Title: General
1General
2Introduction to Fourier Transform Infrared
Spectroscopy
3What is FTIR?
- FTIR stands for Fourier Transform Infra Red, the
preferred method of infrared spectroscopy. - A method for measuring all of the infrared
frequencies simultaneously, rather than
individually as with dispersive instruments
4Why Infrared Spectroscopy ?
- An Infrared spectrum represents a fingerprint of
a sample with absorption peaks which correspond
to the frequencies of vibrations between the
bonds of the atoms making up the material-Because
each different material is a unique combination
of atoms, no two compounds produce the exact same
spectrum, therefore IR can result in a unique
identification of every different kind of
material!
FingerPrint
5Infrared Spectroscopy is simply the study of the
interaction of Infrared light with matter!!
- Most powerful aspect is the ability to identify
complete unknowns - direct correlation between the wavenumbers at
which a molecule absorbs infrared radiation and
its chemical structure
6Infrared Spectroscopy
UV
VIS
NIR
IR
Far IR
Microwave
10-5
10-5
10-4
10-3
10-2
10-1
Wavelength (cm)
Energy (Frequency)
Light, or more properly, electromagnetic
radiation, can be described in terms of
frequency ? or wavelength ? . Energy of
radiation increases with increasing frequency
and decreases with increasing wavelength
7Wavenumbers (cm-1) - standard
IR spectral units
- Modern infrared instruments record the absorption
of energy as a function of decreasing frequency
(decreasing energy) from left to right - W ( cm-1) 1/?
- A frequency-related unit known as a wavenumber
is expressed as reciprocal centimeters- number
of vibrations occurring over 1 cm
8WAVENUMBER SCALE
9Origins of Infrared Absorptions
- All atoms within molecules are in constant
vibration - all bonds with a permanent dipole (slightly
electronegative charge) will absorb infrared
radiation at the appropriate vibrational
frequency (??/ ?X ? 0 ) - the vibrational frequency will depend on the
atoms associated with the bond as well as the
bond strength
10Symmetric stretching
In-plane bending
Antisymmetric stretching
Out-of-plane bending
When the molecule is irradiated with IR
electromagnetic radiation, the vibrating bond
will absorb energy if the frequencies of the
light and the vibration are identical
11Nature of Molecular Vibrations Normal Modes
STRETCHING
VIBRATING BONDS
CONTRACTING
When a molecule absorbs IR radiation, its
chemical bonds vibrate. The bonds can
stretch, contract, and bend. Vibrational motion
can be broken down into a number of constituent
vibrations called normal modes (Rough Guide-
of atoms in a molecule helps
determine of vibrations a molecule
possesses)
12Factors affecting frequency of
molecular vibrations
- Mass of the atoms - heavier molecules will result
in slower movement and result in lower
frequencies - Elasticity of the spring (force constant) - a
stronger spring (bond) will cause the vibration
to be more rapid than a weaker one- HIGHER
FREQUENCY - Bond Strength - increasing of bonds increases
frequency of vibration
13Dispersive IR Instruments
- Infrared light from the source passes through the
sample, then goes through a slit into a grating
to disperse the light into a spectrum of its
component wavenumbers. - The grating is rotated so that different
wavenumber slices pass through the slit and are
sampled by the detector
14A Dispersive Infrared Spectrophotometer
Shimadzu IR-470
15Older Technology- limitations
- Slow Measurements Because dispersive
instruments measure each frequency individually,
a single measurement may take as long as 15
minutes - Relatively Insensitive Older instruments
employed detectors which required a large amount
of sample. Today we can detect as little as 0.01
of a compound in a particular matrix with FTIR - Mechanically Complex- Because dispersive
instruments have lots of moving parts, they are
susceptible to mechanical failure increasing
possible instrument down time and the slits throw
most of the sample beam away at any one time
anyhow
16Advantages of FTIR (technique)
- Universal technique
- sensitivity 10-6 grams
- fast and easy
- relatively inexpensive
- rich information
- sensitive to molecules-anything that contains
chemical bonds - vast majority of molecules in the universe absorb
mid-infrared light, making it a highly useful
tool
17Disadvantages of FTIR
- Cannot detect atoms or monoatomic ions - single
atomic entities contain no chemical bonds - Cannot detect molecules comprised of two
identical atomS symmetric-such as N2 or O2. - Aqueous solutions are very difficult to analyze-
water is a strong IR absorber - Complex mixtures - samples give rise to complex
spectra
18Michelson Interferometer
IR radiation from the source is separated by the
beamsplitter into two beams that travel different
pathlengths.When these two separated beams are
recombined, an interference pattern is generated
and passed through the sample. The FTIR relies on
a mathematical (Fourier Transform) reconstruction
of the spectrum from this interference pattern
19Constructive Interference
? n? Constructive Interference will take
place for any value of ?
