Title: Measurement and Detection of Ionizing Radiation
1Measurement and Detection of Ionizing Radiation
- Ionizing radiation is invisible
- Many methods are available for detection and
measurement, including - Ionization detectors
- Scintillation detectors
- Biological methods
- Thermoluminescence
- Chemical methods free radicals produced
- Measurement of heat- energy dissipated
2Ionization
- Devices contain a gas that can be ionized
- A voltage is applied to the gas
- Specific instrumentation and types of measurement
depend on amount of voltage applied to the gas. - Three types of instruments
- Ion chambers
- Proportional counters
- Geiger-Mueller counters
3Log of electrical signal vs. voltage
4Radiation ionizes the gas. Ions move toward
electrodes, creating current.
http//www.science.uwaterloo.ca/cchieh/cact/nucte
k/interactdetector.html
5Ion chamber continued
- Voltage is high enough that ions reach the
electrodes, produce current. - Proportional to energy the more energy, the more
current. - Generally requires some amplification of the
signal. - Example of use pocket dosimeters
http//www.ludlums.com/images/dosimeter.jpg
6Proportional counters
------- --------------
---------- . . . . . . . . . . . .
. . . . .. . . . . . . . . . . . . . . . .
- Each ionization electron is accelerated by the
voltage so that it ionizes more of the gas. - The higher the energy of the radiation event, the
greater the avalanche, the higher the current - Each ionization event detected separately.
- Used for neutrons
7Geiger Mueller counters
http//www.pchemlabs.com/images/eberline-rm20-geig
er-counter-a.JPG
8How Geiger counters work
- Voltage is high enough that every radiation event
triggers a complete avalanche of ionized gas - Does not discriminate among different energy
levels - Each event is registered
- A quenching agent stops the reaction, resets gas
for next event - Slow response time (comparatively) but simpler
circuitry. - Good for simple, sturdy, instruments
- Best for gamma low efficiency for alpha, beta.
9More Geiger details
Higher voltage leads to constant avalanches
instrument pegs. Improved efficiency with
pancake probe collects more radiation due to
geometry.
10Proper use of Geiger counters as survey meters
- http//orise.orau.gov/reacts/guide/index.htm
- Units of radioactivity and radiation
- Radiation detection instruments and methods
- First check battery and check source
- Enclosed radioactive material of known amount
- Check level of background radiation
- Survey area in question
- Move survey instrument slowly
- Keep constant distance from object being
surveyed do not make contact.
11Solid scintillation counters
- Crystal-based
- Radiation hits crystal which releases visible
photons - Photons amplified by photomultiplier tube,
converts to electrical signal - Zinc sulfide
- Good detection of alpha particles, rapid response
time - Sodium iodide w/ thallium
- Good for detection of gamma
- New ones showing up
12http//www.fnrf.science.cmu.ac.th/theory/radiation
/Radiation20and20Radioactivity_files/image018.gi
f
13Liquid Scintillation counters
- Workhorse in biology labs for many years
- Very useful for beta emitters, some alpha
- Modern equipment
- Computer driven
http//www.gmi-inc.com/Genlab/Wallac20141420LS.j
pg
14Basic principles
- Radioactive sample is mixed with organic solvents
(cocktail) - Toluene replaced with biodegradable solvents
- Detergents allow up to 5 aqueous samples
- Radiation hits solvent, energy is absorbed by
solvent Energy passed to one or more fluors - Fluor emits visible light which is detected
- By fluorescence
- Amplified by photomultiplier, converted to
electrical signal.
15(No Transcript)
16Coincidence circuitry
- Photomultipliers very sensitive
- Inside of instrument completely dark
- Tubes give off thermal electrons
- Result would be very high background counts
- Coincidence circuitry compares results from 2
photomultipliers - Event not detected by both thermal electron
- Ignored
- Event detected by both is affect of beta particle
- Counted.
17Counts and energy discrimination
- As radiation travels through solvent, it gives up
energy - The more energy it has, the more fluor molecules
get excited and release photons - Thus, the higher the energy, the brighter the
flash - The higher the electrical pulse sent from the PMs
- Instruments can be electronically adjusted
- Discriminators set for different pulse height
- Able to count betas from H-3 vs. C-14 vs. P-32
18Beta energy spectra
cpm
Pulse height
19Summary of capabilities
- Pulse height
- From brightness of flash the more energetic the
radiation, the brighter the flash. - Discriminators (gain) in the instrument can be
set so you determine what energy you want
counted. - Number of pulses
- Corresponds to how many flashes, that is how many
radiation events (decays) the amount of
radioactivity.
20Difficulties with LSC
- Static electricity causes spurious high counts,
esp. when humidity is low - dont wipe outside of vials!
- Chemiluminescence chemical reactions in sample,
from overhead lights, glass. - Suspiciously high counts can be redone
chemi-induced high counts subside over time. - Quench
- Anything that interferes with counting
efficiency. - Measured counts per minute (cpm)
- Desired decompositions per minute (dpm)
21Counting efficiency
- Because samples are usually dispersed in clear
containers, geometry is favorable for energy
transfer in all directions and good light
emission - Not all decay events will get registered,
however, because no system is 100 efficient - We seek to know the of decompositions per
minute (dpm) but measure the counts per minute
(cpm). - Using standards helps determine efficiency.
22Effect of Quench
23All about quench
- Chemical quench
- Acids, bases, high salt, any chemical that
interferes with transfer of energy from the
solvent to the fluor. - Result fewer activated fluor molecules, less
intense flash, interpreted as a lower energy
event. - Color quench
- Colored material absorbs visible light from fluor
- Less intense flash, appears as lower energy event
24About quench -2
- Self absorption
- If particulate matter not well suspended, energy
not absorbed by fluor, not detected as well. Both
lowering of cpm and forcing into lower energy
range.
25Counting statistics
- Radioactive decay is a random event
- To be sure results are reliable, a minimum number
of decay events must be recorded. - Reliability depends on total number of counts!
- Example
- Statistical significance is the same in these two
cases - 10 minute count yielding 500 cpm
- 1 minute count yielding 5000 cpm.
- Both have total of 5000 counts
- Instruments have settings for stopping count when
a certain statistical threshold is reached.