Encoding and Image Formation - PowerPoint PPT Presentation

1 / 69
About This Presentation
Title:

Encoding and Image Formation

Description:

... K space The data information is stored in the computer memory location called the K space. Maximum number of lines are 1024 Data collection ... – PowerPoint PPT presentation

Number of Views:22
Avg rating:3.0/5.0
Slides: 70
Provided by: Prin234
Category:

less

Transcript and Presenter's Notes

Title: Encoding and Image Formation


1
Encoding and Image Formation
  • Gradients
  • Slice selection
  • Frequency encoding
  • Phase encoding
  • Sampling
  • Data collection

2
Introduction
  • Encoding means the location of the MR signal and
    positioning it on the correct place in the image
  • RF at precessional frequency of hydrogen applied
    at 900 to B0 resonates and flips the NMV into
    transverse plane.
  • The individual magnetic moments of hydrogen is
    put into phase.
  • The coherent transverse magnetization precesses
    at Larmor frequency in the transverse plane.

3
  • A voltage (signal) is induced in the receiver
    coil placed in the transverse plane
  • This signal has a frequency equal to Larmor
    frequency of hydrogen (at 1.5 T 63.86 MHz)
  • The system must be able to locate the signal
    spatially in three dimensions, so that it can
    position each signal at the correct point on the
    image.
  • First it locates a slice.
  • Then it is located or encoded along both axes of
    the image.
  • This task is performed by magnetic gradients

4
Magnetic Gradients
  • Gradients are alterations to the main magnetic
    field and are generated by coils of wire located
    within the bore of the magnet.
  • The passage of current through a gradient coil
    induces a gradient magnetic field.
  • The gradient field either adds to or subtracts
    from B0.
  • B0 is altered in a linear fashion.

5
  • Magnetic field strength and therefore the
    precessional frequency of the nuclei situated in
    the long axis is deferent and is predictable.
  • This is called spatial encoding

positive
gradient 1 G per cm
negative
A
B
C
2 cm
2 cm
9998 G 42.5614 MHz
10002 G 42.5785 MHz
10000 G 42.57 MHz
6
X,Y,Z Gradient coils
7
  • There are three gradient coils (X,Y,Z) situated
    within the bore of the magnet
  • Z gradient alters the magnetic field strength
    along the Z- (long) axis
  • Y gradient alters the magnetic field strength
    along the Y- (vertical) axis of the magnet
  • X gradient alters the magnetic field strength
    along the X- (horizontal /transverse) axis of
    the magnet

Y
Z
X
8
  • The magnetic isocentre is the centre point of
    the axis of all three gradients, and the bore of
    the magnet.
  • The field strength remains unaltered at the
    isocentre

isocentre
Y
Z
X
9
Steep shallow gradients
  • When a gradient coil is switched on, the magnetic
    field strength is either subtracted from or added
    to B0 relative to the isocentre
  • The slope of the resulting magnetic field is the
    amplitude of the magnetic field gradient and it
    determines the rate of change of the magnetic
    field strength along the gradient axis.
  • Steep gradient slopes alter the magnetic field
    strength between two points more than shallow
    gradient slopes.
  • Steep gradient slopes therefore alter the
    precessional frequency of nuclei between two
    points, more than shallow gradients slopes

10
Slice selection
  • This is done by
  • first switching the appropriate gradient coil to
    alter the field strength and the precessional
    frequency at points along the corresponding axis,
    and
  • then by transmitting a selected band of RF
    frequencies to selectively excite the nuclei
    which precess in that particular frequencies.
  • Resonance of nuclei within the slice occurs
    because RF appropriate to that position is
    transmitted
  • Nuclei situated in other slices does not resonate
    because their precessional frequency is different.

11
  • Z-gradient selects axial slices
  • Y gradient selects coronal slices
  • X gradient selects sagittal slices

Y
Z
X
12
Slice thickness
  • To give each slice a thickness, a band of nuclei
    must be excited by the excitation pulse
  • The slope of the slice-select gradient determines
    the difference in precessional frequency between
    two points on the gradient.
  • Once a certain gradient slope is applied, the RF
    pulse transmitted to excite the slice, must
    contain a range of frequencies to match the
    difference in precessional frequency between two
    points
  • This frequency range is called the bandwidth.
  • As the RF is being transmitted at this point it
    is called the transmit bandwidth.

