Title: ELECTRON SPIN RESONANCE
1ELECTRON SPIN RESONANCE
2Definition
- It is also called electron paramagnetic
resonance (EPR) or electron magnetic resonance
(EMR). - Is a branch of absorption spectroscopy in which
radiation of microwave frequency is absorbed by
paramagnetic substances. - It was invented by Zavoisky in 1944.
3- ESR is concerned with magnetic behaviour of
spinning electrons. - ESR spectrum is obtained by transition from one
spin state to other state of an electron,
transition is induced by radiation of microwave
frequency. - ESR uses reagents called electron spin reagents,
which are stable containing odd electrons have
the capacity to react with aminoacids. - ESR of compounds give information about
structure, viscosity, polarity, phase
transformation and chemical reactivity.
4- Chemical shift is absent.
- The transition b/w electrons occurs upon the
absorption of a quantum of radiation in microwave
region . - The frequency required is 8000MCS-1 .
- Limited application due to systems containing
unpaired electrons.
- Chemical shift is present .
- The transition occurs upon the application of
radiofrequency. - The frequency required is 40MCS-1
5THEORY
- The energy levels are produced by the
interaction of the magnetic moment of an unpaired
electron in a molecule ion with an applied
magnetic field. - ESR spectrum is due to transition b/w these
energy levels by absorbing radiations of
microwave frequency.
6- Like a proton, an electron has a spin, which
gives it a magnetic property known as a magnetic
moment. - When an external magnetic field is supplied, the
paramagnetic electrons can either orient in a
direction parallel or antiparallel to the
direction of the magnetic field - This creates two distinct energy levels for the
unpaired electrons.
7(No Transcript)
8(No Transcript)
9- For an electron of spin s1/2, the spin angular
momentum have values of - ms 1/2 or -1/2
- In the absence of magnetic field, the two values
give rise to doubly degenerate energy levels(same
energy).
10- If a magnetic field is applied, this leads to two
non-degenerate energy levels. - The low energy state will have the spin magnetic
moment alligned with the field (ms -1/2), and - high energy state - opposite to field (ms
1/2).
11(No Transcript)
12(No Transcript)
13- These two states possess energies split up from
original state with no applied magnetic field by
µe H (low energy state) - -µeH (high energy state)
- Where, H magnetic field
- µe magnetic moment
- In ESR, transition b/w different energy levels
take place by absorbing quantum of radiaton of
ferquency in microwave region
14?EhvgßB
h Plancks constant 6.626196 x 10-27
erg.sec n frequency (GHz or MHz) g g-factor
(approximately 2.0) ß Bohr magneton (9.2741 x
10-21erg.Gauss-1) B magnetic field (Gauss or mT)
15Thus ESR spectrum of free electron consist of
single peak corresponding to transition b/w two
levels. When absorption occurs 2 µeH
hv Where , v - frequency of absorbed radiation
(cycles/sec) The energy of transition is given
by ?E hv gßH Where, ß Bohrs
magneton ( a factor for converting angular
momentum into magnetic moment) g
proportionality factor(a function of electron
environment). Also called spectroscopic
splitting factor OR Landes splitting factor.
The value is not constant
16- g is proportionality factor, function of
electrons enviornment,also called as
spectroscopic spliting factor or Landes spliting
factor - g is not constant.
- For a free electron , value is 2.0023.In free
radicals, value is same(electron character is
same).
17- In some crystals, g vary from 0.2 to 0.8.The
reason is that unpaired electron is localised
in a paricular orbital about the atom and the
orbital angular momentum couples with spin
angular momentum give rise to a low value of g
in ionic crystals. - g value depend upon the orientation of the
molecule having the unpaired electron w.r.t
applied magnetic field. - In a crystal the value of g along x,y,z axes
denoted as gx,gy,gz.
18INSTRUMENTATION
19 Most EPR spectrometers are similar to the
original instrumentation used for NMR before the
development of pulsed Fourier Transform
spectroscopy. Most also use electromagnets
rather than permanent magnets or cryomagnets.
