Title: Gas Chromatography Gas chromatography is a technique used
1Gas Chromatography
2- Gas chromatography is a technique used for
separation of volatile substances, or substances
that can be made volatile, from one another in a
gaseous mixture at high temperatures. A sample
containing the materials to be separated is
injected into the gas chromatograph. A mobile
phase (carrier gas) moves through a column that
contains a wall coated or granular solid coated
stationary phase. As the carrier gas flows
through the column, the components of the sample
come in contact with the stationary phase. The
different components of the sample have different
affinities for the stationary phase, which
results in differential migration of solutes,
thus leading to separation
3- Martin and James introduced this separation
technique in 1952, which is the latest of the
major chromatograhpic techniques. However, by
1965 over 18000 publications in gas
chromatography (GC) were available in the
literature. This is because optimized
instrumentation was feasible. Gas chromatography
is good only for volatile compounds or those,
which can be made volatile by suitable
derivatization methods or pyrolysis. Thus, about
20 of chemicals available can be analyzed
directly by GC.
4- Gas chromatography can be used for both
qualitative and quantitative analysis.
Comparison of retention times can be used to
identify materials in the sample by comparing
retention times of peaks in a sample to retention
times for standards. The same limitations for
qualitative analysis discussed in Chapter 26 also
apply for separations in GC. Quantitative
analysis is accomplished by measurement of either
peak height or peak area
5Gas - Solid Chromatography (GSC)
- The stationary phase, in this case, is a solid
like silica or alumina. It is the affinity of
solutes towards adsorption onto the stationary
phase which determines, in part, the retention
time. The mobile phase is, of course, a suitable
carrier gas. This gas chromatographic technique
is most useful for the separation and analysis of
gases like CH4, CO2, CO, ... etc. The use of GSC
in practice is considered marginal when compared
to gas liquid chromatography.
6Gas - Liquid Chromatography (GLC)
- The stationary phase is a liquid with very low
volatility while the mobile phase is a suitable
carrier gas. GLC is the most widely used
technique for separation of volatile species.
The presence of a wide variety of stationary
phases with contrasting selectivities and easy
column preparation add to the assets of GLC or
simply GC.
7Instrumentation
- It may be wise to introduce instrumental
components before proceeding further in
theoretical background. This will help clarify
many points, which may, otherwise, seem vague. It
should also be noted that a detector will require
special gas cylinders depending on the detector
type utilized. The column temperature controller
is simply an oven, the temperature of which can
be varied or programmed
8(No Transcript)
9- Three temperature zones can be identified
- Injector temperature, TI, where TI should allow
flash vaporization of all sample components. - Column temperature, Tc, which is adjusted as the
average boiling points of sample components. - Detector Temperature, TD, which should exclude
any possible condensation inside the detector. - Generally, an intuitive equation can be used to
adjust all three zones depending on the average
boiling point of the sample components. This
equation is formulated as - TI TD Tc 50 oC
10- The Carrier Gas
- Unlike liquid chromatography where wide varieties
of mobile phase compositions are possible, mobile
phases in gas chromatography are very limited.
Only slight changes between carrier gases can be
identified which places real limitations to
chromatographic enhancement by change or
modification of carrier gases
11- A carrier gas should have the following
properties - Highly pure (gt 99.9)
- Inert so that no reaction with stationary phase
or instrumental components can take place,
especially at high temperatures. - A higher density (larger viscosity) carrier gas
is preferred. - Compatible with the detector since some detectors
require the use of a specific carrier gas. - A cheap and available carrier gas is an advantage.
12Longitudinal Diffusion Term
- This is an important factor contributing to band
broadening which is a function of the diffusivity
of the solute in the gaseous mobile phase as well
as the molecular diffusion of the carrier gas
itself. -
- HL K DM /V
- Where DM is the diffusion coefficient of solute
in the carrier gas. This term can be minimized
when mobile phases of low diffusion, i.e. high
density, are used in conjunction with higher flow
rates.
13- The same van Deemter equation as in LC can be
written for GC where - H A B/V CV
- The optimum carrier gas velocity is given by the
derivative of van Deemter equation - Vopt B/C 1/2
- However, the obtained velocity is much greater
than that obtained in LC.
14- The carrier gas pressure ranges from 10-50 psi.
Higher pressures potentially increase compression
possibility while very low pressures result in
large band broadening due to diffusion. Depending
on the column dimensions, flow rates from 1-150
mL/min are reported. Conventional analytical
columns (1/8) usually use flow rates in the
range from 20-50 mL/min while capillary columns
use flow rates from 1-5 mL/min depending on the
dimensions and nature of column. In most cases, a
selection between helium and nitrogen is made as
these two gases are the most versatile and common
carrier gases in GC.
