Signal Averaged ECG - PowerPoint PPT Presentation

1 / 62
About This Presentation
Title:

Signal Averaged ECG

Description:

Signal Averaged ECG . . . ... – PowerPoint PPT presentation

Number of Views:1504
Avg rating:3.0/5.0
Slides: 63
Provided by: asm104
Category:

less

Transcript and Presenter's Notes

Title: Signal Averaged ECG


1
Signal Averaged ECG
S?µ??f?st??? ??e?t???a?d?????f?µa
  • ??t???? S. ?a?????
  • G.?. ?????? ? ??a??e??sµ??

2
(No Transcript)
3
(No Transcript)
4
(No Transcript)
5
(No Transcript)
6
(No Transcript)
7
(No Transcript)
8
High Resolution Electrocardiography
  • A high-resolution electrocardiogram detects very
    low amplitude signals in the ventricles called
    'Late Potentials' in patients with abnormal heart
    conditions.
  • A standard electrocardiogram cannot detect these
    signals.
  • The presence of late potentials is widely
    accepted to have prognostic significance in
    patients after AMI

9
SAECG
  • The ECG is a graphical representation of the
    electrical potentials generated by the heart
  • Based on the resolution of the digital recording
    of analog ECG signals, the instruments
    techniques may be categorized into 2 types ? 1)
    Low-resolution (or standard) ECG, and ? 2)
    High-Resolution ECG (HRECG)
  • A standard 12-Lead ECG is a typical example of a
    widely used low-resolution instrument that
    records 9 sec of cardiac data
  • A SAECG is a typical example of a High-Resolution
    ECG
  • SAECG records ventricular ECG signals of very low
    magnitudes called 'Ventricular Late Potentials'
    (VLP) by averaging a number of signals (QRS)
  • The presence of VLPs is indicative of ?risk for
    subsequent occurrence of arrhythmic events,
    mainly SuVT

10
(No Transcript)
11
(No Transcript)
12
An HRECG instrument consists of 4 key components
1) Amplifiers, 2) Bandpass filters, 3)
Analog/Digital converter, and 4) SAECG Processor.
The SAECG Processor may in turn be functionally
divided into the following components a)
Signal Averager, b) Bidirectional Bandpass
Filter, c) Filtered Vector Magnitude, d) SAECG
Quantifier. In addition, the instrument
includes 7 ECG leads. These leads are bipolar,
orthogonal electrodes comprising X, X-, Y, Y-,
Z, Z-, ground placed in a particular fashion
on the body surface. These electrodes are
usually referred to as XYZ leads.
13
SAECG
  • Signal-averaged electrocardiography (SAECG) is a
    technique involving computerized analysis of
    segments of a standard surface electrocardiogram.
  • It is used for detecting small electrical
    impulses, termed ventricular late potentials,
    that follow the QRS segment.
  • These signals are embedded in the
    electrocardiogram but ordinarily obscured by
    skeletal muscle activity and other extraneous
    sources of "noise" encountered in recording a
    standard electrocardiogram.

14
Late Potentials
  • Ventricular late potentials in patients with
    cardiac abnormalities, especially coronary artery
    disease or following an acute myocardial
    infarction, are associated with an increased risk
    of ventricular tachyarrhythmias and sudden
    cardiac death.
  • Proponents of SAECG claim that it can obviate the
    need for invasive techniques commonly used to
    identify high-risk patients for interventions
    that treat or prevent ventricular tachyarrhythmia
    and sudden death.

15
SAECG
  • The current data on SAECG show relatively
    consistent high negative predictive values, poor
    positive predictive values, and variable
    sensitivity and specificity when the technique is
    used on pts with CM or following a MI
  • The available evidence also indicates that
    combining SAECG with other tests of cardiac
    function is superior to using any single test for
    risk.
  • The utility of SAECG alone as an indicator of
    risk remains to be proven.
  • SAECG combined with other standard tests of risk
    has been demonstrated to have clinical utility in
    patients following an acute myocardial
    infarction.
  • Other patient populations have not been
    conclusively shown to benefit from its use.

