Title: PMT Calibration Results and the MC simulation
1PMT Calibration Results and the MC simulation
- Shigeru Yoshida, Chiba University
- http//www.ppl.phys.chiba-u.jp/
2Outline
- Overview of the IceCube PMT calibration for the
detector MC - Absolute QE/CE
- Uniformity of CE
- Charge Resolution/WaveForm
- The Response Modeling
- A DOM-analog MC prototype
3The DOM/PMT calibration
Gain/Waveform (by S.Yoshida Mons2003)
QE/CE Uniformity (by H.Miyamoto Mons2003/this
meeting)
- DOM 2D Scan
- (Scheduled this year)
4The Absolute QE/CE measurement
Reflectors/Attenuators
Energy Meter
5Attachment for the IceCube PMT
- Photon beam with 13 mm diameter scans the cathode
6CE Uniformity2D scan
PMT
UV LED
R-guide
Rotation-bed
Motor
7The Collection Efficiency Lego plot
8Example The relative Collection Efficiency
SF0030 vs SF0050
SF0016 vs SF0050
9(No Transcript)
10(No Transcript)
11Charge Resolution/Waveform
In the Freezer
Temp 32degree
12PMT Setting in the Freezer
Base circuit
Diffuser attached to UV LED
13SPE Charge Spectrum
14SPE Waveform
15The calibration data for the MC simulation
- The absolute QE/CE data of 4 PMTs now under
measurement - The CE uniformity data of 16 PMTs available
- The SPE spectrum and the waveform _at_ -30C At HV
from 1600V to 2000 V for 4 PMTs - The Gain uniformity data of 1 PMT (possible to
add 3 more PMTs) - We are ready to add more PMT data this summer in
the pre-shipping measurement.
16The DOM response Modeling
Charge Response/Waveform
17SPE Charge Response
- Sspe(q) (1-Pe)exp-q/qt Pe NG(q-q0/qs)
18SPE Charge Response ContdSome GOOD News
- qt/q0, qs/q0 universally scaling
-
19SPE Charge Response ContdSome GOOD News
- Global qt/q0, qs/q0 can reconstruct the response
of ALL PMTs at every gain
20And some BAD News
- GAIN seems to have some position dependences.
21(No Transcript)
22We see it in the CE scan data
CE X Gain
- We may be able to ignore this because
- so small in terms of AREA
- Simulating MPEs response would be OK
- ONLY SPE simulation might be problematic
23SPE Waveform Model
24SPE Waveform Model
- FWHM of a SPE pulse depends on PMT gain.
- But the change is small 200 psec or so
25Some Issues and Proposal
- Uniformity exhibits significant tube to tube
variance. We do not want to implement 4800 tables
in the DOM MC. - SIMULATE the response for the broader light beam
as a function of angle and compare it with the
measurement at UW.
26Some Issues and Proposal (Contd)
- Localized gain variance has been found.
- Use the AVERAGE behavior ?
27Some Issues and Proposal (Contd)
- Localized gain variance has been found.
- Just ignore it!
28DOM Analog MC prototype
DOMVPmtWaveform.cc
DOMVPmtCharge.cc
- (1-Pe)exp-q/qt Pe NG(q-q0/qs)
GEANT4
DOMVPmtQe.cc
DOMVGlass.cc DOMVGel.cc
29Summary
- The PMT calibration data is ready for the
detector MC - Modeling the PMT response was done.
- A 1st prototype PMT MC module is being built
under the root framework - Running this MC would address some concerns like
the tube-tube variance - Please bookmark
- http//www.phys.chiba-u.jp/research/IceCube/PMT
/