Title: Pacemaker Malfunction
1Pacemaker Malfunction
2ECG Signs of Pacemaker Malfunction
- Failure to output
- Failure to capture
- Undersensing
- Inappropriate pacemaker rate
- true malfunctions / pseudomalfunctions
3Failure to Output
- Causes
- Oversensing
- Battery depletion ( of time pacing is required,
thresholds necessary, single- or dual-chamber
pacing, rate-modulation and other features) - Circuit failure (Lead fracture, disconnection of
lead from the PM) - Random component failure (rare)
- Pseudomalfunction
- Crosstalk (dual-chamber PM)
- when the monitor system does not display the
pacemaker stimulus - artifact when it is really present (frequently
with bipolar pacing)
4Oversensing
- Unexpected sensing of an intracardiac or
extracardiac signal. - Intermittent or constant
- Electrical signals that may cause oversensing
include myopotentials, T waves, and P - waves.
- Oversensing of myopotentials in a single-chamber
PM may result in pauses, - whereas oversensing of myopotentials by the
atrial-sensing circuit of a dual- - chamber PM may result in rapid paced
rhythms. - Corrected by reprogramming the sensitivity or by
reprogramming the refractory - period of the channel on which oversensing is
occurring - Oversensing of extracardiac events occurs much
less commonly with bipolar - sensing.
5dual-chamber bipolar PM. Paced ventricular
activity is absent after the third paced atrial
beat because of oversensing
PM programmed to unipolar sensing. Sensing of
myopotentials led to symptomatic pauses.
Reprogramming the pacemaker to a bipolar sensing
prevented myopotential oversensing
6Cross talk
- Not a true malfunction
- In a dual-chamber PM, the PM stimulus in one
chamber is sensed in - the other chamber.
- For the PM-dependent patient, inhibition could
result in ventricular asystole. - Most PMs have two methods of protection
- Interposition of a short 'blanking period' of
refractoriness in the ventricular channel
simultaneous with the atrial output stimulus. - Ventricular "safety pacing," whereby any event
sensed on the ventricular sensing circuit within
a defined early portion of the AV delay initiates
the delivery of a ventricular pacing stimulus.
7DDD pacemaker. After the third atrial pacing
artifact, there is evidence of atrial
depolarization, but there is no ventricular
pacing output. Failure to deliver the ventricular
pacing artifact is due to crosstalk
8Failure to Capture
- Cause
- Dislodgment of the PM lead from the endocardial
surface (usually in the first - few weeks)
- Break in the insulation of the PM catheter allows
some of the current from the - electrode to escape into the surrounding
tissues. - Impending battery depletion
- If the pacing threshold required to depolarize
the myocardium is greater than the programmed
voltage amplitude and pulse duration (Poor lead
position, exit block etc.) - Marked metabolic abnormalities, such as
hyperkalemia, and some cardioactive - drugs, such as flecainide.
- Pseudomalfunction
- Inappropriately low voltage-amplitude and
pulse-duration settings - A PM artifact occurring within the myocardial
refractory period
9DDDR pacemaker. All but one ventricular pacing
artifact fail to result in ventricular
depolarization, that is, failure to capture
Intermittent ventricular failure to capture in a
patient with a dual-chamber pacemaker
10Undersensing
- Intermittent or total
- Rarely an urgent problem
- Results in PM output that is undesirably
competitive with the intrinsic rhythm. - Can result in an unwanted rhythm (eg. atrial
pacing that competes with NSR may - result in AF).
- Competition in the ventricle is possible but is
almost never a problem except when - the fibrillation threshold has been altered by
ischemia, electrolyte imbalance, or - some other metabolic abnormality.
- Undersensing can frequently be corrected by
reprogramming the sensitivity
11- Causes
- Lead dislodgment
- Poor lead position at the time of implantation
- Interruption in the insulation of the pacing
catheter. - Delivery of a low-amplitude P wave or QRS complex
to a normally functioning - pacing system. (eg. concomitant drug therapy,
body position, MI, and - cardiomyopathy)
- Pseudomalfunction
- Magnet application
- Environmental electrical noise
- When a P wave or QRS complex falls within the RP
of PM - Monitor artifact
- In dual-chamber PMs, apparent undersensing may
occur during the initial portion of - the AV interval (blanking period).