16
SAECG Normal (left) and abnormal (right)
results are shown from a patient with prior MI
and VT. Bottom panels Shaded blue areas at the
end of each tracing represent voltage content of
last 40 ms of the filtered QRS integral. The
small shaded area in the abnormal study denotes
prolonged, slow conduction and suggests the
potential for reentrant ventricular arrhythmias.
17
Late Potentials
  • One of the constituents of reentrant ventricular
    arrhythmias in patients with prior myocardial
    damage is slow conduction.
  • Direct cardiac mapping techniques can record
    myocardial activation from damaged areas that
    occurs after the end of the surface ECG QRS
    complex during sinus rhythm.
  • These delayed signals have very low amplitude
    that cannot be discerned on routine ECG and
    correspond to the delayed and fragmented
    conduction in the ventricles recorded with direct
    mapping techniques

18
SAECG
  • Signal averaging has been applied clinically most
    often to detect such late ventricular potentials
    of 1 to 25 µV
  • Criteria for late potentials are
  • (1) filtered QRS complex duration (QRSD) gt114
    120 ms,
  • (2) lt 20 µV of root-mean-square (RMS) signal
    amplitude in the last 40 ms of the filtered QRS
    complex, and
  • (3) the terminal filtered QRS complex remains
    below 40 µV (low amplitude signal-LAS) for longer
    than 38 ms

19
Time-Domain Analysis Results of most studies
have been based on analysis of a vector magnitude
of the filtered leads, vx2y2 z2, called the
filtered QRS complex. The end of the filtered
QRS complex is defined as the midpoint of a 5
msec segment in which mean voltage exceeds the
mean noise level plus 3 times the standard
deviation of the noise sample. The end point and
onset of the filtered QRS complex should be
verified visually, and the system should allow
manual adjustment of the automatically
determined end points.
20
SAECG Criteria
  • Analysis should include determination of
  • 1) the filtered QRS duration
  • 2) root mean square voltage of the terminal 40
    msec of the filtered QRS and
  • 3) amount of time that the filtered QRS complex
    remains below 40 µV
  • For the definition of a late potential and the
    scoring of a high-resolution ECG as normal or
    abnormal
  • Representative criteria include that a late
    potential exists (using 40 Hz high-pass
    bidirectional filtering) when
  • 1) the filtered QRS complex is greater than 114
    msec,
  • 2) there is less than 20 µV of signal in the last
    40 msec of the vector magnitude complex, and
  • 3) the terminal vector magnitude complex remains
    below 40 µV for more than 38 msec

21
(No Transcript)
22
(No Transcript)
23
Late Potentials
  • These late potentials have been recorded in 70
    90 of patients with spontaneous sustained and
    inducible VT after myocardial infarction,
  • in only 0 to 6 of normal volunteers, and
  • in 7 to 15 of patients after myocardial
    infarction who do not have VT

24
Late Potentials
  • Late potentials can be detected as early as 3 h
    after the onset of chest pain, increase in
    prevalence in the first week after MI, and
    disappear in some patients after 1 year.
  • If not present initially, late potentials usually
    do not appear later.
  • Early use of thrombolytic agents may reduce the
    prevalence of late potentials after coronary
    occlusion
  • Patients with BBB or paced ventricular rhythms
    have wide QRS complexes already, rendering the
    technique less useful in these cases.

25
Late Potentials
  • Late potentials also have been recorded in
    patients with VT not related to ischemia, such as
    dilated cardiomyopathies.
  • Successful surgical resection of the VT can
    eliminate late potentials but is not necessary to
    cause tachycardia suppression.
  • The presence of a late potential is a sensitive,
    but not specific, marker of arrhythmic risk and
    thus its prognostic use is limited

26
Late Potentials
  • In specific situations, LPs can be helpful
  • for instance, a patient with a prior inferior
    wall myocardial infarction (normally the last
    portion of the heart to be activated) who has no
    late potential has a very low likelihood of
    having VT episodes.

27
SAECG/ Time Domain Analysis
  • The high-pass filtering used to record late
    potentials meeting the criteria just noted is
    called time domain analysis because the filter
    output corresponds in time to the input signal.
  • Because late potentials are high-frequency
    signals, Fourier transform can be applied to
    extract high-frequency content from the
    signal-averaged ECG, called frequency domain
    analysis.

28
SAECG/ Frequency Analysis
  • A sequence generated by sampling a time-domain
    signal like the ECG can be represented in the
    frequency domain by taking the fast Fourier
    transform.
  • Even though the informational content in time-
    and frequency-domain representations of a
    periodic waveform is equivalent, the extent to
    which each can depict components of interest
    depends on the signal being analyzed.
  • The Fourier transform is a complete description
    of the ECG and contains information that may not
    be seen in the output of a particular fixed-band
    filter.
  • Frequency analysis offers potential advantages
    for identification and characterization of
    signals that differentiate patients with from
    those without sustained ventricular tachycardia.
  • Most studies have calculated the fast Fourier
    transform to estimate scalar-lead spectra of the
    terminal QRS and ST segment of signal-averaged
    Frank X, Y, Z or uncorrected orthogonal
    leads.11-16,47,59
  • The results have often been expressed as indexes
    of the relative contributions of specific
    frequencies that comprise these ECG segments.
  • A window function such as the four-term
    Blackman-Harris window has been used to diminish
    spectral leakage caused by edge discontinuities.