13
  • To achieve thin slices, a steep slice select
    slope and/or narrow bandwidth is applied
  • To achieve thick slices, a shallow slice select
    slope and/or broad transmit bandwidth is applied.

Steep gradient
Shallow gradient
slice select gradient
broad Bandwidth
Transmit bandwidth
Narrow Bandwidth
Thick slice
Thin slice
Thin slice
Thick slice
14
Gradient strength slice thickness
Shallow (weaker gradient)
Steeper ( strong) gradient
15
In Practice
  • The system automatically applies the appropriate
    gradient slope and transmit bandwidth according
    to the thickness of slice required.
  • The slice is excited by transmitting RF at the
    centre frequency corresponding to the
    precessional frequency of nuclei in the middle of
    the slice,
  • The bandwidth and gradient slope determine the
    range of nuclei that resonate on either side of
    the centre.

16
  • The gap between the slices is determined by the
    gradient slope and by the thickness of the slice.
  • In spin echo pulse sequences, the slice select
    gradient is switched on during the application of
    the 900 excitation pulse and during the 1800
    rephasing pulse, to excite and rephase each slice
    selectively.
  • In gradient echo, the slice select gradient is
    switched on during the excitation pulse only.

1800
900
900
Slice select gradient
17
Frequency encoding
  • Once a slice has been selected, the signal coming
    from it must be spatially located (encoded) along
    both axes of the image
  • Locating the signal along the long axis of
    anatomy is done by a process called frequency
    encoding
  • A gradient is applied along the selected axis
  • The precessional frequency of signal along the
    axis is therefore altered in a linear fashion.
  • The signal can now be located along the axis of
    the gradient according to its frequency

18
A
B
C
Nuclei in column A precess at frequency A
Nuclei in column C precess at frequency C
Nuclei in column B precess at frequency B
  • For frequency encoding of
  • Coronal sagittal images use z gradient
  • Axial images use X gradient
  • Axial images of Head use Y gradient

19
  • In practice
  • The frequency encoding gradient is switched on
    when the signal is received and is often called
    the readout gradient

1800
900
900
FID
Echo
FID
rephasing
dephasing
Frequency encoding gradient
peak
The steepness of the slope of the frequency
encoding gradient determines the size of the
anatomy covered Field Of View (FOV) along the
axis during scan.
20
Phase encoding
  • The location of the signal along the remaining
    third axis is achieved by a process called phase
    encoding.
  • This is achieved by applying a gradient along
    this remaining axis
  • A gradient is switched on it alters the speed of
    precession as well as the accumulated phase of
    the nuclei along their precessional path.
  • It produces a phase difference or shift between
    nuclei positioned along the axis.

21
Gradient phase difference
nuclei travel slower
14998 G 63.852 MHz
Loose phase
15000 G 63.86 MHz
Nuclei travel faster
15002 G 63.868 MHz
gain phase
22
  • When the phase encoding gradient is switched off,
    the magnetic field strength experienced by the
    nuclei returns to B0 and the precessional
    frequency of all the nuclei returns to the larmor
    frequency.
  • However the phase difference between nuclei
    remains
  • The nuclei travel at the same speed around their
    precessional paths, but their phases or positions
    are different.
  • This difference in phase between the nuclei is
    used to determine their position along the phase
    encoding gradient (axis).

23
  • In practice
  • The phase encoding gradient is switched on just
    before the application of the 1800 rephasing
    pulse in spin echo sequences.

1800
900
900
Phase encoding gradient
24
Summary of phase encoding
  • The phase encoding gradient alters the phase
    along the short axis of the anatomy
  • In Coronal images x gradient
  • In sagittal images - Y gradient
  • In axial images - Y gradient
  • Axial images of brain x gradient

25
Summary spatial encoding
  • The slice-select gradient is switched on
  • during the 90 and 180 pulses in spin echo pulse
    sequences , and
  • during the excitation pulse only in gradient echo
    pulse sequences
  • The slope of the slice-select gradient determines
    the slice thickness and slice gap (along with
    transmit bandwidth)