- During this interval, the ventricular channel of
the pacemaker is refractory to avoid - sensing of the atrial stimulus and
depolarization. If an intrinsic ventricular event
12ventricular PM programmed to 50 bpm. The second
and third pacing artifacts occur inappropriately
close to the preceding QRS complex (that is, at
less than 1200 ms the programmed rate of the PM)
Functional undersensing. ventricular PM
programmed to 70 bpm. There is failure to capture
with the second and third pacing artifacts. The
third pacing stimulus occurs 850 ms after the
preceding intrinsic QRS (normal sensing). The
fourth pacing stimulus occurs 640 ms after the
preceding intrinsic QRS, indicating that the
preceding QRS complex was not sensed because it
occurred during the pacemaker's ventricular RP
13Inappropriate Pacing Rates
- Runaway pacemaker
- May be lethal for both PM-dependent and
non-dependent patients. - Pacing rates may be very rapid and cause
hemodynamic instability and collapse. - Result from battery failure, random component
failure, or component failure - induced by therapeutic radiation.
- Treatment is disabling the faulty pacemaker
output - Pseudomalfunction
- Sensing abnormalities
- Tracking of atrial fibrillatory or flutter waves
- PM re-entrant tachycardia (seen in dual-chamber
PMs occurs when sensing of a - retrograde atrial depolarization initiates
ventricular pacing, which in turn leads - to retrograde conduction and repetition of the
cycle). - Electromagnetic interference from the patient's
environment may cause the - generator to be reset to a rate different from
that programmed. - Many PMs operate at a slower rate when battery
depletion is imminent.
14Pacemaker re-entrant tachycardia with a
dual-chamber PM. Atrial and ventricular pacing
stimuli precede the first three paced complexes
at a rate of 80 bpm. A PVC follows the third
paced ventricular complex and is conducted in a
retrograde to the atria. The atrial activation is
sensed by the PM and initiates ventricular
pacing. The pacing rate is limited to the
programmed upper rate limit of 110 bpm
15Environmental Causes of Pacemaker Malfunction
- Electrocautery
- Causes temporary sensing problems or
reprogramming (rarely permanent alteration) -
- Transthoracic Defibrillation
- Can cause reversion to back-up mode, transient
increases in capture threshold - and loss of capture as well as destruction of
the PM generator and circuitry - Damage is minimized by positioning paddles
anteroposteriorly and as far from - the pacemaker or lead as possible (ideally
15cm). - Extracorporeal Shock-Wave Lithotripsy
- Usually synchronized to the patient's ventricular
depolarization or to the output - stimulus of the PM.
- In dual-chamber PMs, synchronization of the
lithotriptor with the atrial output - can result in inhibition of ventricular output.
- In patients with an rate-adaptive PM, sensing of
the shock waves can result in - increased pacing rates and damage to the
piezoelectric crystal. - The following guidelines should be followed 1)
Program the PM to the VVI or - VOO mode 2) keep the focal point of the
lithotriptor at least 6 inches away
16- Magnetic Resonance Imaging
- All pacemakers to revert to asynchronous mode
because of reed-switch closure. - Investigations have shown that MRI does not
permanently damage the reed - switch or other pulse generator components.
- The radiofrequency does not alter the acutely
programmed variables, change the - normal magnet rate, or induce pacing in most
pacemakers tested. - A study by Vahlhaus et al. of 32 patients with PM
exposed to MRI at 0.5 Tesla. - Lead impedance, sensing and stimulation
thresholds did not change, battery - current and impedance tended to increase, no
change in programmed data or the - ability to interrogate and program. No
irreversible changes in PM systems -
PACE 2001 24 489-95 - In general, MRI should be avoided in a patient
with an implanted PM. - MRI may be attempted in non-PM-dependent patients
if the device can be - programmed to an output at which there is
failure to capture or OOO mode. - Low magnetic field (0.5 Tesla) is preferable.
