Title: Waveguide for Station 5


1
Waveguide for Station 5
  • M. Yoshida
  • (Osaka Univ.)

2
  • Test assembly
  • 450cm-long
  • Select better fiber canes
  • External waveguide
  • 290cm
  • Internal waveguide
  • 140cm
  • Quality assessment has been done in Osaka Univ.

3
Fiber preparation- Population of measured fiber
brightness -
  • Cut fibers into 450cm canes
  • Glue fibers on test cookies
  • Photo of flash reflection
  • LED trasmission

12 bad fibers rejected
4
Connector parts of waveguide for station 5
  • Bulkhead/Lead connector (1A / 2A)
  • 1.10 mm drilling with countersunk hole on the
    back end (by G-Tech)
  • 22-way connector (3A)
  • 1.08 mm holes by 1.067 mm drill (AC precision)
  • D0 connector
  • 1.067 mm drilling (Fermilab)
  • Flexible conduit
  • Adaptaflex PAFS21 and PAFS28
  • Heat shrink tube

5
External waveguide
6
Internal Waveguide
1A
3A
3B
1
1-20
2
21-42
3
1-128
43-64
4
65-86
5
87-108
6
109-128
1400mm
7
Internal Connector (1A,3A)
1-1
1 - 20
1-2
1 - 22
1-3
1 - 22
1-4
1 - 22
3A Surface
1-5
1 - 22
1-6
1 - 20
1A Surface
8
External Connector(D0,2A)
2A Surface
D0 Surface
9
Mechanical strength of the external waveguide
  • D0 connector boot and flexible conduit can be
    detached by pulling by hand
  • Found that heat shrink tube at the connection
    does not have enough strength
  • Epoxy glue at the connections, and then covered
    by heat shrink tube
  • Also, we will find better heat shrink tube
  • Glue paint on inner wall of the tube

10
Flash light reflection- External waveguide -
  • No fiber has reflection at D0 connector
  • One fiber has reflection at lead connector end
  • Incomplete optical cement / Air bobble (?)
  • Damage of larger diameter fiber (?)
  • Need more careful treatment and feedback

11
Flash light reflection- Internal waveguide -
  • All the fiber have no reflection light
  • Slight damage on cladding layer of the 5th 22-way
    connector
  • The problem seems to be just on the surface
  • Could be recovered by final polishing at Fermilab

12
Channel Assignment
handy flash
  • Inject light from a clear fiber to connector
    surface
  • Scan injection point by hand
  • Take movie
  • Perfect assignment on all the connectors!

waveguide
fiber
scan
camera
13
LED transmission test- External waveguide -
  • All the fiber have good transmission within 10

10
16 13170
14
LED transmission test- Internal waveguide -
  • All the fiber have good transmission within 10

10
11914 (87)
15
Summary
  • The first production of external/internal
    waveguide show good transmission in all the
    channels.
  • 3 bundle with test cookies have been delivered
    (675 fibers in total)
  • 450 cm long
  • Check bad fibers by Takatomi and Hideyuki, and
    send them back to G-Tech in this week.
  • G-Tech will ship 5 sets of waveguides to Fermilab
    in the end of Feb.
  • The fabrication process will be improved by
  • Fixation by glue at the conduit connections
  • Better heat shrink tube with glue painted inside
  • More careful gluing on the connector surface
  • Strict rejection of larger diameter fiber