Title: Physiological Fluid Dynamics
1??????
- Physiological Fluid Dynamics
Arterial Fluid Dynamics ??? ???????????
2Evolution of Arterial Pressure Away from the heart
3Systemic Arteries
- Conduct blood flow from Left ventricle (LV) to
peripheral organs - Aortic valve ? Aortic arch (180 turn)
- Geometry changes Tapering
- Geometry changes Branching
- Mechanical properties changes
4(No Transcript)
5Radius and wall thickness of human arteries
change with age
6Incremental Elastic Modulus of human arteries
change with age
7Stress-Strain relations of rabbits thoracic
aorta
8Fluid Mechanics of Elastic Conduct
- Mass Conservation
- Conservation of momentum
- Conservation of energy
9Background
- Fundamental Variables Pressure? Flow
- Geometrical Variables Size? Thickness ? Length?
Curvature - Mechanical Properties Stiffness ?
Visco-Elasticity
10Equations of Viscous Pipe Flow
- Consider a conduct filled with incompressible
fluid of density ? and pressure p, let u be the
only non-zero velocity component
11Poiseuilles Law (1840)
- Assume steady flow, u u(r ) with no body forces,
the equation of motion
12Laminar Poiseuillean flow
- Rate of flow through the tube
- Mean velocity of flow
- Shear stress at the wall
13Laminar Poiseuillean flow
- Skin friction
- Shear stress in terms of skin friction
14Implication of Poiseuilles Law
- Q is proportional to the fourth power of the
radius. - Q is directly proportional to the pressure
difference. - Q is inversely proportional to the length of the
tube. - If the arteries becomes constricted, the blood
pressure requires to supply the blood flow
adequately will rise substantially,leading to the
state of hypertension.
15Optimum design of Blood Vessel Bifurcation
(Poiseuilles formula)
For a given pressure drop, 1 change in vessel
radius results in a 4 changes in flow
Murray (1926) Rosen (1967)
Work done
Metabolism Energy loss
16Minimum cost function for optimum vessel
configuration
With respect to radius a ? the optimum radius
The optimum vessel radius is proportional flow
to the 1/3 power, and
17Optimum vessel bifurcation that with minimum cost
function
Minimize P at the bifurcation point B
An optimum location B would be
for arbitrary movements of B.
18Let B displaced along A-B direction first
The optimum is obtained when
19Again, let B displaced in the C-B direction
The optimum is obtained when
Similarly, displaced B along D-B direction, we
find
20Similarly, displaced B along D-B direction, we
find
The continuity equation gives
We find
which is often referred to as Murrays Law
??37.5
21Let ao denotes the radius of the aorta, and
assume equal bifurcation in all generation
If the capillary blood vessel has a radius of 5
um and the radius of the aorta is 1.5 cm.
We find n30.
The total number of blood vessel is about
230?109.
Note in fact arteries rarely bifurcation
symmetrically (a1a2). For human, only one
symmetric bifurcation. For dog, there are none.
22Pulsatile Blood Flow
- Consider pulsatile flow in a circular vessel,
pp(x, t) and u u(r, t)
23Pulsatile Blood Flow(2)
- The general solution of the ODE in the form
involves Bessel functions of complex arguments
U(ra)0 (non-slip)
U(r0)finite
24Pulsatile Blood Flow(3)
- Introducing Womersley number ?
- As ??0, the velocity profile becomes parabolic.
- As ???, viscosity is negligible U(r)-i P/??.
25Analysis of Blood Flow using Elastic Theory
- From Poiseuilles Law
-
- the flux is proportional to the pressure
difference (p1-p2). However, the blood flow in
veins are remarkably non-linear. - The flow in elastic conduct gradually attains a
maximum value as the pressure difference
increases and then on longer increases.
26Arterial Flow in Elastic Tube
- Axial velocity, v
- Lumen area, S
27Pressure-Diameter relationship
- Let T denotes the tension of the blood vessel per
unit thickness, wall thickness h, vessel radius a - Let ro be the radius of zero tension state, the
Hookes Law gives elastic constant E as
28Poiseuilles flow in elastic tube
- Consider steady flow in elastic tube of length L,
assume the tube is long and the pressure is
function of axial coordinate z, let P1 and P2
denote the inlet and outlet pressure and the
external pressure surrounding the tube is P0 - Assume the flow through the tube obey
Poiseuilles law, the flow becomes
29Transmission of Pulse wave (Velocity) in elastic
tube
- Consider inviscid and incompressible fluid flow
in elastic tube of lumen area A, - By linearizing the equations
30Transmission of Pulse wave (Velocity) in elastic
tube
- Combining the continuity and momentum equations,
- The wave equations
Pulse Wave Velocity (PWV)
31Analysis of Aortic Diastolic and Systolic
Pressure Waveforms
- Constitutive relationship between aortic volume
and pressure where K is the
volume elasticity of the aorta, and V0 is the
end-systolic volume. - If the aorta is very soft (K is very small), let
I(t) and Q(t) denote the inflow and outflow
rates, we have
32- During diastole, the aortic valve is closed and
there is no flow into the aorta. Hence I(t)
0. where ? is a
non-invasive measure aortic volume elasticity.