20(No Transcript)
21(No Transcript)
22- SOURCE
- Klystron
- most common source operating in MW band region of
3 cm ? - Vacuum pump, can produce MW oscillations centered
on small range of frequency - Frequency on monochromatic radiation determined
by applied voltage to Klystron - Operated at 9500 Mc/sec.
23- b. Isolator
- A non-reciprocal device, a strip of ferrite
material - Minimises vibrations in frequency of MW produced
by Klystron oscillator - Variation in frequency due to backward reflection
in region b/w Klystron circulator - Wavemeter
- b/w isolator attenuator to know frequency of MW
produced by Klystron oscillator - Calibrated in frequency units
- Attenuator
- b/w wavemeter circulator, adjusts the level of
MV power incident upon sample - Possesses an absorption element
24- CIRCULATOR / MAGIC T
- MW radiations enter circulator through a wave
guide by a loop of wire that couples with
oscillating magnetic field sets up a
corresponding field in wave guide - Made up of hollow rectangular copper or brass
tubing with silver or gold plating inside to
produce a highly conducting flat surface
25- SAMPLE CAVITY
- Cavity system is constructed in such a way as to
maximise applied magnetic field along the sample
dimension - 0.15 0.5 ml sample can be used (less DEC)
- Flat cells (0.25 mm thickness) used for samples
(high DEC) 0.05 ml sample - Most ESR spectrometers use dual sample cavities
for sample and reference
26- 4. MAGNET SYSTEM
- A static magnetic field is provided by an
electromagnet with a current- regulated power
supply. - A homogeneous and stable field is required for
best results. - The resonant cavity is placed b/w pole pieces of
electromagnet - A Hall probe, driven from a stable
constant-current power system, with a digital
multimeter (DMM) reading the Hall voltage, is
used to measure the value of the magnetic field
between the poles of the magnet
27- CRYSTAL DETECTORS
- Silicon crystal is most commonly used detector
- Act as a microwave rectifier
- Convert MW power into direct current output
- AMPLIFIER AND PHASE SENSITIVE DETECTOR
- The signal from crystal detector undergo
narrow-band amplification by auto amplifier - Noise present in the amplified signal is removed
by phase sensitive detector - OSCILLOSCOPE AND PEN RECORDER
28- WORKING
- Sample is kept in resonant cavity
- The cavity is a long path-length cell in which
wavas are reflected to fro thousand of times - Klystron oscillator produce MW that pass through
isolator, wave meter and attenuator to reach
circulator through arm 1 - MW power divide b/w arms 2 3 (sample cavity)
- Arm 2 have balancing resistance
- When total resistance on 2 3 are same, MW power
is absorbed completely - When R on arm 3 changes due to some ESR resonance
absorption by the sample, then some MW power
enter into detector through arm 4 -
29- detector convert the MW power into DC
- If the magnetic field around the sample is
changed to the value required for resonance,
absorption takes place by the sample - The current is recorded by the recorder and show
an absorbance peak - If main magnetic field is swept slowly for
several minutes, recorder shows the derivative of
the absorption spectrum - generally absorption spectra are recorded as
first derivative spectra
30(No Transcript)
31- PRESENTATION OF ESR SPECTRUM
- ESR spectrum is obtained by plotting intensity
against the strength of magnetic field - ESR spectrum is better represented by derivative
curve - First derivative (slope) of absorption curve is
plotted against strength of magnetic field - Each negative slope in derv. curve represents a
peak or shoulder in absorption spectrum - Every crossing of the derv. axis with a negative
slope indicates a true maximum and crossing with
positive slope indicates minimum - Thus no. of peaks/shoulders in absorption curve
can be determined from no. of minima or maxima in
the derivative curve.
32- Area covered by either the absorption or
derivative curve is proportional to no of
unpaired electrons in the sample - To calculate no of electrons in the sample,
comparison is made with standard sample having
known no of unpaired electrons and possessing the
same line shape - Most widely used std is 1,1-diphenyl-picrylhydrazy
l (DPPH) free radical
33 DPPH is chemically stable Splitting
factor g 2.0036 1,53 1021unpaired electrons
per gm