29
SAECG/ Frequency Analysis
  • Key issues that affect the spectra of ECG signals
    are being investigated. For example, the
    frequency content of ECG signals is spatially
    variable and thus lead dependent.
  • Indexes derived from spectra of uncorrected leads
    may not be comparable to end points or approaches
    developed using corrected leads.
  • Analysis of multiple segments (spectrotemporal
    mapping) may allow better separation between
    noise and late potentials
  • The value of autoregressive models of spectral
    estimation and analysis of the entire cardiac
    cycle with methods that obviate window functions
    are currently being determined.6263
  • Accordingly, the committee believes it is
    premature to standardize this approach at present

30
SAECG
  • Some data suggest that frequency domain analysis
    provides useful information not available in the
    time domain analysis.
  • Signal averaging has been applied to the P wave
    to determine risk for developing atrial
    fibrillation as well as maintenance of sinus
    rhythm after cardioversion. The overall use of
    the technique remains limited at present.

31
(No Transcript)
32
A 90 year old lady with syncope
33
(No Transcript)
34
(No Transcript)
35
(No Transcript)
36
(No Transcript)
37
(No Transcript)
38
(No Transcript)
39
(No Transcript)
40
(No Transcript)
41
(No Transcript)
42
(No Transcript)
43
(No Transcript)
44
(No Transcript)
45
SAECG/Noise
  • noise be measured in the averaged signal over an
    interval of at least 40 msec in the ST or TP
    segment with a four-pole Butterworth filter.
  • With this approach, noise should be lt 1 µV with a
    25 Hz high-pass cutoff or
  • lt 0.7 µV with a 40 Hz high-pass cutoff as
    measured by the root mean square method from a
    vector magnitude of the X, Y, and Z leads.
  • The segment for noise level analysis should be
    determined automatically
  • The inherent noise level of the recording should
    be low so that adequate noise reduction can be
    achieved by averaging 50-300 beats.
  • Averaging a greater number of beats to obtain
    adequate noise reduction indicates that baseline
    noise is excessive for optimum recording.

46
SAECG / Indications
  • Vulnerability to SuVT ? risk stratification
    post-MI
  • Unexplained syncope
  • Post-operative patients

47
Heart Rate Variability-HRV
  • Heart rate variability has become the
    conventionally accepted term to describe
    variations of both instantaneous heart rate RR
    intervals
  • oscillation in the interval between consecutive
    heartbeats as well as the oscillations between
    consecutive instantaneous heart rates
  • the interval between consecutive beats is being
    analyzed rather than the heart rate per se

Measurement of HRV
  • Time Domain Methods
  • Continuous ECG record each QRS complex is
    detected, so-called normal-to-normal (NN)
    intervals or the instantaneous HR is determined
  • Simple time domain variables mean NN interval,
    mean heart rate, difference between the longest
    shortest NN interval, difference between night
    day heart rate, etc

48
HRV
  • Simplest variable to calculate the SD of the NN
    intervals (SDNN), that is, the square root of
    variance. Since variance is mathematically equal
    to total power of spectral analysis, SDNN
    reflects all the cyclic components responsible
    for variability in the period of recording
  • short-term 5-min recordings nominal 24-h
    long-term recordings appear to be appropriate
    options
  • Other commonly used statistical variables
    SDANN?SD of the average NN intervals calculated
    over 5 min (an estimate of changes in HR due to
    cycles gt5 min), the SDNN index ? mean of the
    5-min SDs of NN intervals calculated over 24 h
    (variability due to cycles lt 5 min)
  • Most commonly used measures derived fm interval
    differences include RMSSD ?square root of the
    mean squared differences of successive NN
    intervals, NN50 ?the number of interval
    differences of successive NN intervals gt50 ms,
    pNN50 ? the proportion derived by dividing NN50
    by the total number of NN intervals. All of these
    measurements of short-term variation estimate
    high-frequency variations in heart rate thus
    are highly correlated

