Title: Gas Chromatography
1Gas Chromatography
- General Design of a Gas Chromatograph
- Separation Processes in Gas Chromatography
- GC Columns
- GC Injectors
- GC Detectors
Gas Chromatography A Practical Course by
Gerhard Schomburg (QD79.C45 S3913 1990), 29-30,
31-36, 38-60, 66-73
2General Design of a GC
3Some of the designdetails
- Gas supplies usually have either in-line or
instrument mounted traps to remove any water,
oxygen, hydrocarbons or other contaminants from
compressed gases - Gas flows can be controlled using either needle
valves or mass-flow controllers (electronic
sensors) - Instruments can have multiple injectors,
detectors or columns - Injectors and detectors usually have their own
temperature controlled zones (small heaters) - The GC oven has a large fan and a vent door to
help with rapid cooling of the oven - Data collection (and integration) can be done
using a chart recorder, integrator or a
computerized data system
4Separation Processes in GC
- Gas Chromatography as it is usually performed is
correctly called gas-liquid chromatography - the analyte is in the gas phase in the GC and
partitions between the mobile phase (carrier gas)
and the liquid stationary phase that is coated on
the inside of an open-tubular capillary column or
on particles inside a packed column - Some packed-column GC uses non-coated solid
stationary phases, in which case one is
performing gas-solid adsorption chromatography - Capillary, open-tubular (WCOT specifically)
column GC is the primary type of GC used in
quantitative analysis - higher resolution greater ability to
discriminate between components - smaller capacity of the column is not important
as long as sufficient analyte is available for
detection - pg/mL (ppt) to ?g/mL (ppm) concentration range
for liquid analytes
5(No Transcript)
6The Objective in Chromatography (all types)
- Separate your analytes (resolution of 1.5 or
better) in the shortest amount of time possible
and detect them. - How can we do this in GC?
- Use different columns for different analyte types
- stationary phase
- diameter of column, stationary phase thickness
- column length
- Use different injection types/temperatures to
optimize the process of loading the sample on the
column - Use different temperature (or pressure) programs
for the column - Select and use a detector that is suitable for
the analyte(s) of interest
7GC Columns (concentrating on open-tubular
capillary columns)
- Column frame constructed of fused silica tubing
- Polyamide coating on the outside gives it
strength - Liquid stationary phases coated or bonded to the
inside of the tubing - 0.1 - 0.53 mm ID, 5-100 meters in length,
stationary phases usually 0.10 to 1.5 ?m in
thickness - Mounted on a wire cage to make them easier to
handle - 5-150 meters long.
8Capillary Column Stationary Phases
9(No Transcript)
10(No Transcript)
11Choosing a GC Column
- Is the column compatible with your analytes
- polar analytes require polar stationary phases so
they will spend some of their time in the
stationary phase - non-polar analytes require non-polar stationary
phases - You usually have to compromise on the stationary
phase to get a good column for your analytes
(which are probably a mix of polar and non-polar) - DB-5, HP-5, EC-5, RTX-5 (5 dimethyl, 95
diphenyl polysiloxane) most common general use
column. - Temperature range, solvent and carrier gas
compatibility - Sample capacity versus resolution
- usually determines packed vs.. capillary
- GCs usually setup for either packed or capillary
- Lets say you choose a capillary column, theres
more to think about!
12(No Transcript)
13(No Transcript)
14(No Transcript)
15(No Transcript)
16For capillary GC columns.
- Increased length greater N, therefore a greater
R - expense is possible band broadening if analytes
are on the column too long! - Increased length leads to longer separations. Do
you have the time? - Increased stationary phase thickness and column
diameter provides increased sample capacity and
can provide increased resolution - tradeoffs are a longer analysis time and more
column bleed with thicker stationary phases - Is the column compatible with the detector?
- Thick stationary phases bleed more and will
contaminate a mass spectrometer. - For most analytical work, a best compromise
column is chosen and other variables (temp, etc.)
are altered to optimize the separation.
17(No Transcript)
18(No Transcript)
19Capillary vs. Packed Columns
- Packed Columns
- Greater sample capacity
- Lower cost (can make your own)
- More rugged
- Most common in process labs or separating/determin
ing major components in a sample (prep GC) - Limited lengths reduces R and N
- Not compatible with some GC detectors
- Capillary Columns
- Higher resolution (R)
- Greater HETP and N
- Shorter analysis time
- Greater sensitivity
- Most common in analytical laboratory GC
instruments - Smaller sample capacity
- Higher cost/column
- Columns more susceptible to damage
20Temperature Programming in GC
- The simplest way to alter the separation in GC
is to alter the temperature program in the oven.
You can also alter the pressure of the carrier
gas, but this is less common (much). - Isothermal constant temperature
- Gradient varied temperature
- By altering the temperature, you vary the rate of
the reaction for any analyte - they spend more or less time in the stationary
phase - the greater the difference in the times between
analytes, the better the separation!
