failed to deal with the problems effectively
  • No communication between all parties
  • Fixes such as sellotaping up keys or removing
    them endorsed
  • Did not make changes demanded by Canadian
    Government after initial accidents

17
Safety Critical Problems
  • Allowed user to continue potentially fatal
    treatments paused instead of stopping
  • No checks on the dose administered
  • No hardware logging often turned off
  • Overflow allowed on gun control flag
  • Error messages were not prioritised or informative

18
Summary
  • Reutilisation of old Software
  • Complacency towards
  • Testing
  • Risk Assessment
  • Design
  • Training
  • Error Management