15Injectors
- Septum type injectors are the most common. These
are composed of a glass tube where vaporization
of the sample takes place. The sample is
introduced into the injector through a
self-sealing silicone rubber septum. The carrier
gas flows through the injector carrying vaporized
solutes. The temperature of the injector should
be adjusted so that flash vaporization of all
solutes occurs. If the temperature of the
injector is not high enough (at least 50 degrees
above highest boiling component), band broadening
will take place.
16(No Transcript)
17Column Configurations and Ovens
- The column in chromatography is undoubtedly the
heart of the technique. A column can either be a
packed or open tubular. Traditionally, packed
columns were most common but fast developments in
open tubular techniques and reported advantages
in terms of efficiency and speed may make open
tubular columns the best choice in the near
future. Packed columns are relatively short
(2meters) while open tubular columns may be as
long as 30-100 meters
18- Packed columns are made of stainless steel or
glass while open tubular columns are usually made
of fused silica. The temperature of the column is
adjusted so that it is close to the average
boiling point of the sample mixture. However,
temperature programming is used very often to
achieve better separations. The temperature of
the column is assumed to be the same as the oven
which houses the column. The oven temperature
should be stable and easily changed in order to
obtain reproducible results.
19Detection Systems
- Several detectors are available for use in GC.
Each detector has its own characteristics and
features as well as drawbacks. Properties of an
ideal detector include
- High sensitivity
- Minimum drift
- Wide dynamic range
- Operational temperatures up to 400 oC.
- Fast response time
- Same response factor for all solutes
- Good reliability (no fooling)
- Nondestructive
- Responds to all solutes (universal)
20a. Thermal Conductivity Detector (TCD)
- This is a nondestructive detector which is used
for the separation and collection of solutes to
further perform some other experiments on each
purely separated component. The heart of the
detector is a heated filament which is cooled by
helium carrier gas. Any solute passes across the
filament will not cool it as much as helium does
because helium has the highest thermal
conductivity. This results in an increase in the
temperature of the filament which is related to
concentration. The detector is simple,
nondestructive, and universal but is not very
sensitive and is flow rate sensitive.
21(No Transcript)
22(No Transcript)
23- Note that gases should always be flowing through
the detector including just before, and few
minutes after, the operation of the detector.
Otherwise, the filament will melt. Also, keep
away any oxygen since oxygen will oxidize the
filament and results in its destruction.
- Remember that TCD characteristics include
- Rugged
- Wide dynamic range (105)
- Nondestructive
- Insensitive (10-8 g/s)
- Flow rate sensitive
24b. Flame Ionization Detector (FID)
- This is one of the most sensitive and reliable
destructive detectors. Separate two gas
cylinders, one for fuel and the other for O2 or
air are used in the ignition of the flame of the
FID. The fuel is usually hydrogen gas. The flow
rate of air and hydrogen should be carefully
adjusted in order to successfully ignite the
flame.
25(No Transcript)
26(No Transcript)
27- The FID detector is a mass sensitive detector
where solutes are ionized in the flame and
electrons emitted are attracted by a positive
electrode, where a current is obtained. - The FID detector is not responsive to air, water,
carbon disulfide. This is an extremely important
advantage where volatile solutes present in water
matrix can be easily analyzed without any
pretreatment.
28- Remember that FID characteristics include
- Rugged
- Sensitive (10-13 g/s)
- Wide dynamic range (107)
- Signal depends on number of carbon atoms in
organic analytes which is referred to as mass
sensitive rather than concentration sensitive - Weakly sensitive to carbonyl, amine, alcohol,
amine groups - Not sensitive to non-combustibles H2O, CO2,
SO2, NOx - Destructive
29Electron Capture Detector (ECD)
- This detector exhibits high intensity for halogen
containing compounds and thus has found wide
applications in the detection of pesticides and
polychlorinated biphenyls. The mechanism of
sensing relies on the fact that electronegative
atoms, like halogens, will capture electrons from
a b emitter (usually 63Ni). In absence of
halogenated compounds, a high current signal will
be recorded due to high ionization of the carrier
gas, which is N2, while in presence of
halogenated compounds the signal will decrease
due to lower ionization.
