Title: Deli Sivekov
1Aberration-Free Akreos MI60 Versus Spherical
Clariflex and Aspherical Tecnis Z9002
Eye clinic of the 3rd Faculty of Medicine,
Prague, Czech Republic
- Deli Siveková M.D., Magdalena Vokrojová M.D.,
- Lucia Wágnerová M.D., Prof. Pavel Kuchynka, CSc.
The authors have no financial interest in the
presentation material
2Purpose
- To compare visual functions after cataract
surgery with implantation of different types of
IOL. - Prospective nonrandomized, non-comparative single
center study.
3Examination
- Preoperatively - UCVA
- - BCVA
- - Wavefront analysis
- Postoperatively- UCVA
- - BCVA
- - Wavefront analysis
- - Contrast sensitivity
-
4Methods
- 65 eyes of 40 patients
-
- 24 (60) women, 16 (40) men
- Mean age 71,6 (64-84) years
- Except for cataract, all patients were in healthy
ocular condition -
- Surgery standard operation technique
- temporal approach
- incision 2,6mm
- Infinity Vision system (Alcon)
- topical anesthesia
5Materials and methods
- In all patients there was a wavefront analysis
and contrast sensitivity. - - wavefront - analysis was performed
monocularly with Wavescan (VISX, AMO) 3 months
postoperatively. - - contrast sensitivity (CSV-1000,
VectorVision) was measured monocularly with the
best distance correction 3 months
postoperatively under photopic light conditions
(85 cd/m2).
63 groups of IOLs
Group A Group B Group C
Akreos AO MI60 (BauschLomb) Clariflex (AMO) Tecnis Z9002 (AMO)
7Akreos AO MI60 Clariflex Tecnis Z9002
Structure Single piece, foldable, 360 posterior ridge 3-pieces, foldable, UV - blocking filter 3-pieces, foldable, UV - blocking filter
Optics Aspheric aberration-free optics Biconvex optic diameter 6,0 mm Biconvex optic diameter 6,0 mm
Material hydrofilic acrylate polymethyl - diphenylsiloxane polydimethyl - diphenylsiloxane
Refractive index 1,45 1,46 1,46
A-constant 118,4 118,0 118,4
8Results(n65)
Akreos AO MI60 Clariflex Tecnis Z9002
UCVA 0,64 (0,5 1,0) 0,93 (0,63 - 1,0) 0,94 (0,63 - 1,0)
BCVA 0,95 (0,63 1,0) 0,97 (0,8 - 1,0) 0,98 (0,8 - 1,0)
9Results(n65)
Akreos AO MI60 Clariflex Tecnis Z9002
RMS error (Z 5,0mm) 0,22 0,05 SD 0,19 0,09 SD 0,15 0,05 SD
SA (Z 5,0mm) 0,11 0,06 SD 0,17 0,06 SD -0,02 0,05 SD
10Comparison
plt0,0001
11Contrast sensitivity
pgt0,05
plt0,0001
plt0,0001
Akreos AO MI60
Tecnis Z9002
Clariflex
12Conclusion
- We have found statistically significant
difference in postoperative measured HOA and
spherical aberration between our groups. - Results of the examination of contrast
sensitivity show that aspheric and aberration
free intraocular lenses are significantly better
than spherical lenses. These results between
aspherical IOL and aberration free IOL was not
found statistically significant. - We did not find any statistically significant
difference in our results of postoperative BCVA.