Title: Radiation Protection in Radiotherapy
1Radiation Protection inRadiotherapy
IAEA Training Material on Radiation Protection in
Radiotherapy
- Part 11
- Medical Exposure Brachytherapy
- Lecture 1 (cont.) Sources, implant techniques
and equipment
24. Brachytherapy equipment
- Design considerations often similar to external
beam therapy - Many points made in part 10 of the course are
also relevant for brachytherapy
Nucletron
3Design Considerations (BSS appendix II.11)
- Equipment used in medical exposure shall be so
designed that - failure of a single component of the system be
promptly detectable so that any unplanned medical
exposure of patients is minimized - the incidence of human error in the delivery of
unplanned medical exposure be minimized
4Defence in depth example Timers
- Need two completely independent timers
- One should count time up, one down
- Should be tested regularly
5Defence in depth example
- Retraction of sources
- Normal power
- Backup battery (must be checked)
- Manual system
6Design Considerations BSS II.13
- Registrants and licensees, in specific
co-operation with suppliers - (a) the equipment conform to applicable standards
of the International Electrotechnical Commission
(IEC) and the ISO or to equivalent national
standards
7BSS appendix II Design criteria specific to
radiotherapy
- II.15. Registrants and licensees, in specific
co-operation with suppliers, shall ensure that - ...
- (e) radioactive sources for either teletherapy
or brachytherapy be so constructed that they
conform to the definition of a sealed source and
- (f) when appropriate, monitoring equipment be
installed or be available to give warning of an
unusual situation in the use of radiation
generators and radionuclide therapy equipment.
8Remote Afterloading Equipment
- The most complex pieces of equipment in
brachytherapy - Low dose rate units
- High dose rate units
- Many important design consideration in IEC
standard
9Low dose rate brachytherapy
- Selectron for gynecological brachytherapy
- 137-Cs pellets pushed into the applicators using
compressed air - Location of active and inactive pellets can be
chosen by the operator to optimize the source
loading for an individual patient - Shown are 6 channels - the red lights indicate
the location of an active source
Nucletron
10Other features
- No computer required
- Two independent timers
- Optical indication of source locations
- Permanent record through printout
- Key to avoid unauthorized use
11HDR brachytherapy units
- Must be located in a bunker
- Have multiple channels to allow the same source
to drive into many catheters/needles
MDS Nordion
12Nucletron HDR unit control
Printout permanent record
Keypad
Emergency off button
Display
Key
Key for source out
Memory card for transfer of the dwell
positions for the treatment of a particular
patient - labelled
13Catheters are indexed to avoid mixing them up
Transfer catheters are locked into place during
treatment - green light indicates the catheters
in use
14Transfer of sources
- Important design issue - connectors and transfer
tubes must be well designed and checked - Should be labelled and easily identifiable
The correct channel of the afterloader must be
connected to the correct applicator/needle in the
patient
15Regular maintenance is required
- Source drive must be working within specified
accuracy (typically 1-2mm) - Emergency buttons must work
- Manual retraction of the source in case of power
failure must work
16Regular maintenance is required
- Maintenance work should follow manufacturers
recommendations - All modifications MUST be documented
- A physicist should be notified to perform
appropriate tests
17LDR and HDR units are not all...
- Other brachytherapy equipment
- PDR (pulsed dose rate) units
- Seed implant equipment
- Endovascular brachytherapy
18LDR and HDR units are not all...
- Other brachytherapy equipment
- PDR units - similar to HDR
- Seed implant equipment - discussed in more detail
in the second lecture of part 6 - Endovascular brachytherapy
19Endovascular brachytherapy
- Details of function was discussed in part 6
- Some similarities to HDR units, however,
different catheters and sources are in use
Courtesy Guidant
20Centering of the source in the catheter
- May be important due to short range of many
radiation sources - Improves dose distribution on all vessel walls
Courtesy Guidant
21For all equipment...
- Operator must be familiar with the unit
- Maintenance staff must be trained and aware of
radiation protection issues - Present course can not replace manufacturer
training
22Auxiliary equipment for brachytherapy
- Diagnostic equipment for localization of the
applicators and/or sources - May be simulator (compare part 10) or other
diagnostic units (more details in the course on
diagnostics)
Courtesy Siemens
23Auxiliary equipment for brachytherapy
- Other equipment may also be required with
modifications for brachytherapy. - Prior to each implant its integrity must be
checked as patient safety and appropriate
treatment delivery can rely on it.