30(No Transcript)
31- Remember the following facts about ECD
- 1. Electrons from a b-source ionize the carrier
gas (nitrogen) - 2. Organic molecules containing electronegative
atoms capture electrons and decrease current - 3. Simple and reliable
- 4. Sensitive (10-15 g/s) to electronegative
groups (halogens) - 5. Largely non-destructive
- 6. Insensitive to amines, alcohols and
hydrocarbons - 7. Limited dynamic range (102)
- 8. Mass sensitive detector
32Gas Chromatographic Columns and Stationary Phases
- Packed Columns
- These columns are fabricated from glass,
stainless steel, copper, or other suitable tubes.
Stainless steel is the most common tubing used
with internal diameters from 1-4 mm. The column
is packed with finely divided particles (lt100-300
mm diameter), which is coated with stationary
phase. However, glass tubes are also used for
large-scale separations.
33- Several types of tubing were used ranging from
copper, stainless steel, aluminum and glass.
Stainless steel is the most widely used because
it is most inert and easy to work with. The
column diameters currently in use are ordinarily
1/16" to 1/4" 0.D. Columns exceeding 1/8" are
usually used for preparative work while the 1/8"
or narrower columns have excellent working
properties and yield excellent results in the
analytical range. These find excellent and wide
use because of easy packing and good routine
separation characteristics. Column length can be
from few feet for packed columns to more than 100
ft for capillary columns.
34Capillary/Open Tubular
- Open tubular or capillary columns are finding
broad applications. These are mainly of two
types - Wall-coated open tubular (WCOT) lt1 mm thick
liquid coating on inside of silica tube - Support-coated open tubular (SCOT) 30 mm thick
coating of liquid coated support on inside of
silica tube - These are used for fast and efficient separations
but are good only for small samples. The most
frequently used capillary column, nowadays, is
the fused silica open tubular column (FSOT),
which is a WCOT column.
35- The external surface of the fused silica columns
is coated with a polyimide film to increase their
strength. The most frequently used internal
diameters occur in the range from 260-320
micrometer. However, other larger diameters are
known where a 530 micrometer fused silica open
tubular column was recently made and is called a
megapore column, to distinguish it from other
capillary columns. Megapore columns tolerate a
larger sample size.
36(No Transcript)
37(No Transcript)
38(No Transcript)
39(No Transcript)
40- It should be noted that since capillary columns
are not packed with any solid support, but rather
a very thin film of stationary phase which
adheres to the internal surface of the tubing,
the A term in the van Deemter equation which
stands for multiple path effects is zero and the
equation for capillary columns becomes - H B/V CV
41- Capillary columns advantages compared to packed
columns - higher resolution
- shorter analysis times
- greater sensitivity
- Capillary columns disadvantage compared to packed
columns - smaller sample capacity
42Solid Support Materials
- The solid support should ideally have the
following properties - Large surface area (at least 1 m2/g)
- Has a good mechanical stability
- Thermally stable
- Inert surface in order to simplify retention
behavior and prevent solute adsorption - Has a particle size in the range from 100-400 mm
43Selection of Stationary Phases
- General properties of a good liquid stationary
phase are easy to guess where inertness towards
solutes is essential. Very low volatility
liquids that have good absolute and differential
solubilities for analytes are required for
successful separations. An additional factor
that influences the performance of a stationary
phase is its thermal stability where a stationary
phase should be thermally stable in order to
obtain reproducible results. Nonvolatile liquids
assure minimum bleeding of the stationary phase
44Weight of liquid stationary phase 100
-
- Loading
- Increasing percent loading would allow for
increased sample capacity and cover any active
sites on the solid support. These two advantages
are very important, however increasing the
thickness of stationary phase will affect the C
term in the van Deemter equation by increasing
HS, and therefore Ht.
Weight of stationary phase plus solid support
45- Generally, the film thickness primarily affects
the retention character and the sample capacity
of a column. Thick films are used with highly
volatile analytes, because such films retain
solutes for a longer time and thus provide a
greater time for separation to take place. Thin
films are useful for separating species of low
volatility in a reasonable time. On the other
hand, a thicker film can tolerate a larger sample
size. Film thicknesses in the range from 0.1 5
mm are common.
46Liquid Stationary Phases
- In general, the polarity of the stationary phase
should match that of the sample constituents
("like" dissolves "like"). Most stationary phases
are based on polydimethylsiloxane or polyethylene
glycol (PEG) backbones
47- The polarity of the stationary phase can be
changed by derivatization with different
functional groups such as a phenyl group.
Bleeding of the column is cured by bonding the
stationary phase to the column or crosslinking
the stationary phase.