when the two beams are in phase
n0,1,2,3,..
20Destructive Interference
Fixed Mirror
Moving Mirror
? ( n 1/2 ) ? Totally destructive
interference takes place when optical path
difference is 1/2 ? or some multiple of it-beams
are completely out of phase
21At all path differences other than ZPD (zero path
difference), a variation of constructive and
destructiveinterference takes place - this
modulated radiation is denoted by how many
times a wave may switch between being light and
dark known as the frequency
22 where Fv Modulation
frequency Fv 2VW V moving
mirror velocity in cm/sec
W wavenumber passing
through interferometer -Both of these
wavenumbers are distinguished by the detector as
a cosine wave, but their frequencies
are different as expected from the equation above
23Interferogram of a Broadband Source
A plot of light intensity versus optical path
difference is an Interferogram These are the
fundamental measurements obtained by an FTIR
24Relationship between Interferogram and Power
Spectrum
25FTIR HhOWHHrinciples
How an Interferogram becomes a Spectrum!
26Fourier Transform of an Interferogram
- The wavelength and intensity data at each point
in the scanned spectrum are encoded so that all
of the spectral information is acquired at once
it is mathematically decoded through Fourier
Transform. - Essentially, wavelength and intensity are related
mathematically to mirror position and velocity.
27 What is Apodization?????? (comes from
Greek word a podi or no feet)
- Calculating A Fourier Transform involves
performing a mathematical integral on the IFG.
Ideally the limits of this integral should be /-
infinity (infinity for OPD and of data points
collected to Fourier Transform properly) - This
is impossible, so the IFG and the integral must
be truncated at some point, the limits being 0
OPD and the maximum OPD - An unfortunate outcome of truncating the signal
is the lineshapes become distorted causing
sidelobes which are sinusoidal undulations in
the baseline. These lobes, or feet are
suppressed by multiplying the IFG by an
Apodization Function
28Main Side-Effect of Apodization!!
- - Spectral Resolution Will be Reduced - AFs
vary in how well they suppress sidelobes, and how
much they degrade resolution - AF that provides the highest resolution, and does
the worst job of suppressing sidelobes, is the
boxcar apodization (only use with gases or where
utmost resolution is required) - AF recommended for Condensed Phase samples is
Happ-Genzel or Medium Beer-Norton
29Relationship between the Interferogram and the
Spectrum
30Multiplex (Fellgett) Advantage
All wavelengths measured all the time-because all
data is measured simultaneously, measurements are
made in seconds rather than several minutes
31Jacquinot ( Throughput) Advantage
'
All infrared radiation passes through the sample
and strikes the detector at once-there are no
slits to restrict the energy increased
sensitivity and reduced noise automatically
32Connes (Frequency Calibration) Advantage
Because the He-Ne laser emission occurs at
exactly 632.8nm, it is used to set the exact
wavenumber(cm-1) scale in the IR spectrum
Spectrum of high wavenumber accuracy is obtained
and you never have to calibrate
33The Signal-Averaging Advantage
- Fast scans with an FTIR enable the coaddition of
many scans in order to reduce the random
measurement noise-use the computer to do this
quite easily - The positive and negative fluctuations in the
random noise level cancel themselves out as more
scans are added together
34 FTIR 8300/8400/8700/8900 SOURCE
- Purpose provide radiant energy in the infrared
region of the electromagnetic spectrum - Ceramic Source/Globar- bright and very intense -
inert solid is heated electrically to 1500-2000 K
-guaranteed for 5000 hours
35(No Transcript)
36BEAMSPLITTER
- KBr is almost universally used as a substrate
material in FTIR beamsplitters-this material does
not split the beam since it transmits in the IR - Instead, a thin coating of Germanium is
sandwiched between two pieces (disks) of KBr, and
it is this Ge coating that splits the beam. So
they are referred to as Ge on KBr beamsplitters - Disadvantage KBrs natural tendency to absorb
moisture-fogging
37FTIR 83/84/87/8900 Interferometer
Mirror is the only
moving part
- FJS -Flexible Joint Support-
unique-patented - Shimadzu system
- Swing system with smooth film support
system. - Base plate swings while keeping parallel against
the - top plate.