26
  • The phase encoding gradient is switched on
  • just before the 180 pulse in spin echo, and
  • between excitation and the signal collection in
    gradient echo.
  • The slope of the phase encoding gradient
    determines the degree of phase shift along the
    phase encoding axis.
  • The frequency encoding gradient is switched on
    during the collection of the signal
  • The amplitude of the frequency encoding gradient
    and the phase encoding gradient determines the
    two dimensions of the FOV

27
Gradient timing in spin echo
TR
1800
900
900
echo
Phase encode
slice select
slice select
Frequency encode
28
Sampling
  • The signal is collected during the frequency
    encoding gradient (readout gradient)
  • The duration of readout gradient is called
    sampling time
  • The system samples up to 1024 frequencies during
    sampling time
  • The rate at which the samples are taken is called
    the sampling rate

29
  • The number of samples taken determines the number
    of frequencies sampled
  • The range of frequencies is called the receive
    bandwidth

Frequency columns in FOV
f1
f2
f4
f3
f5
f6
Frequencies sampled are mapped across the FOV
along the frequency axis
Receive bandwidth
30
  • Sampling time, sampling rate and receive
    bandwidth are linked by a mathematical principle
    called the Nyquist theorem.
  • It states that any signal must be sampled at
    least twice per cycle in order to represent or
    reproduced it acurately.
  • In addition enough cycles must occur during the
    sampling time to achieve enough frequency samples
    ( if 256 samples are to be taken 128 cycles must
    occur during the sampling time)
  • Number of cycles occurring per second is
    determined by the receive bandwidth
  • Receive bandwidth is proportional to the Sampling
    rate

31
  • Sampling time is inversely proportional to
  • The sampling rate
  • The receive bandwidth
  • The receive bandwidth affect the minimum TE (
    because the sampling time is changed)
  • Reducing the receive bandwidth increase the TE
    (sampling time increases) vise versa
  • Usually the receive bandwidth sampling time are
    fixed

32
Nyquist theorum
Sampling once
Reproduced as a straight line
Sampling twice
Reproduced more accurately
33
Bandwidth versus sampling time
Sampling time (8 ms)
Bandwidth
128 cycles occur (256 samples can be taken)
16,000 Hz
64 cycles occur (only 128 samples can be taken)
8,000 Hz
If bandwidth is reduced, the sampling time must
be increased so that the same number of samples
can be taken
34
Data collection
  • Location of individual signals within the image
    by measuring the number of times the magnetic
    moments cross the receiver coil (frequency), and
    their position around their precessional path
    (phase)




Frequency shift
3 cycles/s
2 cycles/s
1 cycle/s
Phase shift
35
K space
  • The data information is stored in the computer
    memory location called the K space. Maximum
    number of lines are 1024

frequency
ve
outer
One line is filled for one phase encoding gradient
central
phase
-ve
36
Data collection step 1
  • During each TR the signal from each slice is
    phase encoded and frequency encoded.
  • A certain value of frequency shift is obtained
    according to the slope of the frequency encoding
    gradient, which is determined by the size of the
    FOV.
  • As the FOV remains unchanged during the scan, the
    frequency shift value remains the same.
  • A certain value of phase shift is also obtained
    according to the slope of the phase encoding
    gradient
  • The slope of the phase encoding gradient will
    determine which line of K space is filled with
    the data from that frequency and phase encoding

37
Phase shift pseudo-frequency
  • The system cannot measure the phase values
    directly
  • It can measure frequency
  • The phase shift values are converted to a sine
    wave
  • The frequency of this sine wave is called a
    pseudo-frequency
  • Different phase shift gradient produce different
    sine waves with different pseudo-frequency

38
The pseudo frequency curve
time
Phase shift value
39
Phase encoding gradient pseudo frequency
  • Steeper gradients results in high pseudo
    frequencies
  • Shallow gradients results in low frequencies

40
  • In order to fill out different lines of K space,
    the slope of the phase encoding gradient must be
    altered after each TR
  • With each phase encoding one line of K space is
    filled
  • Different lines in K space are filled after every
    TR
  • The phase encoding gradient is altered for every
    TR
  • In order to complete the acquisition all the
    lines of selected K space must have been filled
  • The number of lines that are filled is determined
    by the number of different phase encoding slopes
    that are applied

