Title: Quality Control- What, Why, How,
1Quality Control-What, Why, How, How Often
2What?
- Quality control encompasses
- Instrumentation
- Reagents
- Users
- Method(s) used to ensure reproducibility of data
over time.
3Inter/Intra Instrument Contribution
- Instrumentation is variable
- An instrument varies over time
- Alignment drift
- PMT/Log Amp degredation
- Laser changes
- Fluidic problem (salt crystals, debris,
microorganisms) - Instruments from different manufacturers vary
- Analog/Digital electronics
- Sensitivity
- Filter sets used
4Reagent Contribution
- Lot to Lot variation
- ProteinFluorophore ratio
- Tandem conjugate efficiency
- Poor QC on manufacturers part
- Tube variability
- Exposure of fluorochrome to light and air
- Repeat freeze/thaw or cold/warm cycles
- Non Antibody Reagents
- Sheath, Buffers, Media, etc
5User Contribution
- Instrument Parameters
- User doesnt know how to properly use instrument
- Drastically changes voltages settings between
expts. - Doesnt understand compensation
- Staining Parameters
- Hasnt done proper optimization of staining
protocol - Poor bench skill
6Why?
- For a Core
- Instrumentation must work properly
- Providing a service that you charge for
- For clinicians
- Diagnosis/Prognosis based -ages and expression
7Why?
- People carelessly report ?MFI without taking into
consideration above variables. - People may abort experiments if trial doesnt
work.
8How?
- Beads, of course.
- Schwartz, et al1 established a taxonomy of bead
standards used for cytometer quality control and
quantitative flow cytometry - Alignment Particles (Type I)
- Instrument Set-up Particles (Type II)
- Quantitative/Calibration Particles (TypeIII)
9Schwartz A, et al Standardizing flow cytometry
a classification system of fluorescence standards
used for flow cytometry. Cytometry 33106-14
(1998)
10Schwartz A, et al Standardizing flow cytometry
a classification system of fluorescence standards
used for flow cytometry. Cytometry 33106-14
(1998)
11Instrument Performance
- Facility/owner is responsible
- Use Type I beads
- make sure laser alignment/power is stable
- Dichroics and stream are aligned
- No partial clogs/obstructions in stream
- On the whole, system is performing well
- Bead Properties
- Uniform in size and intensities (low CVs-use
linear scale) - Very bright in all channels and excited by many
laser lines - Stable over time (dye contained in bead)
- Not effected greatly by environment
12Instrument Performance Experimental Control
- Facilty and/or User is responsible
- Use Type II beads
- Ensure compensation electronics are functioning
- Establishing window of analysis
- Setting voltages and compensation (e.g. FACSComp)
- Bead Properties
- Sizes and intensities vary (use Log scale)
- Typically are same size as lymphs
- Can be environmentally sensitive, when
fluorochrome-matched
13Instrument Performance Quantitative Flow
- Responsibility of Facility and/or user
- Use Type III beads
- Measure linearity of Log amps
- Dynamic Range of analysis
- Detection Threshold or Resolution (Sensitivity)
- Used for Calibration in Quantitative Flow
- Bead Properties
- Uses actual fluorochrome on bead instead of dye
- Sensitive to environment and handling
- Short shelf-life
- Typically come in a set including particles with
several fluorescence intensities.
14How Often?
6 months or more Monthly Weekly Daily Per Experiment
Change sheath filter Bleach system Check alignment Check alignment Check alignment?
Optimize Laser/Dichroics on analyzers Check tubing/connectors Assess compensation issues Assess comp issues? Run Reference Beads?
Check Linearity Inspect instrument Run Calibration Beads?
Sensitivity
15Case Study
- IACF Flow Cytometry Facility at the University of
Chicago - Instrumentation
- 2 FACSCans Analog Electronics, 1 laser, 5
detector - 1 FACSCalibur Analog Electronics, 2 lasers, 6
detectors - 1 LSR II Digital Electronics, 3 lasers, 10
detectors - 1 FACStar Plus Analog Electronics, 2 lasers, 7
detectors - 1 Cyan LX Digilog Electronics, 3 lasers, 11
detectors - 1 MoFlo - Digilog Electronics, 3 lasers, 11
detectors
16Cell Sorter QC
- Daily and/or Per Experiment Alignment
- Includes laser steering, stream, and dichroics
- Monthly Bleaching of system
- Twice yearly laser alignment- or sooner!
17(No Transcript)
18(No Transcript)
19(No Transcript)
20(No Transcript)
21Benchtop Analyzers
- Weekly alignment check
- Flow Check for 488nm
- Molecular Probes 633nm beads
- Molecular Probes UV beads
- SpectrAlign Beads (good for all laser lines)
- Weekly Compensation Check
- Calibrite Beads (FITC, PE, PerCP, APC)
- Use FACSComp when available
- Monthly Bleaching
- Twice yearly, laser alignment/dichroic check
- Yearly, linearity, sensitivity check, or as
needed
22User Responsibility
- Run Reference beads (Type II) if doing
longitudinal study - Run Calibration beads (Type III) if doing
quantitative analysis - Problem
- Do users know they should be doing this type of
quality control??? - Probably not
23Record Keeping
- Using Type I beads
- Record MFI at specified voltage
- Record CV
- Chart both over time
- Using Type II beads
- Run compensation controls (FACSComp)
- Set up acceptable values for all (before
service is called)
24(No Transcript)
25Discussion Topics
- What is the responsibility of the Core
Facility/Non-Clinical Lab as far as quality
control is concerned? - How do clinical lab protocols differ?
- How often should routine alignment QC be
performed? Linearity? Sensitivity? - What type of beads are most useful for alignment?
- Compensation QCnecessary?
- How do you make users aware of potential
problems with instruments - How do you inform users of the necessity of
running Type II or III beads when doing
inter-experimental comparisons.
26References, Further Reading
- Schwartz A, et al Standardizing flow cytometry
a classification system of fluorescence standards
used for flow cytometry. Cytometry 33106-14
(1998) - Shapiro, H Practical Flow Cytometry, 4th
Edition 353-358. 2003. - Gandler W, Shapiro H. Logarithmic Amplifiers.
Cytometry 11447-450 (1990). - Purvis N, Stelzer G. Multi-Plateform, Multi-Site
Instrumentation and Reagent Standardization.
Cytometry 33156-165 (1998). - Whitby L, et al Quality Control of CD4
T-Lymphocyte Enumeration. Cytometry (Clinical
Cytometry) 50102-110 (2002). - Gratama J.W. Flow Cytometric Quantitation of
Immunofluorescence Intensity Problems and
Perspectives. Cytometry 33166-178 (1998). - Bagwell, C.B., A Simple and Rapid Method for
Determining the Linearity of a Flow Cytometer
Amplification System. Cytometry 10689-694
(1989) - Schmid, I. Conversion of Logarithmic Channel
Numbers Into Relative Linear Fluorescence
Intesity. Cytometry 9533-538 (1988).