49
Relationship between the RMSSD and pNN50 (a)
and pNN50 and NN50 (b) measures of HRV
assessed from 857 nominal 24-hour Holter tapes
recorded in survivors of acute MI before hospital
discharge. The NN50 measure used in b was
normalized in respect to the length of the
recording
50
To perform geometric measures on the NN interval
histogram, the sample density distribution D is
constructed, which assigns the number of equally
long NN intervals to each value of their lengths.
The most frequent NN interval length X is
established, that is, YD(X) is the maximum of
the sample density distribution D. The HRV
triangular index is the value obtained by
dividing the area integral of D by the maximum Y.
When the distribution D with a discrete scale is
constructed on the horizontal axis, the value is
obtained according to the formula HRV
index(total number of all NN intervals)/Y. For
the computation of the TINN measure, the values N
and M are established on the time axis and a
multilinear function q constructed such that
q(t)0 for tN and tM and q(X)Y, and such that
the integral ?08 (D(t)-q(t))2 dt is the minimum
among all selections of all values N and M. The
TINN measure is expressed in milliseconds and
given by the formula TINNM-N.
51
Selected Time Domain Measures of HRV
Since many of the measures correlate closely with
others, the following 4 measures are recommended
(1) SDNN (estimate of overall HRV), (2) HRV
triangular index (estimate of overall HRV), (3)
SDANN (estimate of long-term components of HRV),
and (4) RMSSD (estimate of short-term components
of HRV) 2 estimates of the overall HRV are
recommended because the HRV triangular index
permits only casual preprocessing of the ECG
signal. The RMSSD method is preferred to pNN50
NN50 because it has better statistical properties
52
Frequency Domain Analysis
  • Power spectral density (PSD) analysis provides
    the basic information of how power (variance)
    distributes as a function of frequency.
  • Short-term recordings 3 main spectral components
    are distinguished in a spectrum calculated from
    short-term recordings of 2 to 5 min VLF, LF,
    HF components
  • Measurement of VLF, LF, HF is made in absolute
    values of power (ms squared). LF HF may also be
    measured in normalized units, which represent the
    relative value of each power component in
    proportion to the total power minus the VLF
  • Long-term recordings. Spectral analysis also may
    be used to analyze the sequence of NN intervals
    of the entire 24- period. The result then
    includes a ULF component, in addition to VLF, LF,
    HF components. The slope of the 24-h spectrum
    also can be assessed on a log-log scale by linear
    fitting the spectral values
  • Although time domain methods, esp. SDNN RMSSD,
    can be used to investigate recordings of short
    durations / frequency methods are usually able to
    provide results that are more easily
    interpretable in terms of physiological
    regulations.
  • In general, time domain methods are ideal for the
    analysis of long-term recordings
  • Substantial part of the long-term HRV value is
    contributed by the day-night differences. Thus,
    the long-term recording analyzed by the time
    domain methods should contain at least 18 hours
    of analyzable ECG data that include the whole
    night

53
Spectral analysis of RR interval variability in a
healthy subject at rest during 90 head-up
tilt.
At rest, 2 major components of similar power are
detectable at low and high frequencies. During
tilt, the LF component becomes dominant, but as
total variance is reduced, the absolute power of
LF appears unchanged compared with rest.
Normalization procedure leads to predominant LF
smaller HF components, which express the
alteration of spectral components due to tilt.
Circulation 1996931043-1065
54
Selected Frequency Domain Measures of HRV
55
Example of an estimate of power spectral density
obtained from the entire 24-hour interval of a
long-term Holter recording
56
Approximate Correspondence of Time Domain and
Frequency Domain Methods Applied to 24-Hour ECG
Recordings
57
Physiological Correlates of HRV
  • Vagal activity is the major contributor to the HF
    component.
  • Disagreement exists in respect to the LF
    component. Some studies suggest that LF, when
    expressed in normalized units, is a quantitative
    marker of sympathetic modulations other studies
    view LF as reflecting both sympathetic activity
    and vagal activity. Consequently, the LF/HF ratio
    is considered by some investigators to mirror
    sympathovagal balance or to reflect the
    sympathetic modulations.
  • Physiological interpretation of lower-frequency
    components of HRV (that is, of the VLF and ULF
    components) warrants further elucidation.
  • It is important to note that HRV measures
    fluctuations in autonomic inputs to the heart
    rather than the mean level of autonomic inputs.
    Thus, both autonomic withdrawal and saturatingly
    high level of sympathetic input lead to
    diminished HRV

