Title: CLD


1
CLD
2
  • With dry air produced ozone is mixed with sample
    gas.
  • The mixture enters the reaction chamber, while
    the NO of the sample gas reacts to NO2
  • NO O3 ? NO2 O2 h?
  • A proportion of the reaction energy is emitted as
    infrared radiation (h?).
  • The infrared radiation is detected, enhanced and
    transformed into the detector signal by the
    photomultiplier.
  • The NO2 of the sample gas is converted into NO by
    the integrated converter and then measured in a
    separate reaction chamber.
  • The difference between both measurements
    corresponds to the NO2 content of the sample gas.

3
CLD - Measuring Principle
The chemiluminescence is disturbed by Quenching
effects. They are caused by absorption of the
emitted radiation by other components and by
radiation-free energy loss due to collision
processes. Remedy Reduced pressure (35 mbar) ?
reduction of collision processes. Selection of a
measuring range (650-850 nm) with low foreign
absorption. Due to the high amplification factor
of the photomultiplier (1100.000) emission
flanks can be used.