- Liquid Stationary Phases should have the
following characteristics - Low volatility
- High decomposition temperature (thermally
stable) - Chemically inert (reversible interactions with
solvent) - Chemically attached to support (to prevent
bleeding) - Appropriate k' and a for good resolution
48Bonded and Crosslinked Stationary Phases
- The purpose of bonding and cross-linking is to
prevent bleeding and provide a stable stationary
phase. With use at high temperatures, stationary
phases that are not chemically bonded or
crosslinked slowly lose their stationary phase
due to bleeding in which a small amount of the
physically immobilized liquid is carried out of
the column during the elution process.
Crosslinking is carried out in situ after a
column is coated with one of the polymers
49(No Transcript)
50- In summary, stationary phases are usually bonded
and/or crosslinked and the following remarks are
usually helpful - 1. Bonding occurs through covalent linking of
stationary phase to support - 2. Crosslinking occurs through polymerization
reactions to join individual stationary phase
molecules - 3. Nonpolar stationary phases are best for
nonpolar analytes where nonpolar analytes are
retained preferentially - 4. Polar stationary phases are best for polar
analytes where polar analytes are retained
preferentially
51Gas-liquid chromatography (GLC)
- Packed columns are fabricated from glass, metal,
or Teflon with 1 to 3 m length and 2 to 4 mm in
internal diameter. The column is packed with a
solid support (100-400 mm particle diameter made
from diatomaceous earth) that has been coated
with a thin layer (0.1-5 mm) of the stationary
liquid phase. Efficiency increases with
decreasing particle size as predicted from van
Deemter equation. The retention is based on
absorption of analyte (partition into the liquid
stationary phase) where solutes must have
differential solubility in the stationary phase
52- Open tubular capillary columns, either WCOT, SCOT
are routinely used. In WCOT the capillary is
coated with a thin film (0.1-0.25 mm) of the
liquid stationary phase while in SCOT a thin film
of solid support material is first affixed to the
inner surface of the column then the support is
coated with the stationary phase. WCOT columns
are most widely used. Capillary columns are
typically made from fused silica (FSOT) and are
15 to 100 m long with 0.10 to 0.5 mm i.d.
53- The thickness of the stationary phase affects the
performance of the column as follows - Increasing thickness of stationary phase allows
the separation of larger sample sizes. - Increasing thickness of stationary phase reduces
efficiency since HS increases. - Increasing thickness of stationary phase is
better for separation of highly volatile
compounds due to increased retention. -
54- Much more efficient separations can be achieved
with capillary columns, as compared to packed
columns, due to the following reasons - Very long capillary columns can be used which
increases efficiency - Thinner stationary phase films can be used with
capillary columns - No eddy diffusion term (multiple paths effect) is
observed in capillary columns
55(No Transcript)
56Temperature Programming
- Gas chromatographs are usually capable of
performing what is known as temperature
programming gas chromatography (TPGC). The
temperature of the column is changed according to
a preset temperature isotherm. TPGC is a very
important procedure, which is used for the
attainment of excellent looking chromatograms in
the least time possible. For example, assume a
chromatogram obtained using isothermal GC at 80
oC, as shown below
57(No Transcript)
58(No Transcript)
59(No Transcript)
60(No Transcript)
61The General Elution Problem
- Look at the chromatogram below in which six
components are to be separated by an elution
process using isothermal conditions at for
example 120 oC - Â
62- It is clear from the figure that the separation
is optimized for the elution of the first two
components. However, the last two components have
very long retention and appear as broad peaks.
Using isothermal conditions at high temperature
(say for example 200oC) can optimize the elution
of the last two compounds but, unfortunately,
results in bad resolution of the earlier eluting
compounds as shown in the figure below where the
first two components are coeluted while the
resolution of the second two components becomes
too bad
63(No Transcript)
64One can also optimize the separation of the
middle too components by adjusting the isothermal
conditions (for example at say 160 oC). In this
case, a chromatogram like the one below can be
obtained
65- However, in chromatographic separations we are
interested in fully separating all components in
an acceptable resolution. Therefore, it is not
acceptable to optimize the separation for a
single component while disregarding the others.