58
Changes of HRV Related to Specific Pathologies
  • MI
  • ?HRV after MI may reflect a ? in vagal activity
    directed to the heart, which leads to prevalence
    of sympathetic mechanisms and to cardiac
    electrical instability. In the acute phase of MI,
    the ? in 24-hour SDNN is significantly related to
    LV dysfunction, peak CK, Killip class
  • an ? LF a ? HF were observed during both
    resting controlled conditions 24-hour
    recordings analyzed over multiple 5-min periods
  • These changes may indicate a shift of
    sympathovagal balance toward a sympathetic
    predominance and a ? vagal tone
  • Diabetic Neuropathy
  • a ? in time domain parameters of HRV seems not
    only to carry negative prognostic value but also
    to precede the clinical expression of autono-mic
    neuropathy / the initial manifestation of this
    neuropathy is likely to involve both efferent
    limbs of ANS (no change in spectral analysis)
  • Heart Failure
  • A ? HRV has been observed consistently in pts
    with HF

59
Modifications of HRV by Specific Interventions
  • The rationale for trying to modify HRV after MI
    stems from the multiple observations indicating
    that cardiac mortality is higher among those
    post-MI patients who have a more depressed HRV
  • The inference is that interventions that augment
    HRV may be protective against cardiac mortality
    and sudden cardiac death.
  • Although the rationale for changing HRV is sound,
    it also contains the inherent danger of leading
    to the unwarranted assumption that modification
    of HRV translates directly into cardiac
    protection, which may not be the case
  • The target is the improvement of cardiac
    electrical stability, and HRV is just a marker of
    autonomic activity
  • ß-Adrenergic Blockade and HRV
  • AADs HRV some AADs a/w increased mortality can
    reduce HRV
  • Scopolamine
  • Thrombolysis
  • Exercise training

60
Clinical Use of HRV
  • General consensus of practical use of HRV has
    been reached only in 2 clinical scenarios ? HRV
    can be used as a predictor of risk after AMI as
    an early warning sign of DM neuropathy

Cumulative survival of pts after MI a, Survival
of pts stratified according to 24-h SDNN values
into 3 groups with cutoff points of 50 100 ms
b, Similar survival curves of pts stratified
according to 24-h HRV triangular index values
with cutoff points of 15 20 units, respectively
Data suggest that ? HRV is not a simple
reflection of sympathetic overdrive and/or vagal
withdrawal due to poor LV performance but that it
also reflects ? vagal activity, which has a
strong association with the pathogenesis of
ventricular arrhythmias sudden cardiac death
61
Predictive value of HRV
The predictive value of HRV alone is modest.
Combination with other techniques substantially
improves the positive predictive accuracy of HRV
over a clinically important range of sensitivity
(25 to 75) for cardiac mortality arrhythmic
events
Comparison of positive predictive characteristics
of HRV (solid lines) of combinations of HRV
with LVEF (dashed lines) of HRV with LVEF
ectopic counts on 24-hour ECGs (dotted lines)
used for identification of pts at risk of 1-year
cardiac mortality (a) 1-year arrhythmic events
(SCD and/or symptomatic suVT, b) after acute MI
62
Interpreting HRV
  • Depressed HRV After Acute MI
  • ?HRV is a predictor of mortality arrhythmic
    complications that is independent of other
    recognized risk factors / measured 1 wk pMI
  • HRV assessed from short-term recordings provides
    prognostic information (initial screening ), but
    HRV measured in 24-h recordings is a stronger
    risk predictor
  • Better prognosis c time domain HRV measures (SDNN
    or HRV triangular index). Some other measures,
    e.g, ULF of entire 24-h spectral analysis,
    perform equally well. A high-risk gp may be
    selected by the dichotomy limits of SDNN lt50 ms
    or HRV triangular index lt15
  • Predictive value of HRV alone is modest, but
    higher than that of any other recognized risk
    factor. To improve the predictive value, HRV may
    be combined with other factors
  • Assessment of Diabetic Autonomic Neuropathy (DAN)
  • Characterized by early widespread neuronal
    degeneration of small nerve fibers of both
    sympathetic parasympathetic tracts.
    Manifestations postural hypotension, pers.t
    tachycardia, gustatory sweating, gastroparesis,
    bladder atony, nocturnal diarrhea. Once DAN
    supervenes, 5-y mortality is 50. Thus, early
    subclinical detection is important
  • There are 3 HRV methods from which to choose (1)
    simple bedside RR interval methods, (2) long-term
    time domain measures that are more sensitive and
    more reproducible than the short-term tests, and
    (3) frequency domain analysis performed under
    short-term steady state conditions, which is
    useful in separating sympathetic from
    parasympathetic abnormalities
Write a Comment
User Comments (0)
About PowerShow.com