41
K space
Line 1 phase encode 1 frequency/phase data
Line 2 phase encode 2





Line 128 phase encode 128
42
Fast Fourier Transform (FFT)
  • The data in K space is converted into an image
    mathematically by Fourier Transform.
  • The receive signal is a composite of multiple
    signals with different frequencies and amplitudes
  • The signal intensity/time domain is converted to
    a signal intensity/frequency domain

Amplitude
RF intensity
Frequency
Time
Frequency domain
Time domain
43
Matrix FOV
  • The FOV relates to the amount of anatomy covered
  • It can be square or rectangular
  • Image consists of a matrix of pixels
  • Te number of pixels depends on the number of
    frequency samples and phase encodings
  • Matrix frequency samples x phase encodings

44
Matrix
4 frequency samples
8 frequency samples
8 phase samples
4 phase samples
Coarse matrix 4x4
Fine matrix 8 x 8
45
Data collection - step 2, NSA (NEX)
  • When all the lines of K space is filled the
    acquisition is over
  • But the signal can be sampled more than once with
    the same slope of phase encoding gradient.
  • Doing so each line of K space is filled more than
    once
  • The number of times each signal is sampled with
    the same slope of phase encoding gradient is
    usually called the number of signal averages
    (NSA) or the number of excitations (NEX).
  • The higher the NEX, the more data is stored and
    the amplitude of the signal at each frequency and
    phase shift is greater

46
Scan timing
  • Every TR, each slice is selected, phase encoded
    and frequency encoded.
  • The maximum number of slices that can be selected
    and encoded depends on the length of the TR.
  • E.g.
  • TR of 500ms may allow 12 slices.
  • TR of 2000 ms may allow 18 slices

47
TR number of slices
180
TR
90
echo
Slice 1
TE
Slice 2
Slice 3
Phase encode 1
Slice 4
Slice 1 second TR
Phase encode 2
48
  • The phase encoding gradient slope is altered
    every TR and is applied to each selected slice in
    order to phase encode it.
  • At each phase encode a different line of k space
    is filled. The number of phase encoding steps
    therefore affects the length of the scan
  • E.g. 256 phase encodings require 256 x TR to
    complete the scan.
  • The scan time is also affected by the number of
    times the signal is phase encoded with the same
    phase encoding gradient slope, or NEX . So,
  • Scan time TR x Number of phase encodings x NEX

49
K space filling
  • The negative half of the k space is a mirror
    image of the positive half.
  • The polarity of the phase gradient determines
    whether the positive or negative half is filled
  • Gradient polarity depends on the direction of the
    current through the gradient coil
  • The central lines are filled with data produced
    after the application of shallow phase encoding
    gradients
  • The outer lines are filled with data produced
    with steep phase encoding gradients

50
  • The steepness of the slope of the phase encoding
    gradient depends on the current driven through he
    coil.
  • The central lines of K space are usually filled
    first. (if 256 phase encodings are performed 128
    positive lines and 128 negative lines are filled.
  • The lines are usually filled sequentially either
    from top to bottom or from bottom to top

51
Signal amplitude phase shift gradient
  • The shallow phase encoding gradients have smaller
    phase shifts. The resultant signal therefore has
    a large amplitude
  • The steeper phase encoding gradients have larger
    phase shift along their axis and therefore small
    signal amplitudes

52
Phase encoding slope signal amplitude
Low amplitude
Steeper gradient
medium amplitude
medium gradient
shallow gradient
high amplitude
53
Signal amplitude frequency gradient
  • The vertical axis of k space correspond to the
    frequency encoding
  • The left of the k space is a mirror image of the
    right
  • The centre represents the maximum signal
    amplitude because all the magnetic moments are in
    phase
  • The magnetic moments on either side are either
    rephasing and dephasing and therefore the
    amplitude is less

54
Signal amplitude frequency gradient
Peak
Rephasing
Dephasing
55
K space filling spatial resolution
  • Number of phase encodings determines the number
    of pixels in the FOV along the phase encoding
    direction
  • If the FOV is fixed voxels of smaller dimensions
    result in an image with high spatial resolution
  • The steeper gradients result in high spatial
    resolution (two adjacent points have different
    phase values and can be differentiated)

