Title: Steve Smith
1- Steve Smith
- for
- J. Frisch, T. Borden, H. Loos, T. Montagne, M.
Ross, D. Schultz, J. Wu, et al - April 20, 2006
2Applications of Bunch Length
- Beam longitudinal profile for accelerator
physics - Calibrated profile needed to understand machine
- Measurement can be low rate, invasive
- Bunch length signal for feedback
- Non invasive
- Signal at full repetition rate of beam
- Only need an output which is monotonic and stable
with respect to bunch length tuning phases
3Bunch Length Monitor Requirements
- After BC1
- 80 to 360 microns at 1nC
- 130 600 GHz Gaussian width
- 25 to 120 microns at 0.2nC
- 400GHz 2THz Gaussian width
- After BC2
- 8 to 40 microns at 1nC
- 1.2THz to 6 THz
- 4 to 20 microns at 0.2nC
- 2.4THz to 12 THz
- Bunches not Gaussian ? frequency distribution
will be somewhat different. - Goal for commissioning run
- Instrument and commission BC1
- Gain operational experience
- Discussion here almost entirely for BC1
4Measurement Options
- Temporal
- Works like a high speed oscilloscope.
- Transverse deflection Cavity (LOLA)
- Electro-optical measurement
- Spectral
- Measure power spectrum radiated by beam
- Coherent radiation
- Any kind
- Synchrotron
- Edge
- Diffraction
- Gap
- Spectral measurement does not include phase
- ?information is lost.
5Precision Measurement
- Transverse RF deflection structure (LOLA)
- High resolution
- directly calibrated
- using known phase shifts.
- Measurement from LOLA
- TTF at DESY
- 4 micron resolution (13 fs) demonstrated ???????
- Intercepting
13 femtosecond FWHM spike!
1 picosecond
- LOLA deflection cavity installed in LCLS will be
used as the Gold Standard bunch length
measurement - Beam physics experiments
- Calibration of spectral detectors
- Run LOLA at some slow rate (as needed)
- to maintain calibration of non-intercepting bunch
length monitors
6Coherent Radiation Detectors
- BC1 range is 100GHz to 1THz
- (BC2 to 10THz)
- Corresponds to the 100um to few mm wavelength
range for BC1 - Two approaches
- Waveguides
- Optics
- Standard microwave waveguide techniques difficult
above 10 GHz - near impossible at THz
- Free space quasi-optical techniques difficult at
longer wavelengths (mm) due to diffraction. - Materials absorption not well known in this
frequency range. - Calibrated measurement difficult
- Saved by LOLA
- Use both free-space and waveguide technology at
BC1
7Spectral Measurements
- Detect coherent radiation two ways
- CSR or Edge radiation in a bend
- Coherent radiation from ceramic gap
- Both provide order of a microJoule of energy.
- CSR/edge radiation provides somewhat more power
and lower divergence - easier to collect on the detector.
- Radiation from last bend of BC1 available
- BL11 is CSR/edge radiation detector
- Easy to add ceramic gap downstream.
- BL12 is gap radiation detector
- Calibration of bunch length vs. spectral power
- difficult to do from first principles
- but we have transverse cavity (LOLA)
- As long as signal is monotonic and reproducible,
we can do periodic calibrations - Eliminates the most serious problems with
spectral detection.
8Detectors
- High performance mm-wave detectors are cryogenic.
- Used for astronomy, etc.
- Avoid cryogenics if possible
- Room temperature detectors in principle have an
energy sensitivity of - Ethermal kBT 10-20J.
- Real detectors much worse
- Two common technologies
- Pyroelectric
- Diodes
9Diodes vs. Pyroelectrics
- Diodes limited to 750GHz
- Diodes have better sensitivity
- Diodes have worse dynamic range, 10,0001, but
this is probably not a limit - Diodes more expensive (5K at high frequencies),
- rather than 500 (including preamplifier for pyro
- Diodes are more damage sensitive.