21(No Transcript)
22- The traps of temperature
- If your temperature at a given time is too high,
you can cause the peaks to co-elute - poor resolution vs but a faster separation
- If your temperature at a given time is too low,
you can get still get a good separation - adequate resolution, but a separation that takes
very long - You have to choose a compromise temperature
program
23(No Transcript)
24GC Carrier Gases (the mobile phase)
- Usually inert gases (dont react with analytes
except sometimes in the detector) - Purpose
- sweep sample through the column
- protect column from oxygen exposure at
temperature - assist with function of the detector
- Most common
- Helium (available relatively pure without
extensive purification after it leaves a
compressed gas cylinder) - Nitrogen (usually requires an oxygen and water
trap) - Hydrogen
- normally used only with flame ionization
detectors (FID) since the FID needs it as fuel
for the flame - still rarely used due to safety concerns (and
chromatographic ones)
25(No Transcript)
26GC Injection.
- Samples are injected through a septum
- keeps oxygen out of the column
- provides a seal to keep the carrier gas pressure
up at the head of the column - carrier gas flow rate is determined by the
pressure or the gas at the opening of the column - Many different (mostly proprietary) materials
- red rubber (bleeds at about 250 C)
- Thermogreen (good up to about 300 C)
- High-temperature blue (good a little over 300 C)
- The injector is usually lined with a de-activated
glass liner - prevents metal injector-sample reactions that
would alter analytes or damage the metal of the
injector - Can be cleaned/replaced regularly
27(No Transcript)
28(No Transcript)
29Injection types
30- On-Column Injection
- used widely in packed-column GC, less in
capillary GC - sample is deposited directly on the column
- Good for thermally unstable compounds
- Good for quantitative analysis at low
concentrations - all sample is available to travel to the detector
- BUT, you can inject only a relatively small
amount of sample in capillary GC anyhow. - Splitless Injection
- Sample is vaporized in the injector itself and
ALL of the sample is swept onto the column by the
carrier gas - Again, relatively small samples are injected (10
? L or less in capillary GC) - Sample spends a large amount of time in the
injector - Best for trace (1 -100 ppm range) concentrations
of high boiling point analytes in low boiling
point solvents - extra time in the injector helps volatilize the
analytes.
31- Split Injection
- the injection is split, with only a portion of
the sample (usually 1 - 20) actually making it
to the column - the most common method of injecting samples onto
small diameter, open-tubular columns. - Even if you inject 20 ?L, only a fraction
(adjustable) makes it on to the column - Not good for analytes with a wide range of
boiling points - some may be swept out the split vent before they
are volatilized - Modern capillary GCs come with a Split/Splitless
injectors standard - you switch between modes by changing the split
vent gas flow and using a different injection
liner.
32Dont Forget SPME (Solid Phase Microextraction)
33GC Detectors
- A dozen or more varieties (some obscure)
- Must be
- sensitive to the analytes of interest
- compatible with the column, carrier gas, solvent,
etc. - rugged enough to withstand general unattended
used - Ive run our new GC for 36 hours straight without
touching it! - Should have a known linear range
- if the detector response is very linear, you can
use a response factor instead of a calibration
curve for quantitation! - Usually require separate gas supplies (other than
the carrier gas), have their own temperature
control. - Measure nothing more than a voltage or a current.
34(No Transcript)
35(No Transcript)
36(No Transcript)
37(No Transcript)
38FID
FPD
39Thermal Conductivity (TCD)
- The carrier gas has a known thermal conductivity.
- As the thermal conductivity of the column eluent
(gas flow in) changes, the resistance of the
filament changes. - The presence of analyte molecules in the carrier
gas alter the thermal conductivity of the gas
(usually He) - There is normally a second filament to act as a
reference (the carrier gas is split) - Increased sensitivity with decreasing temperature
(detector), flow rate and applied current. - Filaments will burn out (oxidized) in the
presence of oxygen if hot!
Non-destructive
40(No Transcript)
41FID
- Destructive, sample lost.
- Analytes containing C burn in a hydrogen-oxygen
flame and produce ions - CHO ions are collected on a cathode and the
current they produce results in the signal - WILL NOT detect non-C containing compounds!
- Requires H2 supply (tank or generator) and O2
supply (compressed air) - H2 carrier gas can be used, eliminating the need
for a supply for the detector - A makeup gas can also be required!
42(No Transcript)
43ECD
- Particularly sensitive to halogens nitriles,
carbonyls, nitro compounds - Analytes pass through a cell, in which electrons
are traveling between a 63Ni electrode and a
collector electrode - As analytes with electron capturing ability
pass through the cell, the flow of electrons is
interrupted. - The change in current, due to reduced flow of
electrons, is recorded. - EXTREMELY SENSITIVE TO HALOGENS
- could ruin detector with 1 ppm hexachlorocyclohexa
ne by contaminating it with excess analyte - Widely used for the determination of pesticides,
herbicides and PCBs in environmental samples. - Non-destructive