OVERVIEW OF MQSA INSPECTION FINDINGS - PowerPoint PPT Presentation

1 / 23
About This Presentation
Title:

OVERVIEW OF MQSA INSPECTION FINDINGS

Description:

FINDING LEVEL DISTRIBUTION - Historical Perspective. INSPECTION FINDINGS ... Equipment performance including dose, phantom image, processing, & darkroom fog ... – PowerPoint PPT presentation

Number of Views:108
Avg rating:3.0/5.0
Slides: 24
Provided by: walidg
Category:

less

Transcript and Presenter's Notes

Title: OVERVIEW OF MQSA INSPECTION FINDINGS


1
OVERVIEW OF MQSA INSPECTION FINDINGS
  • 8/22/01
  • Walid G. Mourad, Ph.D., CSO
  • Inspection Support Branch
  • DMQRP/OHIP/CDRH/FDA

2
OVERVIEW OF MQSA INSPECTION FINDINGS
  • BACKGROUND
  • FINDING LEVEL DISTRIBUTION - Historical
    Perspective
  • INSPECTION FINDINGS - Final Regulations
  • L1 L2 FINDINGS - The last two years
  • PROGRAMS UNDERWAY

3
BACKGROUND
  • MQSA (10/92)
  • Interim Regulations (10/94)
  • Final Regulations (published 10/97, effective
    4/99)
  • MQSRA (10/98)
  • Lay summary to all women
  • Release of original mammogram
  • Demonstration Program Inspection Frequency

4
Inspection Scope
  • Equipment performance including dose, phantom
    image, processing, darkroom fog
  • Quality assurance quality control test records
    (including survey report mammography equipment
    evaluations)
  • Consumer complaint mechanism
  • Personnel qualifications
  • Medical reports and lay summaries
  • Medical outcomes audit

5
MQSA Finding Levels
  • Level 1
  • Serious - Warning Letter within 15 days
    subsequent facility response within 15 days
  • Level 2
  • Moderate - 30 day response
  • Level 3
  • Minor - Follow up at next inspection

6
Inspection Findings by Level - FY 95 to FY 01
( of facilities with highest level findings)
  • L 1 L 2 L 3 No Finding
  • FY95 2.6 19.8 47.4 30.0
  • FY96 1.5 12.4 42.2 43.7
  • FY97 0.9 12.4 30.7 55.8
  • FY98 1.1 18.7 21.6 58.4
  • FY99 1.6 23.5 16.2 58.5
  • FY00 3.9 32.6 10.3 53.0
  • FY01 3.5 28.0 10.1 58.2
  • (to 7/31/01)

7
Inspection Findings by Level - FY 95 to FY 01
( of facilities with highest level findings)
FY to 7/31/01
8
of Facilities w/L1 Findings - FY95-FY01
(to 7/31)
9
of Facilities w/L2 Findings - FY95-FY01
(to 7/31)
10
of Facilities w/Highest Finding Levels FY 99
  • L 1 L 2 L 3 No Finding
  • FY99 (total) 1.6 23.5 16.2 58.5
  • FY99 (I) 0.9 18.9 18.3 61.7
  • (Q1-Q3)
  • FY99 (F) 3.6 36.8 10.1 49.2
  • (Q4)

11
of Facilities w/L1 Findings - FY 99
--- FY 99 total (1.6)
12
Inspection Findings Under the Final Regulations
  • Level Changes Consequences
  • L 1 L 2 Findings Two Year Results
  • Future Trend

13
Final Regulations - Perspective
  • For the final regulations, FDA
  • Added new requirements ? new findings at all
    levels
  • Elevated several L3 L2 findings
  • Deleted several L3 findings
  • As a result
  • L3 findings decreased
  • L2 L1 findings increased

14
Number of Potential Findings
15
of Facilities w/L1 Findings FY 00
--- FY 00 total (3.9)
16
of Facilities w/L2 Findings FY 00
--- FY 00 total (32.6)
17
of Facilities w/L1 Finding FY 01
(to 7/31)
--- FYTD 01 total (3.5)
18
of Facilities w/L2 Findings FY 01
(to 7/31)
--- FY 00 total (28)
19
Facilities with Level 1 Findings Recent Results
(10/1/00 - 7/2/01)
  • Number
  • L1 Findings (facilities cited) 245 3.5
  • Processor or Phantom QC 159 2.3
  • Results Communication 57 0.8
  • Initial Qualifications - Any Personnel 34 0.5
  • Other lt 9 lt 0.1
  • No Valid Certificate
  • Unaccredited Unit in Use for gt 1y
  • Phantom score
  • Underprocessing (STEP test)
  • No Survey for 2 years

20
Facilities with Level 2 Findings (as highest)
Recent Results (10/1/00 - 7/2/01)
  • Number
  • L2 Findings (facilities cited) 1899 27.6
  • Processor or Phantom QC 862 12.6
  • L2 Requirements - Any Personnel 602 8.8
  • Med. Reports w/o Results/ID 467 6.8
  • Med. Outcomes Audit System 227 3.3
  • Survey Rep. MEE 299 4.4
  • No S.O.P. Consumer Complaints 199 2.9
  • No S.O.P. Infection Control 110 1.6
  • Other lt 45 lt 0.7
  • Unaccredited Unit in Use for lt 1y
  • Phantom score
  • Underprocessing (STEP test)

21
FUTURE TREND
  • Based on experience to date, we believe that in
    the foreseeable future ( 1 year)
  • Facilities with L1citations lt 2.5
  • Facilities with L2 citations lt 25.0
  • Facilities with L3 citations lt 10.0

22
PROGRAMS UNDERWAY
  • Demonstration Program
  • Scheduled to start May 2002
  • About 300 eligible facilities (14 states) in the
    pool
  • 150 will be inspected once in 2 years
  • New Mammographic Modality FFDM
  • Currently only GEs Senographe 2000D (since 6/00)
    - others expected in the near future
  • Small number of facilities units
  • 8 hours training in FFDM are the only citations
    implemented to date

23
...for more Information on MQSA
  • MQSA Internet home page
  • http//www.fda.gov/cdrh/mammography
  • MQSA facility hotline 1-800-838-7715
  • (DSMA) Facts on Demand 1-800-899-0381
Write a Comment
User Comments (0)
About PowerShow.com