Title: Resolution Enhancement Compression- Synthetic Aperture Focusing Techniques
1Resolution Enhancement Compression-Synthetic
Aperture Focusing Techniques
- Student
- Hans Bethe
- Advisor Dr. Jose R. Sanchez
- Bradley University
- Department of Electrical Engineering
2Motivation
- Ultrasound Imaging is important in medical
diagnosis
Figure 1 Imaging fetus 1
Figure 2 Imaging pancreas 1
3Motivation
- Ultrasound imaging involves exciting transducer
and forming ultrasound pulses to be fired at
internal tissue - Synthetic Aperture Focusing Techniques (SAFT)
beam-forming techniques capable of enhancing
lateral resolution - Resolution Enhancement Compression (REC) coded
excitation (wave shaping) technique employed to
produce excitation signal capable of enhancing
axial resolution - Objectives
- a/ Investigate REC and SAFT techniques through
literature research and simulation - b/ Combine REC and SAFT
4Outline
- I. Ultrasound Imaging System
- II. Functional Requirements
- III. Progress
5I. Ultrasound Imaging System
Excitation (REC)
Image reconstruction system
Transducer
Beam-forming (SAFT)
Figure 3 Block diagram
6Transducer
- Converts signal or energy of one form to another
- In imaging, converts electrical signal to
ultrasound signal
Transducer
Target
Ultrasound pulses
Echoes
Figure 4 Ultrasound emission and reflection
7Image Reconstruction System
excitation
Pre- amplifier
Matched filter
A
Delay Unit
Transducer
Echo
image
A
Apodization
S
8Image Reconstruction System
excitation
Pre- amplifier
Matched filter
A
Delay Unit
Transducer
Echo
image
A
Apodization
S
9Image Reconstruction System
excitation
Pre- amplifier
Matched filter
A
Delay Unit
Transducer
Echo
image
A
Apodization
S
10Image Reconstruction System
excitation
Pre- amplifier
Matched filter
A
Delay Unit
Transducer
Echo
image
A
Apodization
S
11Image Reconstruction System
excitation
Pre- amplifier
Matched filter
A
Delay Unit
Transducer
Echo
image
A
Apodization
S
12Image Reconstruction System
excitation
Pre- amplifier
Matched filter
A
Delay Unit
Transducer
Echo
image
A
Apodization
S
13III. Functional Requirements
- A/ SAFT
- Transducer shall be a linear array comprising 128
elements - SAFT shall be performed through MATLAB Field II
- SAFT mode excite all elements and receive with 1
element person emission - Delay and sum calculations shall be performed
through a GPGPU - Total synthetic aperture processing time shall be
lt 1 second - (Adjustment total processing time shall be about
10-20 seconds) - Signal-to-noise ratio (SNR) of the images shall
be at least 50 dB
14III. Functional Requirements
- B/ REC
- Actual impulse response of system (denoted as
h1(t)) shall have a center frequency f0 of 2 MHz.
- System bandwidth shall be about 83.
- Sampling frequency fs shall be 400 MHz.
- Desired impulse response of imaging system
(denoted as h2(t) ) shall have a bandwidth about
1.5 times the bandwidth of h1(t). - The side lobes associated with compressed pulse
shall be reduced below 40 dB.
15(No Transcript)
16(No Transcript)
17Figure 5 Illustration of convolution equivalence
principle
18REC Mechanism
19REC Mechanism
20REC Mechanism
21REC Mechanism
22Figure 15 Illustration of convolution
equivalence principle
23K0(REC)
K0(CP)
Figure 16 Axial resolution between CP and REC
24QUESTIONS ?
25References
1 Ultrasound images gallery http//www.ultrasoun
d-images.com/pancreas.htm 2 http//sell.bizrice
.com/selling-leads/48391/Digital-Portable-Color-Do
ppler-Ultrasound-System.html 3 J. R. Sanchez
et al., "A Novel Coded Excitation Scheme to
Improve Spatial and Contrast Resolution of
Quantitative Ultrasound Imaging" IEEE Trans
Ultrasonics, Ferroelectrics, and Frequency
Control, vol. 56, no. 10, pp. 2111-2123, October
2009. 4 S. I. Nikolov, Synthetic Aperture
Tissue and Flow Ultrasound Imaging 5 T.
