Gas Chromatography - PowerPoint PPT Presentation

1 / 43
About This Presentation
Title:

Gas Chromatography

Description:

... with flame ionization detectors (FID) since the FID needs it as fuel ... FID. FPD. Thermal Conductivity (TCD) The carrier gas has a known ... FID ... – PowerPoint PPT presentation

Number of Views:389
Avg rating:3.0/5.0
Slides: 44
Provided by: johnandtam
Category:

less

Transcript and Presenter's Notes

Title: Gas Chromatography


1
Gas Chromatography
  • General Design of a Gas Chromatograph
  • Separation Processes in Gas Chromatography
  • GC Columns
  • GC Injectors
  • GC Detectors

Gas Chromatography A Practical Course by
Gerhard Schomburg (QD79.C45 S3913 1990), 29-30,
31-36, 38-60, 66-73
2
General Design of a GC
3
Some of the designdetails
  • Gas supplies usually have either in-line or
    instrument mounted traps to remove any water,
    oxygen, hydrocarbons or other contaminants from
    compressed gases
  • Gas flows can be controlled using either needle
    valves or mass-flow controllers (electronic
    sensors)
  • Instruments can have multiple injectors,
    detectors or columns
  • Injectors and detectors usually have their own
    temperature controlled zones (small heaters)
  • The GC oven has a large fan and a vent door to
    help with rapid cooling of the oven
  • Data collection (and integration) can be done
    using a chart recorder, integrator or a
    computerized data system

4
Separation Processes in GC
  • Gas Chromatography as it is usually performed is
    correctly called gas-liquid chromatography
  • the analyte is in the gas phase in the GC and
    partitions between the mobile phase (carrier gas)
    and the liquid stationary phase that is coated on
    the inside of an open-tubular capillary column or
    on particles inside a packed column
  • Some packed-column GC uses non-coated solid
    stationary phases, in which case one is
    performing gas-solid adsorption chromatography
  • Capillary, open-tubular (WCOT specifically)
    column GC is the primary type of GC used in
    quantitative analysis
  • higher resolution greater ability to
    discriminate between components
  • smaller capacity of the column is not important
    as long as sufficient analyte is available for
    detection
  • pg/mL (ppt) to ?g/mL (ppm) concentration range
    for liquid analytes

5
(No Transcript)
6
The Objective in Chromatography (all types)
  • Separate your analytes (resolution of 1.5 or
    better) in the shortest amount of time possible
    and detect them.
  • How can we do this in GC?
  • Use different columns for different analyte types
  • stationary phase
  • diameter of column, stationary phase thickness
  • column length
  • Use different injection types/temperatures to
    optimize the process of loading the sample on the
    column
  • Use different temperature (or pressure) programs
    for the column
  • Select and use a detector that is suitable for
    the analyte(s) of interest

7
GC Columns (concentrating on open-tubular
capillary columns)
  • Column frame constructed of fused silica tubing
  • Polyamide coating on the outside gives it
    strength
  • Liquid stationary phases coated or bonded to the
    inside of the tubing
  • 0.1 - 0.53 mm ID, 5-100 meters in length,
    stationary phases usually 0.10 to 1.5 ?m in
    thickness
  • Mounted on a wire cage to make them easier to
    handle
  • 5-150 meters long.

8
Capillary Column Stationary Phases
9
(No Transcript)
10
(No Transcript)
11
Choosing a GC Column
  • Is the column compatible with your analytes
  • polar analytes require polar stationary phases so
    they will spend some of their time in the
    stationary phase
  • non-polar analytes require non-polar stationary
    phases
  • You usually have to compromise on the stationary
    phase to get a good column for your analytes
    (which are probably a mix of polar and non-polar)
  • DB-5, HP-5, EC-5, RTX-5 (5 dimethyl, 95
    diphenyl polysiloxane) most common general use
    column.
  • Temperature range, solvent and carrier gas
    compatibility
  • Sample capacity versus resolution
  • usually determines packed vs.. capillary
  • GCs usually setup for either packed or capillary
  • Lets say you choose a capillary column, theres
    more to think about!

12
(No Transcript)
13
(No Transcript)
14
(No Transcript)
15
(No Transcript)
16
For capillary GC columns.
  • Increased length greater N, therefore a greater
    R
  • expense is possible band broadening if analytes
    are on the column too long!
  • Increased length leads to longer separations. Do
    you have the time?
  • Increased stationary phase thickness and column
    diameter provides increased sample capacity and
    can provide increased resolution
  • tradeoffs are a longer analysis time and more
    column bleed with thicker stationary phases
  • Is the column compatible with the detector?
  • Thick stationary phases bleed more and will
    contaminate a mass spectrometer.
  • For most analytical work, a best compromise
    column is chosen and other variables (temp, etc.)
    are altered to optimize the separation.

17
(No Transcript)
18
(No Transcript)
19
Capillary vs. Packed Columns
  • Packed Columns
  • Greater sample capacity
  • Lower cost (can make your own)
  • More rugged
  • Most common in process labs or separating/determin
    ing major components in a sample (prep GC)
  • Limited lengths reduces R and N
  • Not compatible with some GC detectors
  • Capillary Columns
  • Higher resolution (R)
  • Greater HETP and N
  • Shorter analysis time
  • Greater sensitivity
  • Most common in analytical laboratory GC
    instruments
  • Smaller sample capacity
  • Higher cost/column
  • Columns more susceptible to damage

20
Temperature Programming in GC
  • The simplest way to alter the separation in GC
    is to alter the temperature program in the oven.
    You can also alter the pressure of the carrier
    gas, but this is less common (much).
  • Isothermal constant temperature
  • Gradient varied temperature
  • By altering the temperature, you vary the rate of
    the reaction for any analyte
  • they spend more or less time in the stationary
    phase
  • the greater the difference in the times between
    analytes, the better the separation!