The solution of this problem can be achieved by
consecutive optimization of individual components
as the separation proceeds. In this case,
temperature should be changed during the
separation process. This is called temperature
programming gas chromatography (TPGC)
66- First, a temperature suitable for the separation
of the first eluting component is selected, and
then the temperature is increased so that the
second component is separated and so on. The
change in temperature can be linear, parabolic,
step, or any other formula. The chromatographic
separation where the temperature is changed
during the elution process is called TPGC. A
separation like the one below can be obtained
67(No Transcript)
68Temperature Zones in GC
- Three temperature zones should be adjusted before
a GC separation can be done. The injector
temperature should be such that fast evaporation
of all sample components is achieved. The
temperature of the injector is always more than
that of the column, which depends on the
operational mode of the separation. The detector
temperature should be kept at some level so as to
prevent any solute condensation in the vicinity
of the detector body.
69Gas-solid chromatography (GSC)
- Gas-solid chromatography is based upon adsorption
of gaseous substances on solid surfaces.
Distribution coefficients are generally much
larger than those for gas-liquid chromatography.
Consequently, gas-solid chromatography is useful
for the separation of species that are not
retained by gas-liquid columns, such as the
components of air, hydrogen sulfide, carbon
disulfide, nitrogen oxides, and rare gases.
Gas-solid chromatography is performed with both
packed and open tubular columns.
70Molecular Sieves
- Molecular sieves are metal aluminum silicate ion
exchangers, whose pore size depends upon the kind
of cation present, like sodium in sodium aluminum
silicate molecular sieves. The sieves are
classified according to the maximum diameter of
molecules that can enter the pores. Commercial
molecular sieves come in pore sizes of 4, 5, 10,
and 13 angstroms. Molecular sieves can be used to
separate small molecules from large ones.
71Porous Polymers
- Porous polymer beads of uniform size are
manufactured from styrene crosslinked with
divinylbenzene. The pore size of these beads is
uniform and is controlled by the amount of
crosslinking. Porous polymers have found
considerable use in the separation of gaseous
species such as hydrogen sulfide, oxides of
nitrogen, water, carbon dioxide, methanol, etc.
72Quantitative Analysis
- GC is an excellent quantitative technique where
peak height or area is proportional to analyte
concentration. Thus the GC can be calibrated with
several standards and a calibration curve is
obtained, then the concentration of the unknown
analyte can be determined using the peak area or
height. The detector response factor for each
analyte should be considered for accurate
quantitative analysis.
73- Gas chromatographs are widely used as criteria
for establishing the purity of organic compounds.
Contaminants, if present, are revealed by the
appearance of additional peaks. Qualitative
Analysis is usually done by comparison with
retention times of standards, which are very
reproducible in GC, provided good injection
practices are followed. Injection should be done
with a suitable Hamilton type syringe through the
heated septum injector till all needle
disappears, then the needle is drawn back as
steadily and fast as possible. This is important
for reproducible attainment of retention times.
74The Retention Index
- The retention index, RI, was first proposed by
Kovats in 1958 as a parameter for identifying
solutes from chromatograms. The retention index
for any given solute can be derived from a
chromatogram of a mixture of that solute with at
least two normal alkanes (chain length gtfour
carbons) having retention times that bracket that
of the solute. That is, normal alkanes are the
standards upon which the retention index scale is
based.
75- By definition, the retention index for a normal
alkane is equal to 100 times the number of
carbons in the compound regardless of the column
packing, the temperature, or other
chromatographic conditions. The retention index
system has the advantage of being based upon
readily available reference materials that cover
a wide boiling range. The retention index of a
compound is constant for a certain stationary
phase but can be totally different for other
stationary phases.
76- In finding the retention index, a plot of the
number of carbons of standard alkanes against the
logarithm of the adjusted retention time is first
constructed. The value of the logarithm of the
adjusted retention time of the unknown is then
calculated and the retention index is obtained
from the plot. - The adjusted retention time, tR, is defined as
- tR tR - tM
77(No Transcript)
78Interfacing GC with other Methods
- As mentioned previously, chromatographic methods
(including GC) use retention times as markers for
qualitative analysis. However, this
characteristic does not absolutely confirm the
existence of a specific analyte as many analytes
may have very similar stationary phases. GC, as
other chromatographic techniques, can confirm the
absence of a solute rather than its existence.
When GC is coupled with structural detection
methods, it serves as a powerful tool for
identifying the components of complex mixtures. A
popular combination is GC/MS.
79Mass Spectrometry
O
C
H
3
Mass Spectrometer
C
N
C
H
3
C
N
C
H
C
C
N
N
O
H
Typical sample isolated compound (1 nanogram)
194
Mass Spectrum
67
109
Abundance
55
82
42
136
165
94
40
60
80
100
120
140
160
200
180
Mass (amu)
80(No Transcript)