56
  • The outer lines of K space contain data with high
    spatial resolution
  • The central lines of k space contain data with a
    low spatial resolution
  • The central portion of k space contains data that
    has high signal amplitude low spatial
    resolution
  • The outer portion of k space contains data that
    has low signal amplitude and high spatial
    resolution

57
Resolution Amplitude
High spatial resolution
High signal
High spatial resolution
58
Way of filling K space
  • The amplitude of frequency encoding gradient
    determines how far to the left and right K space
    is traversed and this in turn determines the size
    of the FOV in the frequency direction of the
    image
  • The amplitude of the phase encoding gradient
    determines how far up and down a line of K space
    is filled and in turn determines the size of the
    FOV in the phase direction of the image (or the
    spatial resolution when the FOV is square)
  • The polarity of each gradient defines the
    direction traveled through K space

59
K space filling in gradient echo
  • The frequency encoding gradient switches
    negatively to forcibly dephase the FID and then
    positively to rephase and produce a gradient echo
  • Frequency encoding gradient is negative, k space
    traversed from left to right
  • Frequency encoding gradient is positive, k space
    traversed from right to left
  • Phase encode gradient is positive , fills top
    half of K space
  • Phase encode gradient is negative, fills bottom
    half of K space

60
K space filling in gradient echo
Phase encode amplitude determines distance B
Positive gradient traverse from centre through
distance C
Negative gradient traverse from centre through
distance A
Line of k space filled
B
C
A
61
Manipulation of K space filling
  • The way in which K space is filled depends on how
    the data is acquired and can be manipulated to
    suit the circumstances of the scan e.g. in the
    following
  • Rectangular field of view
  • Anti-aliasing
  • Ultra fast pulse sequences
  • Respiratory compensation
  • Echo planar imaging

62
Partial or fractional echo imaging
  • This refers to when only part of the signal is
    read (sampled) during application of frequency
    encoding gradient
  • As the sampling time is reduced minimum TE can be
    reduced
  • This allows maximum T1 and proton density
    weighting and number of slices for a given TR

63
Minimum TE
Readout gradient
Partial echo imaging
Only this half is read
Minimum TE reduced
Only half of the k Space is filled
This extrapolated from filled segment
64
Partial or fractional averaging
  • The negative and positive halves of K space on
    each side of the phase axis are symmetrical and
    mirror image of each other
  • The filling of at least half of the lines is
    adequate to produce an image
  • If 60 of lines are to be filled only 60 of
    phase encodings are required and the remaining
    lines are filled with zeros
  • The scan time is there fore reduced
  • E.g. 256 phase encodings and, 1 TR and ¾ NEX is
    selected
  • This is called partial or fractional averaging

65
Partial averaging
If phase encodings 256 TR 1s NEX1, Scan
time 256 x 1 x 1 256 s
75 of k space is filled with data
If phase encodings 256 TR 1s NEX3/4, Scan
time 256 x ¾ x 1 192 s
25 is filled with zeros
66
PRE-SCAN
  • This is a method of calibration that should be
    performed before every data acquisition. It
    includes
  • Finding the centre frequency on which to transmit
    RF. I.e. Resonant frequency of water protons
    within the area under examination
  • Finding the exact magnitude of RF that must be
    transmitted to generate maximum signal in the
    coil. (to flip the NMV through 900)
  • Adjustment of the magnitude of the received
    signal so that it is not too large nor too small.

67
Reasons for failing pre-scan
  • The coil is not plugged in properly
  • The coil is faulty
  • Chemical saturation techniques are utilized and
    there is an uneven distribution of fat and water
    in the area to be saturated
  • The patient is either very large or very small

68
Types of acquisition
  • Sequential data collected for slice by slice
    (k- space for each slice is filled one by one)
  • Two-dimensional volumetric data collected for
    all the slices simultaneously (line 1 in first
    slice, then line 1 in slice 2)
  • Three-dimensional volumetric (volume
    imaging)-collect data from total volume. The
    excitation pulse is not slice selective, and the
    whole prescribed volume is excited. At the end of
    acquisition the volume is divided into partitions
    by slice select gradient which separates the
    slices according to their phase value along the
    gradient. (This is called slice encoding)

69
End
Write a Comment
User Comments (0)
About PowerShow.com