24A note on intercom systems
- Need to be able to see the patient - is he/she
comfortable? Is she/he moving? - Need to be able to talk to the patient
- Need to be able to hear if the patient is in
distress
255. Radiation protection issues in brachytherapy
- Patients, a variety of staff and potentially
visitors are involved - Use of sources with high activity
- Preparation of sources
- Insertion of sources
- Removal of sources
26Training and information staff
- Training of staff is essential
- This applies to radiation workers and others,
such as domestic or maintenance staff
27Information for patients
- Patients must be aware of the particulars of
their treatment prior to it commencing (and
consent to it) - It is often a shock for patients waking up after
an operation with catheters and needles in place
28Source change in High Dose Rate (HDR)
Brachytherapy
- Source change is required about every 3 to 4
months due to 192-Ir decay - Is also required if any variations of source
movement from the planned treatment are noticed
or if a source stuck in the patient...
29HDR brachytherapy source change
- Must be done by suitably qualified personal
- Calibration of the new source is essential - it
is NOT appropriate to trust the source
certificate only - Different calibration methods are possible - more
in the next lecture
30HDR brachytherapy source change
- Need to ensure source dimensions are unchanged
(fits in all applicators) - Need to verify source movement e.g. using a
suitable jig or a video camera for observation
31A source stuck in the patient
- Maybe due to kink in catheter or other problems
- In the case of HDR, the source MUST be removed
within lt 1min - This requires a suitably trained person (RSO,
medical physicist) to be present - Emergency procedures must be established
- Relevant equipment must be present (e.g.
radiation monitor, crank to rewind the source
drive cable manually)
32A source stuck in the patient
- Maybe due to kink in catheter or other problems
- In the case of HDR, the source MUST be removed
within lt 1min - This requires a suitably trained person (RSO,
medical physicist) to be present - Emergency procedures must be established
- Relevant equipment must be present (e.g.
radiation monitor, crank to rewind the source
drive cable manually)
The procedure must be trained regularly
33Transfer of sources to the patient in
afterloading procedures
- All source channels in the patient must be easily
identifiable - Mix-up of channels is a serious accident
- Transfer pathways should be kept as short as
possible
For HDR units the transfer time may need to be
considered for the treatment plan
34After a temporary implant is completed
- One must ensure that ALL sources are removed from
the patient - This requires
- accounting of sources
- monitoring of the patient
35After a temporary implant is completed
- One must ensure that ALL sources are stored
safely - This requires
- accounting of sources
- monitoring of the patient
- monitoring of the room
- Room monitoring may include the bed. However, as
only sealed sources are used in brachytherapy,
contamination of linen and cloths is typically
not a problem
36Summary
- Due to the use of radioactive sources in direct
contact with the patient, brachytherapy has the
potential of radiation protection problems - Written procedures, protocols and adherence to
good design can eliminate many sources of safety
issues
37Typical Radiation Levels
- Iodine-125 Prostate Implant
- 100 seeds of 0.36 mCi/seed 36 mCi
- less than 0.25 mR/h at 1m ? 0.0025 mSv/h
- 17 days for 1 mSv (Background)
- Gold-198 seed lip implant
- 8 seeds of 3 mCi/seed 24 mCi
- 6 mR/h at 1m ? 0.06 mSv/h
- 17 hours for 1 mSv (Background)
38Typical Radiation Levels
- Selectron LDR (Cs-137) Cervix insertion
- 10 pellets of 15 mCi/seed 150 mCi
- 20 mR/h at 1m ? 0.2 mSv/h
- 5 days for 1 mSv (Background)
- this is inside the room!
- microSelectron HDR (Ir-192) turned ON!
- 10 Ci source 10 000 mCi
- 4700 mR/h at 1m ? 47 mSv/h
- 1.3 minutes for 1 mSv (Background)
- door interlock ensures that no-one is in room
39Brachytherapy
- Additional points covering safety associated with
brachytherapy equipment are made in the lectures
on shielding (part 7) and on brachytherapy (part
6).
40Any questions?
41Question
- Please provide a list of equipment which you
would expect to find in a hot lab used for
preparation of sources for brachytherapy
42Issues and items for the Hot Lab
- Good ventilation and lighting
- Close to brachytherapy treatment area
- Lockable
- Shielding as required
- Appropriate signs and documentation
- Equipped with
- Telephone
- Radiation Monitor
- Well counter for calibration (or equivalent)
- Shielded workbench
- Safe for storage of sources
- Safe for waste
- Mobile storage safe
43Acknowledgement
- Mamoon Haque, Royal Prince Alfred Hospital,
Sydney - Craig Lewis, LRCC, London, Ontario