Let Td be the duration of diastolic phase, the
aortic pressure (Pd) at the end of this phase or
just prior to ejection is given by - The volume elasticity that depicts the
exponential drop of aortic pressure is given by
33Reynolds Strouhal Womersley
- Reynolds number
- Strouhal number
- Womersley number
34Flows under the action of Oscillating pressure
gradient
35(No Transcript)
36Wave propagation in Blood Vessel
- Pulse wave propagation in arteries
- A(x, t) depends on transmural pressure,
Here c is the wave propagation velocity.
37For thin walled elastic tube
- Consider the elasticity of the tube, arterial
diameter ? blood pressure
For a thick walled elastic tube
38Balance of Force in Arterial Wall
39Resonant vibration of flow in a circular tube
- When the tube length is equal to the half wave
length
- This is called the fundamental frequency of the
natural vibration. - Hemodynamics Effects of Frequency on the
Pressure-flow relationship of Arterial tree
40Boundary conditions
41Pressure-Flow
42Mean Velocity Profile (Dog Aorta)
43Velocity waveform at the upper descending aorta
of a dog
44Effect of Womersly number on the velocity
distribution
45(No Transcript)
46Blood PressureEvolution
47(No Transcript)
48Effect of sinusoidal pressure wave speed of
various frequencies on the instantaneous aortic
pressure
49(No Transcript)
50Distribution of Atherosclerotic Sites in Human
51Stress Concentration Conditions
52Atherosclerotic disease at the carotid bifurcation
Stress contours in arterial branching
53Whats the blood flows in reality?
- Unsteady
- Non-uniform geometry
- Bifurcations
- Non-Newtonian
- Viscoelastic wall
- Fluid-solid interactions
54In vitro measurement of artery pressures and flows
55(No Transcript)
56(No Transcript)
57Ultrasonic flowmeters
58Electromagnetic flowmeters
- Electromagnetic flowmeters have existed for
measurement of blood flow rate outside the body
during open heart surgery.
This miniature probe is for acute and chronic,
low flow measurements in small animals and
rodents. Sizes from 1 to 10 mm internal
circumference
59Electromagnetic flowmeters
60Electromagnetic flowmeters
- Faraday's principle of electromagnetic induction
can be applied to any electrical conductor
(including blood) which moves through a magnetic
field. The electromagnetic blood flowmeter is
sometimes used during vascular surgery to measure
the quantity of blood passing through a vessel or
graft, before during or after surgery. A circular
probe with a gap to fit the vessel is fitted
around the vessel. This probe applies an
alternating magnetic field across the vessel and
detects the voltage induced by the flow via small
electrodes in contact with the vessel.
61Electromagnetic flowmeters
- Alternating magnetic fields (typically at 400 Hz)
are used since the induced voltages are in the
microvolt region and d.c. electrode potentials
may cause significant errors with unchanging
magnetic fields. A number of probes are required
to fit the various diameters of blood vessel. - An alternative design carries the sensing device
on the tip of a special catheter which passes
inside the vessel and generates a magnetic field
in the space around it and has the electrodes on
its surface.
62SQUARE-WAVE ELECTROMAGNETICBLOOD FLOWMETERS
63Pulse Oximetry
- Takuo Aoyagi(1974) developed the principle of
pulse oximetry. The next year, Nihon Kohden
introduced the world's first ear oximeter,
OLV-5100, which used pulse oximetry to
noninvasively measure saturated blood oxygen
without the need to sample blood. All pulse
oximeters today are based on Dr. Aoyagi's
original principle of pulse oximetry.
64(No Transcript)
65CCA Root
66(No Transcript)
67(No Transcript)
68(No Transcript)
69(No Transcript)
70Variation of velocity waveforms across the
arterial vessel
- Common Carotid Artery (CCA)
71CCA
72CCA
73CCA
74CCA
75CCA
76CCA
77CCA
78CCA
79CCA
80CCA
81CCA
82CCA
83CCA
84CCA
85Radial Artery
0.23 cm in diameter
86Brachial Artery
87Brachial Vein