- Mirror is mounted in front of the base.
- Mirror runs parallel with theTop plate
- Advantage
- Long endurance and precision because of
noncontact - free from wear
- High linearity and precision almost equal to an
air - bearing system
38(No Transcript)
39Friction-less Mirror Drive Requires No
Maintenance
- Unique, patented Shimadzu design employs a
friction-less electro-magnetic moving mirror
drive supported by two tough, durable polyimide
sheets (a flexible joint support system). - Stability and linearity of response is equivalent
to the Air-bearing Drive without the need for
pure, dry air
40FTIR83/84/87/8900 DETECTORS
- DLATGS (deuterated L-Alanine doped tri-glycine
Sulfate) - Shimadzu patented design - pyroelectric bolometer - changes in the amount of
infrared radiation striking the detector cause
the temperature of the DTGS element to change,
measured as a voltage change - -Temperature controlled high sensitivity
detector-better sensitivity than traditional DTGS
detector - MCT Detector (optional) - AIM Microscope required
- ----mercury cadmium telluride (HgCdTe)- consists
of an alloy of 3 elements, and it is a
semi-conductor. Energy is detected by use of a
bandgap (photons exceeding the energy bandgap
will be detected)
41He-Ne Laser
- Laser acts as an internal wavenumber standard. -
He-Ne gives off light at precisely 15,798.637
cm-1. All infrared frequencies are measured
relative to it. - Laser is used to track the position of the moving
mirror. - so the optical path difference can be
measured precisely and any adjustments will be
made instantaneously to the fixed mirror
positioning
42Dynamic Alignment System
43Dynamic Auto-Alignment for Maximum Reproducibility
- Dynamic feedback loop - maintains the optimum
alignment conditions during measurement, a
consistently stable and reproducible spectrum is
obtained. - Continuous laser referencing - operates
continuously at 5,000 to 10,000 Hz - Auto-Adjust Function - provides quick and easy
fine tuning of the interferometer. It eliminates
the need for tedious mechanical adjustments
44Dynamic Auto-Alignment (contd)
- Recalls last position at power-down
- Minimizes start-up time
- 5 min or less for Qualitative Analysis
- 30 min for Quantitative Analysis
- Eliminates the need to keep the FTIR on
continuously
45Triple Protection of the Interferometer
46Interferometer Optical Component Protection
features
- A high-tech proprietary anti-moisture coating is
applied directly to the KBr beamsplitter surface - Optical compartment is sealed and desiccated with
Calcium Oxide - will not re-release moisture back
into the optical compartment - KRS-5 Exit Window does not react with water, it
is impervious to moisture and will never fog-
Disadvantage only 70 transmissivity
47Common Causes of Instrument Problems?
- Beam blocked by something
- Low throughput accessory
- Poor (vibration,humidity) operational environment
- Failing Source
- Fogged Beamsplitter
- Poorly Aligned (tuned) moving mirror
- Failing detector
- Electronic Failures
48Initial Suggestions to Follow
- Follow infra red beam (Are laser spots in the
sample compartment visible on your hand?) - Make
sure no obstructionsin light path/ Does problem
exist with Accessories removed? - Turn power on/off - Look for laser lights. If not
present, replace laser tube/laser power supply -
49More Suggestions !
- Inspect beamsplitter for fogging, and replace if
necessary - Check the Moving Mirror in Hayato 105 (Service
Mode) and see if it moves freely-it should since
we are forcing it to move- If it does not then
inspect with different scanning speeds to observe
movement- if none- replace the CPU boards- FTIR
Main/Digital II
50(No Transcript)