21
(No Transcript)
22
  • The traps of temperature
  • If your temperature at a given time is too high,
    you can cause the peaks to co-elute
  • poor resolution vs but a faster separation
  • If your temperature at a given time is too low,
    you can get still get a good separation
  • adequate resolution, but a separation that takes
    very long
  • You have to choose a compromise temperature
    program

23
(No Transcript)
24
GC Carrier Gases (the mobile phase)
  • Usually inert gases (dont react with analytes
    except sometimes in the detector)
  • Purpose
  • sweep sample through the column
  • protect column from oxygen exposure at
    temperature
  • assist with function of the detector
  • Most common
  • Helium (available relatively pure without
    extensive purification after it leaves a
    compressed gas cylinder)
  • Nitrogen (usually requires an oxygen and water
    trap)
  • Hydrogen
  • normally used only with flame ionization
    detectors (FID) since the FID needs it as fuel
    for the flame
  • still rarely used due to safety concerns (and
    chromatographic ones)

25
(No Transcript)
26
GC Injection.
  • Samples are injected through a septum
  • keeps oxygen out of the column
  • provides a seal to keep the carrier gas pressure
    up at the head of the column
  • carrier gas flow rate is determined by the
    pressure or the gas at the opening of the column
  • Many different (mostly proprietary) materials
  • red rubber (bleeds at about 250 C)
  • Thermogreen (good up to about 300 C)
  • High-temperature blue (good a little over 300 C)
  • The injector is usually lined with a de-activated
    glass liner
  • prevents metal injector-sample reactions that
    would alter analytes or damage the metal of the
    injector
  • Can be cleaned/replaced regularly

27
(No Transcript)
28
(No Transcript)
29
Injection types
30
  • On-Column Injection
  • used widely in packed-column GC, less in
    capillary GC
  • sample is deposited directly on the column
  • Good for thermally unstable compounds
  • Good for quantitative analysis at low
    concentrations
  • all sample is available to travel to the detector
  • BUT, you can inject only a relatively small
    amount of sample in capillary GC anyhow.
  • Splitless Injection
  • Sample is vaporized in the injector itself and
    ALL of the sample is swept onto the column by the
    carrier gas
  • Again, relatively small samples are injected (10
    ? L or less in capillary GC)
  • Sample spends a large amount of time in the
    injector
  • Best for trace (1 -100 ppm range) concentrations
    of high boiling point analytes in low boiling
    point solvents
  • extra time in the injector helps volatilize the
    analytes.

31
  • Split Injection
  • the injection is split, with only a portion of
    the sample (usually 1 - 20) actually making it
    to the column
  • the most common method of injecting samples onto
    small diameter, open-tubular columns.
  • Even if you inject 20 ?L, only a fraction
    (adjustable) makes it on to the column
  • Not good for analytes with a wide range of
    boiling points
  • some may be swept out the split vent before they
    are volatilized
  • Modern capillary GCs come with a Split/Splitless
    injectors standard
  • you switch between modes by changing the split
    vent gas flow and using a different injection
    liner.

32
Dont Forget SPME (Solid Phase Microextraction)
33
GC Detectors
  • A dozen or more varieties (some obscure)
  • Must be
  • sensitive to the analytes of interest
  • compatible with the column, carrier gas, solvent,
    etc.
  • rugged enough to withstand general unattended
    used
  • Ive run our new GC for 36 hours straight without
    touching it!
  • Should have a known linear range
  • if the detector response is very linear, you can
    use a response factor instead of a calibration
    curve for quantitation!
  • Usually require separate gas supplies (other than
    the carrier gas), have their own temperature
    control.
  • Measure nothing more than a voltage or a current.

34
(No Transcript)
35
(No Transcript)
36
(No Transcript)
37
(No Transcript)
38
FID
FPD
39
Thermal Conductivity (TCD)
  • The carrier gas has a known thermal conductivity.
  • As the thermal conductivity of the column eluent
    (gas flow in) changes, the resistance of the
    filament changes.
  • The presence of analyte molecules in the carrier
    gas alter the thermal conductivity of the gas
    (usually He)
  • There is normally a second filament to act as a
    reference (the carrier gas is split)
  • Increased sensitivity with decreasing temperature
    (detector), flow rate and applied current.
  • Filaments will burn out (oxidized) in the
    presence of oxygen if hot!

Non-destructive
40
(No Transcript)
41
FID
  • Destructive, sample lost.
  • Analytes containing C burn in a hydrogen-oxygen
    flame and produce ions
  • CHO ions are collected on a cathode and the
    current they produce results in the signal
  • WILL NOT detect non-C containing compounds!
  • Requires H2 supply (tank or generator) and O2
    supply (compressed air)
  • H2 carrier gas can be used, eliminating the need
    for a supply for the detector
  • A makeup gas can also be required!

42
(No Transcript)
43
ECD
  • Particularly sensitive to halogens nitriles,
    carbonyls, nitro compounds
  • Analytes pass through a cell, in which electrons
    are traveling between a 63Ni electrode and a
    collector electrode
  • As analytes with electron capturing ability
    pass through the cell, the flow of electrons is
    interrupted.
  • The change in current, due to reduced flow of
    electrons, is recorded.
  • EXTREMELY SENSITIVE TO HALOGENS
  • could ruin detector with 1 ppm hexachlorocyclohexa
    ne by contaminating it with excess analyte
  • Widely used for the determination of pesticides,
    herbicides and PCBs in environmental samples.
  • Non-destructive
Write a Comment
User Comments (0)
About PowerShow.com