Vet ABC Troubleshooting
2
CHANGING THE PACK
  • Service menu Change Pack Remove the
    old pack etc. (on-screen instructions)
  • Remove the Red caps from the new pack
  • Install new pack ensuring arrows have disappeared
  • Prime the new pack into an open vessel
  • Re-install waste pipe onto new pipe
  • Write on the pack the installation date

3
Operational Problems
  • Basic understanding of how the ABC works
  • Function of the main components
  • How to spot potential problems
  • How to carry out minor repairs

4
(No Transcript)
5
(No Transcript)
6
STARTUP FAILED CHECK REAGENTS
  •   Blank values high PLT high, pack can be
    contaminated, reagent too cold
  •    Perform concentrated cleaning or change pack
  •   Electromagnetic noise (X-rays)
  •    Isolate Vet ABC and run start-up

7
STARTUP FAILED CHECK REAGENTS
  •   If other parameters are OK, photometer blank
    failed
  •    Check photometer (see below)
  •   Pack was shaken while in transport (bubbles)
  •    After Transport by a carrier, wait 1-2 hours
    before installation
  •   Tube on valve, needle, chamber or dispenser
    block is loose   Check tube connections and
    repair if possible

8
SENSOR DEFECT OR REAGENT EMPTY
  • Pack wasnt installed properly
  • Re-seat the pack ensuring arrows have disappeared
  • Valve on base of pack has collapsed inwards
  • Adjust the position of the valve
  • Pack is empty
  • Change the pack

9
SENSOR DEFECT OR REAGENT EMPTY
  • Pack is distended with excess air
  • Burp the pack
  • Red caps were not removed
  • Remove the Red caps
  • Waste tube is blocked or bent
  • Check tube is free of obstructions

10
ERROR MOTOR PRESSURE SYRINGE
  • Pack is distended with excess air
  • Burp the pack
  • Piston is pushed out of chamber by high pressure
  • Re-install the piston
  • Piston is sticking
  • Check movement Service menu Mechanic
    Pressure Syringe.

11
ERROR LIQUID SYRINGE MOTOR
  • Piston or gear is damaged
  • Check movement Service menu Mechanic
    Liquid Syringe
  • Dilution Block is damaged
  • Check for leaks Call Woodley Equipment.

12
START-UP FAILED HGB REFERENCE FAILURE
  • WBC Chamber is contaminated
  • Run concentrated clean
  • Reagent is contaminated
  • Change pack

13
START-UP FAILED HGB REFERENCE FAILURE
  • No reagent is available
  • Prime the pack check that reagent appears in
    chambers
  • Photometer is out of adjustment
  • Run concentrated clean Check photometer
    adjustment Service menu Technician
    Adjustments

14
(No Transcript)
15
IF IN ANY DOUBT CALL WOODLEY!