10Waveguide Attenuation
- Waveguides available as small as WR-0.51
- Internal Dimensions 130um X 65um
- Frequency 1.4-2.2 THz
- Attenuation can be very high for small waveguide
- 3dB/M at 100GHz
- 17dB/M at 300GHz
- 100dB/M at 1THz
- (attenuations from empirical fit to data)
- Limits use of waveguide at high frequency
11Waveguide vs. Free space
Compare Rayleigh length for free space with
sigma .5cm relative to length for 10dB
attenuation in Waveguide Approximate
cross-over At 400GHz
12Coherent Synchrotron Radiation
- Narrow opening angle, large transverse size at
end of magnet suggest use of free space optics to
image onto detector. - Expect order of 1uJ collected on detector.
- gt1000X Pyroelectric sensor sensitivity.
- No advantage to diodes here
- Since free space optics works well at high
frequencies, this seems a good solution for
frequencies gt250GHz
13Conceptual Design
Diagnostics
Focusing
200mm
DR
10mm
Bend
ER
Mirror
Focusing f 200mm
SR
38mm
200mm
14BL11 Bunch length monitor
- Use CS/edge radiation
- free space
- pyroelectric detector.
- Systems like this already in use
- M. Hogan at SPPS
- Retractable mirror in vacuum.
- Use flat mirror
- Off axis parabola would collect slightly more
signal - but has difficult alignment issues
- Slight modification of existing vacuum chamber
and insertion design - Hole for beam passage
- Small optical table for detector components
- Insertable wavelength filters.
- Alignment diode
- has phase space similar to mm-wave radiation
15BL 11 Quasi-Optical / Pyroelectric Monitor
- Image coherent synchrotron edge radiation on
pyroelectric detector
16BC1 Radiation Distribution
- Wavelength 1mm
- 200mm downstream of BC1
- Near field integration of acceleration field
- Edge length ??²
- Mainly ER from both bend edges, 4x larger than SR
- Radiation from Entrance edge hits vacuum chamber
Horizontal Pol.
Vertical Pol.
17Propagate Gauss-Laguerre Modes
- Use Gauss-Laguerre modes with radial mode number
1 for field of each polarization - Needs ?/2 transverse modes to get correct far
field distribution
Horizontal polarization at magnet edge
? 1cm ? 500
18CER Transmission Through Optics
For one polarization, normalized to total 2p
emission
3 cm-1
at detector
15 cm-1
3 cm-1
15 cm-1
19Transverse Profile Through Optics
3 cm-1
15 cm-1
20Is Interference of CER CDR a Problem?
- Get field at detector for CER and CDR
- CDR is not focused on detector
- Wave front curvature differs from CER
- Intensity at detector shows narrow fringe pattern
- Fringes much faster than changes in form factor
- Conclusion
- CDR can be ignored
21Predicted Detector Signal vs Bunch Length
22Pyroelectric Detectors
- Crystal which converts thermal directly to
electrical output - LiTaO3
- physics is fast nanosecond
- coatings can slow down the detectors.
- Integrate all input energy (DC-gamma rays)
- Very good linearity up to damage threshold.
- Act as current sources, approximately 1uC/J
- Noise limited by preamplifier.
23Pyroelectric Detector Sensitivity
- ELTEC420m3
- 5mm diameter detector (20mm2).
- 0.3 uC/Joule sensitivity
- Detector capacitance Cd 100 pF
- A good charge preamplifier (Amptek A250F) should
see 300 electrons RMS noise - based on 100pf capacitance
- Corresponds to 0.15nJ detector noise.
- Parts cost
- Detector 75
- Pre-amplfier about 500.
- Threshold sensitivity 7.5pJ/mm2
24BL11 Bunch Length MonitorDevelopment Plan
- Use of flat mirror in vacuum and existing chamber
/ mover design minimizes engineering before
installation - Optics and detectors on table can be modified as
needed - only humidity proof cover required
- Serves as a model for the BC2 bunch length
monitor - Short bunch length / high frequencies requires
pyro detectors - allows for easy use of free space optics .