Misaridis and J. A. Jensen, Use of Modulated
Excitation Signals in Medical Ultrasound IEEE
Trans. Ultrason. Ferroelectr. Freq. Control, vol.
52, no. 2, February 2005. 6 M. L. Oelze,
Bandwidth and Resolution Enhancement Through
Pulse Compression, IEEE Trans. Ultrasonics,
Ferroelectrics, and Frequency Control, vol. 54,
no. 4, April 2007.
26References
7 J. R. Sanchez and M. L. Oelze, An Ultrasonic
Imaging Speckle-Suppression and
Contrast-Enhancement Technique by Means of
Frequency Compounding and Coded Excitation, IEEE
Trans. Ultrasonics, Ferroelectrics, and Frequency
Control, vol. 56, no. 7, Julyl 2009. 8 M.
Oelze, Improved Axial Resolution Using
Pre-enhanced Chirps and Pulse Compression, 2006
IEEE Ultrasonics Symposium 9 Tadeusz
Stepinski, An Implementation of Synthetic
Aperture Focusing Technique in Frequency Domain,
IEEE transactions on Ultrasonics, Ferroelectrics,
and Frequency control, vol. 54, no. 7, July
2007 10 J. A. Zagzebski, Essentials of
Ultrasound Physics
27Apodization
- Process of varying signal strengths in
transmission and reception across transducer - Reduces side lobes
- Signal strength will become progressively weaker
with increasing distance from the center - Control beam width gt improve or degrade lateral
resolution
Center
Figure 5 Illustration of apodization
28Beam width and lateral resolution
- Lateral resolution capability of imaging
system to distinguish 2 closely spaced objects
positioned perpendicular to the axis of
ultrasound beam - Larger beam width gt greater likelihood of
pulses covering objects gt echoes from reflectors
more likely to merge gt degrade lateral resolution
beam axis
transducer
beam
objects
1
2
3
Figure 6 Illustration of the effect beam width
has on lateral resolution
29II. Theoretical Background
30SAFT
- In synthetic aperture focusing techniques
(SAFT), a single transducer element is used both,
in transmit and receive modes - Each element in the transducer emits pulses one
by one
1 2 3
Pulse
Echo
target
Figure 7 Illustration of SAF
31The essence of SAFT is delay-and-sum (DAS)
operation
Transducer
L6
L3
L1
L9
pulses
Target
Figure 8 Illustration of DAS
32The essence of SAFT is delay-and-sum (DAS)
operation
Transducer
L6
L3
L1
L9
echoes
pulses
Target
Figure 8 Illustration of DAS
33The essence of SAFT is delay-and-sum (DAS)
operation
Transducer
L6
L3
L1
L9
echoes
pulses
Target
Figure 8 Illustration of DAS
34The essence of SAFT is delay-and-sum (DAS)
operation
Delay unit
Transducer
Transducer
L6
L3
L1
L9
echoes
pulses
Target
Figure 8 Illustration of DAS
35The essence of SAFT is delay-and-sum (DAS)
operation
Delay unit
Sum
Transducer
Transducer
L6
L3
L1
L9
echoes
pulses
Target
Figure 8 Illustration of DAS
36Figure 9 Illustration of delay-and-sum 4
37REC
- Before REC, conventional pulsing (CP) was used
- CP proved ineffective in term of image resolution
Figure 10 Resolution Comparison 3
Figure 11 Background-target separation 3
38WHY REC?
- To enhance image resolution by CP, increase
excitation voltage gt produces excessive heating
gt hazardous to patients gt a better excitation
technique is needed gt gave rise to the
investigation of REC - Advantages of REC
- a/ Improves axial resolution without increasing
acoustic peak power - b/ Offers the capability to obtain the optimal FM
chirp to increase the bandwidth of imaging system
39- REC a coded excitation technique (wave shaping)
- Employs Convolution Equivalence Principle to
generate pre-enhanced chirp excitation signal - Excitation by pre-enhanced chirp increases
bandwidth of imaging system gt produce
shorter-duration pulses gt increases axial
resolution - (axial resolution ability of imaging system to
distinguish objects closely spaced along - the axis of the beam)
objects
transducer
beam
beam axis
Figure 12 Illustration of axial resolution
40objects
echoes
Figure 13 Effect pulse duration has on axial
resolution
41Figure 16 Comparison between CP and REC