- The patient should be monitored by pulse
oximetry, blood pressure and ECG
17- Transcutaneous Electrical Nerve Stimulation
- Appears to be safe in most patients with
permanent PMs. - It is not known how close to the PM the
stimulator can be placed, and it is best - to avoid applying the stimulator to a vector or
path that would be parallel to the - pacing lead.
- In patients with VD or DDD PMs, it may result in
an increased ventricular rate. - Therapeutic Radiation
- Failure of various battery components or
accelerated battery depletion - Changes in sensing capability, failure of
telemetry function, runaway function - and complete shut down may all occur
- No specific prediction relative to dose can be
made. - Particularly patients undergoing radiation for
thoracic / chest wall malignancy. - Precautions - Position the field of radiation at
an angle oblique to the PM, total - accumulated dosage limit of 2 rad, shielding of
the PM with a 1cm margin may - be required.
- If this is not possible, the PM should be
explanted and moved to another site .
18- RFA
- PM should be checked before and after ablation.
- Rate response function should be turned off.
- RF applications should be as brief as possible
and remote from the electrode tip. - If the patient is not dependent, the pacemaker
can be programmed to OOO or - VVI at a lower rate than the intrinsic heart
rate. - If the patient is dependent, the PM should be
programmed to VOO mode - Uncommon sources of EMI
- dental instruments, including ultrasound scalers
and cleaners, and electrosurgical instruments can
cause transient inhibition of PM output - Certain cardiac monitoring systems can cause
inappropriate rate changes in - patients with rate adaptive pacing systems that
use a minute ventilation sensor. - Antitheft devices may cause EMI with pacemakers.
- Acoustomagnetic systems may cause reversion to
asynchronous pacing and - rapid ventricular pacing.
19Nonmedical Equipment and Devices
- Cellular phones
- May interact with PM function by inhibiting the
pacing output, asynchronous - pacing and ventricular triggering.
- Hayes DL et al in a multicenter study studied 980
patients. - The incidence of all types of interference was
20. - Ventricular tracking of signals sensed on the
atrial channel, noise reversion and - inhibition of ventricular output were the most
commonly observed. - The incidence of overall clinical significant
interference was 6.6 . - Interference that was definitely clinically
significant occurred in only 1.7 of - tests and only when the phone was held over the
pacemaker. - Interference was more common in dual chamber
systems (25.3 ) than in single - chamber systems (6.8 ) and in digital
telephones (24) compared to analog - telephones (3 ).
- Patients who are PM dependent should use an
analog type cellular phone - system. Carrying the phone on the same side of
the body as the implanted - PM may cause interference. When using the phone,
it should be held at - least 15 cm away from the PM and on the opposite
ear. - N Eng J Med 1997 336 1473-9
20- Potentially significant restrictions exist for a
small subset of patients - who work in environments with equipment capable
of causing - significant electromagnetic interference eg.
internal combustion - engines, arc welding equipment, degaussing
equipment and induction - ovens.
- The use of bipolar leads can minimize or
eliminate the problem. - Microwaves should not cause any problem. Metal
detectors could - theoretically cause inhibition of a single beat,
but significant clinical - sequelae should not result.
21Triboelectric simulation
- Static electricity or triboelectric phenomena
occur more commonly during - very cold weather and very low relative
humidity. - Triboelectric signals are usually wider and more
irregular than pacemaker - stimuli
- Relatively prolonged overshoot is typical of
electrostatic discharge
22 23Pacemaker Syndrome
- First described in 1969 by Mitsui et al.
- Incidence ranges from 2 to 83.
- Defined as the symptoms associated with right
ventricular pacing relieved with the - return of A-V and V-V synchrony.
- The symptoms include DOE, PND, orthopnea,
hypotension, pre-syncope, and - syncope.
- Additional symptoms include easy fatigability,
malaise, headache, and the sensation - of fullness and pulsations in the head and neck
- Symptoms are most severe when intact V-A
conduction is intact. - Signs of congestive heart failure, cannon
A-waves.