25Radiation from Gap
- 1 nC, 200micron bunch, 1cm gap gives about 2 uJ
total energy - Calculations fro Juhao Wu
- Radiation is distributed over a wide area
difficult to collect. - Corresponds to about 1.6nJ/mm2 for a 2cm radius
gap - Pyroelectric detectors (7.5pJ/mm2) marginal
(especially for a 0.2nC bunch). - Diode detectors (.03 to 0.4 pJ/mm2 depending on
wavelength) look OK. - RF horn will gain 10-20dB in diode detectors
- but probably lose 10dB in waveguide at high
frequencies - Looks reasonable, limited by waveguide, and diode
frequency response to frequencies below about
500GHz.
26 BL12 Bunch Length Monitor
- Located just after the BL11 monitor
- Uses gap and diode detectors
- Only vacuum component is conventional ceramic gap
- Initially instrument with 100GHz diodes
- Add higher frequency diodes as needed
- Diodes used in pairs to reduce effect of beam
motion - 20cm waveguide used to disperse pulse (1ns),
keep peak power reasonable on diodes. - 20dB gain horns on diodes
27BL12 Waveguide / Diode Monitor
28RF Diode Detectors
- Very fast diode to rectify the input signal
- Vout ?? Pin
- for input voltages lt diode drop
- Typically modest output impedance ( few KOhm).
- Linear output range limited to 100mV.
- Use waveguide dispersion to stretch mm-wave pulse
to keep diode in linear range. - Very high sensitivity 1V/W, or 1mC/J .
- Typically connected to waveguide
- Many vendors for Flt130GHz
- Few (only Virginia diodes found so far) for
higher frequencies up to 800GHz.
29RF Diode 100 GHz
- Millitech DXP-10
- WR10 input waveguide
- Active area 3mm2.
- 20dB gain horn available.
- Approx 2KOhm output impedance
- Output charge 0.15mC/J.
- Capacitance small 1pf.
- Assume A250F charge amplifier
- expect 100 electrons noise
- corresponds to 0.1pJ detector noise
- Maximum linear signal 500pJ
- Cost 1K, preamplifier 500
- Threshold sensitivity 0.03 pJ/mm2 (energy
density)
30Microwave Diode at 300GHz
- Virginia Diodes WR-2.2ZBD
- WR-2.2 input waveguide
- Active area 0.16mm2
- 20dB gain horn avaialble
- Output impedance 3KOhm
- Output charge 1mC/J
- Capacitance small 1pf.
- With A250F charge amplifier, expect 100
electrons noise, corresponds to 0.016 pJ detector
noise. - Max linear signal 0.1nJ
- Cost 5K, preamplifier 500
- Sensitivity 0.1pJ/mm2
- Note, for 750GHz diode get threshold sensitivity
0.4pJ / mm2
31BL12 Development Plan
- Similar diodes operating at 100GHz tested in End
Station A. - Additional test in end station A in April 2006
- using same electronics as LCLS
- Will use pair of diodes to check measurement
noise. - Initial test in LCLS will be done with a pair of
100GHz detectors. - As shorter bunch length measurements are
required, additional diodes and waveguide can be
added - Use of optical breadboard makes installation of
new diodes (on optical clamp mount)
straightforward. - Will try using a pyroelectric detector mounted
next to the gap. - Should be able to measure total mm-wave power
- compare with toroid current measurement to get
bunch length signal - Very simple and inexpensive system if it works.
- In principal extends to very short bunch lengths
- Must be calibrated with LOLA
32Controls Interface
- Pyroelectric detectors, and diodes will use very
similar nuclear physics type charge sensitive
preamplifiers - Signals can be read with a conventional GADC
(gated ADC). - Initially will use SLC CAMAC ADC
- Existing software for control and histories
- Can provide slow feedback to main LCLS EPICS
control system for feedback tests. - For high bandwidth feedback convert to EPICS GADC
in VME. - Other controls interface is straightforward
- pneumatic actuators
- temperature monitoring
33Summary
- Two coherent radiation bunch-length monitors for
BC1 - Measure bunch length every pulse
- non-intercepting
- BL11
- Quasi-optical
- Pyroelectric readout
- BL12
- Waveguide
- Diode readout
- Both are calibrated by transverse deflecting
cavity